FC2000流量计算机中文PPT
- 格式:ppt
- 大小:2.97 MB
- 文档页数:32
The FL-2000 Series offers a wide variety of precision flowmeters for use in medical, industrial, chemical, and laboratory applications at an economical price. Units are available with or without valves.Acrylic FlowmetersB-15aFL-2000 SeriesSpeciFicationSaccuracy:Models FL-2001–FL-2025: ±5% F.S. Models FL-2031–FL-2069: ±3% F.S. Models FL-2071–FL-2128: ±2% F.S.Float: Black glass stainless steel Body: Clear acrylicSeals: Buna “O” Rings with brass or PVC fittings FKM “O”-Rings with stainless steel fittingspressure: 100 psig max @ 21°C (70°F)temperature:65°C (150°F) max @ 0 psigFittings: Brass std; stainless steel optional except for FL-2071 through FL-2128, which have 1 NPT PVC fittings onlyValves: Models FL-2001 throughFL-2069: brass standard; stainless steel cartridge type (optional)FL-2071 through FL-2128: Optional plastic in-line gateU e asy-to-Read english or Metric Scales U W ater Ranges from 4 ccM to 20 GpM, air Ranges from 40 ccM to 4000 LpM U t hreaded Brass inserts for Quick installation U e asy Disassembly and assembly for Maintenance U D urable one-piece clear acrylic construction U S table, easy-to-Read Float U S uperior Quality appLicationS U a ir Sampling equipment U a quaculture U D esalinization equipment U G as analyzers U M edical Systems U p hoto processing equipment U W ater treatment andDistribution SystemsFL-2013 air, shownsmaller than actual size.FL-2066-nV Water, shownsmaller than actual size.B(13.5)B-15bFL-2091 through FL-2128 DimensionsFL-2097, shown smaller than actual sizeTo order with plastic integral gate valve, add suffix “-V” to model number for additional cost for FL-2090 Series, and FL-2120 Series.For optional 10-point NIST certificate add suffix, “-NIST” to the model number, for additional cost and two weeks to the standard lead time.Ordering Example: FL-2095, flow meter , 100 to 1400 LPM Air FL-2127-V, flow meter , 4 to 36 LPM water, with valves.Units are standard without valves.To order with plastic integral gate valve, add suffix “-V” to model number for additional cost.For optional 10-point NIST certificate add suffix, “-NIST” to the model number for additional cost and two weeks to the standard lead time.Ordering Examples: FL-2075, flow meter valve, 100 to 1400 LPM air.FL-2080, flow meter valve, 2 to 19 LPM water.FL-2041-nV, shown smaller than actual size.FL-2053, Water,shown smaller thanactual size.FL-2066-nV, shown smaller than actual size.To order with stainless steel valve, add suffix “-SS” to model number for additional cost.To order without a valve, add suffix “-NV” to model number and subtract from cost.For optional 10-point NIST certificate add suffix, “-NIST” to the model number, for additional cost and two weeks to the standard lead time.Ordering Examples: FL-2036, economical flow meter , with brass valve, 14 to 150 SCFH Air.FL-2036-NV , economical flow meter ,without brass valve,14 to 150 SCFH Air.BFL-2060, air, shown smaller than actual size.Dual scales supplied std: SCFM/SCFH, GPM/GPH and LPM/LPH To order with stainless steel valve, add suffix “-SS” to model number for additional cost.To order without a valve, add suffix “-NV” to model number and subtract from cost.For optional 10-point NIST certificate add suffix, “-NIST” to the model number, for additional cost and two weeks to the standard lead time.Ordering Examples: FL-2060, flow meter with brass valve, 0.5 to 5 SCFM. FL-2069-NV , flow meter no valve 2 to 20 LPM.FL-2091, air, shown smaller thanactual size.B-15cFL-2021-nV, Water, shown larger than actual size.Units come standard with brass valves and operator’s manual.To order with stainless steel valves, add suffix “-SS” to model number for additional cost. To order without a valve, add suffix “-NV” to model number and subtract from cost.For optional 10-point NIST certificate add suffix, “-NIST” to the model number, for additional cost and two weeks to the standard lead time.Ordering Examples: FL-2005, economical flow meter with brass valve, 2 to 20 SCFH air. FL-2005-NV , economical flow meter without valve, 2 to 20 SCFH air.accuracy: ±5% Full Scale panel MountB-15d。
附件1报警,警告和故障码如果有报警发生的情况下,当选中报警页后,会显示报警、警告和故障码。
在报警发生和被复位的时间和日期的下边,会以清单的方式列出报警的功能码。
在下列项中的N将表示所测的管路编号,1,2或者是3。
●非责任报警的参数符号qN low未修正的流量值低于Lo qNpresN High压力值高于PressureN High的值。
presN Low 压力值低于PressureN Low的值。
tempN High 温度值高于T emperatureN High的值。
tempN Low 温度值低于T emperatrueN Low的值。
只对涡轮流量计turbN lf如果有涡轮流量计报警发生(叶轮故障)(低频)只对涡轮流量计usN eff.1% 1声道的发射效率低于预设的%限制usN eff.2% 2声道的发射效率低于预设的%限制usN eff.3% 3声道的发射效率低于预设的%限制usN eff.4% 4声道的发射效率低于预设的%限制usN eff.5% 5声道的发射效率低于预设的%限制●责任报警的参数符号presN max 压力高于pressureN max的值presN min 压力低于pressureN min的值presN df. 正在使用的压力值是默认值pressureN dfPresN hart 流量计算机和压力变送器通讯失败,或者是压力变送器处在报警状态tempN max温度高于temperatureN max的值tempN min 温度低于temperatureN min的值tempN df. 正在使用的温度值是默认值pressureN dfT empN hart流量计算机和温度变送器通讯失败,或者是温度变送器处在报警状态qN high未修正的流量值高于HiN q. 的值超过1分钟znN high基准的压缩因子超出允许的znN high值znN low基准的压缩因子超出允许的znN low值只对涡轮流量计turbN hf如果有涡轮流量计报警发生(叶轮故障)(高频)只对涡轮流量计usN paths 声道号定义错误usN secur流量计的安全报警位被设置usN level1 精度减小报警usN level2 精度减小报警usN comms 在最近的5秒钟内没有一个有效的通讯信号●温度报警tempN alrm温度低于t-alarm的值,并且qb高于Lo.q超过1小时在下列项中的X将表示接、插板槽的编号1,2,3,4或者是5。
Main Totals总累计量页面+Vb1工况总累计量+Vn1标况总累计量+E1 能量总累计量+M1 质量总累计量Current Hour n-acc 当前小时总累计量页面+Vb.ch.na1工况当前小时累积量+Vn.ch.na1标况当前小时累积量+E.ch.na1 能量当前小时总累计量+M.ch.na1 质量当前小时总累计量Current Day n-acc 当前天总累计量页面+Vb.cd.na1工况当前天累积量+Vn.cd.na1标况当前天累积量+E.cd.na1 能量当前天总累计量+M.cd.na1 质量当前天总累计量Current Month n-acc 当前月总累计量页面+Vb.cm.na1工况当前月累积量+Vn.cm.na1标况当前月累积量+E.cm.na1 能量当前月总累计量+M.cm.na1 质量当前月总累计量Previous Hour n-acc 前一小时总累计量页面+Vb.Ph.na1工况前一小时累积量+Vn.Ph.na1标况前一小时累积量+E.Ph.na1 能量前一小时总累计量+M.Ph.na1 质量前一小时总累计量Previous Day n-acc 前一天总累计量页面+Vb.Pd.na1工况前一天累积量+Vn.Pd.na1标况前一天累积量+E.Pd.na1 能量前一天总累计量+M.Pd.na1 质量前一天总累计量Previous Month n-acc 前一月总累计量页面+Vb.Pm.na1工况前一月累积量+Vn.Pm.na1标况前一月累积量+E.Pm.na1 能量前一月总累计量+M.Pm.na1 质量前一月总累计量Line Condition>>>GeneralLine Condition工况QLine.1工况瞬时流量Qn.1标况瞬时流量ed.1压力在用值ed.1温度在用值Calculations 压缩因子计算Zn.1标况压缩因子Z.1工况压缩因子K.1压缩系数CFV.1放大系数VOS Diagnostic 声速核查Spd of Sound.1 测量声速dSOS.1 理论声速Gas vel.1 气体流速Line Condition>>>Stream1Hourly Flow Rates小时瞬时流量QLine.1工况瞬时流量Qn.1标况瞬时流量qE.1能量瞬时流量qM.1质量瞬时流量Daily Flow Rates瞬时流量以天计Flow Rate Per Second瞬时流量以秒计Presure压力Pres.sens1.1压力变送器数值Pres.keypad.1压力替代值ed.1压力在用值Temperature温度Temp.sens1.1温度变送器数值Temp.keypad.1温度替代值ed.1温度在用值Z Factor 压缩因子计算Zn.1标况压缩因子Z.1工况压缩因子K.1压缩系数CFV.1放大系数Ultrasonic Data超声波数据Num Path.1超声波通道数4个通道Samp.Rate.1采样率(10~15正常)Spd of Sound.1测量声速Gas vel.1气体流速Meter Type.1流量计型号Ultrasonic Path 1超声波通道1 Valsamples1.1 采样率(10~15正常)Efficiency 1.1效率因子(80%以上)Cpp 1.1测量声速Vpp 1.1气体流速Agclevel_A 1.1探头A电压增益Agclevel_B 1.1探头B电压增益Agclimit_A 1.1探头A电压增益上限Agclimit_B 1.1探头B电压增益上限Ultrasonic Path 2超声波通道2以下以此类推Gas Data>>>Preset 预设值、替代值Received接收值Used 在用值Edit 编辑模式—密码11111、改组分替代值Edit—1111—Stream1—Gas Dataa上下键挪动光标到要改的组分上,并按回车b输入所改的组分值,并按回车c重复a和b直到把所有组分更改完毕d按F4键,选择Exit,选择Exit & Save,完毕。
FC2000-IA流量计算机产品样本北京博思达新世纪测控技术有限公司概述FC2000-IA流量计算机是一款对多个流量测量点集中进行数据采集、高精度补偿运算、数据显示存储以及运用网络实现通讯功能的新一代以工业586计算机为内核的计量仪表,它的问世填补了我国流量测量领域的一项空白,可广泛应用于石油、石化、化工、冶金、电力、轻工、医药及城市燃气、供热等行业的贸易计量和工厂计量管理网络。
FC2000-IA依据有关国际标准与建议、国家与行业标准,针对不同介质和流量计类型建立了多种数学模型和相应计算软件。
一台流量计算机可同时完成多种补偿运算,接入的补偿量可多达5个(温度、压力、湿度、密度、组分等)。
对节流式流量计的流出系数C、流速可膨胀系数ε、压缩系数Z等参数作为动态量进行实时逐点运算。
本机所使用的流量计算软件已通过国家权威部门认证。
FC2000-IA流量计算机在天然气测量上,天然气的压缩因子的计算上严格按《GB/T17747.1-1999 天然气压缩因子的计算,第1部分:导论和指南》,《GB/T17747.2-1999天然气压缩因子计算 第2部分:用摩尔组成进行计算》(又称为AGA8-92DC计算方法)及《GB/T17747.3-1999天然气压缩因子计算 第3部分:用物性值进行计算》(又称为SGERG-88计算方法)进行计算。
经济发达国家普遍采用能量计量机制,SY/T6143-2004也提出了遵照GB/T11062-1998计算天然气的发热量,因此FC2000-1A增加了能量的计算。
FC2000-IA具有强大的通讯功能,对现场仪表除可适配4~20mA信号、脉冲信号外,还可以适配HART、Modbus等数字信号;对上位机可采用包括程控电话网、以太网等方式构成计算机网络应用系统,实现远程监督管理和建立集散式计量管理系统。
FC2000-IA特有的历史事件记忆、历史事件存储、双重口令限制、监控现场仪表的误操作和人为的非法操作(如仪表断电、非法修改参数设置、对现场仪表进行断路和短路)等管理功能使之成为能源计量管理与贸易仲裁的理想工具。
流量计算机报警列表非责任报警列表:NON-ACCOUNTABLE ALARMS名称实例警告示意q1 low工况流量低于已设定的低流量限值. The uncorrected flow rate is below the value of Lo qN.qN lowpres1.1 High压力值高于压力警告上限. The pressure used is above the values of PressureN High.presN Highpres1.1 Low压力值低于压力警告下限. The pressure used is below the values of PressureN Low.presN Lowtemp1.1 High温度值高于温度警告上限. The temperature used is above the values of TemperatureN High.tempN Hightemp1.1 Low温度值低于温度警告下限. The temperature used is below the values of TemperatureN Low.tempN Lowturb1 lf叶轮故障(针对涡轮表-低频). An lf turbine meter alarm (blade failure) has occurred.turbN lfus1 eff.1%第1声道效率因数低于预设值. Meter path 1 efficiency is below pre-set % limit.usN eff.1%us1 eff.2%第2声道效率因数低于预设值. Meter path 2 efficiency is below pre-set % limit.usN eff.2%us1 eff.3%第3声道效率因数低于预设值. Meter path 3 efficiency is below pre-set % limit.usN eff.3%us1 eff.4%第4声道效率因数低于预设值. Meter path 4 efficiency is below pre-set % limit.usN eff.4%us1 eff.5%第5声道效率因数低于预设值. Meter path 5 efficiency is below pre-set % limit.usN eff.5%责任报警列表:ACCOUNTABLE ALARMS名称实例警告示意presN max pres1.1 max压力值高于压力报警上限. Pressure sensor 1 is above the values of pressureN max.presN min pres1.1 min压力值低于压力报警下限. Pressure of sensor 1 is below the values of pressureN min.pr1.N unit pr1.1 unit正在使用的压力变送器单位错误. The selected pressure units do not match the units indicated by pressure sensor 1.presN hart pres1.1 hart压力变送器故障或通信失败.A communications failure has occurred between the Model 2000 and pressure sensor 1.tempN max temp1.1 max温度值高于温度报警上限. Temperature sensor 1 is above the values of temp.N max.tempN min temp1.1 min温度值低于温度报警下限. Temperature sensor 1 is below the values of temp N min.te1.N unit te1.1 unit 正在使用的温度变送器单位错误. The selected temperature units do not match the units indicated by temperature sensor 1.tempN hart temp1.1 hart温度变送器故障或通信失败.A communications failure has occurred between the Model 2000 and temperature sensor 1.qN high q1 high 工况流量高于已设定的高流量限值.(1 min) The uncorrected flow rate is above the value of Hi qN. For more than 1 minute.znN high zn1 high基准压缩因子超过高限.Base compressibility outside tolerance znN high.znN low zn1 low基准压缩因子低于低限. Base compressibility outside tolerance znN low.znN calc zn1 calc基准压缩因子目前无法计算. Base Compressibility cannot be calculated.zN calc z1 calc相对压缩因子目前无法计算. Compressibility cannot be calculated.Calibrate Calibrate 系统目前工作在标定模式下.Unit is in the calibrate mode.turbN hf turb1 hf 叶轮故障(针对涡轮表-高频).An hf turbine meter alarm (blade failure) has occurred.usN paths us1 paths 声道号定义错误.(多为流量计中有个别声道停止工作)Wrong number of acoustic paths specified.usN secur us1 secur 流量计安全报警位打开.(流量计处于可编辑状态)Meter security alarm bit set.usN level1 us1 level1 精度降低报警.Reduced accuracy alarm.usN commsus1 comms流量计与计算机间通讯故障.(停止通信5s 以上)No valid communications from the meter for last 5 seconds.tempN alrm temp1 alrm温度低于t-alarm 值,且工况流量超过1小时. Temperature below t-alarm and qb above Lo.q for more than a preset time interval.警告码列表:WARNING CODES以下列项中X表示插板槽的编号1,2,3,4,或5.名称实例警告示意Warn ipXWarn ip1插槽 X 上丢失输入信号.Missing Input board in user position X.Warn opXWarn op1插槽 X 上丢失输出信号.Missing Output board in user position X.Warn comXWarn com1插槽 X 上丢失通讯信号.Missing Comms. board in user position X.Warn brdXWarn brd1错误的插件安装在插槽X 上.Wrong board type in user position X.CalDefaultCalDefault系统在使用默认的校准数据.Unit is currently using Default Calibration data.op1 brdXop1 brd1插槽X 上的脉冲输出1超出预设范围.Pulse Output 1 on Output Board in position X is Overrange.op2 brdXop2 brd1插槽X 上的脉冲输出2超出预设范围.Pulse Output 2 on Output Board in position X isOverrange.op3 brdXop3 brd1插槽X 上的脉冲输出3超出预设范围.Pulse Output 3 on Output Board in position X is Overrange.op4 brdXop4 brd1插槽X 上的脉冲输出4超出预设范围.Pulse Output 4 on Output Board in position X is Overrange.op5 brdXop5 brd1插槽X 上的脉冲输出5超出预设范围.Pulse Output 5 on Output Board in position X is Overrange.op6 brdXop6 brd1插槽X 上的脉冲输出6超出预设范围.Pulse Output 6 on Output Board in position X is Overrange.op7 brdXop7 brd1插槽X 上的脉冲输出7超出预设范围.Pulse Output 7 on Output Board in position X is Overrange.op8 brdXop8 brd1插槽X 上的脉冲输出8超出预设范围.Pulse Output 8 on Output Board in position X is Overrange.op9 brdXop9 brd1插槽X 上的脉冲输出9超出预设范围.Pulse Output 9 on Output Board in position X is Overrange.op10 brdXop10 brd1插槽X 上的脉冲输出10超出预设范围.Pulse Output 10 on Output Board in position X is Overrange.op11 brdXop11 brd1插槽X 上的脉冲输出11超出预设范围.Pulse Output 11 on Output Board in position X is Overrange.op12 brdXop12 brd1插槽X 上的脉冲输出12超出预设范围.Pulse Output 12 on Output Board in position X is Overrange.故障码列表:FAULT CODES以上列项中X表示插板槽的编号1,2,3,4,或5.名称实例警告示意Fault ipXFault ip1插槽X 故障或丢失输入板.Faulty or missing Input board in user position X.Fault opX Fault op1 插槽 X 故障或丢失输出板.Faulty or missing Output board in user position X.Fault comXFault com1插槽X 故障或丢失通讯板.Faulty or missing Communication board in user position X.Fault brdX Fault brd1 插槽 正在使用错误的功能板.Wrong board type in user position X.Fault datFault dat没有或者在用不完整的组态设置数据.No or incomplete data for operation.另附: FC2000面板LED 状态灯 Indicator LED’s on the Front panel电源指示灯,常绿;责任报警灯,常红,即便现场工作回复正常,此灯也只能在确认报警后才能消失;非责任报警灯,常红,若现场工作回复正常,此灯自动熄灭;故障灯,常红;低流量(管线停输)或高流量(流量持续超过流量上限1分钟)报警灯,常黄.。
一、控制面板电源接通绿色责任报警红色非责任报警红色错误操作红色流量上下限报警黄色上翻键用来把菜单上的高亮条向上移动到所需的数据行上.下翻键用来把菜单上的高亮条向下移动到所需的数据行上返回或者回车键2.2.1.3 信息键信息键用于显示附加的屏幕帮助或者是信息。
任何时候都可以按下这个键,帮助和信息都会以文本框的形式显示出来。
指数键在允许的地方,指数键将用来表示指数类型的一些数字。
二、主菜单 F4总量(TOTALS)管线工况(LINE CONDITIONS)气体数据(GAS DATA)设置(SETTING )预设置数据(PRESET DATA)编辑(EDIT)报警(ALARMS)显示(DISPLAY)系统(SYSTEM)一般信息(GENERAL INFO)1 总量 totalsVb --未修正的体积总量(工况)Vb-c—未对流量计的非线性进行过修正的体积总量Vn—修正后的体积量(标况)M—质量流量E—能量流量次级菜单:主总量(Main Totals)报警总量(Alarm Totals)当前一小时(日、月)累计总量:(Current Hour, Day and Month accumulated Totals)●前一小时(日、月)累计总量(Previous Hour, Day and Month accumulated Totals)当前一小时(日、月)非累计总量(Current Hour, Day and Month non-accumulated Totals)●前一小时(日、月)非累计总量(Previous Hour, Day and Month non-accumulated Totals)当前一小时(日、月)报警累计总量(Current Hour, Day and Month accumulated Alarms Totals)前一小时(日、月)报警累计总量(Previous Hour, Day and Month accumulated Alarms Totals)当前一小时(日、月)报警非累计总量(Current Hour, Day and Month non-accumulated Alarms Totals)前一小时(日、月)报警非累计总量(Previous Hour, Day and Month non-accumulated Alarms Totals)2 管线工况(Line Condition)这一项是只读的显示项,一旦被选中将会显示所有的在这台机器里的管线工况,包括:流量波峰流量和当时发生的时间和日期管线压力和温度管线工况和基准状况下的压缩因子修正系数流量计信息这些条目将分成数页来进行显示,每一页最多只显示 4 条,相关参数Line coudition1、hourly flowe rates 每小时流量 1/32qn.1=_m3/hr2、daily flowe rates 每日流量 3/32qn.1=_m3/day3、flowe rates per second 每秒流量 5/32qn.1=_m3/sec4、pesk flow rates 波峰流量 7/325、Pressure 压力 10/326、pressure averages 压力平均值 13/327、corrected pressure 修正压力 14/328、temperature 温度 15/329、pressure temperature 平均温度 18/3210、z.factor 压缩因子 20/323 主菜单功能项::气体数据(Gas Data)这一项是只读的显示项,一旦被选中将会显示所有的在这台机器里的气体数据,包括:相对密度高位和低位热值按摩尔百分数表示的气体成分这些条目将分成数页来进行显示,每一页最多只显示4 条,并附有适当的符号刻度因子、和单位。
FC2000型号流量计算机调试指南一、首先使用FC2000调试数据线连接电脑和FC2000。
打开安装好的FC2000调试软件,出现如下截图。
输入用户名:user,密码:password。
点击OK按钮。
二、进入程序后点击左上角新建图标(红框圈出),弹出如下一个新窗口。
在新弹出的窗口中起个名字,下拉菜单选择1 Stream Ultrasonic,然后点击OK。
注意:通常情况下该软件会自动识别Comms Port,如未能识别请到设备管理器中查询。
如仍未能找到USB-Serial Port相关内容,请再次确认数据线是否连接好,然后重装安装盘内的驱动。
三、选中刚才新建的连接,然后点击左上角打开按钮(红框圈出)。
然后弹出选择文件的对话框,选中对应的配置文件,点击右下角“打开”按钮。
四、打开后弹出如下对话框,点击OK。
五、根据现场情况配置这台流量计算机。
双击左上角Hardware图标,选中Slot 2 – Input 2 Board,出现温变、压变设置页。
重要:如果温变和压变走两个Hart Loop,相关接线见文末附图1,则该配置不用改动。
如果温变和压变走同一个Hart Loop,即在上诉情况下将温变+和压变+并成一根,温变-和压变-并成一根,则该配置需要做如下改动。
首先设置压变,在图示的框中填1,同时需用手抄器将对应压变的地址改为1.然后设置温变,HART Loop1保持不变,更改Transmitter2,将左侧内容拉入First 的框中,在Address框中填2,同样需用手抄器将对应温变地址改为2。
注意:最后将HART Loop2中的内容清除。
六、左侧主菜单中选择Stream 1-- Ultrasonic,然后在右侧红框中输入与这台流量计算机对应的流量计的最大流量值。
该值在流量计的铭牌上读取。
七、配置文件基本已经配置完毕,下面需要将该配置上载到对应的流量计算机中。
点击左侧Update图标出现进度条,等待进度条读完就完成了上载。