CFA-1100多通道连续流动分析仪技术要求
- 格式:doc
- 大小:40.50 KB
- 文档页数:4
安捷伦1100液相色谱仪各项性能指标悬赏分:5|解决时间:2010-10-16 22:13|提问者:4329211最佳答案朋友,直接致电agilent的800-820-3278免费电话就可以得到Agilent的满意答复了系列Agilent 1100泵系统●电子流控阀(EFC)控制的毛细液相泵系统,精度高、流速范围广柱流速范围:1-20ul/min;10-100ul/min(可选件)0.001-2.5m1/min(EFC关闭状态)●高压制备泵系统,单元或双元高压制备泵流速范围:0.001-100m1/min●分析型泵系统流速范围:单元泵:0.001-10m1/min二元泵:0.001-5m1/min四元泵:0.001-10m1/min品种齐全的Agilent 1100系列进样系统●标准手动进样器(分析型或制备型)●标准自动进样器样品瓶容量:可达100个(2mlx100)进样量:0.1-100ul(0.1-1800ul)可选件●微盘式自动进样器样品瓶容量:2x96(孔板),2x386(孔板)或100x2ml进样量:0.1-100ul(标准件)0.1-1500ul可选件●微量标准自动进样器/微盘式自动进样器进样量:0.01-8ul(标准);0.01-40ul(可选)●恒温标准自动进样器/微盘式自动进样器温度范围:4-40℃可设定步进1℃● 220型微孔板式自动进样器-组合化学样品管理系统样品瓶容量:各种规格试管多达12个微孔板(96孔板,384孔板)进样量:0.1-5ul;0.1-20ulAgilent 1100系列检测器●可变波长扫描紫外检测器(VWD)波长范围:190?600nm●多波长检测器(MWD)波长范围:190?950nm(双灯源)●二极管阵列检测器(DAD)波长范围:190?950nm(双灯源)●荧光检测器(FID)激发波长:200-700nm;发射波长:280-900nm光谱存储:全光谱●示差折光检测器(RID)温控:室温+5℃至55℃内置自动吹扫阀和自动溶剂循环阀●电化学检测器(ECD)● LC/MS四极杆质量检测器(MSD)● LC/MS离子阱质量检测器(Trap MSD)柱温箱和脱气机组件●柱温箱温度范围:室温下10℃至80℃可选件:柱切换阀●真空在线脱气机安捷伦1120液相色谱仪技术参数泵:•流速精度:≤0.07% RSD•流速范围:0.001-10.0ml/min,以0.001ml/min的增量自动进样器:•进样量范围:0.1-100ul,以0.1ul的增量•进样精度:<0.25% RSD(5-100ul),<1% RSD(1-5ul)•样品容量:100个2ml样品瓶;或者2个半托盘,每个40个2ml样品瓶或15个6ml样品瓶柱温箱:•温控范围:室温上10℃到60℃•温控稳定性:±0.15℃VWD:•噪音:±0.75×10 -5AU•波长范围:190-600nm主要特点仪器介绍特点介绍可选5种配置系统,工厂测试,开箱即用和各种检测器兼容:示差折光检测器(RID),荧光检测器(FLD),蒸发光散色检测器(ELSD)易于使用的EZChrom Elite Compact软件,提供英中日界面LMD软件在故障发生前就可以实时通报新的最佳性能色谱柱提供可靠重现的结果附加服务:1120 液相色谱仪有五种集成的一体化配置可供选择:1.等度泵手动进样VWD2.梯度泵手动进样VWD3.梯度泵手动进样柱温箱VWD4.梯度泵ALS 柱温箱VWD5.等度泵ALS 柱温箱VWD*其中包括了硬件、软件、色谱柱和LMD分析方法此仪器尚未上传分析方法相关资料1.安捷伦1120液相色谱仪详细技术参数与资料!!相关仪器1.Agilent1200液相色谱仪(1200)2.液相色谱仪LC2100(LC2100)相关耗材1.安捷伦C8、C18液相色谱柱(规格:C8、C18)2.安捷伦氘灯(规格:Agilent 1100)3.顶空进样瓶、自动进样瓶(规格:20ml)4.色谱微量进样器、液体进样器、气体进样器(规格:ul)5.安捷伦液相色谱住(规格:C8,C18)6.Agilent ZORBAX StableBond SB-C18液相色谱柱(规格:Agilent SB-C18)7.真空泵、无油真空抽滤泵(规格:QP-01)8.液相色谱柱温箱(规格:JK-LCagilent1200安捷伦液相色谱仪技术参数方便地利用四种溶剂进行等梯度或梯度分析,适用于方法快速开发和快速配备流动相,以及冲洗HPLC 系统流量范围宽,最高流速10 mL/min ,延迟体积800-1100 ?L,适用于微径柱、标准柱和半制备应用通过Instant Pilot 组件或安捷伦化学工作站或EZChrom 软件,可以容易地进行编程和控制包含微量真空脱气机,脱气效率高,实现了操作无障碍和最高性能,完全不需要通氦气直接从前面板快速更换维护部件通过自我诊断、内置日志和预编程的测试方法,快速判断问题早期维护反馈(EMF)长期不间断地监测仪器使用和用户预设的限度,一旦超出限度,将提供反馈信息在所有安捷伦1200 系列HPLC 组件范围内可升级和扩展主要特点agilent1200安捷伦液相色谱仪特点介绍同时安捷伦1200系列快速高分离液相色谱系统同样和常规高效液相色谱仪一样耐用,也保留了常规液相色谱仪的方法和工作原理。
关于agilent安捷伦液相色谱仪 1100技术参数的文章Agilent安捷伦液相色谱仪 1100技术参数引言:\n在现代科学研究和工业生产中,液相色谱仪是一种非常重要的分析仪器。
它可以用于分离、鉴定和定量分析各种化合物。
Agilent安捷伦液相色谱仪 1100是一款性能卓越的仪器,具有出色的技术参数,为科学家和工程师提供了强大的分析能力。
1. 高灵敏度检测器:\nAgilent安捷伦液相色谱仪1100配备了高灵敏度的检测器,包括紫外可见(UV-Vis)检测器和荧光检测器。
紫外可见检测器可以在200-800nm范围内进行波长选择,荧光检测器则可以实现更高的灵敏度和选择性。
这些检测器可以满足不同样品的分析需求,并提供准确可靠的结果。
2. 多样化的进样方式:\nAgilent安捷伦液相色谱仪1100支持多种进样方式,包括自动进样器、手动进样和微量进样。
自动进样器可以实现高通量分析,提高实验效率;手动进样适用于少量样品的分析;微量进样则适用于对样品浪费要求较高的分析。
这些进样方式的灵活性使得仪器可以适应不同实验需求。
3. 高精度的流量控制:\nAgilent安捷伦液相色谱仪1100具有高精度的流量控制系统,可以实现0.001-10mL/min范围内的流量调节。
这种高精度的流量控制可以确保分析结果的准确性和重复性,尤其对于需要精确定量分析的实验非常重要。
4. 多种柱温控制方式:\nAgilent安捷伦液相色谱仪1100提供了多种柱温控制方式,包括常温、恒温和梯度温控。
常温模式适用于一般分析;恒温模式可以保持柱温恒定,提高分离效果;梯度温控模式则可以在不同时间段内改变柱温,实现更复杂的分离。
这些柱温控制方式使得仪器具有更广泛的应用范围。
5. 强大的数据处理能力:\nAgilent安捷伦液相色谱仪1100配备了先进的数据处理软件,可以实现数据采集、分析和报告生成等功能。
该软件具有友好的用户界面和强大的数据处理能力,可以帮助用户快速准确地分析样品,并生成专业的报告。
163高锰酸盐指数(COD Mn )是指水中有机物质在酸性或者碱性条件下,被高锰酸钾作为氧化剂分解所需的化学氧化剂量。
作为反映水体中有机物氧化能力和水质污染程度的重要指标之一,一般作为地表水、生活饮用水、天然饮用矿泉水的测定指标,其数值越高说明水体中有机物越难以被氧化分解,水质污染程度越严重。
酸性高锰酸钾滴定法是在2023年10月1日实施《生活饮用水标准检验方法》(GB5750-2023)中所采用的测定方法,但该方法存在分析时间长、操作步骤复杂、易引入人为误差等问题,且难以保证大批量样品分析的质量要求。
使用连续流动分析技术,通过气泡间隔与在线加热,实现自动化加热、反应、出峰等操作,能有效减少操作误差,该技术的优点是灵敏度高、再现性好、试剂使用量少和分析速度快等[1-6]。
1 连续流动分析的原理连续流动分析(CFA)是利用蠕动泵压缩内径不同的弹性泵管,将取样器自动吸取的样品和试剂按照一定比例吸入管路系统中并在一定条件下混匀、分离、反应、显色完全后测定吸光度,用峰高定量,计算机处理分析结果的一种现代分析技术。
对样品类型和试剂的要求不高,适用范围相对较广,样品和试剂在连续流动的管路中进行完全的物理混合并发生化学反应。
所使用的管路普遍较粗(通常≥1mm)。
不同于流动注射分析(FIA)的是气泡间隔连续流动注射分析的过程中会向管路注入空气或氮气气泡,将样品流分隔成不同的片段流,使反应更加直观,利于观察操作的正确性,每一段样品的片段流均发生相同的反应,以此来提高分析精密度。
气泡的引入起到了间隔和清洗作用,降低了管路的壁源效应,减小了样品之间的交叉污染的可能性,并获得足够长的时间使化学反应完全,从而确保灵敏度和精密度达到检测要求。
2 连续流动分析在生活饮用水检测中的优势以生活饮用水中挥发酚和氰化物的测定为例,传统手工检测方法主要包括样品蒸馏、调节pH、显色反应等步骤,样品前期处理过程繁琐、处理工作量大,样品检测消耗时间较长。
连续流动分析仪连续流动分析仪(Continuos Flow Analyzer,CFA)是一种先进的化学分析仪器,可自动、高效地进行大规模样品的化学分析。
它利用连续流动的技术,通过一系列连续的化学反应和检测步骤,实现对不同成分的快速测定。
本文将介绍连续流动分析仪的原理、应用和优势。
连续流动分析仪的工作原理基于恒定流动,通过将连续流动的载流液中的样品连续引入分析系统,然后与一系列预先加入的试剂进行反应,最终通过某种检测方法进行信号转换和测量。
其核心部分是连续流动分析器头,由工作单元、泵浦单元、计量仪单元和检测仪单元组成。
在分析过程中,连续流动分析仪可实现高度自动化的操作,大大提高了分析效率和准确性。
连续流动分析仪广泛应用于水质分析、环境监测、农药残留检测、生物化学分析等领域。
在水质分析中,连续流动分析仪可用于测定水中的氨氮、硝态氮、亚硝态氮等成分;在环境监测中,连续流动分析仪可用于测定大气颗粒物、大气污染物等;在农药残留检测中,连续流动分析仪可用于测定农产品中的农药残留量;在生物化学分析中,连续流动分析仪可用于测定血清中的葡萄糖、脂质、蛋白质等。
与传统手工分析方法相比,连续流动分析仪具有多项优势。
首先,连续流动分析仪具有较高的自动化程度,大大降低了操作人员的工作强度,提高了分析效率。
其次,连续流动分析仪具有较高的精确度和准确度,可实现对样品进行连续、快速和精确的分析。
此外,连续流动分析仪还具有较小的样品量需求,有效节约了化学试剂和样品的使用成本。
最重要的是,连续流动分析仪可以充分利用化学反应的连续性和稳定性,提高了分析结果的可靠性和可重复性。
尽管连续流动分析仪具有许多优势,但也存在一些局限性。
首先,连续流动分析仪对于样品的前处理要求较高,需要对样品进行预处理、前处理和样品制备等步骤,以保证结果的准确性和可靠性。
其次,连续流动分析仪的应用范围相对有限,主要适用于水质分析、环境监测和生物化学分析等领域,对于复杂样品和矩阵干扰较大的样品需要结合其他分析方法进行分析。
流动注射-4-氨基安替比林法测定水中挥发酚摘要:连续流动分析仪测定水中挥发酚操作步骤简单快捷,所需试剂种类少,试剂、样品消耗量小,检测速度快,自动化程度高,校准曲线线性好,灵敏度高、精密度准确度好,安全可靠,适用于大批量样品分析。
关键词:流动注射分析仪;挥发酚1 引言根据酚类能否与水蒸气一起蒸出,分为挥发酚和不挥发酚。
挥发酚通常是指沸点在230℃以下的酚类,通常属于一元酚。
酚类为原生质毒,属高毒物质。
人体摄入一定量时,可出现急性中毒症状;长期饮用被酚污染的水,可引起头昏、出诊、瘙痒、贫血及各种神经系统症状。
水中含低浓度(0.1~0.2mg/L)酚类时,可使生长鱼的鱼肉又异味,高浓度(>5mg/L)时则造成中毒死亡。
含酚浓度高的废水不宜用于农田灌溉,否则,会使农作物枯竭或减产。
水中含微量酚类,再加氯消毒时,可产生特异的氯溴酚。
酚类主要来自炼油、煤气洗涤、炼焦、造纸、合成氨、木材防腐和化工等废水。
酚类的分析方法较多,各国普遍采用4-氨基安替比林分光光度法。
连续流动分析技术是建立在平衡体系的基础上的一种较新的分析技术,它通过蠕动泵吸入反应的试样及试剂进管道,在管道中把连续流动的试样用空气按一定的间隔有规律的分隔开,再按反应顺序和一定比例混入试剂,通过反应模板的管道及混合线圈完成反应过程,消解后进入检测器检测。
本实验的反应流程和反应基体较为复杂,分析有前处理的挥发酚分析中,采用了CFA技术,该技术具有灵敏度高、准确度高、精密度好和回收率好,节省试样、试剂等特点,并提高了操作的安全性、方便性和工作效率,是替代手工操作的一种理想分析方法。
2 方法原理分析方法:4-氨基安替比林分光光度法在铁氰化钾存在的情况下,1羟基酚与4-氨基安替比林偶合,生成红色染料,使用光谱仪测量505nm波长的吸光度。
化合酚经过水解和蒸馏(酸性条件下)后测量,对位酚不参与化学反应或部分参与反应,并不能完全被测量,二羟基和多羟基酚不能用此方法蒸馏。
AA3型连续流动分析仪(CFA)同时测定橡胶叶全氮、全磷、全钾的方法研究贝美容;罗雪华;杨红竹【摘要】通过对AA3型连续流动分析仪所附试剂配制等方法的改进,研究AA3同时测定橡胶叶片氮、磷、钾的方法.结果表明,氮、磷、钾混合标准曲线相关系数均达0.9999以上;氮、磷变异系数均小于0.5%(n=10),当钾浓度相当于标准曲线中间浓度范围时(即6~10 mg/L),变异系数为0.97%~1.13%,重现性良好;氮、磷、钾加标回收率分别为99.7%~101.1%、99.3%~100.7%、98.4%~102.8%,待测液中氮、磷、钾检出限分别为0.019、0.004、0.072 mg/L.橡胶叶片氮、磷、钾测定结果与常规法相比无显著差异.该方法用于橡胶叶片H2SO4-H2O2消煮后氮、磷、钾同时测定,快速高效,一天半可测300个样品待测液,提高工效10多倍.%Total nitrogen (N), phosphorus (P) and potassium (K) in rubber leaf samples by AA3 Continuous Flow Analyzer (CFA) were simultaneously assayed by improving the reagents preparation methods gave by CFA. The results showed that the linear correlation coefficients of N, P and K mixed standard curve reached 0.999 9 above measured by AA3 simultaneously; when the variation coefficient was less than 0.5% (n=10) for both nitrogen and phosphorus and the potassium concentration was equivalent to the intermediate concentration range of the standard curve (6-10 mg/L) with a variation coefficient of 0.97-1.13%, the data repeated well; the recovery of nitrogen, phosphorus, potassium was 99.72%~101.1%, 99.3%~100.7%, 98.4%~102.8%, repectively; and the detection limit of them was 0.019mg/L, 0.004 mg/L and 0.072 mg/L, respectively; the content of nitrogen,phosphorus and potassium in rubber leaf samples were not different significantly by CFA and standard methods. Used for the nitrogen, phosphorus, potassium simultaneous determination of the rubber leaves digested by H2SO4-H2O2, 300 samples could be measured in one and a half days, and the efficiency could be improved by more than a dozen times.【期刊名称】《热带作物学报》【年(卷),期】2011(032)007【总页数】7页(P1258-1264)【关键词】连续流动分析仪;橡胶叶片;氮磷钾同时测定【作者】贝美容;罗雪华;杨红竹【作者单位】中国热带农业科学院橡胶研究所,海南儋州 571737;农业部橡胶树生物学重点开放实验室,海南儋州 571737;海南省热带作物栽培生理学重点实验室,海南儋州571737【正文语种】中文【中图分类】S794.1随着橡胶树营养诊断指导施肥技术在生产上的推广应用[1],橡胶叶片养分,特别是氮(N)、磷(P)、钾(K)三大元素含量的分析工作量逐渐增加,而常规分析方法中用比色法测定氮磷时,必须分别吸取待测液,分别显色,分别测定,且显色液必须放置1~3 h,等反应完全后才能测定,显色时间长,如钼蓝比色法测定磷需显色1 h,钒钼黄比色法测定磷需显色30 min;纳氏比色法测定氮需显色2 h,靛酚蓝比色法测定氮需显色3 h;氮的蒸馏法虽然是经典法但效率低,蒸馏1个样约需5 min,同样不适宜批量化分析;钾的测定还得另外进行[2-3]。
连续流动分析仪招标项目要求⒈设备名称、数量及用途设备名称:连续流动分析仪数量:1套,包含总氰/氰化物,挥发酚,总氮,总磷,硫化物,阴离子洗涤剂分析通道用途:用于各种水样中总氰/氰化物,挥发酚,总氮,总磷,硫化物,阴离子洗涤剂的全自动分析。
2.主要配置2.1带显示屏和操作面的FUTURA/E主机6套2.2※完全独立的蠕动泵5个2.3完全独立的双光束双波长检测器6个2.4主机电源2个2.5※360位三维取样器1套2.6总氰/氰化物分析模块(含在线外置独立紫外消化、高温蒸馏系统)1套2.7挥发酚独立分析模块(含在线外置独立高温蒸馏系统)1套2.8总氮独立分析模块(含在线自动消化)1套2.9总磷独立分析模块(含在线自动消化)1套2.10硫化物独立分析模块(含在线自动氮吹系统)1套2.11阴离子洗涤剂独立分析模块(含在线萃取系统)1套2.12供仪器正常使用一年耗材(见下列清单)1批2.13CFM V2工作站(含软件)1套3.技术参数及要求:3.1分析主机:3.1.1※配置六个独立主机,含六个独立及以上分析通道,六个通道可同时进行分析,也可单独分析,每个通道主机均含有一个独立的蠕动泵、一个独立的检测器、一个独立的液晶显示屏和操作键盘、一个独立的试剂柜,具备自动开机和自动关机功能,蠕动泵具有自动升降功能;3.1.2※气泡注入方式:采用高精度时蠕动泵直接平稳地将气泡泵入化学反应模块中,不需要借助其他辅助设备。
所有的分析方法必须采用气泡间隔连续流动分析方法。
3.1.3※数量:六套完全独立主机,每套主机含1个及其以上分析通道。
3.2三维自动进样器:3.2.1三维随机自动进样器,样品杯数量:≥360位,采用移动式清洗池,死体积小于10ul;3.2.2样品杯容量:具有5ml,10ml等规格样品杯;3.2.3一次进样,同时检测出分析结果;3.2.4数量:一个。
3.3蠕动泵:3.3.1高精度比例蠕动泵,12根不锈钢泵滚轴,带检漏装置,可自动排出漏液、报警并自动停止运行主机;3.3.2※每个蠕动泵泵管位数≥13位,泵管位总数量≥65通道;可控制泵自动升降,自动压紧与放松,每个泵含有≥5通道自动清洗阀,可由马达自动进行试剂/清洗的切换,分析结束后可编程自动进行对化学模块的清洗,自动松开泵盖;3.3.3※数量:各6个。
解析连续流动分析仪独特之处分析仪如何操作CFA—系列全自动连续流动分析仪是艾塔科仪自主研发一款流动分析仪,广泛应用于水质监测、食品分析、酒类分析和土壤溶液等领域,其检测分析的结果精准明确牢靠,符合国际认证标准。
为浩繁科研分析者供应了样品分析新的选择。
CFA—系列流动分析仪接受先进的连续流动化学分析技术,并拥有多项zhuan li及技术创新:自有zhuan li的气泡注入和气泡分别技术可实现样品的完全反应,避开样品间的干扰;脱蒸馏技术、长流通池装置、公用光源流通池装置、膜过滤装置以及光学分光脱气装置替代传统的手工操作的化学分析,便利快捷,能极大提高监测的效率和测量的精准性,在保护环境、掌控水污染的工作具有紧要作用。
CFA—流动分析仪自有zhuan li的在线吹脱蒸馏技术能快速完成样品的蒸馏收集工作,在线待检样品处理工作,支持多通道同时检测而其它通道可以不受影响地连续操作,提高了工作效率,多通道全谱直读CCD检测器,无需滤波片选择波长也使得灵敏度高,稳定性好。
多路独立膜过滤技术和蒸馏技术充分各项检测模块任意搭配使用,多路独立加热控温模块也充分各检测项目单独控温需求;全自动中文数据处理工作站可实时监控每个分析过程并能输入新的任务恳求,有利于分析操作的精准性。
激光粒度分析仪的特点和应用都有哪些呢随着激光和微电子技术应用到粒度测量领域,粒度分析在大大减轻劳动强度的同时,加快了样品的检测速度,提高了检测结果的质量。
激光粒度分析方法,因测量速度快、精度高及精准度好等特点被人们普遍认同。
激光粒度分析仪顾名思义是利用激光具有的单色性和极强的方向性等特性对颗粒进行检测的仪器,是全球范围内公认的先进,较快捷的颗粒测试仪器。
激光粒度分析方法是近年来进展较快的一种测试方法,其紧要特点是:1)测量的粒径范围广,可进行从纳米到微米量级如此宽范围的粒度分布。
约为:20nm ~2000μm , 某些情况下上限可达3500μm;2)适用范围广泛,不仅能测量固体颗粒 ,还能测量液体中的粒子;3)重现性好,与传统方法相比,激光粒度分析仪能给出精准牢靠的测量结果;4)测量时间快,整个测量过程1—2分钟即可,某些仪器已实现了实时检测和实时显示,可以让用户在整个测量过程中察看并监视样品。
CFA-1100多通道连续流动分析仪技术要求
一、应用要求:
本仪器可以用来分析地下水、地表水和工业污水等连续大批量分析样品,全部采用计算机自动化操作,测量项目包括:挥发酚、氰化物、阴离子合成洗涤剂。
*工作原理:多通道连续流动分析仪是利用连续流,从各种试剂容器吸出试样溶液,在流路系统中将试样与试剂混合,用蠕动泵将空气、试样和试剂分别吸引到已确定的流路中,最终到达检测器完成检测分析过程。
整个过程完全自动化。
二、工作环境:
2.1 环境温度:10-30℃.
2.2 环境湿度:< 80% (无冷凝)
2.3 电源:220V,50 Hz
三、技术参数要求:
3.1、仪器配置:
3.1.1、自动进样器 1套
3.1.2、化学反应模块 3套
3.1.3、多通道蠕动泵 1套
3.1.4、检测器 2通道
3.1.5、数据处理系统 1套
3.2、各系统技术参数:
3.2.1、自动进样器
3.2.1.1、工作模式: XYZ三维机械臂式采样器,步进电机走位。
*3.2.1.2、样品位数: 10个标准位,180个样品位。
3.2.1.3、样品架: 2个样品架。
3.2.1.4、采样时间:多种时间可设定。
3.2.1.5、样品容量:标准样品容量为50ml.测量样品容量为9ml或以上。
3.2.1.6、进样针:双针模式,采用碳纤维材料制作。
3.2.1.7、数据接口:可选用多种数据接口,COM端口,USB接口,22针串行控制等。
3.2.2、化学反应模块
3.2.2.1、化学反应通道数:配置为氰化物、挥发酚、阴离子合成洗涤剂3通道。
3.2.2.2、加热单元: 3个加热单元。
3.2.2.3、温度设定:调节范围从室温到250度。
3.2.2.4、蒸馏单元: 2个蒸馏及馏分收集单元,快速吹脱蒸馏,蒸馏冷凝过程控制在3秒钟内。
3.2.2.5、紫外消化:采用紫外线石英管,便于紫外线透过,达到最佳的消化效果。
*3.2.2.6、混合圈:采用玻璃混合圈,内径为2mm,具有非常好的通过性和化学惰性。
*3.2.2.7、气泡注入:4路气泡注入装置,气泡的大小和速度可调节,以达到管路平衡及最佳的气泡间隔。
3.2.3、多通道蠕动泵
3.2.3.1、通道数: 24通道蠕动泵。
3.2.3.2、转速: 0-100转,任意调节,分辨率为0.1转。
3.2.3.3、流量范围:根据不同泵管流量范围从0.19ml到200ml每分钟。
3.2.3.4、显示:大屏幕液晶显示流量,转速,方向等参数。
3.2.3.5、适用泵管: PVC, Viton,硅胶管,等等。
3.2.3.6、通信接口: 485 信号接口。
3.2.4、检测器
*3.2.4.1、类型: CCD 数字检测器。
3.2.
4.2、通道数: 2通道同时检测。
3.2.
4.3、波长范围: 380-800nm全范围。
3.2.
4.4、带宽: 2nm。
3.2.
4.5、积分时间: 0-1000毫秒任意设定。
3.2.
4.6、积分次数: 0-100次,任意取值。
*3.2.4.7、流通池: 标配包括 1cm、5cm、20cm三种规格,其他多种规格可选用。
3.2.
4.8、光源: LED灯。
3.2.5、数据处理系统
3.2.5.1、操作界面:全中文操作界面,windows 7/XP 操作系统。
3.2.5.2、通道数:双通道数据处理工作站。
3.2.5.3、参数设置:积分时间及次数,波峰间距及范围和采样次数均可设置。
3.2.5.4、控制功能:可自主设置工作流程,实现自动进样器控制,检测器数据采集,分析,图形处理。
3.2.5.5、报表:可自动生成PDF工作报表,可导入到Excel进行计算分析.自动生成校准曲线,自动检测样品峰值。
3.3、测试项目技术要求:
3.3.1、氰化物
3.3.1.1、分析方法:异烟酸-巴比妥酸分光光度法
3.3.1.2、特殊要求:在线蒸馏
3.3.1.3、检测极限:0.0007mg/L
3.3.1.4、测量范围:0.0007 ~0.1mg/L
3.3.1.5、样品分析频率:20样/小时
3.3.2、挥发酚:
3.3.2.1、分析方法:4-氨基安替比林分光光度法
3.3.2.2、特殊要求:在线蒸馏
3.3.2.3、检出极限:0.001mg/L
3.3.2.4、测量范围:0.001 ~0.1mg/L
3.3.2.5、样品分析频率:20样/小时
3.3.3、阴离子合成洗涤剂
3.3.3.1、分析方法:亚甲基蓝分光光度法
3.3.3.2、特殊要求:一次膜分离,一次重力分离
3.3.3.3、检出极限:0.01mg/L
3.3.3.4、测量范围:0.01 ~5mg/L
3.3.3.5、样品分析频率:20样/小时
3.4、系统控制应用软件和相关资料
3.4.1、产品说明书及操作软件
3.4.2、品牌计算机和网络打印机,正版windows 7/XP 软件,19寸液晶显示屏:激光打印机
四、配置要求:
4.1、主机:氰化物模块1套,挥发酚模块1套,阴离子合成洗涤剂模块1套
4.2、180位双针自动进样器1套
4.3、标样管/盖子10个
4.4、样品管(玻璃)200个
4.5、样品架2个
4.6、24通道蠕动泵1套
4.7、启动工具包1个
4.8、品牌台式电脑1台
4.9、品牌打印机1台
4.10、说明书1本
4.11、方法文件1本
五、技术服务要求:
5.1、自验收之日起,对整机向用户提供一年保修,终身维护。
对操作软件终生免费升级。
5.2、货物到达后,在收到买方要求安装3天内,卖方派技术人员免费安装调试及培训。
5.3、仪器在安装校准调试过程中我方不少于2位工作人员进行现场培训,同时对2-3名工作人员免费进行厂家集中培训。
5.4、供应商负责提供技术支持,保证仪器的正常工作。
5.5、提供一年耗材,含在投标报价中。