OLYMPUS-奥林巴斯-内镜维修方式介绍-2014年03
- 格式:pptx
- 大小:3.05 MB
- 文档页数:28
Maintenance Manualfor maintenance engineerCX31/41Thank you for purchasing the Olympus microscope CX31/41.In order to fully utilize its performance and secure optimum condition, please read this manual before maintenance work.Please also keep it at hand during maintenance as well as for future reference.All rights reserved, Reproduction in whole or in part without written permission is prohibited.The purpose of this manual is to satisfy any requirements for maintenance material that maybe considered as necessary to professionals in the maintenance field, such as Maintenance engineer in Hospitals.It is intended to be used as a tool for performing basic maintenance procedures if needed or when required as per the recommended maintenance schedule.The sections from this manual only cover procedure pertaining maintenance ’s that are considered to be easily performed. For more involved maintenance’s or repairs, it is recommended that you contact a qualified service engineer from your local Authorized Olympus dealer.Maintenance parts, grease, and other items specified in the manual can be ordered from your Authorized Olympus dealer, and subject to change without notice.The recommended maintenance schedule is shown below as reference. ( * Necessary item)Portion Cleaning Optical/mechanical check Maintenance schedule Optical components1)Outer surfaceEyepiece, objective,condenser lens, filter,photo eyepiece *Once in a year(If dirt is conspicuous or oilimmersion objective is used,cleaning should be made afterevery use.)2)Inner part Prism,internal lenses*Once in a yearAppearance Microscope frame, mechanical part *Once in a year(If dirt is conspicuous, cleaningshould be made after everyuse.)Observation tube*Optical adjustment:1)Optical axis (standard)2)Left/right optical axis3)Revolving axis4)ParfocalityOnce in a yearMechanical part Focusing unit, stage, revolving nosepiece, aperture/field iris diaphragm*Mechanical movement:Grease replacementMechanical adjustmentOnce in two to three yearsCONTENTSCHAPTER 1MAINTENANCE PROCEDURE (1)1.Maintenance of Microscope (1)2.Guide to Maintenance (2)2-1Overview of maintenance (2)2-2Cleaning method for the optical components (3)3.Preparing for Inspection (6)4.CX31/41 Inspection Sheet (7)CHAPTER 2INSPECTION PROCEDURE (8)1.Checking Performance of Microscope (8)2.Checking Dirty Portion (8)2-1Image influence caused by dirt on each component (8)2-2How to find dirty portion through observation (9)2-3How to check cleaning condition (10)CHAPTER 3REPAIR PROCEDURE (11)1.Optical Adjustment (11)2.Mechanical Adjustment (14)2-1Preparation for the tension adjustment of Y-wire (14)2-2Adjustment method for the tension of Y-wire (15)2-3Confirmation of the Y-wire tension (15)2-4Final adjustment (16)3.Replacing Grease for Fine Focus Adjustment Knob Ass’y (17)4.Electrical Adjustment (18)4-1CX31/41 wiring diagram (18)4-2Replacing circuit board / rheostat ass’y (18)4-3Preparation (19)4-4Voltage adjustments (19)CHAPTER 4JIGS AND TOOLS / GREASES AND ADHESIVES (20)1.List of Jigs and Tools (20)2.List of Greases (20)3.List of Adhesives (20)CHAPTER 5MAINTENANCE PARTS (21)1.List of Maintenance Parts (21)1.Maintenance of Microscope1)Fundamental handlinga.Read the instruction manual thoroughly, handle the microscope correctly.b.Be sure to make a usual cleaning, especially after every use of microscope.c.When handling the microscope, do not expose it to shock, moisture, heat and dust.d.If the problem occurs, do not treat it in self-judgement.2)Using and storing conditionsa.The microscope should be used under no vibration.If it is placed in such environment as vibration, this causes disturbance and fatigue inobservation and affects the photomicrography.b.It should not be stored in high humidity.Such condition causes fungus, corrosion on lens and rust on metal part.Therefore, special caution is exercised if stored in a long period of time.c.Intense temperature change should be avoided.Be careful not to place the microscope near window exposed to direct sunlight and airconditioner. When bringing it into warm room from cold storage location in winter,condensation occurs on lens and metal part, it causes fungus and rust.d.It should be kept clean.The microscope is required to keep away from dust that causes contamination, funguson lens and failure of sliding part in the frame.Be careful not to place it in the environment where the corrodent chemicals such ashydrogen sulfide, hydrogen fluoride and acid are handled.<Reference> fungus: occurrence conditions2.Guide to Maintenance2-1Overview of maintenance2-2Cleaning method for the optical componentsImportant notes:1)Never rub the lens surface strongly.2)Do not use the same lens tissue to clean more than one lens .3)Do not moisten the lens tissue with an excessive amount of cleaning solution.4)When cleaning with tweezers, be careful not to protrude its tip from the lens tissue.3.Preparing for Inspection4.CX31/41 Inspection SheetModel :Check Date :S/N :Checking by:Check Point Check Contents Result Ref. Page 1.Electrical unit1)When the power switch is turned on, the lamp is litand the brightness can be varied by adjustmentknob.OK / NO18, 191)The coarse/fine focus adjustment knob issmoothly moved without any stress orunevenness.OK / NO172)The tension of coarse focus adjustment knob canbe adjusted by the adjustment ring.OK / NO2.Coarse/fine focusadjustment knob3)CX31:The upper limit is changed by turning thering of stopper on the front upper side.CX41:The coarse upper limit lock is effective.OK / NO1)The stage should not fall spontaneously.OK / NO2) A specimen is held securely by the specimenholder.OK / NO3.Stage3)The X/Y movement is smooth withoutunevenness, backlash or slipping.OK / NO15, 161)The interpupillary distance adjustment can beoperated smoothly in working range.OK / NO2)When changing the interpupillary distance, thedisplacement of optical axis is not apparent.OK / NO3)The diopter adjustment ring is moved smoothly inworking range.OK / NO4)The optical axis of left side coincides with that ofright side.OK / NO11, 12, 135)CX41:The tilting angle is easily adjusted and itsangle is stable. (when the U-TBI3 is used:)OK / NO4.Observation tube6)CX41:When changing the tilting angle, theoptical axis displacement is not apparent.(when the U-TBI3 is used:)OK / NO5.Revolvingnosepiece 1)The revolving nosepiece can be rotated smoothlyand stops at the click position.OK / NO 1) The vertical movement of condenser is smooth.OK / NO6.Condenser2)The centering of field iris diaphragm can beadjusted with the centering knobs of attachmentlens(CX-AL).OK / NO7.Illumination1)The field/aperture iris diaphragm ring is movednormally.OK / NO1)Observation image is normal.Without flares /ghosts / vignetting /unevenilluminationOK / NO2)When photographed, there is no unevenness orshading on the film surface.OK / NO8.Visibility(Observation)3)Dust and dirt are not noticeable in observation orphotomicrography.OK / NO8, 9 Remarks:1.Checking Performance of MicroscopeUsing the CX31/41 inspection sheet (P.7), check the electrical unit, mechanical and optical performance.2.Checking Dirty Portion2-1Image influence caused by dirt on each componentThe following figure shows the influence of image on each optical component if stains or dust is adhered to that portion.In general, the microscope image or photographing is largely affected by dirt adhered on the nearer portion to a specimen and image surfaces.Therefore, the optical components should be kept clean and dust-free.A:Dirt is clearly seen.B:Blurred contours of dirt is seen.C:Dirt is seen when the aperture iris diaphragm is stopped down.D:Dirt is not directly seen, but contrast of image deteriorates2-2How to find dirty portion through observation1)Close the aperture iris diaphragm.(When the aperture iris diaphragm is closed, it facilitates finding the dirt particlesbecause the depth of focus increases and the dirt position bring into focus. However, very small dirt particle may not be found in this method.)2)Observe a specimen through the eyepiece.If dirt is seen by observing it, look for the portion where stains or dust is adhered bymoving the following components as well as a specimen.For TV camera, check it by method shown below.Note:If dirt particles do not move by moving the above components, it is assumed that internal lenses are contaminated.In this case, please contact your Authorized Olympus dealer.2-3How to check cleaning condition1)When a large lens is checked, look at the lens while putting it toward bright side orbreathe on the lens and observe the condition that the haze on the whole surface ofthe lens disappears evenly.Dust becomes conspicuous when looking at it with the lens inclination changed.If there is a dirty part or a remained part that is not cleaned completely, the haze of this part will disappear slower than that of the other part.2)For a small lens such as top lens of objective, check it by magnifier.An eyepiece can be substituted formagnifier by turning the eyepiece upside down.1.Optical Adjustment*If this adjustment could not be done, inside mechanism may be damaged. Please contact your Authorized dealer.2.Mechanical Adjustment2-1Preparation for the tension adjustment of Y-wireIf a specimen image is moved when the stage is brought into the desired position of specimen, it is necessary to adjust the wire tension of stage.2-2Adjustment method for the tension of Y-wire1)Loosen the screws (*1) securing the holder.Screws:CUK 3X6SA (*1)2pcs. (clamping)2)Adjust the tension of the wire by turning thescrew (*2) clockwise to tight or counter-clockwise to release.Screw:CSK3X6SA (*2)1pc.(adjustment)3)Temporarily tighten the screws (*1).2-3Confirmation of the Y-wire tension1)Check that the Y-knob (A) is rotated following the movement of lower stage (B) whenmoving the lower stage in stroke by a hand.2)After confirming the above condition, Install the stage to the frame•C set themicroscope at observation state and check that the image is brought into the desired position without backlash (within 2 microns). If it is out of standard, perform theadjustment referring to 2-4 on page16.3)Repeat the above procedures until the condition is satisfied.After adjustment, tighten the screws (*1) firmly and apply adhesive to three screws(*1, 2).Adhesive: OT1378 (Solvent-based adhesive)Note:1)If the wire is too tight, image backlash may occur. If the wire is loose, slip may occurbetween the knob and wire deteriorating the image movement.2)In case where the stage movement is heavy due to hardening of grease or the X- wireadjustment is necessary, please contact your Authorized Olympus dealer because disassembling the stage is required for grease replacement or that adjustment.2-4Final adjustment(B)Image backlash adjustment:1)Under observation state (with 100X objec-tive), move the stage to the desired imageposition by turning the Y-knob (A).At that stop position, check image backlash.If it is over 2 microns, conduct the followingadjustment.2)When adjusting the Y-movement, loosen thetwo screws (*1) and turn the Y-knob (A) tobring backlash within 2 microns.*After turning the knob and temporarilytighten the screws, check image backlash inthe observation state. Repeat theadjustment until image backlash is within thestandard.Screws:AWU3X4SA (*1)2pcs.3)For the X-movement, check that the backlashcan be adjusted by turning the X-knob (B) inthe same manner as the Y-knob adjustment.*If it could not be adjusted, contact yourAuthorized dealer because X-wire need tobe adjusted.Screws:AWU3X4SA (*2)2pcs.3.Replacing Grease for Fine Focus Adjustment Knob Ass’yIf the fine focus adjustment knob is not turned smoothly, replace grease on the shaft of fine focus adjustment ass’y in the following procedure.(In case where the coarse focus adjustment knob is not turned evenly, please contact your Authorized Olympus dealer because it is necessary to disassemble the coarse focus adjustment knob ass’y and/or guide unit.)1)Peel off the plates on the left and right sides of fine focus adjustment knob.Insert Allen wrenches (2.5mm) into the both side screws.Screws:ABS3X8SA2pcs. (*1)2)Turn the screw (*1) on the right side with Allen wrench while the left side screw is held withit, and pull out the fine focus adjustment knob ass’y.(At this time, the fine focus adjustment knob ass’y (right ) is removed as an assembly of shaft (a), gear (b) and fine focus adjustment knob (c).)3)Confirm that the gear is not damaged. If it is damaged, replace the gear or the fine focusadjustment knob ass’y as a whole.4)Remove grease (OT2008) on the shaft by cleaning solution and replace it.*When applying the grease, refer to the explanation shown below.5)Assemble the components in reverse order of disassembly.Fine focus adjustment knob ass'yA004*1A002 (a)*Apply grease(OT2008) to the shaft.Adhesive(OT1378) has been applied to the screw thread of the shaft on the right side.A005 (c)A003 (b)Plate A001H Applying grease (OT2008) to the shaft (a) of the fine focus adj.knob ass’y and the sliding surface:Apply grease to the outer surface of the shaft. First, spread the grease by inserting from the left while rotating the fine focus adj. knob ass’y. After removing the fine adj. knob ass’y, place the shaft at the opposite side mouth (to the right). Again apply grease to the shaft and then insert into the right side. Use the lens tissue to wipe away any grease exuding from the left and right.4.Electrical Adjustment4-1CX31/41 wiring diagramH Voltage AdjustmentThe circuit board ass’y (AQ802400) consists of the circuit board (DZ290300) and rheostat ass’y (DZ290400). In case where the above parts are replaced as AQ802400, the voltage adjustments for the circuit board ass’y are not necessary. However, voltage adjustments are necessary when individually replacing either the circuit board (DZ290300) or rheostat ass’y (DZ290400). The following explains procedures for replacing, preparation and adjustment of the minimum and maximum voltages.4-2Replacing circuit board / rheostat ass’yCUK3X4SA(*1) 5 pcs.CUKHWB3X6SA(*2) 1 pc.(No need toCUK3X4SA(*3) 3 pcs.CUKHWB3X6SA(*4) 1 pc.5)Replace the CIRCUIT BOARD (B) /RHEOSTAT ASS’Y (D).(Rheostat ass’y (D): Please contact yourAuthorized dealer.)4-3Preparation1)Connect the cables to each connector(CN1-CN3).2)Connect the cables to the power switchinlet terminals.Upper side: black Lower side: white3)Set the digital multimeter so that thevoltage of the CN2 2-pin can be measured.H Use a new Halogen bulb when makingvoltage adjustment4-4Voltage adjustmentsMinimum Voltage Adjustment1)Turn ON the power. Turn the light intensity controlknob (A) counterclockwise to lower the lampbrightness to its lowest level.2)Rotate the circuit board’s trimmer VR21 to adjustso that the lamp output voltage between the CN2’s1 and2 pins is within the standard shown belowusing a digital multimeter.Standard:DC1.10~1.15V(adjustment target: 1.123V)Maximum Voltage Adjustment1)Turn ON the power. Turn the light intensity controlknob (A) clockwise to increase the lamp brightnessto its highest level.2)Rotate the circuit board’s trimmer VR23 to adjustso that the lamp output voltage between the CN2’s1 and2 pins is within the standard shown belowusing a digital multimeter.*Do not touch the inlet terminals withpower turned on because it could causean electric shock.Be sure to turn off power duringassembly and reassembly.*Do not turn the trimmer VR22 because itis adjusted to the prescribed currentvalue.(VR22 is set for overcurrent protection.)Standard:DC 5.65~5.75V(adjustment target: 5.70V)CX31/41 MAINTENANCE MANUAL JIGS AND TOOLS / GREASES AND ADHESIVES1.List of Jigs and ToolsNo.Description Ref. page Cleaning tools2, 3, 4, 511, 12, 13 *1WHB10X(It is necessary to set the cross micrometer disk.)Digital multimeter19Philips screwdriver12, 13, 14, 15, 18Allen screwdriver (2.5mm)14Allen wrench (2mm, 2.5mm)16, 17Precision screwdriver19Dust cover*1 Contact your Authorized dealer.2.List of GreasesNo.Description Ref. pageOT2008Grease (medium)173.List of AdhesivesNo.Description Ref. pageOT1378Solvent-based adhesive (transparent)15, 17(For the fine focus adjustment knob ass’y, use adhesiveonly when it is disassembled.)CX31/41 MAINTENANCE MANUAL MAINTENANCE PARTS1.List of Maintenance PartsIndex No.Order No.Description Ref. page*1CX31LBSF Instruction manual*1CX41 Instruction manual17 A001AD071500PlateA002AC732500ShaftA003AA782600GearA004AQ153900Fine focus adjustment knob ass’y(when replaced as an assembly:)A005AB467500Fine focus adjustment knob (right)A006AB059300Fine focus adjustment knob (left)A007AD411200Holder (for Y-wire)1516 A008AC888300X-knobA009AC888400Y-knob18 A010DZ290300Circuit boardA011DZ290400Rheostat ass’yA012AQ802400Circuit board ass’y(when replaced as a set:)*16V30W halogen bulb19H The index No. are shown on the figures in “REPAIR PROCEDURE” instead of parts No.*1 Contact your Authorized dealer.OLYMPUS OPTICAL CO., LTD1-22-2, Nishi shinjuku shinjuku-ku, Tokyo, JapanIssued by Marketing Dept.Printed in Japan 2001 11 SK9214。
OLYMPUS AMERICA INC.3500 Corporate ParkwayP.O. Box 610November 12, 2012Re: Revised GI Endoscope Reprocessing ManualDear Health Care Professional:Olympus is pleased to announce validated improvements in endoscope reprocessing and the availability of a revised reprocessing manual for the following GI endoscopes:• GIF-N180• GIF-Q180• GIF-H180• GIF-XP180N• GIF-H180J • GIF-2TH180 • CF-Q180AL/I • CF-H180AL/I • CF-H180DL/I • PCF-Q180AL/I • PCF-H180AL/I • TJV-Q180V In an effort to improve reprocessing efficiency, Olympus has validated the use of water to perform endoscope precleaning at the patient’s bedside. This eliminates the need to prepare and hold detergent in the procedure room. Other improvements in the reprocessing procedure include reductions in the quantity of fluids required to flush the auxiliary water channel during bedside precleaning. In addition, the Olympus Flushing Pump,commonly known as the OFP, has been validated for automated flushing of the auxiliary water channel as an alternative to a manual syringe flushing.These new procedures, plus improvements in illustrations and additional informational material, are contained in the revised GI endoscope reprocessing manual now available from Olympus. The following is a brief description of the revisions that you will find in the revised reprocessing manual.• The precleaning procedure was revised to eliminate the use of detergent in the procedure room and toprovide the option for automated flushing of the auxiliary water channel using the OFP, Olympus Flushing Pump. The revised instructions specify the use of water to perform precleaning. Please note that thevalidation of the change to precleaning with water for the auxiliary water channel was performed with the Olympus MAJ-855 only.• Reprocessing accessories are no longer required to be detached after detergent cleaning and after rinsingprior to high-level disinfection.• Work flow charts were added to the manual to identify:o Equipment used only for patient procedures, o Equipment used only for reprocessing,o Equipment that has a role in both patient procedures and reprocessing, o Where to find instructions for reprocessing the equipment.• New illustrations were added to further clarify reprocessing instructions, and additional detail was added toexisting diagrams.• Additional instructions were added regarding:o The reprocessing of accessories.o The storage of sterilized endoscopes and accessories.• Finally, additional warnings and cautions were added to identify potential pitfalls in reprocessing practicesand policies.One caution/warning statement in the revised manual that we would like to call to your attention addresses the issue of equipment repair provided by a non-Olympus repair facility or Independent Service Organization (ISO). Olympus does not support or authorize ISOs to repair Olympus equipment, nor does Olympus provide training or parts to ISOs to enable them to repair Olympus equipment to Olympus’ standards. As a result, Olympus has no knowledge of the specifications of materials used by ISOs to repair Olympus equipment, nor the quality of such repairs. Olympus conducts a rigorous testing program to validate both compatibility of our instruments with specific reprocessing methods and agents, as well as the efficacy of such methods for reprocessing the device. Without knowledge of the materials used or the final quality of repair being provided by ISOs, Olympus is unable to validate the compatibility or reprocessing efficacy of the Olympus instructions provided in the product-specific reprocessing manual. As a result, the instructions provided in the product-specific instruction manual are invalid for Olympus equipment that has been repaired by an ISO. If your Olympus equipment was or is being repaired by a non-Olympus facility, please contact the repair facility to obtain instructions on device compatibility and reprocessing to ensure that your facility is following safe and validated reprocessing instructions.Although the new reprocessing instructions describe bedside precleaning with water, the former precleaning procedure using detergent remains a validated process and facilities may continue this practice if they so choose. In addition, the revised instructions are applicable to reprocessing the older model endoscopes that are listed in the table below. If you have questions regarding how to apply these new instructions to the older model Olympus endoscopes that you own, please contact Olympus.†Use channel cleaning brush specified in product-specific Reprocessing Manual.OLYMPUS AMERICA INC.3500 Corporate ParkwayP.O. Box 610OLYMPUS AMERICA INC.3500 Corporate ParkwayP.O. Box 610An electronic version of the newly revised manual is available on the Olympus Connect customer portal(). Hard copies may be obtained by contacting Olympus Customer Service. If you have any questions regarding the revised reprocessing manual, please contact your local Olympus sales representative or the Olympus Technical Assistance Center. To contact the Olympus Customer Service and Technical Assistance Center, please call 1-800-848-9024. Thank you.Sincerely,Mary Ann Drosnock, MS, CICAssociate Manager of Infection Control Clinical Affairs Department Olympus America, Inc.****************************。
November 3, 2014RE: Guidelines for Returning Olympus Endoscopes and Electronic Equipment for Repair in the United StatesDear Healthcare Practitioner:The purpose of this letter is to provide guidance to customers on our requirements when sending Olympus endoscopy equipment to Olympus service centers located in the United States for repair. All equipment, including flexible endoscopes, rigid endoscopes, and electronic equipment must be rendered safe to handle prior to being returned to Olympus for repair. To accomplish this, all endoscopes must be cleaned AND high-level disinfected or sterilized prior to shipment to a service center for repair.Olympus Electronic Equipment must be wiped with an appropriate surface disinfectant, such as Sani-Cloth® Plus Germicidal Disposable Cloths manufactured by PDI.In addition, Olympus requires that the cleaning AND high-level disinfection or sterilization process be verified prior to initiation of repair. To expedite repair of your equipment, please contact Olympus at 800-848-9024 to generate a service RMA. Please complete the Reprocessing Verification section on the equipment service request form provided. You may also generate an RMA and complete the Reprocessing Verification on-line at/serviceportal.Please include the equipment service request form or portal confirmation with your equipment. If Olympus equipment is received without a Reprocessing Verification, Olympus will contact the customer to verify proper reprocessing prior to initiating repair. Endoscopes that have not been cleaned AND high-level disinfected or sterilized will be returned to the customer without being repaired.Mandatory Surface Disinfection of Electronic EquipmentPrior to returning Olympus electronic equipment for repair, wipe all surfaces of the electronic equipment (e.g., the video system center or light source) with a disinfectant wipe, such as Sani-Cloth® Plus Germicidal Disposable Cloths manufactured by PDI. Follow the product’s instructions for use regarding how to wipe and the required contact time for the product to be effective. For example, Sani-Cloth® Plus Germicidal Disposable Cloths require a 5-minute contact time.Mandatory Cleaning and Disinfection or Sterilization of Rigid Endoscopes:Prior to sending rigid endoscopes to Olympus for repair, reprocess them according to the product-specific instruction manual and/or Endoscopy System Guide (if applicable), which includes the steps of cleaning, high-level disinfection and/or sterilization.Mandatory Cleaning and Disinfection or Sterilization of Flexible Endoscopes:For flexible endoscopes, refer to the reprocessing manual that accompanies each endoscope for instructions on proper manual cleaning, high level disinfection and/or sterilization, which is required prior to returning the equipment to Olympus for repair.The following information provides guidance on how to reprocess damaged flexible endoscopes to render them safe to handle. Damaged endoscopes that are still watertight should be reprocessed according to instructions provided in the Olympus reprocessing manual. However, if the endoscope has developed a leak, routine reprocessing may lead to further damage. Therefore, Olympus has developed special guidelines for flexible endoscopes that have failed leakage testing.The key to preventing further damage to a leaking endoscope is to provide positive pressure to the endoscope during all phases of the reprocessing cycle by connecting the endoscope to the maintenance unit (Olympus model MU-1) or light source. This will ensure that the endoscope is pressurized throughout the entire reprocessing cycle and will help to prevent fluid from entering the endoscope.A leak in the endoscope will be indicated by a continuous series of bubbles emerging from a location on the endoscope. Before removing the endoscope from the water, identify and make note of the location of the leak. With the maintenance unit or light source still turned on and the leakage tester still connected, remove the endoscope from the water. If the maintenance unit or light source is turned off while the endoscope is immersed, water may invade the internal spaces and further damage the endoscope.Mandatory Manual Cleaning for Leaking Endoscopes1. For a leak detected in the covering of the insertion tube, bending section, or universalcord, turn the maintenance unit OFF. Detach the leakage tester from the maintenance unit or the light source. Wait 30 seconds or until the covering of the bending section contracts to its pre-expansion size. Detach the leakage tester from the endoscope. Thoroughly drythe identified location of the leak on the outer area of the scope using alcohol and a clean lint-free cloth. Carefully apply a piece of electrical tape over the location of the leak prior to immersing the endoscope in detergent solution. Wrapping the tape too tightly mayresult in damage to the endoscope. For leaks detected in other locations (e.g., internalchannel), proceed with the instructions provided below.Note: Please be advised that Olympus recommends use of a colored tape, such as red,yellow, or blue to allow for easier visual identification of a scope in need of repair afterleak testing and cleaning. It is further recommended that customers develop apath/segregation technique for endoscopes requiring repair to minimize the possibility of reuse prior to being returned to Olympus for repair.2. Fill a basin with detergent solution at the temperature and concentration recommendedby the detergent manufacturer. Use a basin that is at least 40 cm by 40 cm (16” by 16”) in size and deep enough to allow the endoscope to be completely immersed.3. Insert the leakage tester connector into the output socket of the maintenance unit or thelight source and turn the maintenance unit or the light source ON. Set the light source’sairflow regulator switch to “HIGH” or “3.”4. Connect the leakage tester’s connector cap to the venting connector of the water-resistant cap. For 190-series endoscopes, the leakage test connects directly to the venting connector.5. Fully immerse the endoscope in the detergent solution.6. Perform manual cleaning according to the instructions provided in the reprocessingmanual. Minimize unnecessary flexion of the insertion tube and universal cord duringcleaning.Mandatory High-Level Disinfection or Sterilization for Leaking Endoscopes1.Sterilization:Following manual cleaning, the preferred method of rendering a leaking endoscope safe to handle is ethylene oxide sterilization. Ethylene oxide sterilization should be performedaccording to the instructions provided in the endoscope’s reprocessing manual. If electrical tape was applied to a leak detected in the endoscope’s external surface, remove the tape and wipe with 70% ethyl or isopropyl alcohol prior to ethylene oxide sterilization.If ethylene oxide sterilization is not possible, perform high-level disinfection or STERRADsterilization according to the instructions provided below.Sterilization in a STERRAD sterilization system is also an acceptable alternative practice provided the endoscope is listed as materially compatible with the STERRAD sterilization system. Those endoscope models which are compatible with STERRAD sterilization may be sterilized using the normally recommended STERRAD model, cycle and conditions (e.g., check booster requirements) as long as the following requirements are met: 1) The endoscope has been properly cleaned and dried, and 2) Fluid has not invaded the endoscope. Retained fluid will cause the STERRAD cycle to abort and may contribute to additional repairs.2.Manual High-Level Disinfection:a.Fill a basin with disinfectant solution at the temperature and concentration recommendedby the disinfectant manufacturer. Use a basin that is at least 40 cm by 40 cm (16” by 16”) in size and deep enough to allow the endoscope to be completely immersed.b.Insert the leakage tester connector into the output socket of the maintenance unit or thelight source and turn the maintenance unit or the light source ON. Set the light source’sairflow regulator switch to “HIGH” or “3.”c.Connect the leakage tester’s connector cap to the venting connector of the water-resistantcap (140/160/180-series) or endoscope directly (190-series).d.Immerse the endoscope in the disinfectant solution.e.Perform the high-level disinfection procedure described in the endoscope’s reprocessingmanual. Minimize unnecessary flexion of the insertion tube and universal cord duringreprocessing.3. Automated High-Level DisinfectionAutomated Endoscope Reprocessors (AERs) circulate high-pressure fluid through the internal channels of the endoscope, which may result in fluid invasion and further damage to aleaking endoscope. As a result, Olympus recommends performing the manual high-level disinfection procedure on leaking endoscopes.However, some AERs are designed to maintain positive pressure to the internal cavities of the endoscope in order to prevent fluid invasion during the reprocessing cycle. Check with your AER manufacturer to determine whether your AER is intended to reprocess leakingendoscopes and whether positive pressure can be maintained during the disinfection cycle without aborting the cycle.If you have any additional questions, please contact your local Olympus sales representative or the Olympus Technical Assistance Center at 1-800-848-9024 (United States). Thank you. Sincerely,Mary Ann Drosnock, MS, CIC, CFER, RM (NRCM)Manager of Infection ControlOlympus America Inc.****************************。
olympus内窥镜系统的介绍本文将简要介绍Olympus内窥镜系统以及它在医疗领域中的重要性。
Olympus内窥镜系统是一种先进的医疗设备,用于在人体内部进行内窥镜检查和治疗。
它由日本著名医疗器械制造商___开发和生产。
内窥镜是一种用途广泛的医疗工具,用于检查和诊断人体内部的病变或异常情况。
Olympus内窥镜系统采用了最新的光学和图像处理技术,能够提供清晰、高分辨率的图像,帮助医生准确判断疾病并制定适当的治疗方案。
Olympus内窥镜系统在医疗领域中的重要性不言而喻。
它为医生提供了一种非侵入性的检查和治疗工具,可以避免传统手术的创伤和风险。
通过内窥镜检查,医生可以直观地观察患者体内的病变情况,如消化道、呼吸道、泌尿系统等,从而提供准确的诊断结果。
此外,Olympus内窥镜系统还具有一些独特的功能和特点,增加了其在医疗领域中的价值和应用范围。
例如,它可以进行光谱分析,帮助医生鉴别肿瘤的类型,从而指导治疗方案的选择。
此外,它还支持显微镜功能,使医生能够更加精细地观察和处理病变部位。
总之,Olympus内窥镜系统作为一种先进的医疗设备,在诊断和治疗过程中发挥着重要作用。
它的高分辨率图像、非侵入性检查和诊断能力,以及独特的功能和特点,使其成为医生们不可或缺的工具之一。
Olympus内窥镜系统是一种先进的医疗设备,用于进行内窥镜检查和治疗操作。
它由多个组件组成,每个组件都有特定的功能,以提供高质量的图像和操作性能。
内窥镜内窥镜是Olympus内窥镜系统的核心组件。
它是一种细长的管状工具,具有高分辨率的摄像头和灯光系统。
内窥镜可插入人体内部,通过可视化组织和器官的图像来帮助医生进行诊断和治疗。
光源光源是Olympus内窥镜系统的另一个重要组件。
它提供所需的光线来照亮内窥镜的摄像头和视野,以确保医生能够清晰地观察器官和组织的细节。
Olympus内窥镜系统使用先进的光源技术,以提供高亮度和均匀的光照。
图像处理图像处理是Olympus内窥镜系统的关键功能之一。
奥林巴斯290系列电子内镜的常见故障及预防
关珊珊; 管强; 于亮; 樊宽章
【期刊名称】《《医疗装备》》
【年(卷),期】2018(031)007
【摘要】目的 290系列电子内镜是结构精良、功能复杂、成本较高的仪器设备,其易损坏,维修率较高。
通过故障分析,探讨预防维护方法,降低故障率、减少维修费用。
方法通过研究内镜各组件基本功能、操作方法及洗消保养过程,对故障现象进行分析。
结果通过总结内镜各部件常见故障原因,可制定有针对性的预防性维护方法。
结论完整的故障分析和维修方法,可应对设备的突发故障。
良好的预防性维护可保
证设备处在最佳工作状态中,故障率低,确保临床安全使用。
【总页数】3页(P131-133)
【作者】关珊珊; 管强; 于亮; 樊宽章
【作者单位】青岛大学附属医院医学工程科山东青岛 266000; 青岛大学附属医院综合部山东青岛 266000
【正文语种】中文
【中图分类】TH77
【相关文献】
1.电子内镜常见故障预防及处理 [J], 吴桂萍;夏云
2.奥林巴斯GIF-260系列电子胃镜的故障与预防 [J], 龚希;朱屹
3.奥林巴斯290系列胃肠镜的常见故障及预防 [J], 徐乐;聂静静
4.奥林巴斯290系列胃肠镜的常见故障及预防 [J], 徐乐;聂静静
5.奥林巴斯CV-290电子胃肠镜的常见故障分析和预防及维修实例 [J], 林益凯因版权原因,仅展示原文概要,查看原文内容请购买。
第42卷㊀第6期2023年㊀12月北京生物医学工程BeijingBiomedicalEngineeringVol 42㊀No 6December㊀2023作者单位:江苏省人民医院(南京㊀210029)通信作者:毕欣雨㊂E⁃mail:1500870517@qq com奥林巴斯CV-260内窥镜故障维修2例毕欣雨㊀刘亚文关键词㊀内窥镜;CCD;分色盘;维护保养DOI:10 3969/j.issn.1002-3208 2023 06 017.中图分类号㊀R318 6㊀㊀文献标志码㊀B㊀㊀文章编号㊀1002-3208(2023)06-0655-01本文著录格式㊀毕欣雨,刘亚文.奥林巴斯CV-260内窥镜故障维修2例[J].北京生物医学工程,2023,42(6):655.㊀㊀内窥镜是具有观察目的的医学仪器,是一种光学装置,可以深入人体自然腔道或通过外科手术打开的孔道进行检查㊁诊断或治疗的医疗器械㊂奥林巴斯CV-260是用于胃肠道检查或者治疗的软性内窥镜,分别由主机㊁光源㊁软性检查镜和台车组成一套系统㊂本文总结了2例常见故障,以供交流㊂1㊀故障一故障现象:光源有光亮,但无图像显示㊂故障分析与排除:电子胃肠镜是经由光源发出色温6000K左右的高亮度的白光,通过内窥镜内部的导光束照射到组织,再由镜头先端部的CCD部件捕获图像,传输到主机处理图像信息㊂分别检查显示屏的视频线以及主机和光源间的电缆线接触是否良好,视频信号通道设置是否正确,检查确认无误㊂而后在检查内窥镜先端部的时候,发现弯曲橡皮处有细小的破口,且在镜头内部看到有污渍的痕迹,怀疑是内窥镜内部进水,导致CCD短路损毁㊂返厂检测,确定是CCD故障㊂内窥镜在使用的过程中由于一些人为的因素或者部件老化,会出现细小破口㊂在对内窥镜做清洗消毒的过程中,若没有及时察觉,一旦发生内窥镜内部进水,会出现导光束变黄折断,钢丝生锈等故障,甚至造成CCD部件的短路,产生高额的维修费用㊂所以在每一次对内窥镜进行清洗消毒之前需要进行测漏,及时发现破口,能大幅度减少维修费用㊂2㊀故障二故障现象:图像颜色显示异常,并伴有纹波㊂故障分析与排除:奥林巴斯CV-260电子胃肠镜,是通过光源内部的分色盘交替发出红㊁绿㊁蓝三种颜色的光照射在组织上,再由CCD部件捕获到三种颜色下的组织图像,传输到主机进行整合处理后形成一幅完整的组织信息图像㊂图像颜色异常首先检查主机到光源㊁主机到显示器的调光电缆接触是否良好,对内窥镜进行白平衡校准,发现颜色还是未恢复正常㊂用万用表测量其导通情况匀正常㊂接下来用手轻拿内窥镜的先端部,匀速晃动,不能清晰地看出红㊁绿㊁蓝三种颜色,说明是分色盘的工作异常㊂拆开光源机壳,发现光源内部灰很多,特别是分光盘上面有明显的灰尘㊂用吸尘器将光源内部的灰尘处理干净,再用拭镜纸将分色盘上的灰尘轻轻擦干净㊂装上机壳后试机,图像颜色恢复正常,故障排除㊂3㊀小结内窥镜在清洗消毒之前做好测漏工作,能提早发现工作部的细小破口,避免对内窥镜本身造成更大的损坏,也能及时寻根溯源,避免由于人为因素重复出现类似的故障㊂内窥镜是内镜中心使用频次最高的设备,在平时的使用过程中,更应该按照操作规程去操作,才能延长使用寿命㊂维修人员在平常应勤做维护保养,减少故障频次,提高医院的经济效益和社会效益㊂(2021-10-18收稿,2021-12-03修回)。