3. Uncertainty on reading Equipment No.: 32143
Smallest division on ruler = 0.01mm, assume rectangular distribution Standard uncertainty= 0.0029 mm
Equipment No.:
Combined uncertainty = Coverage factor = Expanded uncertainty =
0.0215566 mm 2 0.04 mm
Reporting:
2.03±0.04
mm
with k = 2 at 95% confidence level
3 of 3
Equipment No.:
30019
Smallest division on ruler = 0.02mm, assume normal distribution Standard uncertainty= 0.0067 mm
2 of 3
Uncertainty of Cords Thickness Test Data worksheet
Vernier Calliper 30019 2016/8/5 8Ʋm with k=2 0.0040 weight 23101 2015/8/26 32mg with k=2 mm
16 mg 1.00E-06 mm
The thickness uncertainty caused by weight uncertainty is:
1.88 1.95 1.97 1.86 2.01 1.91 2.02
1.95 1.88 1.97 1.99 2.08 1.87 2.04
2.05 1.82 2.09 2.05 2.01 1.85 2.02
Calculation of uncertainty Uncertainty component Equipment 32143 Equipment 39419 Equipment 30014 Equipment 30019 Equipment 23101 Reproducibility Lab staff U 0.00493 0.00416 0.00833 0.00777 1.00E-06 0.00698 0.01563 C 1 1 1 1 1 1 1
Note : We take average thickness of the measurements as the result. Source of Uncertainty 1. Environment: The environment impact caused by the instrument error of measurement uncertainty is assumed to be 0. 2. Equipment Measuring Instrument Used: Equipment No. Data of Measurement: Expanded uncertainty of equipment: Standard uncertainty = 0.0040 Digital Calliper 32143 2016/10/5 8Ʋm with k=2 mm
Measuring Instrument Used: Equipment No. Data of Measurement: Expanded uncertainty of equipment: Standard uncertainty =
Digital Calliper 39419 2016/9/18 6Ʋm with k=2 0.0030 mm
39149
Smallest division on ruler = 0.01mm, assume rectangular distribution Standard uncertainty= 0.0029 mm
Equipment No.:
30014
Smallest division on ruler = 0.02mm, assume normal distribution Standard uncertainty= 0.0067 mm
1 of 3
ቤተ መጻሕፍቲ ባይዱ
Uncertainty of Cords Thickness Test Data worksheet
Measuring Instrument Used: Equipment No. Data of Measurement: Expanded uncertainty of equipment: Standard uncertainty = Measuring Instrument Used: Equipment No. Data of Measurement: Expanded uncertainty of equipment: Standard uncertainty =
Uncertainty of Cords Thickness Test Data worksheet
Test Method Tension the cord under test with a force of (25± 2)N. Measure the thickness of the cord at five locations along its length with a suitable device having an accuracy of ± 0,1mm. Calculate the mean thickness of the cord. Model T=(T1+T2+……T5)/5 where T= The average cord thickness of the five measurements T(1-5)= Measurement thickness at 5 different location on cord Result of measurement 2.02 The average is: 2.03 2.04 mm 2.02 2.04 2.02
Measuring Instrument Used: Equipment No. Data of Measurement: Expanded uncertainty of equipment: Standard uncertainty =
Vernier Calliper 30014 2016/5/10 10Ʋm with k=2 0.0050 mm
S.D. = Standard uncertainty =
0.03492
mm 0.0070 mm
5. Uncertainty of lab staff
1.96 1.93 2.14 1.77 2.11 1.98 2.02 S.D. = Standard uncertainty = 0.09248
2.12 1.93 2.09 2.11 2.08 1.98 2.04 mm 0.0156 mm
4. Reproducibility of measurement 2.03 2.08 1.99 2.00 2.02 2.08 2.02 1.98 1.98 2.04 1.97 1.99 2.01 2.00 2.02 2.01 2.05 1.99 2.00 2.04 2.09 2.00 1.98 2.08 2.02