AU480标准操作文件
- 格式:doc
- 大小:631.50 KB
- 文档页数:16
1212.1 主菜单...................................................................................12-2 12.2 固定显示的按钮配置和功能...................................................12-3功能选择按钮组..................................................................12-3系统控制按钮组...........................................................................12-3 报警器相关按钮组.......................................................................12-3 其他按钮组..................................................................................12-312.3 菜单按钮浏览.........................................................................12-4 12.4 常规菜单................................................................................12-5 12.5 定标菜单................................................................................12-6 12.6 质控菜单................................................................................12-7 12.7 参数菜单................................................................................12-8 12.8 维护菜单................................................................................12-9 12.9系统菜单................................................................................12-10目录树本章节阐述了适用于AU480选项的参考。
AU480生化分析仪简易操作规程1使用操作1.1开机前准备1.1.1检查纯水桶中液量是否充足,出水阀门应处于打开状态。
1.1.2检查B清洗液(洗涤原液)桶中清洗液量是否充足,不足应加满。
1.2开机1.2.1若设备处在在待机状态,按机身右侧的绿色(ON)键。
若设备处在总关机状态,先按机身右侧的白色(RESET)键,在听到冰箱启动的声音后(约10秒)再按绿色键。
1.2.2启动后计算机屏幕显示登陆对话框,输入用户名,单击“OK"登录。
1.2.3登陆后屏幕显示INDEX对话框,单击“OK”确认。
程序进入“Start Condition”菜单,仪器自动进入“WARM UP"状态.“WARM UP”状态下可进行项目编辑、试剂样本准备等操作。
“WARM UP”状态结束后,系统转入“STANDBY”状态,可随时开始分析测定。
1.3项目编辑1.3.1建立项目1.3.1.1在主页上点击“Menu List"→“Parameters"→“Common TestParameters”,在Test Name页面按“Edit",输入项目名称,按“Confirm”确定。
1.3.1.2在Group of Test页面按“Edit",点击要做的项目名称(显示蓝色表示选中),按“Confirm”确定。
1.3.2设置实验参数1.3.2.1在主页上点击“Menu List”→“Parameters”→“Sepcific Test Parameters”,在General页面按“Edit”,编辑输入样本量(Sample Volume)、试剂量(Reagent Volume)、波长(Wave length)、方法学(Method)、反应方向(Reaction Slope)和读数点(Measuring Point),按“Confirm”确定。
注意参数限制:R1试剂量(10~250μl),R2试剂量(0,10~250μl),样本量(1~25μl),读点First(0~26),读点Last(1~27)。
新项目的设定1.输入试验名称,在线注册2.输入参数3.定标参数4.QC参数5.试剂位置一、输入名称和在线注册1.Parameters Common Test Parameters Test Name Edit,在空白的地方输入项目名称,如在NO.60输入Name为“UA1”,Reagent ID “123”点击Confirm确认。
2.Parameters Common Test Parameters Group of Tests Edit Test Item Setting,点击“UA1”,使其变为蓝色,Confirm。
二、输入参数Parameters Specific Test Parameters General EditTest Name UA1 ;Sample Volume 6ul ;Dilution 0 ;Pre-Dilution Rate 1 ; Reagent Volume R1(R1-1) 200Operation Yes ; Lag Time Check YesWave Length Pri. 520 nm ;Sec. 600 nmMethod END ; Reaction Slope +Measuring Point-1 First 0 Last 27Min.OD -2.0 ; Max.OD 2.5First Low -2.0 ; High 2.5Last Low -2.0 ; High 2.5Dynamic Range Low 0 ; High 1200Correlation Factor A 1 ; B 0Factor for Maker A 1 ; B 0Parameters Specific Test Parameters RangeRange Edit Set Decimal Places Decimal Places 0(小数位数)Unit 单位umol/LConfirm三、定标参数Parameters Calibration Parameters CalibratorsNO.Cup NO.1 Name ID11(位置) Randox 1Parameters Calibration Specific General Edit Test Name”UA1”, Calibration Type “AB” ; Formula Y=AX=B ;Counts 2Calibrator ConcPoint-1 1.Randox (名称) 369(浓度)标准品的位置、名称及浓度要看具体的标准。
简明操作手册(AU全自动生化仪培训资料Ver 1.0)适用机型:贝克曼AU480贝克曼AU680每天工作流程:一、开机前检查;二、开机;三、修改日期索引;四、检查、添加试剂;五、做质控,查看质控结果;六、如有必要,做定标;七、做定标项目的质控,查看结果;八、做标本;九、插入急诊标本;十、查看标本结果;十一、传输标本结果;十二、关机;十三、常见符号;十四、常见报警。
仪器主要单元介绍仪器主要界面介绍主要按键介绍一、开机前检查1、检查样品配送器、试剂配送器是否泄漏;2、检查清洁剂蠕动泵装置是否泄漏;3、检查主清洁剂的数量并补给;4、检查并清洗样品探针、试剂探针和搅拌棒;二、开机开启ON键,仪器主机以及操作电脑将会自动启动,操作电脑进入“修改日期索引”界面。
三、修改日期索引1、开机后,仪器自动进入以下界面2、点击OK,仪器将自动建立一个日期索引。
注:每天必须修改日期索引,并且每天只能修改一次,如果当天第二次开机,则选择“Current Index”,然后点击OK,仪器将不建立新索引。
四、检查、添加试剂1、点击“主菜2、点击3、点击“Reagent”点击“Reagent”后,进入以下界面点击点击“Details”后,进入以下界面1、在此查看各试剂剩余测试数,如果哪个试剂少,请添2、点击“Reagent Check”。
点击“Reagent Check”,进入以下界面1、如果添加的试剂品种很多,就使用第一个选择;2、如果添加试剂品种只有几个,使用第二个;3、第三个选择几乎不用;4、试剂状态显示红色的时候,使用“试剂复位”。
选择好后,点击“Start”,仪器将对试剂进行检查。
五、做质控,查看质控结果1、做质控:将质控品放在“绿色”样品架相应的位置,将“绿色”样品架放置于仪器进样轨道上,点仪器“开始”键。
仪器自动做设置好的项目的质控。
2、查看质控结果:在中文报告系统,点击“质控”“质控图”,即可查看质控结果。
六、定标1、点击“主菜2、点击3、点击“Rack点击“Rack Requisition”后,进入以下界面点击点击“Calibrarion”后,进入以下界面2、选择需要定标的3、点击“Entry”4、点击“Display CupSet”以查看定标液放置的1、点击“StartEntry”将定标液放置于“黄色”定标架指定位置,在“蓝色”试剂空白架的“1号”位置放水。
AU480生化分析仪简易操作规程1使用操作1.1开机前准备1.1.1检查纯水桶中液量是否充足,出水阀门应处于打开状态。
1.1.2检查B清洗液(洗涤原液)桶中清洗液量是否充足,不足应加满。
1.2开机1.2.1若设备处在在待机状态,按机身右侧的绿色(ON)键。
若设备处在总关机状态,先按机身右侧的白色(RESET)键,在听到冰箱启动的声音后(约10秒)再按绿色键。
1.2.2启动后计算机屏幕显示登陆对话框,输入用户名,单击“OK”登录。
1.2.3登陆后屏幕显示INDEX对话框,单击“OK”确认。
程序进入“StartCondition”菜单,仪器自动进入“WARM UP”状态。
“WARM UP”状态下可进行项目编辑、试剂样本准备等操作。
“WARM UP”状态结束后,系统转入“STANDBY”状态,可随时开始分析测定。
1.3项目编辑1.3.1建立项目1.3.1.1在主页上点击“Menu List”→“Parameters”→“CommonTest Parameters”,在Test Name页面按“Edit”,输入项目名称,按“Confirm”确定。
1.3.1.2在Group of Test页面按“Edit”,点击要做的项目名称(显示蓝色表示选中),按“Confirm”确定。
1.3.2设置实验参数1.3.2.1在主页上点击“Menu List”→“Parameters”→“Sepcific Test Parameters”,在General页面按“Edit”,编辑输入样本量(Sample Volume)、试剂量(Reagent Volume)、波长(Wavelength)、方法学(Method)、反应方向(Reaction Slope)和读数点(Measuring Point),按“Confirm”确定。
1.3.2.2注意参数限制:R1试剂量(10~250μl),R2试剂量(0,10~250μl),样本量(1~25μl),读点First(0~26),读点Last(1~27)。
作业指导书标题:Beckman Coulter AU480 全自动生化分析仪标准操作规程修改记录目录一、目的二、仪器的使用要求三、检测系统概述3.1 系统组件3.1.1 仪器3.1.2 试剂3.2 样本检测过程概述四、仪器的操作4.1仪器主要操作界面4.2标本传送装置4.3开机操作规程4.4试剂装载程序4.5校准操作程序4.6质量控制操作程序4.7常规标本测定程序4.8重复测定操作程序五、关机与维护保养操作程序5.1.关机程序5.2.仪器校准程序5.3.维护保养程序六、高级操作程序6.1项目参数的设定程序6.2删除实验项目七、相关文件一、目的为了正确使用检测设备,维护和保持检测设备的技术性能,以Beckman Coulter AU480全自动生化分析仪用户手册为依据编写了此规程,供指导日常常规操作使用。
二、仪器的使用要求仪器的安全使用要求,下面列出了警告标签。
三、检测系统概述3.1 系统组件3.1.1 仪器外部结构见下图(图3-1)图3-1 仪器外部结构图3.1.2 试剂AU480能使用的有60ml、30ml和15ml规格的试剂瓶3.2 样本检测过程概述一项实验请求就是一条对一份样品进行特定实验的指令。
当一个样品被放入本设备时,实验需求的信息就已经与样本所需要的实验建立了联系。
样本通过进样模块进入到仪器,通过轨道被运输至采样位置。
试剂探针吸取R1试剂,加入反应杯。
混匀器混匀R1后,样本探针随即吸取样本加入到相应反应杯中,混匀器混匀样本和R1并且完成孵育后,试剂探针吸取R2试剂,加入反应杯中。
仪器从添加R1开始,反应期间每间隔18秒进行一次测量,这期间测定的值和特定实验参数定义的波长可以计算出被测物浓度。
四、仪器的操作4.1仪器主要操作界面A.主界面(图4-1)图4-1 仪器主界面主界面主要由3部分组成A:主控制区B:菜单区C:警告区B. 菜单界面(图4-2)图4-2 菜单进入菜单键后,左边是主菜单,主要由常规(Routine)、定标(Calibration)、质控(QC)、参数(parameters)、维护保养(Maintenance)、系统(System)组成;右边是相应的子菜单,点击可进入相应的操作界面4.2标本传送装置试管架:AU480全自动生化分析仪采用“排列式”的进样方式,传输轨道将装载有样品管(杯)的样品架传输到加样位置进行加样操作。
AU480 用户使用手册eXXXXXXXX, 2008 分析前准备分析前的准备55.1 启动系统.........................................................................................5-25.1.1 启动系统............................................................................................5-25.1.2 登陆到系统........................................................................................5-35.1.3 更新索引............................................................................................5-45.2 使用系统前的准备工作...................................................................5-65.2.1 确认分析仪和试剂状态......................................................................5-65.2.2 检查打印机(选配)和纸张...............................................................5-75.2.3 运行日常保养.....................................................................................5-85.2.4 电解质分析仪(选配)准备……........................................................5-95.3试剂的准备……………...................................................................5-15 5.4 准备分析样品.................................................................................5-195.4.1给样品架贴条形码标签.....................................................................5-19 5.4.2样品条形码规格................................................................................5-20 5.4.3准备样品杯.......................................................................................5-23 5.4.4给样品杯粘贴条形码.........................................................................5-25 5.4.5准备样品...........................................................................................5-26 5.4.6将样品杯/比色杯放入样品架............................................................5-27 5.4.7 把样品架放到进架器上….................................................................5-325.5 定标测试…….....................................................................................5-345.6 执行质量控制(QC )........................................................................5-385.7 样品申请:输入数据及选择测试……………......................................5-425.7.1 组设置..............................................................................................5-435.7.2 手动输入申请....................................................................................5-425.7.3 输入批申请........................................................................................5-445.7.4 从主机下载申请…….........................................................................5-455.7.5 输入急诊样品…….............................................................................5-46本章会详细介绍使用系统前的准备工作:5.1 启动系统本节主要介绍初次使用AU480内容:5.1.1 启动系统............................................................................................5-25.1.2 登陆到系统........................................................................................5-35.1.3 更新索引............................................................................................5-4 5.1.1 启动系统系统有两个电源开关:●主电源开关,这是整个系统电流的总开关。
简明操作手册(AU全自动生化仪培训资料Ver 1.0)适用机型:贝克曼AU480贝克曼AU680每天工作流程:一、开机前检查;二、开机;三、修改日期索引;四、检查、添加试剂;五、做质控,查看质控结果;六、如有必要,做定标;七、做定标项目的质控,查看结果;八、做标本;九、插入急诊标本;十、查看标本结果;十一、传输标本结果;十二、关机;十三、常见符号;十四、常见报警。
主要按键介绍一、开机前检查1、检查样品配送器、试剂配送器是否泄漏;2、检查清洁剂蠕动泵装置是否泄漏;3、检查主清洁剂的数量并补给;4、检查并清洗样品探针、试剂探针和搅拌棒;二、开机开启ON键,仪器主机以及操作电脑将会自动启动,操作电脑进入“修改日期索引”界面。
三、修改日期索引1、开机后,仪器自动进入以下界面2、点击OK,仪器将自动建立一个日期索引。
注:每天必须修改日期索引,并且每天只能修改一次,如果当天第二次开机,则选择“Current Index”,然后点击OK,仪器将不建立新索引。
四、检查、添加试剂1、点击“主菜单”2、点击“Roution”3、点击“Reagent”点击“Reagent”后,进入以下界面点击“Details”点击“Details”后,进入以下界面点击“Reagent Check”,进入以下界面1、在此查看各试剂剩余测试数,如果哪个试剂少,请添加。
2、点击“Reagent Check”。
1、如果添加的试剂品种很多,就使用第一个选择;2、如果添加试剂品种只有几个,使用第二个;3、第三个选择几乎不用;4、试剂状态显示红色的时候,使用“试剂复位”。
选择好后,点击“Start”,仪器将对试剂进行检查。
五、做质控,查看质控结果1、做质控:将质控品放在“绿色”样品架相应的位置,将“绿色”样品架放置于仪器进样轨道上,点仪器“开始”键。
仪器自动做设置好的项目的质控。
2、查看质控结果:在中文报告系统,点击“质控”“质控图”,即可查看质控结果。
AU480用户使用手册66.1 启动分析.......................................................................................6-2 6.2 分析监测...................................................................................................6-3 6.3 终止一项测试...........................................................................................6-9 6.4 结果检查..................................................................................................6-116.4.1 检查测试结果....................................................................................6-11 6.4.2 显示反应监测...................................................................................6-11 6.4.3 检查定标和试剂空白.........................................................................6-15 6.4.4 错误标记和警报检查.........................................................................6-22 6.4.5质量控制检查....................................................................................6-236.5 急诊样品处理..............................................................................6-296.5.1 执行急诊台分析……........................................................................6-31 6.5.2 简易急诊模式…...............................................................................6-36 6.5.3 红色样品架的使用...........................................................................6-376.6 打印结果.....................................................................................6-39 6.7 重测............................................................................................6-426.7.1 准备运行手动重测…........................................................................6-43 6.7.2 在急诊台上执行重测........................................................................6-46 6.7.3 使用橙色样品架执行重测…….........................................................6-46 6.7.4 检查重测数据……...........................................................................6-476.8 传输数据到其他.计算机................................................... ...........6-49 6.9 暂停分析操作................................................................... ..........6-50 6.10 终止加入样品架................................................................ ..........6-52 6.11 系统关闭......................................................................... ............6-53 6.12 运行紧急终止……………………………………………….....…....6-54 6.13 编辑质量控制数据........................................................ ..............6-55 6.14 编辑分析数据............................................................... ..............6-586.14.1 患者样品数据的覆写......................................................................6-59 6.14.2 修正患者样品数据..........................................................................6-61 6.14.3 使用一条变化的定标曲线来重新计算分析数据 ………….………...6-63 6.14.4 联机传输编辑好的数据...................................................................6-646.15 确认电解质分析仪(选配)状态......................................... ........6-65 6.16 试剂情况管理…………………………………………………. ........6-68执行分析本章描述了执行不同原理分析的方法6.16.1 检查试剂状况……………………………………………………….….6-686.16.2 试剂状况编辑管理……………………………………………………..6-706.16.3 设定日常分析所需试剂量……………………………………………..6-76 AU480用户使用手册你必须确保本系统做好了进行分析的准备.在分析开始启动以后,从分配好样品到计算出第一个结果大约需要8 分40 秒。
四.仪器的操作Beckman AU480是一台临床生化分析仪。
此仪器可以测定人血清、血浆、尿或其它体液的临床生化和一些免疫学项目。
4.1开机前准备工作a.检查水源AU480要求使用去离子水,其最大耗水量(全部使用浓缩双试剂)20L/小时。
开机前应检查去离子水贮备量。
可用下面公式估算:贮水量(L)=预计测试总量/480×42×0.7或装备产水速度大于20L/小时的去离子水制水设备。
打开贮水容器的出水阀门。
b.检查电源AU480建议配备容量3.5KV A以上的断电不间断电源(UPS)。
检查UPS电源应该处在断电保护(Inverter)及AC灯亮的工作状态。
c.检查清洗原液量检查( A )、B清洗液桶内液面,如不足请加满。
d.检查样品、试剂分配器及样品针、试剂针、搅拌棒是否正常。
e.检查样品架收集槽内有无样品架如有样品架请移空。
f.放试剂盘请在开机前将放有试剂瓶的试剂盘放入相应的试剂仓中。
如果开机后再放试剂盘,将无法使用选择性查试剂方式,仅可使用ALL(全查)。
4.2开机AU480有三种开机方式:总关机后开机、待机状态下开机和自动开机。
a. 总关机后开机总关机是指按“EM STOP”键关机后再开机。
此状态下开机需先按仪器右侧的白色(RESET)键,按键后可听见冰箱启动声音。
再按白键旁边的绿色(ON)按键,仪器及计算机将启动,系统进入待机的过程需要105分钟。
b. 待机状态下开机关机后保持冰箱继续开启为待机状态。
此状态下开机仅需要按“ON”按键,仪器及计算机将启动总共大约需要最多20分钟。
c. 自动开机AU480可以在系统参数内或者在关机窗口内预先设定每天或第二天的自动开机时间。
自动开机必须在待机状态下才可以执行。
在设定的时刻仪器及计算机将自动启动。
d. 仪器初始化在任一种方式下开机启动后,计算机屏幕出现<Program download to the analyzer>(程序下载)后将自动进入到AU480程序的<Start Condition>(起始菜单)。
9错误标记在能够产生错误结果的情况下,系统会提示错误标示,从轻微错误提示到需要立即注意的严重错误警告的所有情况均包括在内。
对于操作者来说,熟知每个错误提示及其产生的根本原因并且采取相应的对应措施是非常必要的。
如果出现未知或未见的标示,系统将不会报告结果。
当对结果产生怀疑时,建议重新测定样品,必要时可对其进行稀释。
本章列举了全部的错误标示,并按优先级、产生原因及对应措施进行了排列。
内容9.1 错误标示概要....................................................................................9-29.2 错误标示详解....................................................................................9-4d (用户拒绝接纳质量控制).................................................................................... 9-5e (用户所编辑的数据) .......................................................................................... 9-5 ( (清洁剂预计用量不足) ....................................................................................... 9-5 Wa (使用错误比色杯得出的分析结果) ................................................................. 9-5 R (试剂不足) ........................................................................................................ 9-6 S# (样品不足) ...................................................................................................... 9-6 % (发现凝块) ....................................................................................................... 9-7 ? (无法计算结果) ................................................................................................. 9-7 n (LIH 检测无法运行) .......................................................................................... 9-8 l=(结果可能受脂血影响) ...................................................................................... 9-8 i=(结果可能受黄疸影响) ...................................................................................... 9-8 h (结果可能受溶血影响) ...................................................................................... 9-8 Y (试剂空白OD高于分光测定最高点) .................................................................. 9-8 U (试剂空白OD低于分光测定最低点) ………………………….............................. 9-9 y (试剂空白/常规OD 为分光测定起始点高值) ..................................................... 9-9 u (试剂空白/常规OD 为分光测定起始点低值) ……………………........................ 9-9 @ (OD 值高于3.0) ………………….................................................................. 9-10 $ (确定反应线性数据不足)................................................................................... 9-10 D (反应OD值高于最大测定范围) ……………………………………...................... 9-11 B (反应OD值低于最小测定范围) ......................................................................... 9-11* (线性估值错误) ................................................................................................. 9-12 & (前带测试数据异常) ........................................................................................ 9-12 Z (前带错误) ....................................................................................................... 9-13 E (估值化验发觉反应过度) ................................................................................. 9-13 Fx ( (OD) 值结果高于动态范围) ......................................................................... 9-13 Gx ( (OD) 值结果低于动态范围) ........................................................................ 9-13 ! (无法计算浓度) ................................................................................................. 9-14 ) ( 样品分析所用试剂批号与定标分析所用不符)................................................. 9-14 a (试剂过期) ....................................................................................................... 9-14 ba (定标过期) ..................................................................................................... 9-15 bh (无有效定标) .................................................................................................. 9-15 bn (使用主要曲线) .............................................................................................. 9-15 bz (使用前带数据定标曲线) ................................................................................ 9-15F (结果高于动态范围) ......................................................................................... 9-16G (结果低于动态范围) ........................................................................................ 9-16 ph (结果高于高危值) ........................................................................................... 9-16 pl (结果低于低危值) ............................................................................................ 9-16 T (化学检查发现异常性) …………………………….............................................. 9-17 P (阳性) ............................................................................................................... 9-17 N (阴性) ............................................................................................................... 9-17 H (结果高于参考范围) ......................................................................................... 9-17 L (结果低于参考范围) .......................................................................................... 9-17 J (结果高于重复确认范围) ................................................................................... 9-18 K (结果低于重复确认范围) .................................................................................. 9-18 fh (结果高于重复反射范围)................................................................................... 9-18 fl (结果低于重复反射范围) ................................................................................... 9-18 Va (多次测量结果离散检测是NG)……………………………………...................... 9-19 xQ (多重法则中的一项失败QC) ........................................................................... 9-19 1Q (QC数据超出输入的单一检查级别) …………………………………………...... 9-20 2Q (QC 数据超出13SD 控制范围) ................................................................... 9-20 3Q (QC 数据持续超过2 SD 控制限制 .............................................................. 9-21 4Q (QC 数据超出R4S 控制范围) ...................................................................... 9-21 5Q (QC 数据超出41S 控制范围) ...................................................................... 9-22 6Q (连续QC结果预设值与平均值不符)…………………………………………….... 9-22 7Q (连续QC结果显示稳定上升或下降值)...............................................................9-23 S (重复取样,重复试验结果取代原始结果) .......................................................... 9-23 / (未决试验) .......................................................................................................... 9-23 r (数据传输到主机) ............................................................................................... 9-23 c (结果经用户修正) ............................................................................................... 9-249.1 Summary of Error Flags错误标示概要下表以优先级顺序对错误标示进行概述标示原因d 被用户从质量控制中排除e 用户编辑的数据( 缺少污染清洁剂参数Wa 在错误的比色杯分析的结果R 试剂不足# 样品不足% 凝块检查? 未能计算结果n 乳糜,黄疸,溶血测试未执行l 可能受乳糜影响的结果i 可能受黄疸影响的结果h 可能受溶血影响的结果Y 试剂空白光密度在最后光度测定点上过大U 试剂空白光密度在最后光度测定点上过小y 试剂空白/常规光密度在初始光度测定点上过大u 试剂空白/常规光密度在初始光度测定点上过小@ 光密度大于3$ 测定反应线性的数据不足D 反应光密度超过最大光密度范围B 反应光密度小于最小光密度范围* 速率法线性错误& 前区试验数据异常Z 前区错误E 检测速率分析过度反应Fx 结果(光密度)超出动态范围Gx 结果(光密度)未及动态范围! 未能计算浓度) 样品分析所用试剂批号与定标分析所用的不同a 试剂过期ba 定标过期bh 未使用有效定标bn 使用通用曲线bz 使用前区数据的定标曲线F 结果超出动态范围G 结果未及动态范围Tx T-Hb或/和HbA1c结果超出动态范围ph 结果大于最高值pl 结果小于最低值T 在微观化学检查中发现异常P 阳性标示原因N 阴性H 结果超出参考范围J 结果未及参考范围K 结果超出重复决策范围L 结果未及重复决策范围Fh 结果超出重复运行反射范围fl 结果未及重复运行反射范围Va 组合测量隔离检查结果未通过Xq 多规则质量控制中单一控制失败1Q 质量控制数据超出单一检测水平范围2Q 质量控制数据超出13S控制范围3Q 质量控制数据持续超出2SD控制范围4Q 质量控制数据超出R4S控制范围5Q 质量控制数据超出41S控制范围6Q 连续质量控制结果的预设值落在平均值的一侧7Q 连续质量控制结果显示稳定增加或降低的数值S 样品重复并且重复结果替代原始结果/ 未定或未分析的测试r 数据传送至主机c 用户校正数据9.2 Error Flag Details错误标示详解d (被用户从质量控制中排除)质量控制数据已被手工排除在计算之外。
作业指导书标题:Beckman Coulter AU480 全自动生化分析仪标准操作规程修改记录目录一、目的二、仪器的使用要求三、检测系统概述3.1 系统组件3.1.1 仪器3.1.2 试剂3.2 样本检测过程概述四、仪器的操作4.1仪器主要操作界面4.2标本传送装置4.3开机操作规程4.4试剂装载程序4.5校准操作程序4.6质量控制操作程序4.7常规标本测定程序4.8重复测定操作程序五、关机与维护保养操作程序5.1.关机程序5.2.仪器校准程序5.3.维护保养程序六、高级操作程序6.1项目参数的设定程序6.2删除实验项目七、相关文件一、目的为了正确使用检测设备,维护和保持检测设备的技术性能,以Beckman Coulter AU480全自动生化分析仪用户手册为依据编写了此规程,供指导日常常规操作使用。
二、仪器的使用要求仪器的安全使用要求,下面列出了警告标签。
三、检测系统概述3.1 系统组件3.1.1 仪器外部结构见下图(图3-1)图3-1 仪器外部结构图3.1.2 试剂AU480能使用的有60ml、30ml和15ml规格的试剂瓶3.2 样本检测过程概述一项实验请求就是一条对一份样品进行特定实验的指令。
当一个样品被放入本设备时,实验需求的信息就已经与样本所需要的实验建立了联系。
样本通过进样模块进入到仪器,通过轨道被运输至采样位置。
试剂探针吸取R1试剂,加入反应杯。
混匀器混匀R1后,样本探针随即吸取样本加入到相应反应杯中,混匀器混匀样本和R1并且完成孵育后,试剂探针吸取R2试剂,加入反应杯中。
仪器从添加R1开始,反应期间每间隔18秒进行一次测量,这期间测定的值和特定实验参数定义的波长可以计算出被测物浓度。
四、仪器的操作4.1仪器主要操作界面A.主界面(图4-1)图4-1 仪器主界面主界面主要由3部分组成A:主控制区B:菜单区C:警告区B. 菜单界面(图4-2)图4-2 菜单进入菜单键后,左边是主菜单,主要由常规(Routine)、定标(Calibration)、质控(QC)、参数(parameters)、维护保养(Maintenance)、系统(System)组成;右边是相应的子菜单,点击可进入相应的操作界面4.2标本传送装置试管架:AU480全自动生化分析仪采用“排列式”的进样方式,传输轨道将装载有样品管(杯)的样品架传输到加样位置进行加样操作。
AU480 操作规程一、开机:如果是正常关机,则只需按ON键。
如果是异常关机(如紧急停机),则先RESET,再按ON键。
每次开机时会提问是否建立一个新的Index, 每一天的第一次开机选择YES。
当天如果有再次开机,不需要建立一个新的Index, 可以选择下面的一行已经建立的Index。
这样的话当天的病人资料都在一个Index里,方便查询。
仪器侧面的主电源一般不关。
仪器背面还有几个开关(给冰箱供电、给电脑供电),一般也不关。
二、关机选择EXIT—Power off---Yes即可。
仪器及电脑自动关,不需按任何开关。
三、测试项目设置(原装有条码)任何测试项目都必须完成以下的设置,才可以进行后续的试剂转载,定标和样本测试。
这些设置都在设置菜单(Setup Menu)里进行。
项目设置按以下顺序逐项进行:1,公共测试参数设置(Common Tset Parameters)设置菜单(Setup Menu)--参数设置(Parameters)--公共参数设置(Common Parameters) --项目名称设置(Tset Name)点击编辑(Edit),A. 输入项目简称(Short name),项目全称(Long Name)-注:可以不输入,B. 输入试剂ID *** 注:3位数字,本ID在项目手册上。
C. 输入剩余测试量提醒。
--注:默认值为测试数小于32时,有报警提示。
D. 选择Multi-switch 为Multi.---同一测试试剂,如果有2盒及以上(同一试剂最多可以同时装到5盒)第一盒的R1用完,R2还有剩余,那么第一盒的R2可以和第二盒的R1混用。
反之亦然。
以此类推。
重复A-D,完成所有项目的共用参数设置。
按Confirm(F1),保存并退出。
2,设置测试菜单(Group of Test)测试菜单Group,共有3组,Group1,Group2,Group3。
每组最多可以配置63个测试项目(含3个ISE)这也是一起同时可以检测的最大项目数。
. . . 简明操作手册(AU全自动生化仪培训资料Ver 1.0)适用机型:贝克曼AU480贝克曼AU680每天工作流程:一、开机前检查;二、开机;三、修改日期索引;四、检查、添加试剂;五、做质控,查看质控结果;六、如有必要,做定标;七、做定标项目的质控,查看结果;八、做标本;九、插入急诊标本;十、查看标本结果;十一、传输标本结果;十二、关机;十三、常见符号;十四、常见报警。
仪器主要单元介绍仪器主要界面介绍主要按键介绍一、开机前检查1、检查样品配送器、试剂配送器是否泄漏;2、检查清洁剂蠕动泵装置是否泄漏;3、检查主清洁剂的数量并补给;4、检查并清洗样品探针、试剂探针和搅拌棒;二、开机开启,仪器主机以及操作电脑将会自动启动,操作电脑进入“修改日期索引”界面。
三、修改日期索引1、开机后,仪器自动进入以下界面2、点击仪器将自动建立一个日期索引。
注:每天必须修改日期索引,并且每天只能修改一次,如果当天第二次开机,则选择“Current Index”,然后点击仪器将不建立新索引。
四、检查、添加试剂2、点击“Roution”点击“Reagent”后,进入以下界面点击“Details”点击“Details”后,进入以下界面1、在此查看各试剂剩余测试数,如果哪个试剂少,请添加。
点击“Reagent Check ”,进入以下界面1、如果添加的试剂品种很多,就使用第一个选择;2、如果添加试剂品种只有几个,使用第二个;3、第三个选择几乎不用;4、试剂状态显示红色的时候,使用“试剂复位”。
选择好后,点击“Start”,仪器将对试剂进行检查。
五、做质控,查看质控结果1、做质控:将质控品放在“绿色”样品架相应的位置,将“绿色”样品架放置于仪器进样轨道上,点仪器“开始”键。
仪器自动做设置好的项目的质控。
2、查看质控结果:在中文报告系统,点击“质控”“质控图”,即可查看质控结果。
六、定标点击“Rack Requisition”后,进入以下界面点击“Calibrarion”点击“Calibrarion”后,进入以下界面2、选择需要定标的项目将定标液放置于“黄色”定标架指定位置,在“蓝色”试剂空白架的“1号”位置放水。
作业指导书标题:Beckman Coulter AU480 全自动生化分析仪标准操作规程修改记录目录一、目的二、仪器的使用要求三、检测系统概述3.1 系统组件3.1.1 仪器3.1.2 试剂3.2 样本检测过程概述四、仪器的操作4.1仪器主要操作界面4.2标本传送装置4.3开机操作规程4.4试剂装载程序4.5校准操作程序4.6质量控制操作程序4.7常规标本测定程序4.8重复测定操作程序五、关机与维护保养操作程序5.1.关机程序5.2.仪器校准程序5.3.维护保养程序六、高级操作程序6.1项目参数的设定程序6.2删除实验项目七、相关文件一、目的为了正确使用检测设备,维护和保持检测设备的技术性能,以Beckman Coulter AU480全自动生化分析仪用户手册为依据编写了此规程,供指导日常常规操作使用。
二、仪器的使用要求仪器的安全使用要求,下面列出了警告标签。
三、检测系统概述3.1 系统组件3.1.1 仪器外部结构见下图(图3-1)图3-1 仪器外部结构图3.1.2 试剂AU480能使用的有60ml、30ml和15ml规格的试剂瓶3.2 样本检测过程概述一项实验请求就是一条对一份样品进行特定实验的指令。
当一个样品被放入本设备时,实验需求的信息就已经与样本所需要的实验建立了联系。
样本通过进样模块进入到仪器,通过轨道被运输至采样位置。
试剂探针吸取R1试剂,加入反应杯。
混匀器混匀R1后,样本探针随即吸取样本加入到相应反应杯中,混匀器混匀样本和R1并且完成孵育后,试剂探针吸取R2试剂,加入反应杯中。
仪器从添加R1开始,反应期间每间隔18秒进行一次测量,这期间测定的值和特定实验参数定义的波长可以计算出被测物浓度。
四、仪器的操作4.1仪器主要操作界面A.主界面(图4-1)图4-1 仪器主界面主界面主要由3部分组成A:主控制区B:菜单区C:警告区B. 菜单界面(图4-2)图4-2 菜单进入菜单键后,左边是主菜单,主要由常规(Routine)、定标(Calibration)、质控(QC)、参数(parameters)、维护保养(Maintenance)、系统(System)组成;右边是相应的子菜单,点击可进入相应的操作界面4.2标本传送装置试管架:AU480全自动生化分析仪采用“排列式”的进样方式,传输轨道将装载有样品管(杯)的样品架传输到加样位置进行加样操作。
样品架以颜色区分不同功能,以“Rack ID”(架子条码)为架子编号a)白色样品架:用于放置常规样品b)红色样品架:用于放置插入急诊样品c)橙色样品架:用于放置手动重测样品d)蓝色样品架:用于放置试剂空白样品e)黄色样品架:用于放置定标品f)绿色架:用于放置质控样品4.3开机操作规程A.开机前检查程序a)检查供水与供电:1.检查去离子水电源、纯水机工作状态、水箱水量,确保仪器工作状态水箱有足够的水量,AU480耗水速度最大为360L/h,水质要求电导率<2us/cm2.检查UPS电源工作状态,UPS电源应该处于断电保护及AC灯亮的工作状态b)检查仪器1.检查清洗剂B原液量(图4-4),桶内液面不低于1/3,如不足需添加2.检查样品针、试剂针、搅拌棒是否有结晶黏附、弯曲,是否处于正常状态。
如有结晶黏附,用乙醇擦拭干净,如弯曲,则校正至正常状态3.检查样品和试剂分配器是否有渗漏(图4-5)4.检查清洗液蠕动泵是否有渗漏5.检查ISE试剂分配是否有渗漏(图4-5)6.检查打印机和打印纸7.检查样品针清洗液(W1位),Det1位置放置2%清洗液B50ml,Det2放有效氯量为0.5%的次氯酸钠消毒液50ml。
如不够则添加8.检查样品架收集槽内有无样品架,有则去除9.确认仪器台面清洁、无杂物10.确认以上各项均正常后,执行开机程序11.填写检查记录表图4-4 检查清洗液B图4-5 检查各分配器B.开机程序a)启动电源:按仪器前面右侧的电源开关“ON”启动仪器(图4-6)b)仪器自动初始化过程:当计算机显示“Program download to the analyser”后仪器将询问用户是建立新的索引还是使用上一次的索引(建议建立新的索引),仪器将自动运行INITIAL(初始化)程序使仪器各部分做复位动作后,仪器转到WARM UP(预温)状态,WARM UP状态约需22min结束(正常开机需22min,如紧急停机后启动需90min左右)图4-6 开机,紧急停机,重置键4.4 试剂装载程序A.Beckman Coulter AU480全自动生化分析仪配置1个试剂仓,试剂仓共可装载76个(种)试剂。
试剂必须使用配套的试剂瓶才能装载在试剂仓中,可供使用的试剂瓶规格有60ml、30ml、15ml共3种,使用15ml、30ml试剂瓶时必须使用相应的“适配器,挡片”固定,以防止试剂瓶倾倒。
试剂瓶尽量不要重复使用,或者清洗干净后再次使用。
Beckman Coulter AU480全自动生化分析仪配置ID条形码识别方式和固定试剂瓶位置两种试剂识别方式。
将试剂设定为ID条形码方式时,仪器通过扫描试剂瓶上的条形码自动识别试剂的种类。
这种识别方式必须使用原装配套试剂。
对于非原装试剂或在某些情况下ID条形码识别方式不能正常使用时,必须采取固定试剂位置方式,指定试剂种类的位置B.装载试剂:以装载总蛋白(TP)测试试剂为例说明装载试剂的方法和步骤a)原装试剂1.打开试剂仓的盖子,将总蛋白试剂1(R1,30ml瓶装)、总蛋白试剂2(R2,30ml瓶装)放入试剂仓任意位置,加装挡片固定,盖上试剂仓盖子,并确定盖子位置正确2.点击“Home”(主界面)——选择“Reagent Management”(试剂管理)——进入“ReagentManagement ”(试剂管理)选择“Main”(主要)——点击“Reagent Check”(试剂检查)——点击“Details”(详细信息)进入“Details”(详细信息)界面观察试剂详细信息b)非原装试剂1.打开试剂仓的盖子,将总蛋白试剂1(R1,30ml瓶装,无条码)、总蛋白试剂2(R2,30ml瓶装,无条码)分别放入试剂仓的第4、5号位置,加装挡片固定,盖上试剂仓盖子,并确定盖子位置正确2.点击“Home”(主界面)——选择“Reagent Management”(试剂管理),右下角有一个“Reagent Check”(试剂检查),点击它等待试剂检查完成——点击“Details”(详细信息)——在“Reagent Display”(试剂显示)处选“Position”(位置),——上下翻页——选择需要放置试剂的位置(本例TP的R1和R2分别放置位置为4、5号位置,可以发现此处试剂显示为“”)——点击“Position Setting”(位置设置)——选择为“Fixed Reagent”(固定试剂)——点击“Edit”编辑相应的试剂名称和相关信息(TP,30ml等)——点击“Confirm”保存并退出,可以发现固定位置试剂左方有一个“*”3.然后再选择“Reagent Check”(试剂检查)对刚才设定试剂液量进行检查C.更换或添加试剂a)点击“Home”(主界面)——选择“Reagent Management”(试剂管理)——进入“Reagent Management ”(试剂管理)选择“Main”(主要)——点击“Reagent Check”(试剂检查)——点击“Details”(详细信息)进入“Details”(详细信息)界面观察试剂详细信息。
其中显示当前剩余测试数、在机有效期、批号、瓶号等b)根据用量添加或者更换试剂(一台仪器同一个项目最多可以放置5套相同试剂),应尽可能的采用更换试剂方式,避免重复使用试剂瓶。
打开试剂仓盖,根据试剂信息显示,更换或添加所需试剂;点击做下角的Initialize Onboard Stability(初始化在机稳定期)D.检查试剂a)Beckman Coulter AU480全自动生化分析仪要求更换试剂或者添加试剂后必须执行一次“Reagent Check”(试剂检查)程序,更换或者添加的试剂才被确认b)点击“Home”(主界面)——选择“Reagent Management”(试剂管理)——进入“Reagent Management ”(试剂管理),在右下角有一个“Reagent Check”(试剂检查),点击进行试剂检查。
c)试剂检查方式有多种:开机后第一次试剂检查或者更改过试剂参数后只能选择“Check all position”(检查所有位置),将在机的所有试剂进行检查。
后面更换或者添加试剂,可以根据具体情况选择具体的试剂检查方式。
有如下方式:1.Reset only(仅复位):执行此操作只做试剂仓复位2.Check all position(检查所有位置):检查全部项目的试剂3.Check specificd position(检查指定位置):检查指定试剂位置,选择此项后弹出项目窗口,用鼠标选择需要检查的试剂的具体位置4.Check specific position(With ID)(检查指定条码项目):检查指定试剂位置,选择此项后弹出项目窗口,用鼠标选择需要检查的试剂的具体位置5.Check changed position(With ID):检查开机后位置改变了的条码项目d)查看试剂信息:点击“Home”(主界面)——选择“Reagent Management”(试剂管理)——进入“Reagent Management ”(试剂管理)选择“Main”(主要)——点击“ReagentCheck”(试剂检查)——点击“Details”(详细信息)进入“Details”(详细信息)界面观察试剂详细信息。
其中显示当前剩余测试数、在机有效期、批号、瓶号等。
确认正确后,退出并返回主“Home”(主界面)。
填写《试剂添加与更换记录表》4.5 校准操作程序仪器在使用过程中,应按分析项目操作规程的校准周期要求对仪器项目进行校准A.设置校准参数a)设置校准品1.点击“Menu List”(菜单列表)——选择“Parameters”(参数)——点击“CalibratorParameters”(定标品参数)——选择“Calibrators”(定标品)——进入“Calibrators”(定标品)界面2.点击“Edit”(编辑)设置定标液的名称,定标液放在样本架的位置——点击“Confirm”保存并退出b)设定校准参数1.点击“Menu List”(菜单)——选择“Parameters”(参数)——点击“CalibratorParameters”(定标参数)——选择“Calibrator Specific”(定标具体参数)——进入“Calibrator Specific”(定标具体参数)界面2.点击“Edit”(编辑)设置按参数文件设定相应定标液的具体值,限制范围。