Alltima HPLC columns信息
- 格式:pdf
- 大小:442.47 KB
- 文档页数:4
General Instructions for Care and UseContents of Package1 ZIC ®-c HILIC HPLC Column 1 Certificate of Analysis 1 Use and Care InstructionIntroductionThe ZIC ®-c HILIC column has a phosphorycholine zwitterionic stationary phase covalently attached to porous silica. The permanent and hydrophilic zwitterion functionality makes the column suitable for HydrophilicInteraction Liquid Chromatography (HILIC). Weak electrostatic interactions between charged analytes and the neutral zwitterionic stationary phase results in a unique selectivity, and especially suitable for analytes that are poorly retained on reversed phase columns.The ZIC ®-c HILIC column can be used as a tool to change the selectivity or to improve peak resolution for polar and hydrophilic compounds such as carbohydrates, metabolites, acids and bases,organic and inorganic ions, metal complexes, amino acids, peptides, and protein digests.Column Hardware and Chemical CompatibilityThe capillary and microbore columns (0.3 - 1.0 mm ID) are made from stainless steel that has been internally glass lined, and has 10 - 32 UNF female fittings. The nano columns (0.1 mm ID) are made from poly(ether-ether ketone) [PEEK] sheeted fused silica tubing, and has 10 - 32 UNF female fittings. The frits have a porosity to retain 3 - 5 μm particles and are made from stainless steel.The ZIC ®-c HILIC column can be operated in the pH range 2 to 8, while strongly alkaline solutions and washing with sodium hydroxide should beavoided. The ZIC ®-c HILIC column can be heated and operated up to 70 °C.Cleaning and RegenerationIf the backpressure increases or a shift in selectivity is observed, use thefollowing recommended column wash procedure:30 column volumes of deionised water 30 column volumes of 0.5 M NaCl 30 column volumes of deionised waterAn initial washing with deionised water is used to remove organic solvent and polar impurities, followed by a flush with a 0.5 M sodium chloride solution. Finally remove the salt solution with sufficient water andequilibrate the column with 80 % (v/v) acetonitrile. On-line cleaning of the column can be accomplished by increasing the aqueous portion of the eluent close to 100 % while keeping a sufficient ion strength. Thisprocedure can be included at the end of a gradient run and is especially useful for routine applications with samples containing appreciable amounts of salts, such as urine and plasma.StorageThe column is delivered filled with 80 % (v/v) acetonitrile in ammonium acetate buffer (5 mM, pH 6.8) and that is also the recommended solvent for long term storage. Connect the end stop plugs when the column is removed from the system.Store columns as shipped:Acetonitrile / NH 4Ac 5 mM, pH 6.8; 80 : 20 (v/v)Dispose the column according to local authorities and regulations.Sample Solvent and Solvent StrengthSample solvents should consist of 60 - 100 % organic solvent, orinitialeluent composition. Water should be minimized. Weak HILIC solvents such as acetonitrile are favoured. It is recommended to have about 5 % water in the auto sampler wash solution. THF is normally a poor solvent for HILIC separations due to its strong hydrogen bonding ability.The relative solvent strength for HILIC is:Acetone<Acetonitrile<Isopropanol<Ethanol<Methanol<WaterMobile Phase ConsiderationsTo obtain reproducible results, maintain at least 3 % water in the mobile phase, in order to ensure sufficient hydration of the stationary phase particles.Suitable buffer systems for HILIC separations are formate and acetate, due to their excellent solubility even in very high concentrations of organic solvent. Avoid phosphate, and other low solubility buffers, to prevent precipitation on the column bed. A buffer concentration in the range 5 - 20 mM is recommended for most analytes, with an upper limit of 200 - 300 mM, depending on the solubility in the eluent. TFA and other ion pair reagents should be avoided, as they can interfere with the HILIC separa-tion mechanism, and suppress MS signals.Typical Elution ProtocolsIsocratic elution: 80 : 20 (v/v) acetonitrile / NH 4Ac, (5 - 20 mM) or other suitable buffer salt.Gradient elution:90 % to 40 % acetonitrile in 20 minutes (~2.5 %/min).Equilibrating the ColumnEquilibrate the column with 7 - 8 column volumes of initial mobile phase composition. Note that equilibration can be performed at higher flow rates to save time.Flow-rate and Injection VolumeNever exceed the maximum backpressure listed in the table below.The low viscosity of acetonitrile can enable very high flow-rates before maxpressure is reached. A suitable injection volume is about 1 % of the total column volume.Table: Flow-rate, backpressure and injection volumeColumn Injection Flow-Backpressure I.D. (mm)volume (nL)rate (µL /min)Expected (MPa)Max (MPa)0.1 0.3 1.02 - 20 20 - 200 200 - 20000.15 - 6 1 - 40 10 - 4001 - 5 1 - 5 1 - 540 40 40TrademarksZIC ® is a registered trademark of Merck KGaA, Darmstadt, Germany. All rights reserved 2012.WarningUse of the product in applications not specified, or failure to followinstructions contained in this information insert, may result in improper functioning of the product, personal injury, or damage to property or the product.SeQuant ®ZIC ®-c HILIC HPLC ColumnNano, Capillary, and Microbore ColumnsMerck KGaA, 64271 Darmstadt, Germany, Tel. +49(0)6151 72-2440/HPLCEMD Millipore Corporation, 400 Summit DriveBurlington MA 01803, USA, Tel. +1-978-715-4321Sigma-Aldrich Canada Co. or Millipore (Canada) Ltd. 2149 Winston Park, Dr . Oakville, Ontario, L6H 6J8 Phone: +1 800-565-1400Status: 2020-05-13Made in Germany。
液相色谱柱型号名称液相色谱技术是一种广泛应用于实验室和工业生产中的分离、纯化和分析技术。
而液相色谱柱作为液相色谱技术中不可或缺的组成部分,不同柱材质、柱型号都对色谱分离过程产生着重要影响。
以下是常用的液相色谱柱型号名称列表及其特点。
1. C18柱C18柱是一种经典的反相柱,在反相液相色谱分离中被广泛应用。
其最基本的特点是使用碳端基(-C18H37O)覆盖的硅胶基质,使得C18柱能够与极性化合物获得良好的相互作用,从而在分离过程中实现良好的分离效果。
C18柱的使用寿命长,稳定性高,适用于大多数样品分离。
2. C8柱C8柱与C18柱相似,但其具有更弱的相互作用能力。
因此,C8柱适用于一些更具极性的化合物分离,例如酚类、酮类和醛类等。
3. CN柱CN柱是一种极性的正相柱,在正相液相色谱分离中被广泛应用。
CN柱使用的是氰基覆盖的硅胶基质,能够与酸性和极性化合物获得良好的相互作用,对一些醇类、醛类和酮类等分子分离效果较好。
4. NH2柱NH2柱是一种偏极性正相柱,在正相液相色谱分离中应用较广。
NH2柱使用的是氨基覆盖的硅胶基质,能够与极性分子发生氢键作用。
NH2柱适用于如醇类、酮类、醛类、糖类等极性分子分离。
5. Phenyl柱Phenyl柱是一种反相柱,与C18柱相似,但其具有苯环的结构。
该结构赋予了Phenyl柱更强的π-π堆积相互作用,使得其中的芳香族化合物和芳基侧链能够得到更好的分离和识别。
6. C30柱C30柱是一种新型反相柱,在C18和C8柱之间。
C30柱的使用范围更广,能够处理更多的化学样品,例如脂肪酸、天然产物和药物等。
总之,不同的液相色谱柱型号具有不同的柱材质、柱层厚度、柱孔径、填充剂种类等特点,使用场合不同,选择的柱也不同。
选择合适的液相色谱柱型号来分离样品,将会获得更好的分离结果。
HPLC的常用术语等第一部分色谱曲线1、色谱图(chromatogram):色谱柱流出物通过检测器系统时所产生的响应信号对时间或流动相流出体积的曲线图,或者通过适当的方法观察到的纸色谱或薄层色谱斑点、谱带的分布图。
2、(色谱)峰(chromatographic peak):色谱柱流出组分通过检测器系统时所产生的响应信号的微分曲线。
3、峰底(peak base):峰的起点与终点之间的连接的直线(图1 中的CD)。
4、峰高(h ,peak height):色谱峰最大值点到峰底的距离(图1 中的BE)。
5、峰宽(W ,peak width):在峰两侧拐点(图1 中的F ,G)处所作切线与峰底相交两点的距离(图1中的KL)。
6、半高峰宽(W h/2 ,peak withd at half height):通过峰高的中点作平行于峰底的直线,此直线与峰两侧相交两点之间的距离(图1 中的HJ)。
7、峰面积(A ,peak area):峰与峰底之间的面积(图1中的CHEJDC)。
8、拖尾峰(tailing peak):后沿较前沿平缓的不对称的峰。
9、前伸峰(leading peak):前沿较后沿平缓的不对称的峰。
(又叫伸舌峰、前延峰)10、假峰(ghost peak):除组分正常产生的色谱峰外,由于仪器条件的变化等原因而在谱图上出现的色谱峰,即并非由试样所产生的峰。
这种色谱峰并不代表具体某一组分,容易给定性、定量带来误差。
(又叫鬼峰)11、畸峰(distrorted peak):形状不对称的色谱峰,前伸峰、拖尾峰都属于这类。
12、反峰(negative peak):也称倒峰、负峰,即出峰的方向与通常的方向相反的色谱峰。
14、原点(origin):纸或薄层板上滴加试样部位的中心点(图2)。
15、斑点(spot):平面色谱法中,组分在展开和显谱后呈现近似圆形或椭圆形的色区(图2)。
16、区带(zone):在色谱柱、纸或薄层板上被分离组分所占的区域。
hplc色谱柱
HPLC(高效液相色谱)是一种常用的分离技术,广泛应用于
化学、生物、环境等领域。
而HPLC色谱柱是HPLC系统中
的核心组成部分,起着至关重要的作用。
HPLC色谱柱是一种用于分离和分析混合物的柱状装置。
它通
常由柱体、填料和保护层组成。
柱体是色谱柱的主体部分,填料则是用于分离混合物的介质,而保护层则用于保护填料和延长色谱柱的使用寿命。
在HPLC色谱柱中,填料的选择非常重要。
不同的填料具有
不同的化学性质和分离能力,因此可以根据需要选择合适的填料来实现对目标物质的有效分离。
常见的填料包括正相填料、反相填料、离子交换填料等。
正相填料是指填料表面具有亲水性质,适用于分离亲水性物质。
反相填料则是指填料表面具有疏水性质,适用于分离疏水性物质。
离子交换填料则是指填料表面具有带电性质,适用于分离带电物质。
除了填料的选择外,色谱柱的尺寸也是需要考虑的因素之一。
色谱柱的长度和内径会直接影响到分离效果和分离速度。
一般来说,较长的色谱柱可以提供更好的分离效果,但也会增加分
析时间。
而较小的内径则可以提高分离速度,但也会降低分离效果。
此外,色谱柱还需要进行适当的保养和维护,以确保其长期稳定的工作。
常见的保养措施包括定期清洗色谱柱、更换保护层和填料等。
总之,HPLC色谱柱是HPLC系统中不可或缺的组成部分,它的选择和使用对于保证分析结果的准确性和可靠性至关重要。
因此,在进行HPLC分析时,我们应该根据样品特性和分析要求选择合适的色谱柱,并且合理使用和维护色谱柱,以获得满意的分离和分析效果。
气相色谱柱指南目录SH-Rxi™-1ms Columns (熔融石英)SH-Rxi™-5ms Columns (熔融石英)SH-Rxi™-5Sil MS Columns (熔融石英)SH-Rxi™-17 Columns (熔融石英)SH-Rxi™-17Sil MS Columns (熔融石英)SH-Rxi™-624Sil MS Columns (熔融石英)SH-Rxi™-5HT Columns (熔融石英)SH-Rtx™-1 Columns (熔融石英)SH-Rtx™-1 PONA (熔融石英)SH-Rtx™-5/Rtx™-5MS (熔融石英)SH-Rtx™-5 Columns (熔融石英)SH-Rtx™-5MS—Low-Bleed GC-MS Columns (熔融石英)SH-Rtx™-5Sil MS (熔融石英)SH-Rtx™-35 MS (熔融石英)SH-Rtx™-50 Columns (熔融石英)SH-Rtx™-200/Rtx™-200MS (熔融石英)SH-Rtx™-200 (熔融石英)SH-Rtx™-200MS—Low-Bleed GC-MS Columns (熔融石英)SH-Rtx™-1301 (G43) Columns (熔融石英)P.4P.4P.5P.5P.6P.6P.7P.7P.8P.8P.8P.9P.9P.9P.10P.10P.10P.11P.11P.12P.12P.13P.13P.14P.14P.14P.15P.15P.15P.16P.16P.17P.17P.17P.18P.18P.19 SH-Rtx™-624 Columns (熔融石英)SH-Rtx™-1701 Columns (熔融石英)SH-Rtx™-Wax Columns (熔融石英)SH-Stabilwax™ Columns (熔融石英)SH-Rtx™-BAC1 Columns (熔融石英)SH-Rtx™-OPPesticides2SH-Rtx™-OPPesticides2 Columns (熔融石英)SH-Rtx™-CLPesticidesSH-Rtx™-CLPesticides Columns (熔融石英)SH-Stabilwax™-DA Columns (熔融石英)SH-MXT™-1 Columns(Siltek™-treated 不锈钢)SH-MXT™-5 Columns(Siltek™-treated 不锈钢)SH-Rt™-Alumina BOND/KCl Columns(熔融石英 PLOT)Molecular Sieve 5A PLOT ColumnsSH-Rt™-Msieve 5A Columns (熔融石英 PLOT)Porous Polymer ColumnsSH-Rt™-Q-BOND Columns (熔融石英 PLOT)Innovative Integra-Guard™ Columns4• 通用型色谱柱,适用于助燃剂,精油,碳氢化合物,农药, PCB同系物 (e.g., Aroclor mixes), 含硫化合物, 胺类, 溶剂杂质, 模拟蒸馏, 含氧化合物,汽油分析,有机气体、炼厂气。
色谱柱选择原则 (11)选择固定相 (12)色谱柱直径 (20)柱长 (22)色谱柱膜厚 (24)方法指南 (27)环境应用 (40)石化应用 (52)食品、调味品和香精香料应用 (55)工业化学品应用 (59)生命科学应用 (63)GC毛细管色谱柱 (67)GC/MS色谱柱 (68)优级聚硅氧烷色谱柱 (77)聚乙二醇(PEG)色谱柱 (94)专用色谱柱 (101)PLOT 色谱柱 (110)毛细管柱安装快速参考指南 (118)色谱柱性能下降的原因 (121)问题评估 (128)故障排除指南 (130)GC方法开发基础 (133)目录寻找最佳载气平均线速度 (134)进样器缺省设置 (135)柱温箱温度 (135)开发升温程序 (137)安捷伦J&W高性能气相色谱柱 (1)GC应用 (39)色谱柱的安装和故障排除 (117)气相色谱基础 (2)什么是气相色谱? (2)毛细管柱 (4)术语和条件 (5)安捷伦J&W高性能气相色谱柱2000年,发明熔融石英GC毛细管柱的公司—安捷伦科技公司与第一个用交联硅氧烷聚合物制造GC固定相的制造商—J&W Scientific合并。
现在,因为他们的合作,您可以找到属于安捷伦名下的著名的HP和DB色谱柱系列。
安捷伦科技是一个有着40多年气相色谱经验的公司,能为您提供所需的所有服务。
流失性最低的色谱柱可获得高灵敏度和高性能色谱柱流失会降低质谱图完整性、缩短正常运行时间和色谱柱的寿命。
但是安捷伦J&W 色谱柱有最小的低流失标准而它的固定相具有优异的惰性和高温上限的特点(特别是对于离子阱MS用户)。
更高的精密度可获得更好的结果安捷伦J&W色谱柱符合严格的保留因子(k)性能指标,提供一致的保留和分离性能。
他们的保留指数窄和每米理论塔板数高,确保窄的色谱峰和改善相邻流出峰的分离度。
业界最严格的质量控制性能指标安捷伦的严格测试可确保可靠的定性和定量结果,甚至对于最具挑战性的化合物也可保证。
漠迩化学色谱柱
漠迩化学(MORHCHEM)公司是色谱柱相关耗材的专业生产商,总部位于美国加州。
经过30多年的不断革新,漠迩化学已经研发出一系列HPLC色谱柱,例如Caprisil,NexHgel,ItolSep,Guard column,Synersorb,Spherigel,HyperStar,Prep,CapriChiarl等。
这些色谱柱具有优秀的性能和特点,如良好的峰形,适用于大多数化合物及条件的变化;杰出的PH稳定性;耐酸碱,使用寿命长;以及适合分离痕量杂质等。
例如,漠迩化学的CE553-C18-TH色谱柱具有非常出色的性能,通过优化键合工艺,使键合相更稳定,不容易脱落,具有更长的使用寿命和更好的PH稳定性(pH范围1.5-10.5)。
此外,这些色谱柱也应用于不同的领域和实验中。
例如,CE553-C18-TH色谱柱可以用于分离痕量杂质,适用于大多数化合物及条件的变化。
这些色谱柱还具有耐酸碱的特性,因此可以用于长时间的分析和分离。
总的来说,漠迩化学的色谱柱具有优秀的性能和特点,可以满足不同领域和实验的需求。
这些色谱柱是高效液相色谱法进行物质分离和检测的重要工具,为科学研究、工业生产和质量控制等领域提供了强有力的支持。
气相色谱柱基础知识及安捷伦气相色谱柱介绍主要内容1.气相色谱柱(GC基础知识2.安捷伦GC柱产品及选择指南3.GC柱辅助产品气相毛细管色谱柱的柱材料•石英•不锈钢玻璃最常用、弹性好、惰性好耐高温、不易折断、惰性差、不易截取易碎、惰性差石英毛细管色谱柱其他一些重要概念柱容量色谱柱对一种溶质可容纳的最大量,超过此值,该溶质的峰形会发生畸变(WCOT柱表现为前沿,PLOT柱表现为拖尾)与柱容量相关因素:固定相与溶质极性匹配性(相似相溶原理)膜厚柱内径溶质的保留色谱柱温度极限如60 C至240/260 C温度下限恒温温度上限程序升温温度上限低于该温度使用柱效会降低,但不会损伤色谱可在此温度长时间使用不可超过此温度,且在此温度使用不得超过10分钟主要内容1.气相色谱柱(GC基础知识2.安捷伦GC柱产品及选择指南3.GC柱辅助产品安捷伦GC柱产品及选择指南安捷伦气相色谱柱发展简介Agilent WCOT气相色谱柱介绍Agilent PLOT气相色谱柱介绍安捷伦GC柱产品及选择指南安捷伦气相色谱柱发展简介Agilent WCOT气相色谱柱介绍Agilent PLOT气相色谱柱介绍其他一些重要概念交联和键合固定相交联是将聚合物链通过共价键进行联接键合是将其通过共价键与管壁表面相连交联键合固定相的热稳定性和溶剂稳定性大大提高固定相流失固定相(聚硅氧烷类及聚乙二醇类趋于热力学稳定的降解反应或热分解反应柱流失过程ill /Z =207 m/Z =281in/Z =331柱流失的规律叨流失与温度成正比妬极性固定相比非极性固定相易流失妬对于一定的固定相•流失水平与色谱柱中固定相的绝对量成正比: 流失曲线a长柱b大口径c厚膜的色谱柱产生的流失更大低流失色谱柱的优势吻提高灵敏度增强信号-高效、惰性的色谱柱有利于活性化合物产生尖锐色谱峰降低噪音-低流失色谱柱的基线低、噪音小,从而降低了基线的干扰。
叨温度上限高由于增加了聚合物的稳定性,用户可以提高色谱柱的操作温度,从而能够缩短分析时间、延长色谱柱使用寿命。
Introduction to GPCColumns, Distributions,Sample Prep., Calibration,What’s NewPolymer Analysis TechniquesHigh Performance Liquid Chromatography (HPLC) (mainly Size Exclusion Chromatography, SEC)Mass Spectroscopy (MS)Thermal Analysis (TA)Rheometry, Nuclear Magnetic Resonance spectroscopy (NMR), Fourier Transform Infrared spectroscopy (FTIR)Introduction to GPC OutlineWhat is GPC?GAP of AdditivesWhat’s New?What is GPC?▪Gel Permeation Chromatography (GPC) separates samplemolecules based upon their relative size in solution▪Size Exclusion Chromatography (SEC)▪Gel Filtration Chromatography (GFC)▪GPC is an isocratic mode of separation▪GPC is well-suited for polymer analysis –provides a “molecularweight distribution”Particles arePorous, rigidpolymericmaterialsBig ones are not sloweddown because they aretoo big to go into thepores, so they elute firstWhat is GPC?▪The elution profile represents the molecular weight distribution based upon the relative content of different molecular weights…▪Based on size in solution Elution Volume (retention time)largest smallest Mn Mw Mz Mz+1B ig O nes C ome O ut F irst M o l e c u l a r W e i g h tSome definitionsMolecular weight averages are used to provide numerical differences between samples –Mn: Number average molecular weighto At this point in the curve the number of molecules in the sample to left is equal to the number of molecules to the right–Mw: Weight average molecular weighto At this point in the curve the weight of the molecules to the left is equal to the weight of the molecules to the right–Mz and Mz+1: these values are calculated based on molecular weight and abundance (obtained by ultracentrifugation and GPC software computation)o The values are used for “comparison” purposes•Known samples to unknown samplesThese averages are statistical moments calculated from the molecular weight distribution curveMolecular Weight AveragesGPC Delivers all MW information with one experiment GPC calculates the MW distributionof the polymer; this distribution canbe measured for:–Mn can affect a polymer’s brittleness,flow and compression properties.–Mw is related to strength properties,and impact resistance–Mz is related to elongation andflexibility, (Gumby -rubber)–Mz+1 is related to die swell,(extrusion parameter)Molecular Weight/Physical Property Correlations P r o p e r t y /P r o c e s s s P a r a m e t e r E f f e c t o f H i g h M W E f f e c t o f L o w M W I m p a c t S t r e n g t h M e l t V i s c o s i t y P r o c e s s i n g T e m p F l e x L i f e B r i t t l e n e s s D r a w a b i l i t y S o f t e n i n g T e m p S t r e s s -c r a c k R e s i s t a n c e M e l t F l o w Properties of PolymersGPC Process –Separates by Size in Solution Sample PolymerVo Vt“Bank” of3 GPC Columns Molecular Weight Distribution Chromatogram B igO nesC omeO utF irstWhat is happening inside a column?BOCOF Separation by size –Usually no chemical Retention mechanismBasic Column InformationColumns are put together in seriesto form a bank (>2)Always put the highest pore sizecolumn first (from the injector), andsmallest pore size column last–This reduces back pressure on themost fragile column (LMW column)Ramp the flow up slowly –0.1ml/minute –1.0 ml/minute insmall increments*Change over solvent at a0.1ml/minute flow rate over nightSome applications atHigh Temperature >150°CGPC Column TypesOrganic vs. Aqueous GPC columnsDifferent Pore SizesAnalytical vs. Preparative GPC columns Conventional vs. Solvent efficient, and High speed GPC columns.Care and Use of GPC ColumnsBasic GPC –typically a bank of 3 columns of different pore sizes to cover a broad MW range (as needed)Never use methanol or acetonitrile with Organic GPC columns because certain polar solvents will shrink the column, causing it to void–Note: Make sure to flush the complete system with solvent to be used before connecting the columns to the systemLife time of the columns can be as long as ~ 1 year or moreStore the columns in the solvent used with the column bank kept togetherMake sure that the end fittings on the column are tight to keep the column packing from drying out Be careful not to drop the columns, because they are fragileFilter sample solutionsPrevent air bubbles from getting into columnsGPC RulesPolymer is dissolved in a solvent at a low concentration (<0.10% w/v)Polymer solution passes through crosslinked, organic gel columns packed with controlled pore size particles.Larger molecules do not fit into pores (they are excluded) and elute firstVo : Exclusion volumeVt : Total volume or Permeation volumeExclusion principle assumes no adsorption of polymer molecules on packing materialEvery polymer will be eluted between Vo and VtDetectors in Polymer ChromatographyConcentration–Response ~ concentration, (C)–Refractometer; ∆N = (dn/dc) CStructure-selective–UV/Vis Detector (Also can use as concentration detector if sample has UV response), Lambert-Beer Law –IRMolecular weight sensitive–Response ~ C x f(M)–Light scattering: f(M) = M; M(C)–Viscometer: f(M) = [η] = kMα; [η]C•Where k, alpha are the Mark-Houwink constants–Mass Spec.: f(M) = 1/M; C/MRI’s were among the first detectors used in LC/GPC, (late 1960’s)Typically referred to as a “Universal Detector”Detects all dissolved solutes –“non -specific”Refractive index of any optical medium is defined as the ratio of the speed of light in a vacuum to the speed of light in the mediumDetection based on the refractive index of a given analyteMeasures the difference in RI from the eluent to the dissolved sample, (differential type)–The greater the dRI, the stronger the signal Sensitivity increase as RI difference increases Refractive Index (RI) Definition RSUV/Vis DetectionHigh SensitivityHigh SelectivityGain information on chemical compositionCan be used in gradient modeExcellent for most polymer additivesLinear response over a wide absorption range:–A = x l x C (Beer/Lambert)Sometimes UV/Vis detection is used for–Higher detection sensitivity.–Copolymer analysis (usually coupled with RI detector)–Under gradient elution condition.Polystyrene StandardsUV Detection at 260nmA U 0.000.010.020.030.040.050.060.070.080.090.100.110.120.130.140.150.16Minutes0.5 1.0 1.5 2.0 2.5 3.0 3.5 4.0 4.5 5.0 5.5 6.0 6.57.06.0mm X 15cm RT MB-M Column (1)(Same Pore Size as Conventional Column Set)Flowrate: THF at 0.60 ml/min Injection Volume: 5 µl Detector: 2487 UV @260nm PS Standards3,840,0002,890,0001,260,000775,000422,000186,00042,80016,7005,5702,980890474Evaporative Light Scattering DetectorEvaporative Light Scattering Detection for polymer characterization–ELSD is a concentration detector.–ELSD is not affected by solvent changes–Most appropriate detector for Gradient Analysis of Polymers (GAP) or Gradient Polymer Elution Chromatography (GPEC)–Good alternative to dRI for compounds having a low dn/dcEvaporative Light Scattering Detection for additives–ELSD is a universal detector–Compounds without UV-chromophore groups will be detectedEvaporative Light Scattering DetectorPolymers are combined with low molecular compounds (0.1-3%)Additives have several key functions :–Protection : light stabilizers, antioxidants, anti-UV–Safety : flame retardants–Processing : plasticizers, slip agentsFull characterization of synthetic polymers involves detection of additivesMost of slip agents do not absorb UV light. ELS Detector is a good alternative to UV detectionAdditives MixtureComparison of Detection ModesCrodamide (oleamide) is not detected in UV mode5.0010.0015.0020.00T i n u v i n 312T i n u v i n PB H T L uw i n o x 44 B 25S u c c o n o x 16N a u g a r d 445S u c c o n o x 18T i n u v i n 328I r g a n o x 1330I r g a n o x 1076I r g a f o s 1680.005.0010.0015.0020.00T i n u v i n 312T i n u v i n PL u w i n o x 44B 25S u c c o n o x 16N a u g a r d 445S u c c o n o x 18C r o d a m i d eT i n u v i n 328I r g a n o x 1010I r g a n o x 1330I r g a n o x 1076I r g a f o s 168UV Detection (220 nm)ELS DetectionI r g a n o x 1010GPC Calibration -Narrow Standards A calibration curve is built with low dispersity (narrow) standards with known molecularweight; (ideal if same structure as unknown polymer)(PS, PMMA...).This calibration curve may be used toquantitate a polymer of different nature (PC, PMMA...). Then results are expressed in PSor PMMA equivalents, (or relative to PS orPMMA –incorrect for the polymer sample of interest).The chromatographic process is based on hydrodynamic volume (H)(size in solution), and not molecular weight.Two different polymers with the same MW will elute at different retention volumes.Vo Excluded V Total MW*RangeNoresolutionABOVEthis MW Resolutionrangeis created bydifferences inelution timeElution Time or Volume NoresolutionBELOWthis MW*MW is Log scaleSame “Pore Size” ColumnsVo Excluded V Total MW RangeNoresolution ABOVE this MWResolution range is created by differences in elution timeElution Time or VolumeNoresolution BELOW this MWVo Excluded V TotalMW RangeNoresolution ABOVE this MWResolution range is created by differences in elution timeElution Time or VolumeNoresolution BELOW this MW2 Columns of the SAME Type-SAME MW Range -More ResolutionDifferent “Pore Size” ColumnsVo ExcludedV TotalMW RangeNoresolution ABOVE this MWResolution range is created by differences in elution timeElution Time or VolumeNoresolution BELOW this MWVo Excluded V TotalNew MW RangeNoresolution ABOVE this MWResolution range is created by differences in elution timeElution Time or VolumeNoresolution BELOW this MW2 Columns of the Different Type -EXPANDED MW Range -More ResolutionActual Calibration Curves for the Different Pore SizeHT ColumnsH igh T emperatureHT 2 for Low MWHT 6 for High MWHT 6E for BlendedGPC Calibration 1 –Narrow StandardsCalibration:ually >10 standards usedbracketing MW range2.Standards may be injected as amixture3.Mixtures should be 1 order ofmagnitude different (1000, 10000,100000)To build a calibration curve:Narrow dispersity standards (PD<1.1)Elution volume at peak heightCurve : Log(M) = f(Ve)2.62.83.03.23.43.63.84.04.24.44.64.85.05.25.45.65.86.06.26.418192021222324252627Log Mol WtTime (min)11000001900009100Log(M)Elution volumeGPC Calibration –Narrow StandardsOrganic Polymers Aqueous PolymersPolystyrene Poly(ethylene oxides)Polybutadienes Poly(ethylene glycols)Poly(methylmethacrylates)PullulansPolyisoprenesNarrow Standards -Preparation ConsiderationsNarrow standards may be mixed together to develop a relative calibration curve –No more than 3 in one “cocktail” –be careful of concentrations–MW’s should be one decade apartNarrow standards should only be swirled gentlyNo need to filter narrow standardsAdd antioxidant if high temperature applicationSample Preparation ConsiderationsSample may be mixed to facilitate dissolution–Be careful of shear for high MW (>1M) samplesIn some cases, sample solution should be filtered–Presence of microgels, fillers, any other insolublesAllow enough time for complete dissolutionFor certain crystalline polymers, high temperature may be needed–Example: Isotactic polypropylene requires 2 hours at 170C in an external oven –The PP may then be run at 145CSample Concentration GuideThe more dilute the polymer solution, the better–This will prevent viscosity effects and non-reproducible retention–A dilute solution will allow the polymer to open up into its most relaxed conformation o No chain entanglemento No microgel formationInjection volume no more than 100 µL per columnFor very high MW polymers, flow rate may have to be lowered–HMW columns may be needed as well to prevent shearSample Concentration GuideMW < 1,000MW 1,000 –10,000MW 10,000 –100,000 MW 100,000 –500,000 MW 500,000 –1MMW >1M 0.20 –0.30% 0.15 –0.20% 0.10 –0.15% 0.05 –0.10% 0.01 –0.05% 0.005 –0.01%Above concentrations assumeno more than100 µL injection per columnGPC Column Operation Techniques Switch solvents by flushing to the column at0.1ml/min overnight in the new solventIncrease flow rates at 0.1 ml/minute/minute to the columns specified flow ratesRecommended to keep the flow through the columns at low flow, 50 ~ 100 m l/minute when idlingFor TCB at 140 -150 C -purge the toluene out of the columns at 0.1 mL/min. overnight at ~80C and thenramp up to temperature over 300 min after 3 column volumes of TCB have passed throughDetection–RI–Viscometry–UVPolymer DistributionsA polymer is a mixture of different size chain lengths of the same monomer. Tomeasure this distribution of sizes we use molecular weight averages.Slices are made to the chromatogram, where the height of each slice (H i), represents the population of molecules at that chain length or MW.iHlouvmeulEtionPolymer DistributionsMv High MWLow MWMw MnMv is derivedfrom ViscometryMP Mz Mz+1GPC and Flow Rate PrecisionMnMwVeLC Peak Identification (Narrow Standard)Based on retention timeMolecular Weight DeterminationGel permeation chromatography (GPC)Based on volume of solvent flowing through the column Essential for the flow rate to be absolutely constant To calibrate, plot Log MW of standards vs volume (time)Any flow variability will result in large MW errorsGPC and Flow Rate PrecisionPrecise solvent flow is essential for precise GPC…MwMnVe▪Comparisons of Molecular Weight Distributions of different samples can highlight even small MW shift information (good vs. bad PP’s below)▪Determine "good from bad" -e.g. QC of resin batches… these are typical formulators –use Mw, Mn, Mz and polydispersityPractical use –Fingerprinting the Polymerd w t /d (l o g M )0.000.200.400.60 3.504.00 4.505.00 5.506.00 6.50“Bad”“Good”Degradation of Polyethylene Terephthalate –HFIPAged PET Virgin PETd w t /d (l o g M )0.00.10.20.30.40.50.60.70.80.91.01.11.2Log Mol Wt2.83.03.23.43.63.84.04.24.44.64.85.05.25.45.6PET Exposed To Oils And Refrigerants For Several Days At Elevated TemperatureVirgin PETIntroduction to GPC OutlineWhat is GPC?GAP of AdditivesWhat’s New?Gradient Analysis of Polymers (GAP)In recent years there has been increased interest in using gradient HPLC techniques, such as Gradient Polymer Elution Chromatography (GPEC), with polymers for determining the compositional drift of copolymers, the composition of polymer blends, or for the analysis of polymer additives. Depending upon the gradient conditions and columns selected for analysis, separations may be obtained dependent on molecular weight or based upon precipitation, or adsorption mechanisms. The use of an Evaporative Light Scattering Detector (ELSD) allows one to perform solvent gradients with a universal mass detector and observe both UV absorbing and non-UV absorbing polymer samples without baseline disturbances from the solvent gradient. The addition of a Photodiode Array Detector (PDA) allows for compositional analysis across the molecular weight distribution of many copolymers, can be useful for the identification of components in a polymer blend, and also is invaluable for the quantitation of polymer additives and other small molecules in traditional reverse phase separations.Experimental ConditionsSystem:Waters Alliance 2690 Separations Module with column heater at 30 ºCDetector 1:Waters 996 Photodiode Array DetectorDetector 2:Alltech Model 500 ELSD with LTA Adapter (Drift Tube at 40º C, 1.75 Liters/min Nitrogen)Data System:Waters Millennium 32 Chromatography Manager Column:As listed in Figures, 30 ºCFlow Rate:1mL/minSamples:10 - 25 µl injections of 0.2 - 0.5% samplesGradient:Linear gradient, conditions and mobile phases as listed in Figures.GPC Analysis of a Polymer BlendGradient Analysis of a Polymer BlendStyrene-Acrylonitrile (25% Acrylonitrile)PolystyreneStyrene-ButadieneRubber (50% Styrene)% THF0102030405060708090100Minutes246810121416Symmetry Shield™ C8 Column (3.9x150mm)100% ACN to 100% THF over 20 minutes ELSD DetectionGradient Analysis of Narrow PS Standards% T H F102030405060708090100Time (Minutes)1234567891011Symmetry Shield C8 (3.9x150mm)100% ACN to 100% THF over 10 minutes ELSD DetectionStyrene Oligomers2890910043,900190,000355,000706,0001,090,0002,890,000Analysis of Styrene-Butadiene-Styrene Block% T H F102030405060708090100Minutes1011121314151617181920100% S t y r e n e40% S t y r e n e30% S t y r e n e22% S t y r e n e100% B u t a d i e n ePrototype DVB/Vinylpyrolidone Column (3.9x150mm)100% ACN to 100% THF over 20 minutes ELSD DetectionCalibration of %Styrene in SBR'sGradient Analysis of Low Molecular Weight WaxesTime (Minutes)68101214161820221112131415161718m V"Slack wax""C18 wax"Nova-Pak ®C18 (3.9x150mm)100% ACN to 100% THF over 30 minutes ELSD DetectionPolymer Additives –Tinuvins, (UV Stabilizers)4681012141618202224262814161820222426283032mVMinutesTinuvin 440Tinuvin 900Tinuvin 328ELSD DetectionPolymer Additives –Phthalate Plasticizers5.05.56.06.57.07.58.08.59.09.510.010.511.014161820222426283032343638mVMinutesDcHPDOP (DEHP)DIDPUDPELSD DetectionPolymer Additives -Slips and Antistats204060801001201401601802002201416182022242628303234mVMinutesOleamideErucamideStearic AcidELSD Detection。
General Instructions for Care and UseContents of Package1 ZIC ®-HILIC HPLC Column 1 Certificate of Analysis 1 Use and Care InstructionIntroductionThe ZIC ®-HILIC column has a zwitterionic stationary phase covalently attached to porous silica. The permanent and hydrophilic zwitterionfunctionality makes the column suitable for Hydrophilic Interaction Liquid Chromatography (HILIC). Weak electrostatic interactions between charged analytes and the neutral zwitterionic stationary phase results in a unique selectivity, and especially suitable for analytes that are poorly retained on reversed phase columns.The ZIC ®-HILIC column can be used as a tool to change the selectivity or to improve peak resolution for polar and hydrophilic compounds such as carbohydrates, metabolites, acids and bases,organic and inorganic ions, metal complexes, amino acids, peptides, and protein digests.Column Hardware and Chemical CompatibilityThe column is made from poly(etherether ketone) [PEEK] coated stainless steel and has 10 - 32 UNF female fittings. The frits have a porosity toretain 3 - 10 µm particles and are made from Titan. PEEK generally shows excellent chemical resistance to a wide range of organic solvents com-monly used in HILIC applications, e. g., acetonitrile, formic acid, and alcohols. Swelling of PEEK material may, however , occur after prolonged exposure to solvents like THF , methylene chloride or DMSO.The ZIC ®-HILIC column can be operated in the pH range 3 to 8, while strongly alkaline solutions and washing with sodium hydroxide should be avoided. The ZIC ®-HILIC column can be heated and operated up to 70 °C.Cleaning and RegenerationIf the backpressure increases or a shift in selectivity is observed, use the following recommended column wash procedure:30 column volumes of deionised water 30 column volumes of 0.5 M NaCl 30 column volumes of deionised waterAn initial washing with deionised water is used to remove organic solvent and polar impurities, followed by a flush with a 0.5 M sodium chloride solution. Finally remove the salt solution with sufficient water andequilibrate the column with 80 % (v/v) acetonitrile. On-line cleaning of the column can be accomplished by increasing the aqueous portion of the eluent close to 100 % while keeping a sufficient ion strength. Thisprocedure can be included at the end of a gradient run and is especially useful for routine applications with samples containing appreciable amounts of salts, such as urine and plasma.StorageThe column is delivered filled with 80 % (v/v) acetonitrile in ammonium acetate buffer (5 mM, pH 6.8) and that is also the recommended solvent for long term storage. Connect the end stop plugs when the column is removed from the system.Store columns as shipped:Acetonitrile / NH 4Ac 5 mM, pH 6.8; 80 : 20 (v/v)Dispose the column according to local authorities and regulations.Sample Solvent and Solvent StrengthSample solvents should consist of 60 - 100 % organic solvent, orinitialeluent composition. Water should be minimized. Weak HILIC solvents such as acetonitrile are favoured. It is recommended to have about 5 % water in the auto sampler wash solution. THF is normally a poor solvent for HILIC separations due to its strong hydrogen bonding ability.The relative solvent strength for HILIC is:Acetone<Acetonitrile<Isopropanol<Ethanol<Methanol<WaterMobile Phase ConsiderationsTo obtain reproducible results, maintain at least 3 % water in the mobile phase, in order to ensure sufficient hydration of the stationary phase particles.Suitable buffer systems for HILIC separations are formate and acetate, due to their excellent solubility even in very high concentrations of organic solvent. Avoid phosphate, and other low solubility buffers, to prevent precipitation on the column bed. A buffer concentration in the range 5 - 20 mM is recommended for most analytes, with an upper limit of 200 - 300 mM, depending on the solubility in the eluent. TFA and other ion pair reagents should be avoided, as they can interfere with the HILIC separation mechanism, and suppress MS signals.Typical Elution ProtocolsIsocratic elution: 80 : 20 (v/v) acetonitrile / NH 4Ac, (5 - 20 mM) orother suitable buffer salt.Gradient elution:90 % to 40 % acetonitrile in 20 minutes (~2.5 %/min).Equilibrating the ColumnEquilibrate the column with 7 - 8 column volumes of initial mobile phase composition. Note that equilibration can be performed at higher flow rates to save time.Flow-rate and Injection VolumeNever exceed the maximum backpressure listed in the table below.The low viscosity of acetonitrile can enable very high flow-rates before maxpressure is reached. A suitable injection volume is about 1 % of the total column volume.Figure: Column plate height (u ) and backpressure (¨ ) vs. volumetric flow rate. Cytosine injected on a 50 x 4.6 mm ID ZIC ®-HILIC column at k’ 1.3 using an eluent with 80 : 20 acetonitrile /bufferTable: Flow-rate, backpressure and injection volumeColumn Injection Flow-Backpressure I.D. (mm)volume (µL)rate (mL /min)Expected (MPa)Max (MPa)2.1 4.6 7.50.5 - 5 5 - 50 10 - 1500.05 - 2 0.25 - 10 0.6 - 251 - 5 1 - 5 1 - 555 50 40TrademarksZIC ® is a registered trademark of Merck KGaA, Darmstadt, Germany. All rights reserved 2012.WarningUse of the product in applications not specified, or failure to followinstructions contained in this information insert, may result in improper functioning of the product, personal injury, or damage to property or the product.Flow Rate (mL/min)P l a t e H e i g h t (m m )C o l u m n B a c k p r e s s u r e (M P a )7.75458.0006-xxxxxxxxxx msp.SeQuant ®ZIC ®-HILIC HPLC ColumnAnalytical PEEK coated ColumnsMerck KGaA, 64271 Darmstadt, Germany, Tel. +49(0)6151 72-2440/HPLCEMD Millipore Corporation, 400 Summit DriveBurlington MA 01803, USA, Tel. +1-978-715-4321Status: 2020-05-13Made in Germany。
This data sheet contains impaortant information about the product.Operating instructions forAstec CHIROBIOTIC® LC Strationary PhasesColumn Installation 柱安装CHIROBIOTIC columns are shipped in methanol . Before starting to use a new column , wash with 20ml HPLC grade methanol at 1ml/min.The column test standarad, 5-methyl-5- phenylhydantoin, can be injected at this stage to establish performance .CHIROBIOTIC色谱柱用甲醇装运。
开始使用一个新的色谱柱之前,用色谱纯甲醇以1ml/min 的流速洗涤20毫升。
色谱柱测试标准,5-甲基-5- phenylhydantoin在这个阶段被注入确立表现。
New columns can take longer to equilibrate, but once baseline stability is achieved it is consistent. New threaded hardware improves column performance. Do not overtighten. This threaded hardware also allows for direct connection to a guard column.新柱子需要更长的时间达到平衡,但一旦基线达到稳定,是一致的。
新的螺纹硬件提高柱子的性能。
不要拧得过紧。
该螺纹硬件也可直接连接预柱。
非常感谢您选择Waters XBridge™色谱柱。
XBridge™填料设计精良,能在酸性和碱性流动相条件下获得良好的峰形、达到高柱效并展现出色的稳定性。
所有XBridge™填料均由通过cGMP和ISO 9001: 2000认证的工厂使用超纯试剂生产。
每批XBridge™填料都采用酸性、碱性和中性分析物进行了色谱测试,结果均符合严格的合格标准,可确保色谱柱具有出众的重现性能。
每根色谱柱都经过单独测试,并随附性能测试色谱图以及合格证书。
目录I. 入门指南a.色谱柱安装b.色谱柱平衡c.初始柱效测定II. 色谱柱使用a. 保护柱b.样品制备c. pH操作限值d.溶剂e.压力f.温度III. 放大/缩小等度方法IV. 故障排除V. 色谱柱清洗、再生和储存a.清洗与再生b.储存VI. 将色谱柱连接到HPLCa.色谱柱接头和系统管路注意事项b.测量系统谱带展宽体积和系统方差c.测量梯度延迟(或延迟体积)VII. 其他信息a.使用窄径(3.0 mm内径)色谱柱b.谱带展宽体积对2.1 mm内径色谱柱性能的影响 c.未优化和优化LC/MS/MS系统对比:系统配置更改建议d.沃特世小粒径(2.5 µm)色谱柱 - 快速色谱分离e. XBridge HILIC色谱柱入门f. XBridge Amide色谱柱入门XBridge色谱柱I. 入门指南每根XBridge™色谱柱都附带一份COA报告和一幅性能测试色谱图。
COA报告为XBridge™色谱柱中所用的填料批次所特有,包括批号、未键合颗粒和键合填料分析结果,以及色谱结果和检测条件。
性能测试色谱图提供每根色谱柱的以下信息:批号、色谱柱序列号、USP理论塔板数、USP拖尾因子、保留因子以及色谱条件。
这些数据应妥善保存,以备将来参考。
a. 色谱柱安装注:下述步骤给出的流速适用于粒径5 µm、内径4.6 mm的典型色谱柱。
用户可根据所安装XBridge™色谱柱的内径、柱长、粒径 和柱压相应地提高或降低流速。
英文回答:The Waters HPLC column is a sophisticated and advanced liquid chromatography apparatus, specifically engineered to facilitate the separation and analysis of a diverse spectrum ofpounds within various samples. Crafted from top-tier materials, it is adaptable to a wide array of solvents and buffers. This column is available in a variety of dimensions and particle sizes to cater to the specific demands of a plethora of applications. Installation and operation are seamless, with the added benefit of ensuring exceptional resolution and reproducibility, thereby engendering accurate and dependable results. It is the epitome of excellence for employment in pharmaceutical, environmental, food and beverage sectors, as well as various other industries, for both routine analysis and research endeavors.Waters HPLC柱是一个精密和先进的液相色谱仪,专门设计用于促进各种样品中各种磅的分离和分析。
沃特斯液相色谱柱型号沃特斯(Waters)公司是世界领先的分析测试仪器和化学试剂供应商,其液相色谱(HPLC)柱广泛应用于生命科学、化学、医药等领域的研究和生产中,具有高效、高精度、高灵敏度、高重复性等优点,成为现代科学技术的重要组成部分。
沃特斯液相色谱柱型号是指沃特斯公司生产的液相色谱柱产品的唯一标识符,它包含了柱直径、柱长度、填充材料、填充粒径和填充密度等信息,可以为用户量身定制和推荐最适宜的柱型,满足不同实验需求和分析目的。
沃特斯公司的液相色谱柱型号分为以下几类:1. Symmetry列:Symmetry列是沃特斯公司最早推出的柱系列之一,其特点是高效、高灵敏度、高可重复性和高选择性。
Symmetry列包括Symmetry C18、SymmetryC8、Symmetry Shield RP18等柱型。
2. XBridge列:XBridge列是沃特斯公司全新推出的柱系列,其特点是高效、高分辨率、高选择性、高稳定性和低背景噪声。
XBridge列包括XBridge C18、XBridge C8、XBridge Shield RP18等柱型。
3. Atlantis列:Atlantis列是沃特斯公司推出的一系列针对生命科学应用的柱系列,其特点是适用于多种样品类型和复杂样品基质,并具有高效、高灵敏度、高选择性等优点。
Atlantis列包括Atlantis dC18、AtlantisT3、Atlantis HILIC等柱型。
4. Acquity UPLC BEH列:Acquity UPLC BEH列是沃特斯公司研发的一系列超高效液相色谱柱,其互相作用特性高度统一,电荷配属一致,优化了固定相的表面性质,让分析分离更加可靠。
Acquity UPLC BEH列包括BEHC18、BEH Phenyl、BEH Shield RP18等柱型。
5. Nova-Pak列:Nova-Pak列是沃特斯公司历史上推出的一系列柱,可用于分离小分子化合物和大分子生物分子。
月旭Ultimate®XB-C18色谱柱说明书一、色谱柱简介Ultimate®XB-C18色谱柱采用超纯硅胶为基质,进行彻底的双封尾键合工艺;具有宽的pH使用范围(1.5-10),适应于大部分药物,环境和化工行业的化合物分离,对碱性和极性化合物具有最佳的峰形,是目前市场上最通用的C18分析柱之一,可替代Waters Symmetry C18,Agilent Zorbax XDB C18,Phenomenex Luna C18,Supelcosil LC-18-DB,YMC ODS-AM,Alltima C18,GL-science Inertsil ODS-2等。
二、色谱柱特点超纯全多孔球形硅胶,纯度〉99.999%独特的固定相键合、双封尾技术高柱效:理论塔板数〉90000/米良好的峰形对称性:拖尾因子控制在0.95-1.05较宽pH范围:适合pH值1.5-10.0较长的使用寿命优良的批与批重现性三、色谱柱参数键合相C18粒径3µm,5µm,10µm,孔径120Å,300Å比表面积320m2/g(120Å),90m2/g(300Å)载碳量17%(120Å),8%(300Å)封尾双封尾pH值稳定性 1.5-10.0四、新柱活化采用80%甲醇0.5ml/min冲洗4小时,再换成分析流动相平衡;如果流动相中含有缓冲盐,请使用过渡流动相过渡后再换分析流动相平衡;五、日常维护1、建议检测前样品和流动相进行过滤;2、建议每天做完样品后及时进行清洗;3、常规检测,测试完后直接把色谱柱反向连接采用90%有机相冲洗45min,最后保存;4、使用缓冲盐条件:1)等度条件:使用缓冲盐之前和之后都用过渡流动相以分析流速冲洗45min;2)梯度条件:使用缓冲盐之前与初始流动相组成相同的过渡流动相以分析流速冲洗45min;注意:过渡流动相是指有机相和水相比例与分析流动相相同比例,只是不含有缓冲盐;3)缓冲盐冲洗干净后,采用90%有机相反向冲洗60min,最后保存;注意:缓冲盐溶液不能存留色谱柱中过夜;5、色谱柱的保存1)短时间内色谱柱的保存如果使用时间间隔不超过四天,保存到最后的有机溶剂水里面即可。