当前位置:文档之家› MW3000微波消解方法库

MW3000微波消解方法库

Multiwave 3000 Microwave Sample Preparation System

Classified List of Applications

All used reagents must be of appropriate purity and following concentration:Explanation of the table

y: yes n: no

o: optional, digestion possible with T-Sensor or p-Sensor (in combination with IR Sensor)

H2O2: 30 %H3PO4: 80 %

All methods in this list are conservative recommendations. It is the users resposibility to optimize the methods, with respect to the sample material. Due to different reaction behaviour of the sample, it could be useful to lower the sample weight, vary the reagent mixture or apply softer power profiles in order to avoid spontaneous reactions or to achieve proper results.

When digesting sample materials, which tend to react spontaneously like many organic materials and even metals, it is

recommended to start preliminary experiments with a lower sample weight. The program must start with a slow ramp and the pressure increase rate should be reduced to 0.3 bar/sec.

For some materials which tend to float on the surface, we recommend to use the stirring option in order to improve the contact between sample and acid.

When digesting sample materials which do not react spontaneously, the weight could be increased and the digestion time reduced in further optimizations.

Before starting with decompostions read the Manual very carefully!

HNO3:65 %HCl: 36 %

HF: 45 %H2SO4: 96 %

Methods without p/T-Sensor should not be changed and must not be applied for samples with different or unknown reaction behaviour.

Prior to digestion, check regularly the power and temperature calibration of the instrument to guarantee proper function and reproducible results. The operator is liable to the consideration and - if applicable - the appropriate matching of all parameters.

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Cleaning

Cleaning

SOLV-MF-T8

y

y

y

n

12.5 mL Acetone Temp

1105

51Additionally / optionally the liners can be treated at 12.5 mL Hexane

15

3150°C to 180°C in a drying oven, standing upside down Cleaning EVAP-MF-8y n n n 10 mL HNO3Power

6005101For rotor EVAP-4 up to rotor EVAP-8

15

3Cleaning, 4 to 8HF100-8y y y n 6 mL HNO3Power 10005151For rotor HF100-4 up to rotor HF100-8 using p/T-sensor unit 0

15

3Cleaning, 8 to 16HF100-16y y y n 6 mL HNO3Power 12005151For rotor HF100-8 up to rotor HF100-16 using p/T-sensor unit 0

15

3Cleaning, 4 to 8MF100-8y y y n 6 mL HNO3Power 10005151For rotor MF100-4 up to rotor MF100-8 using p/T-sensor unit 0

15

3Cleaning, 8 to 16MF100-16y y y n 6 mL HNO3Power 12005151For rotor MF100-8 up to rotor MF100-16 using p/T-sensor unit 0

15

3Cleaning, up to 4XF100-4y n y n 6 mL HNO3Power 8005151for up to 4 vessels XF100

15

3Cleaning, 4 to 8XF100-8y n y n 6 mL HNO3Power 12005151For rotor XF100-4 up to rotor XF100-80

15

3XQ80-4y n y n 6 mL HNO3Power 8005151for up to 4 vessels XQ80

15

3XQ80-8y n y n 6 mL HNO3Power 12005151For rotor XQ80-4 up to rotor XQ80-80

15

3MF100-8y n n n 6 mL HNO3Power 10005101For rotor MF100-4 up to rotor MF100-8 without using p/T-sensor 500151unit!

15

3MF100-16y n n n 6 mL HNO3Power 14005151For rotor MF100-8 up to rotor MF100-16 without using p/T-sensor 500101unit!

15

3

Cleaning, 4 to 8Cleaning, 8 to 16Sensors Program Cleaning, up to 4Cleaning, 4 to 8

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program HF100-8

y

n

n

n

6 mL HNO3

Power

10005

101For rotor HF100-4 up to rotor HF100-8 without using p/T-sensor 500151unit!

15

3HF100-16y n n n 6 mL HNO3Power 14005151For rotor HF100-8 up to rotor HF100-16 without using p/T-sensor 500101unit!

15

3Coal

Activated Carbon 250XF100-8y n y 1 5.5 mL HNO3Power 140015201Method can be used either with Rotor XF100 or Rotor XQ80 0.5 mL HCl

15

3Rotor XQ80 should be used for analysis of e.g. Hg, Se, As Ash -1100XF100-4y n y 18 mL HNO3 Power 8001545

1

Next step: Complexation with 18 mL H3BO3!2 mL HCl Do not perform digestion without using the IR-sensor!

3 mL HF 6 mL H2SO4

153Ash - 2100XF100-4y n y 1 5 mL HNO3 Power

80010

601Next step: Complexation with 12 mL H3BO3!

2 mL HF

153Carbon Fibre 200XF100-4y n y 1 6 mL HNO3 Power

80015401Method can be used either with Rotor XF100 or Rotor XQ802 mL H2O2

153Rotor XQ80 should be used for analysis of e.g. Hg, Se, As Coal-1250XF100-4y n y 1 4 mL HNO3Power

80010501Next step: Complexation with 12 mL H3BO3!2 mL HF 0

15

3

2 mL H2O2

Coal-2500XF100-4y n y 1 3.5 mL H3PO4Power 80010501Do not perform digestion without using the IR-sensor!

3.5 mL H2SO40

15

3

4 mL HF 0.

5 mL H2O 0.5 mL HClO4

Coke-1100XF100-4y n y 1 6 mL HNO3Power 80010451Leaching procedure

2 mL H2O2

15

3

Method can be used either with Rotor XF100 or Rotor XQ80Rotor XQ80 should be used for analysis of e.g. Hg, Se, As

Cleaning, 4 to 8

Cleaning, 8 to 16

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program Coke-2

100

XF100-4

y

n

y

1

6 mL HNO3Power

80010

451Complete digestion

2 mL H2O20

15

3

Next step: Complexation with 12 mL H3BO3!

2 mL HF

Flyash 300MF100-16y y y 1 6 mL HNO3Power 140010301Next step: Complexation with 12 mL H3BO3!

1 mL HCl 0

15

3

2 mL HF

Complexation

MF100-8y y y n xmL H3BO3Power 10005151For rotor MF100-4 up to MF100-8 using p/T-sensor unit!0

15

3

Add 6 mL saturated boric acid solution for each mL used HF,to complex free fluorides.

MF100-16y y y n xmL H3BO3Power 14005151For rotor MF100-8 up to MF100-16 using p/T-sensor unit! 0

15

3

Add 6 mL saturated boric acid solution for each mL used HF,to complex free fluorides.

HF100-8y y y n xmL H3BO3Power 10005151For rotor HF100-4 up to HF100-8 using p/T-sensor unit! 0

15

3

Add 6 mL saturated boric acid solution for each mL used HF,to complex free fluorides.

HF100-16y y y n xmL H3BO3Power 14005151For rotor HF100-8 up to HF100-16 using p/T-sensor unit! 0

15

3

Add 6 mL saturated boric acid solution for each mL used HF,to complex free fluorides.

XF100-4y n y n xmL H3BO3Power 8005151For rotor XF100-4 !

15

3

Add 6 mL saturated boric acid solution for each mL used HF,to complex free fluorides.

XF100-8y n y n xmL H3BO3Power 14005151For rotor XF100-8 !

15

3

Add 6 mL saturated boric acid solution for each mL used HF,to complex free fluorides.

Drying

10 g DRY-1n n n n Power 100011Use dishes to place samples in the rotor!

25020

1010

10

2

Complexation XF100-4Complexation XF100-8Complexation MF100-8Complexation MF100-16Complexation HF100-8Complexation HF100-16Apple

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 1 g

DRY-1

n

n

n

n

Power

300011Use dishes to place samples in the rotor!

65020

1010

10

21 g DRY-1n n n n Power 100011Use dishes to place samples in the rotor!

25020

101if powdered material, use dishes covered with watch glasses 0

10

210 g DRY-1n n n n Power 100011Use dishes to place samples in the rotor!25020

1010

10

210 g DRY-1n n n n Power 100011Use dishes to place samples in the rotor!25020

1010

10

21 g DRY-1n n n n Power 100011Use dishes to place samples in the rotor!

25020

101If powdered material, use dishes covered with watch glasses 0

10

21 g DRY-1n n n n Power 300011Use dishes to place samples in the rotor!

65020

101If powdered material, use dishes covered with watch glasses 0

10

210 g DRY-1n n n n Power 100011Use dishes to place samples in the rotor!25020

1010

10

250 g DRY-1n n n n Power 100011Use dishes to place samples in the rotor!25010

301Depending on the water content,

10

2 the second program phase could be extended!500 g DRY-1n n n n Power 300011Place sample in large evaporation dishes!65020

1010

10

210 g DRY-1n n n n Power 10001115030

1010

10

2

Fish Lobster Cucumber Marl Potato Sludge Soil Starch Clay

Cod Muscle

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 10 g

DRY-1

n

n

n

n

Power

100011Use dishes to place samples in the rotor!

25020

1010

10

2Environment

500HF100-16y y y n 6 mL HNO3Power 140010101Reduce the pressure increase rate to 0.3 bar/sec o

15

3100XF100-8y n y n 8 mL HNO3Power 110015451Next step: Complexation with 18 mL H3BO3!4 mL HCl 14005

151Do not perform digestion without using the IR-sensor!3 mL HF 0

15

3

6 mL H2SO4

100XF100-8y n y n 3 mL HNO3Power 120010501Next step: Complexation with 18 mL H3BO3!2 mL HCl 12000

301Do not perform digestion without using the IR-sensor!

3 mL HF

15

3

500XF100-8y n y n 4 mL HNO3Power 120010301Method can be used either with Rotor XF100 or Rotor XQ802 mL H2O2

15

3

Rotor XQ80 should be used for analysis of e.g. Hg, Se, As Sample weight depends strongly on modification Reduce the pressure increase rate to 0.3 bar/sec

500HF100-8y y y n 6 mL HNO3Power 140010201Reduce the pressure increase rate to 0.3 bar/sec

o

15

3500MF100-16y y y n 3 mL HNO3Power 140010201Sample weight depends strongly on modification

o

3 mL HCl

15

345 g MF100-T8y y y n 5 mL HNO3Temp 1651000aqueous sample, water, 17010

01for 4 and 8 vessels!

20

245 g MF100-T16y y y n 5 mL HNO3Temp 1650100aqueous sample, water,17010

01 for 16 vessels!

20

20.25-0.50 g MF100-T8y y y n 10 mL HNO3Temp 1755:3000sediments, soil, sludge, contaminated soil and oils (oil <0,25g!)1804:30

01 for 4 and 8 vessels!

20

2

Compost EPA 3015EPA 3015EPA 3051Cellulose Fibres Cellulose Lye Ashes-1Ashes-2Yogurt

Algae

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 0.25-0.50 g MF100-T16

y

y

y

n

10 mL HNO3

Temp

17505:300sediments, soil, sludge, contaminated soil and oils (oil <0,25g!) 1804:30

01for 16 vessels!

20

2EPA 30520.1 - 1.0 g MF100-T16y y y n 9 mL HNO3Temp 180005:300siliceous and organically based matrices for 16 vessels!1800

09:301reactive samples should be digested with longer heating ramp 0

20

2

Reagent optimization: add 0-5mL HF, 0-4mL HCl or 0-2mL H2O2

500XF100-8y n y n 4 mL HNO3Power 140010301Method can be used either with Rotor XF100 or Rotor XQ800.5 mL HCl 0

15

3

Rotor XQ80 should be used for analysis of e.g. Hg, Se, As 2 mL H2O

Sample weight depends strongly on modification Reduce the pressure increase rate to 0.3 bar/sec

300MF100-4y y y n 6 mL HCl Power 800102512 mL HNO3

15

3500HF100-16y y y n 5 mL HNO3Power 140010301Next step: Complexation with 12 mL H3BO3!

1 mL HCl 0

15

3

2 mL HF

150XF100-4y n y n 2 mL HNO3Power 80010301Method can be used either with Rotor XF100 or Rotor XQ800.5 mL HCl 0

15

3

Rotor XQ80 should be used for analysis of e.g. Hg, Se, As

2 mL H2O22 mL H2O

500MF100-16y y y n 5 mL HNO3Power 140010501Decompose TiO2 filler by opening vessel after cooling 2 mL HCl 0

15

3

add 3 mL H2SO4 and restart the same method again

2 mL H2SO4

500MF100-16y y y n 5.5 mL HNO3Power 140010101Silicates remain undissolved!

0.5 mL HCl

15

3400HF100-16y y y n 4 mL HNO3Power 140010101Reduce the pressure increase rate to 0.3 bar/sec

o

2 mL H2O2

15

3

Filter, microfibre polypropylen

Filter, PES-Membrane Grass, Leaching Oyster Tissue Filter, Cellulose Nitrate

Filter Dust, Leaching Filter, glass fibre EPA 3051

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 250

HF100-8

y

y y

n

6 mL HNO3Power

90010

201Reduce the pressure increase rate to 0.3 bar/sec

o

1 mL HCl

15

3400MF100-16y y y

n

5 mL HNO3Power

140010101Reduce the pressure increase rate to 0.3 bar/sec o

1 mL HCl 0

15

3

2 mL HF

250HF100-16y y y n 1,5 mL HNO3Power 14005301Next step: Complexation with 6 mL H3BO3!4.5 mL HCl 0

15

3

For samples with higher inorganic content.

1 mL HF

250HF100-16y y y n 4.5 mL HNO3Power 14005301Next step: Complexation with 6 mL H3BO3!1.5 mL HCl 0

15

3

For samples with higher organic content.

1 mL HF

1 g MF100-16y y y n 3 mL HNO3Power 140053019 mL HCl

15

3500HF100-8y y y n 2 mL HNO3Power 12005351Next step: Complexation with 18 mL H3BO3!6 mL HCl 0

15

3

carbonacious soil, low organic content

3 mL HF

500HF100-8y y y n 6 mL HNO3Power 12005351Next step: Complexation with 18 mL H3BO3!2 mL HCl 0

15

3

humus soil, high organic content

3 mL HF

250MF100-8y y y n 5 mL HNO3Power 80010251Next step: Complexation with 6 mL H3BO3!2 mL HCl 0

15

3

Reduce the pressure increase rate to 0.3 bar/sec

1 mL HF

500HF100-8y y y n 5 mL HNO3Power 90010351Next step: Complexation with 6 mL H3BO3!o

2 mL HCl 0

15

3

Reduce the pressure increase rate to 0.3 bar/sec

1 mL HF

500MF100-8y y y n 5 mL HNO3Power 80010251Reduce the pressure increase rate to 0.3 bar/sec

2 mL HC

15

3500HF100-T8y y y n 6 mL HNO3Temp 24010401Next step: Complexation with 12 mL H3BO3!

2 mL HF

15

3

Sediment, Total-1Sediment, Total-2Sediment, Leaching Soil, Total-1Soil, Total - 2Sludge, Total-1Sludge, Leaching Straw Peat

Pine Needles Sludge, Total-2

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 5 mL

MF100-16

y

y

y

n

2.5 mL HNO3Power

140010

201Reduce the pressure increase rate to 0.3 bar/sec

7.5 mL HCl

15

35 mL MF100-16y y y n 5 mL HNO3Power

140010201Reduce the pressure increase rate to 0.3 bar/sec 1 mL HCl 0

15

3

1 mL H2O2

800HF100-16y y y n 8 mL HNO3Power 140015151Reduce the pressure increase rate to 0.3 bar/sec

15

3Evaporation

50 mL EVAP-4y n n n 0.1 mL HNO3Power 10002111Evaporation with 4 vessels, fill the unoccupied positions with empty vessels 0

10

3concentration of 50 mL solution to 1 mL

50 mL EVAP-4y n n n 0.1 mL HNO3Power 1000251Evaporation with 4 vessels, fill the unoccupied positions with empty vessels 0

10

3concentration of 50 mL solution to 25 mL

25 mL EVAP-4y n n n 0.1 mL HNO3Power 1000251Evaporation with 4 vessels, fill the unoccupied positions with empty vessels 0

10

3concentration of 25 mL solution to 1 mL 50 mL EVAP-8y n n n 0.1 mL HNO3Power 12002151Evaporation with 8 vessels

10

3concentration of 50 mL solution to 1 mL 50 mL EVAP-8y n n n 0.1 mL HNO3Power 1200281Evaporation with 8 vessels

10

3concentration of 50 mL solution to 25 mL 25 mL EVAP-8y n n n 0.1 mL HNO3Power 1200281Evaporation with 8 vessels

10

3concentration of 25 mL solution to 1 mL 8EVAP-8y n n n Power 80010101fume off acids after previous digestion

10

3(typically 6 to 10 mL acid will be reduced to 1-2 mL)EVAP-8y n n n 3 mL HNO3Power 80010151add reagent and continue fuming off HF content 0

103repeat this step if necessary!

EVAP-4y n n n 5 mL HNO3Power 200101Evaporation with 4 vessels, fill the unoccupied positions with empty vessels

4001010

15

3

Water, Seeping

Water waste Wood Chips Water, 50mL to 1mL HNO3 Evaporation Water, 50mL to 1mL Water, 50mL to 25mL Water, 25mL to 1mL Water, 25mL to 1mL Water, 50mL to 25mL Fume off HF step1Fume off HF step2

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program Extraction

2 g

SOLV-MF-T8

y

y

y

1

15 mL n-Hexane Temp

1155

151Only applicable in MW3000 with built in Safety Module SOLV!15 mL Acetone

15

3

Only applicable in MW3000 with built in Safety Module SOLV! 5 g

SOLV-MF-T8

y

y

y

n

12.5 mL Acetone Temp

1105

5:301Extraction of soils, clays, sediments, sludge and waste,12.5 mL n-Hexane

15

3

The use of the magnetic stirrer option is recommended!

Only applicable in MW3000 with built in Safety Module SOLV!

5 g

SOLV-MF-T8

y

y

y

2

12.5 mL Acetone Temp

1105

10112.5 mL n-Hexane

15

3

Only applicable in MW3000 with built in Safety Module SOLV!

500

SOLV-MF-T8

y

y

y

1

2.5 mL

Methylenechlorid Temp 11510101IR-Limit: 130°C!7.5 mL Acetonitril

15

3

Yellow extraction phase!

Only applicable in MW3000 with built in Safety Module SOLV!

200

SOLV-MF-T8

y

y

y

2

22.5 mL

Dichlormethane Temp 1205201The sample is gel-like and centrifugation does not work well 7.5 mL Cyclohexane

15

3

With a water pump 1mL of extract can be filtered through a 0.45 μm teflonfilter

Only applicable in MW3000 with built in Safety Module SOLV!

500

SOLV-MF-T8

y

y

y

1

17.5 mL n-Hexane,Temp

1205

2017.5 mL Acetone

15

3

Only applicable in MW3000 with built in Safety Module SOLV!

2 g

SOLV-MF-T8

y

y

y

2

15 mL Cyclohexane Temp

905

901Versatile method for extraction of stabilizers.

15 mL 2-Propanol

15

3

Only applicable in MW3000 with built in Safety Module SOLV!

2 g

SOLV-MF-T8

y

y

y

2

15 mL Cyclohexane Temp

905

901Versatile method for extraction of stabilizers.

15 mL 2-Propanol

153Food

1 g HF100-16y y y n 8 mL HNO3Power

140010101Reduce the pressure increase rate to 0.3 bar/sec

o

0.5 mL HCl

153300HF100-16y y y n 4 mL HNO3Power

14005201Next step: Complexation with 12 mL H3BO3!2 mL HF

15

3

Animal Feed for Organochloride pesticides

Animal Feed Mineral nutrient

EPA 3546

Milkpowder

Mussel Tissue

Plastics for Phtalat

Polypropylene, unground

Polyethylene, unground

Soil Extraction

Animal Feed

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 500

XQ80-4

y

n

y

n

6 mL H2SO4

Power

80015

451Use IR-sensor to avoid overheating!

15

3500XQ80-4y n y n 10 mL HNO3Power 800154510

15

3500HF100-16y y y n 10 mL HNO3Power 120015101Reduce the pressure increase rate to 0.3 bar/sec 0

15

3500HF100-8y y y n 5.5 mL HNO3Power 80010201Reduce the pressure increase rate to 0.3 bar/sec 0.5 mL HCl

15

3500HF100-8y y y n 6 mL HNO3Power 800102010

15

3800HF100-T16y y y n 5.5 mL HNO3Temp 220103010.5 mL HCl

15

3500HF100-T16y y y n 7 mL HNO3Temp 180103012 mL HCl

15

3500HF100-16y y y n 5 mL HNO3Power 140010151Reduce the pressure increase rate to 0.3 bar/sec o

2 mL H2O2

15

3500HF100-T16y y y n 6 mL HNO3Temp 24010201Method can be used either with Rotor XF100 or Rotor XQ804 mL HCl

Rotor XQ80 should be used for analysis of e.g. Hg, Se, As 0

15

3500XF100-8y n y n 4 mL HNO3Power 140010301Reduce the pressure increase rate to 0.3 bar/sec

0.5 mL HCl 0

15

3

4 mL H2O2

350HF100-16y y y n 5 mL HNO3Power 140010151Reduce the pressure increase rate to 0.3 bar/sec

o

2 mL H2O2

15

3500HF100-16y y y n 5 mL HNO3Power 140010151Reduce the pressure increase rate to 0.3 bar/sec

o

2 mL H2O20

15

3

1 mL HCl

Biscuits Bruschetta with tomato Bouillon Camomile Carrots Casein Cheese Powder Betain

Brazil Nut Bread additives Bread brown, BCR 191Butter

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 500

HF100-16

y

y y

n

6 mL HNO3

Power

140010

101o

15

3500HF100-16y y y n 6 mL HNO3Power 140010101o

15

3500HF100-16y y y n 4 mL HNO3Power 140010101Reduce the pressure increase rate to 0.3 bar/sec

o

2 mL H2O20

15

3

0.5 mL HCl

500HF100-16y y y n 4 mL HNO3Power 140010101Reduce the pressure increase rate to 0.3 bar/sec

o

2 mL H2O20

15

3

0.5 mL HCl

500MF100-16y y y n 5 mL HNO3Power 140010101Reduce the pressure increase rate to 0.3 bar/sec

o

2 mL H2O2

15

3500XQ80-8y n y n 5 mL HNO3Power 400151011 mL H2O2

80010

1010

15

31000HF100-T16y y y n 5 mL HNO3Temp 24010101Reduce the pressure increase rate to 0.3 bar/sec

2 mL H2O2

15

3600XQ80-8y n y n 5 mL HNO3Power 1400151013 mL H2O20

15

3

0.5 mL HCl,

800XQ80-8y n y n 8 mL HNO3Power 75020101Reduce the pressure increase rate to 0.3 bar/sec

120010

2010

15

3800XQ80-8y n y n 6 mL HNO3Power 6005101Reduce the pressure increase rate to 0.3 bar/sec

0.5 mL HCl

140010

1010

15

3800XQ80-8y n y n 6 mL HNO3Power 140015301Reduce the pressure increase rate to 0.3 bar/sec

1 mL HF

15

3

Fat Fish Flour Coffee beans, milled Cotton Seed Oil Curly Kale Edible Oil Chocolate

Chilli Paste Citrus leaves Cocoa

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 800

XF100-8

y

n

y

n

7 mL HNO3Power

300551Reduce the pressure increase rate to 0.3 bar/sec

1 mL HCl 500551

2 mL H2O

1000101510

153600HF100-16y y y n 5.5 mL HNO3Power

140010

101Reduce the pressure increase rate to 0.3 bar/sec o

0.5 mL HCl

15

3800HF100-T8y y y n 6 mL HNO3Temp

22015301Reduce the pressure increase rate to 0.3 bar/sec 1 mL HCl

15

31000HF100-8y y y n 4 mL HNO3Power

120015301o

2 mL H2O20

15

3

0.5 mL HCl

450HF100-T16y y y n 4 mL HNO3Temp 220101512 mL H2O20

15

3

0.5 mL HCl

300XF100-8y n y n 6 mL HNO3Power 140010301Method can be used either with Rotor XF100 or Rotor XQ801 mL HCl

15

3Rotor XQ80 should be used for analysis of e.g. Hg, Se, As 400XF100-8y n y n 2 mL HNO3Power 140010301Method can be used either with Rotor XF100 or Rotor XQ804 mL H2O20

15

3

Rotor XQ80 should be used for analysis of e.g. Hg, Se, As

0.5 mL HCl

600HF100-16y y y n 5 mL HNO3Power 1400102011 mL HCl

15

3500HF100-16y y y n 6 mL HNO3Power 8001551Reduce the pressure increase rate to 0.3 bar/sec

12005

510

15

3500HF100-4y n n n 8 mL HNO3Power 500901Set the IR-Limit to 200 °C!

3000

2510

15

3300HF100-16y y y n 4 mL HNO3Power 140010101Reduce the pressure increase rate to 0.3 bar/sec 2 mL H2O2

15

3

Molasses Gelatine

Meat Milkpowder Lettuce Linseed, milled Liver Lobster SRM Honey Lactose Milkpowder, only with IR sensor

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 1000

XF100-4

y

n

y

n

4 mL HNO3Power

80015

301Method can be used either with Rotor XF100 or Rotor XQ802 mL H2O2

15

3Rotor XQ80 should be used for analysis of e.g. Hg, Se, As 500XF100-8y n y n 8 mL HNO3Power

140015201Method can be used either with Rotor XF100 or Rotor XQ802 mL H2SO4

15

3Rotor XQ80 should be used for analysis of e.g. Hg, Se, As 250XQ80-4y n y n 6 mL HNO3Power

80015151Reduce the pressure increase rate to 0.3 bar/sec

15

3

500XF100-8y n y n 4 mL HNO3Power 1400153011 mL HCl 0

15

3

2 mL H2O2

500HF100-T16y y y n 7 mL HNO3Temp 18010301o

2 mL HCl

15

3500XQ80-8y n y n 2 mL HNO3Power 140015301Reduce the pressure increase rate to 0.3 bar/sec

0.5 mL HCl 0

15

3

4 mL H2O2

500XF100-8y n y n 6 mL HNO3Power 1400102014 mL HCl

15

3500MF100-16y y y n 5 mL HNO3Power 140010151Reduce the pressure increase rate to 0.3 bar/sec

o

2 mL H2O2

15

3500MF100-16y y y n 6 mL HNO3Power 140010151Reduce the pressure increase rate to 0.3 bar/sec o

2 mL HCl

15

3800XQ80-8y n y n 8 mL HNO3Power 75020101Reduce the pressure increase rate to 0.3 bar/sec 120010

1010

15

3500XQ80-8y n y n 3 mL HNO3Power 1400101513 mL H2O20

15

3

0.5 mL HCl

Saffron Soybean Muscle SRM

Orange Oil Peas Mussel Spinach Tea Pesto all′arrabiata Pumpkin Seed Oil Sugar

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 600

MF100-16

y

y y

n

7 mL HNO3Power

140010

101o

2 mL HCl

15

3500XF100-4y n y

1

8 mL HNO3Power

7001501Reduce the pressure increase rate to 0.3 bar/sec

0.5 mL HCl

8005

1510

15

3500MF100-16y y y n 5 mL HNO3Power 140010101o

1 mL H2O20

15

3

0.5 mL HCl

800HF100-T16y y y n 6 mL HNO3Temp 22010301Reduce the pressure increase rate to 0.3 bar/sec

0.5 mL HCl

15

3Geology

300HF100-16y y y n 1 mL HNO3Power 14005251Next step: Complexation with 24 mL H3BO3!

1 mL HCl 0

15

3

4 mL HF

200XF100-4y n y n 4 mL H2SO4Power 80010451Next step: Complexation with 12 mL H3BO3!4 mL H3PO40

15

3

Do not perform digestion without using the IR-sensor!

2 mL HF 2 mL H2O

200HF100-16y y y n 4 mL H3PO4Power 140010301Do not perform digestion without using the IR-sensor!

2 mL HNO30

15

3

0.5 mL HCl

200HF100-16y y y n 1.5 mL HNO3Power 140010301Next step: Complexation with 24 mL H3BO3!

0.5 mL HClO40

15

3

4 mL HF

200HF100-16y y y n 6 mL HCl Power 140053010

15

3200HF100-16y y y n 4 mL HNO3Power 14005201Leaching procedure, silcates remain in solution.

2.5 mL HCl

15

3300HF100-16y y y n 3.5 mL HF Power 14005251Next step: Complexation with 21 mL H3BO3!2.5 mL H2SO4

15

3

Do not perform digestion without using the IR-sensor!Ba-Ferrite Feldspar Al-Silicate

Bauxite Wheat Crackers Wheat Flour Tomato Paste

Vegetable Oil Kaolin Hematite Iron Ore

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 500

HF100-16

y

y

y

n

2 mL HNO3Power

140010

101Next step: Complexation with 6 mL H3BO3!

6 mL HCl 0

15

3

1 mL HF

2g HF100-T16y y y n 10 mL HF, 5 mL H2O Temp 240103510

15

3250HF100-16y y y n 1 mL HNO3Power 14005301Next step: Complexation with 18 mL H3BO3!

2 mL HCl 0

15

3

3 mL HF

250HF100-16y y y n 6 mL HNO3Power 14005301Next step: Complexation with 12 mL H3BO3!

2 mL HF

15

3250XF100-4y n y n 3 mL HNO3Power 80010451Next step: Complexation with 18 mL H3BO3!

1 mL HCl 0

15

3

3 mL HF

200HF100-T4y y y n 4 mL HNO3Temp 24015301Next step: Complexation with 12 mL H3BO3!

2 mL HCl 0

15

3

2 mL HF

Material

200XF100-8y n y n 3 mL H2SO4Power 140015301Do not perform digestion without using the IR-sensor!

3 mL H3PO40

15

3

4 mL H2O

200XF100-8y n y n 3 mL H2SO4Power 110010501Next step: Complexation with 12 mL H3BO3!3 mL H3PO40

15

3

Do not perform digestion without using the IR-sensor!

4 mL H2O 2 mL HF

200HF100-8y y y n 4 mL HNO3Power 90015451Do not perform digestion without using the IR-sensor!

4 H3PO40

15

3

2 HClO4

200HF100-T16y y y n 8 mL HNO3Temp 20015151Next step: Complexation with 24 mL H3BO3!

0.5 mL HCl 0

15

3

2 mL H2O2

Bitumen Alumina Magnetite

Quartz-Sand River Sediment Rock, Basaltic Rock, Granite Talcum Al2O3Al2O3 with SiO2

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 500

XF100-4

y

n

y

n

3 mL HNO3Power

80010

301Next step: Complexation with 18 mL H3BO3!

2 mL H2SO40

15

3

3 mL HF

500HF100-16y y y n 1 mL HNO3Power 140010301Next step: Complexation with 24 mL H3BO3!

1 mL HCl 0

15

3

4 mL HF

700HF100-T16y y y n 5 mL HNO3Temp 1505101Reduce the pressure increase rate to 0.3 bar/sec

2 mL H2O2

24010

1010

15

3300HF100-T8y y y n 6 mL HNO3Temp 24015151Reduce the pressure increase rate to 0.3 bar/sec

1 mL HF 0

15

3

1 mL H2O2

50XF100-4y n y n 6 mL HNO3Power 80010901Next step: Complexation with 6 mL H3BO3!

6 mL HCl 0

15

3

1 mL HF 3 mL H2O2

50XF100-4y n y n 8 mL H2SO4Power 80010901Next step: Complexation with 24 mL H3BO3!5 mL HF 0

15

3

Do not perform digestion without using the IR-sensor!

4 mL HCl

200HF100-4y y y n 3 mL HNO3Power 80015301Next step: Complexation with 15 mL H3BO3!1 mL HCl 0

15

3

Do not perform digestion without using the IR-sensor!

4 mL H3PO42 mL HClO4

250XF100-8y n y n 3 mL H2SO4Power 120015451Next step: Complexation with 12 mL H3BO3!3 mL H3PO41200601Do not perform digestion without using the IR-sensor!

2 mL HF

15

3100XF100-8y n y n 4 mL H2SO4Power 120015451Next step: Complexation with 12 mL H3BO3!4 mL H2O 1200601Do not perform digestion without using the IR-sensor!

2 mL HF

15

3

Cardboard Catalyst, spent-2 PGM loaded

Boron Nitrite

Boron Silicate Glass Cement, Ca-aluminat Ceramic-1Ceramic-2Carpet, Vinyl Catalyst, spent-1 PGM loaded

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 200

XF100-8

y

n

y

n

4 mL HNO3Power

140010

301Next step: Complexation with 12 mL H3BO3!

2 mL HCl 0

15

3

2 mL HF

1000XF100-4y n y n 6 mL HF Power 800206010

15

3

500XF100-8y n y n 5 mL HNO3Power 140010201Method can be used either with Rotor XF100 or Rotor XQ801 mL HCl 0

15

3

Rotor XQ80 should be used for analysis of e.g. Hg, Se, As

4 mL H2O2

400HF100-T4y y y n 5 mL HNO3Temp 22010251HF is needed for digestion of filling materials

1 mL HCl 0

15

3

1 mL H2O2(1-

2 mL HF)

1000XF100-8y n y n 5 mL HNO3Power 5005151Next step: Complexation with 6 mL H3BO3!

0.5 mL HCl 90010

1511 mL HF

15

3

800XQ80-8y n y 17 mL HNO3Power 1400152512 mL H2O2

15

3500MF100-8y y y 1 6 mL HNO3Power 1000152512 mL H2O2

15

3200HF100-4y y y n 4 mL HNO3Power 80015451Next step: Complexation with 18 mL H3BO3!3 mL HF 0

15

3

Do not perform digestion without using the IR-sensor!

3 mL H2SO4

200HF100-8y y y n 3 mL HNO3Power 120010151Next step: Complexation with 12 mL H3BO3!2 mL HF 0

15

3

Do not perform digestion without using the IR-sensor!

2 mL H2SO4

250XF100-8y n y n 2 mL HNO3Power 140020301Next step: Complexation with 24 mL H3BO3!4 mL HF 0

15

3

Do not perform digestion without using the IR-sensor!

1.5 mL H3PO4

Paper Rutile Ore Clinker

Glass Paper Leather Peat Porcelain Materials Paint, with filler Hair

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 100

XF100-8

y

n

y

n

7 mL HNO3Power

140015

901Next step: Complexation with 36 mL H3BO3!6 mL HF 0

15

3

Do not perform digestion without using the IR-sensor!

3 mL H2SO4

50XF100-4y n y n 15 mL HCl Power 800159012 mL HNO30

15

3

1 mL Br2

300HF100-T16y y y n 5 mL HNO3Temp 100051Next step: Complexation with 12 mL H3BO3!

0.5 mL HClO424010

3012 mL HF

15

3300HF100-T8y y y n 4 mL HCl Temp 21015251Next step: Complexation with 12 mL H3BO3!

2 mL HF 0

15

3

6 mL HNO3

Metals

500HF100-8y y y n 1 mL HNO3Power 12005301Next step: Complexation with 6 mL H3BO3!

6 mL HCl 0

15

3

1 mL HF

500HF100-8y y y n 2 mL HNO3Power 120053018 mL HCl

15

3500HF100-16y y y n 3 mL HNO3Power 1400101013 mL HCl 0

15

3

5 mL H2O

500HF100-8y y y 210 mL HCl Power 1200102010

15

3300HF100-8y y y n 3 mL H2SO4Power 80010301Reduce the pressure increase rate to 0.3 bar/sec 3 mL H3PO40

15

3

Next step: Complexation with 6 mL H3BO3!

1 mL HF Do not perform digestion without using the IR-sensor!

1 mL H2O 1 mL HClO4

300HF100-8y y y n 6 mL HNO3Power 80010301Next step: Complexation with 18 mL H3BO3!

3 mL HF 0

15

3

1 mL HClO4

Zeolite Cr-Oxide-2Cr-Oxide-1Al-Alloy, containining Si Brass Slag, combustion SiC

Soil containing PGM Co-Oxide Al-Alloy, without Si

Multiwave 3000 Microwave Sample Preparation System

Category

Sample

Amount Rotor Stirrer Reagents Profile

Notes

[mg]

IR

T

p

Temp / Power

Ramp

Hold

Fan

Sensors Program 1000

XF100-4

y

n

y

n

2 mL HNO3Power

80010

201Next step: Complexation with 6 mL H3BO3!

1 mL HF 0

15

3

6 mL HCl

500MF100-16y y y n 6 mL HNO3Power 1400103012 mL HCl

15

3200HF100-16y y y n 6 mL HNO3Power 140053012 mL HCl

15

3400XF100-8y n y n 5 mL HNO3Power 140015301Next step: Complexation with 3 mL H3BO3!

0.5 mL HF 0

15

3

1 mL HCl

400XF100-4y n y n 1.5 mL HNO3Power 8005351Next step: Complexation with 30 mL H3BO3!

5 mL HF 0

15

3

2.5 mL HCl

600HF100-16y y y n 5 mL HCl Power 14005301Next step: Complexation with 18 mL H3BO3!

3 mL HF 0

15

3

5 mL H2O

1000XF100-4y n y n 5 mL HNO3Power 80015601Next step: Complexation with 12 mL H3BO3!

5 mL HCl 0

15

3

2 mL HF 4 mL H2O

600HF100-16y y y n 4 mL HCl Power 140010101Next step: Complexation with 12 mL H3BO3!

2 mL HNO30

15

3

2 mL HF 2 mL H2O

250XF100-4y n y 1 2.5 mL HNO3Power 80010601AgCl remains!

7.5 mL HCl

15

3For Ag determination, a seperate run using HNO3 only is needed!

400HF100-16y y y n 1.5 mL HNO3Power 140053014.5 mL HCl

15

3

Gold/Copper Fe-Oxide Cu-Oxichloride Cu-Alloy Cr-Si-Alloy, Cr-Ni-Alloy, Cr-Mo- Alloy

Fe-Si-Alloy Fe-Alloy Fe-Cr-Alloy Fe-Mn Gold-Silver-Alloy

相关主题
文本预览
相关文档 最新文档