IDEAL-2000臭氧在线检测仪英文说明书
- 格式:doc
- 大小:285.50 KB
- 文档页数:16
IDEAL-2000
Ozone concentration detector
Manual
United States-owned IDEAL Computer(ZIBO) CO,.LTD.
Dear users:
Thank you very much for using our products.
Before using this product, please read these instructions carefully.
The products described in this manual only for trained qualified workers. This product's adjustment, repair and maintenance must be authorized officers specified by the manufacturer.
For use this product correctly and safety, you need to operate in accordance with this manual.
The manufacturer recommendations:
Maintain the integrity of the product label;
Do not disassemble this product;
All maintenance and calibration of this product needs to operate in accordance with the requirements of the Instruction Manual.
The manufacturer has no responsibility to any damage and injure caused by improper use of this product.
The manufacturer can improve the product specifications or modify the contents of the manual without notice.
If you find any errors in this manual, please contact us, we are very grateful.
DO NOT copy or imitation of all or part of this manual without The
m anufacturer’s Permission!
Directory
1.Outline
2.Technical Parameters
3.Structure
4.Function Introduction
munication protocol
6. Maintenance
7.Notes
8.Store
9.date of validity
1. Outline
IDEAL-2000-type Ozone concentration detector mainly use for the export’s ozone concentration detection of ozone generator,UV lamp as light source system,Life is 30,000 hours or more,Stable measurements, High precision;Use Integration Layer Stream light pool structure,No leakage、absorb high pressure、absorb high flow sample gas’s impact、easy to clean、easy to maintenance、easy to operate. The product will correction zero automatically during using, ensure accuracy of concentration and avoid zero drift,have 4-20mA and RS-485 output signal。IDEAL-2000-type Ozone concentration detector is application of the photoelectric conversion detection technology, use the microprocessor control technology,long life pulse mode light source,use the characteristics of UV absorption of ozone,accordance with The Beer-Lambert Law.
2. Technical Parameters and Installation 1.Technical Parameters:
Measuring range:0~200mg/L;
Display resolution:0.1mg/L;
Display digits:4;
margin of error:±3%;
Communication method:RS-485;Communication Parameters:9600,8,N,1;Output method:differential 4-20mA,Linear;Weight:1.5kg;
Input gas flow:1.0±0.3L/min;
Input gas Pressure:MAX 1bar(0.1MPa);
Power supply:AC 220V±10% 50Hz;
Size:160mm(W)×80mm(H)×250mm(L); Gas input port、output port ID 4mm,OD 6mm;Embedded instrumentation size:152mm×76mm;Cycle of back to manufacturer to check:1.5 years.