2014年PCT血糖测定标准操作规程SOP
- 格式:doc
- 大小:125.50 KB
- 文档页数:7
血糖测定流程及评分标准及整改措施英文回答:Blood glucose testing is a crucial procedure for monitoring and managing diabetes. It involves measuring the level of glucose in the blood to determine if it is within the target range. The process typically includes the following steps:1. Preparing the equipment: Before conducting the test, gather all the necessary supplies, including a blood glucose meter, lancet device, test strips, alcohol swabs, and a sharps container.2. Washing hands: Cleanse your hands with soap and water or use an alcohol-based hand sanitizer to ensure they are free from any substances that could interfere with the test.3. Inserting a test strip: Insert a test strip into themeter according to the manufacturer's instructions. The meter will usually turn on automatically.4. Pricking the finger: Use a lancet device to prick the side of your fingertip gently. This will create a small drop of blood.5. Applying blood to the test strip: Touch the blood droplet to the designated area on the test strip. The meter will display a countdown or a flashing icon to indicatethat it is processing the sample.6. Reading the result: After a few seconds, the meter will display the blood glucose level. Record the result in a logbook or on a mobile app if necessary.7. Disposing of the lancet and test strip: Safely dispose of the used lancet and test strip in a sharps container to prevent accidental needlestick injuries.Scoring criteria for blood glucose testing may vary depending on the specific context and guidelines providedby healthcare professionals. However, some common evaluation factors include:1. Accuracy: The test results should be within an acceptable range of the actual blood glucose level. Deviations from the target range may indicate the need for adjustment in medication or lifestyle.2. Consistency: Consistent blood glucose readings over time demonstrate good control of diabetes. Fluctuations or erratic patterns may indicate the need for further investigation or intervention.3. Frequency: Regular monitoring of blood glucose levels is essential for effective diabetes management. The frequency of testing should align with the individual's treatment plan and healthcare provider's recommendations.4. Documentation: Keeping a record of blood glucose results allows for better tracking of trends and patterns. It facilitates discussions with healthcare professionals and helps in making informed decisions about treatmentadjustments.To improve the blood glucose testing process and meet the desired standards, the following measures can be implemented:1. Education and training: Provide comprehensive education and training to individuals with diabetes on proper blood glucose testing techniques, including hand hygiene, equipment handling, and accurate result interpretation.2. Quality control: Regularly calibrate and maintain the accuracy of blood glucose meters and test strips. Conduct internal quality control checks to ensure reliable results.3. Standardized protocols: Develop standardized protocols and guidelines for blood glucose testing based on evidence-based practices. Ensure that these protocols are easily accessible and regularly updated.4. Supportive technology: Encourage the use of mobile apps or digital platforms that enable easy recording and analysis of blood glucose data. This can help individuals and healthcare providers identify patterns and make informed decisions.中文回答:血糖测定是监测和管理糖尿病的关键步骤。
快速测血糖操作流程
姓名:科别:得分:监考人:
目的及注意事项见背面
目的:监测患者血糖水平,评价代谢指标,为临床治疗提供依据
注意事项: 1. 测血糖前,确认血糖仪上的号码与试纸号码一致。
2. 确认患者手指消毒剂待干后实施采血。
3. 避免试纸发生污染。
4. 采血时弃去第一滴血,取第二滴,以免影响结果。
5. 让血液测试盒的尖端充分接触血液。
6. 对需长期监测血糖的患者,告知穿刺部位应轮换,在指尖或耳垂侧边采血疼痛较轻,且血量充足。
另外,采血时不要挤压采血的手指,太用力挤压手指会导致血液稀释,影响检测结果。
并指导血糖监测的方法,告知血糖正常值:空腹血糖:3.9~6.1 mmol/L,餐后2h血糖:6.9~11.2 mmol/L。
7. 低血糖诊断标准为:男<2.78 mmol/L,女<2.5 mmol/L(饥饿72小时后正常男性,女性
最低值);高血糖的诊断标准为: 空腹血糖>7.0 mmol/L,餐后血糖>11.1 mmol/L。
一般处理:遵医嘱服用药物,调节饮食,注意控制血糖水平,寻找原因,对症处理。
8. 遵循查对制度,符合无菌技术、标准预防原则。
附血糖监测考核评价标准
Ⅱ级表示动作熟练、规范、有1-2处缺项,与病人沟通不够自然;
Ⅲ级表示动作欠熟练、规范、有1-2处缺项,与病人沟通较少;
Ⅳ级表示动作欠熟练、有4处以上缺项,与病人没有沟通。
检验项目标准操作规程(SOP)-1-检验标本的采集一、标本的正确采集标本采集必须符合2个条件,即必须满足检测结果正确性的各项要求和检测结果必须能真实地反映检验对象当前病情,避免干扰因素的存在。
二、标本的贮存标本采集后尽快送至实验室,若不能及时送检,已采集的标本要按检验规定的贮存条件,如室温、冰浴、温浴或防腐贮存,将标本直立置于稳定、干燥、避光、密闭的环境中,避免振摇,以免标本遗洒或溶血影响检测结果。
三、标本的运送必须保证运送后标本所分析的结果与刚采集标本后分析的结果一致。
四、标本的签收临床工作人员从口才采集标本并将标本从临床运送到实验室及实验室人员接收临床标本,均应按标准化要求进行,做到认真核对,包括标本来源、标本属性、检查项目、标本采集和运送是否合乎要求等,标本送出人员和标本接收人员都要做认真的记录并签字存档。
五、标本的处理1、实验室接收标本后应及时正确地予以处理,否则会影响检测结果的准确性。
2、如果取血后未尽快转送或分离血清、血浆,血清与血块簪时间接触可发生变化。
3、实验室接收标本后处理应注意事项:(1)、时间:实验室接收标本后应尽快予以分类和离心。
①、促凝标本应尽早处理,可在采血5-15分钟后离心;②抗凝标本可采血后立即离心;③非抗凝(无促凝)标本采血30-60分钟后离心;④抗凝全血标本(全血细胞分析、ESR等)不需要离心。
(2)、温度:一般标本为室温(最好是22-25℃)放置;冷藏标本(对温度依赖性分析物)应保持在2-8℃直到温度控制离心。
(3)、采血管放置:应管口(盖管塞)向上,保持垂直立位放置。
(4)、采血管必须封口:管塞移去后会使血PH改变,影响检测结果,封口可以减少污染、蒸发、喷洒和溢出等。
六、分析前的可变因素1、生物因素:可引起所检测物质在体内的变化,此种变化与检测方法无关,分为可变的和固定的生物因素。
2、干扰因素:在收集和分析标本过程中,干扰因素常导致分析结果与被测物真实浓度不符。
选手班级: 选手姓名: 选手学号: 标本号:血糖测定(葡萄糖氧化酶法)竞赛时间:15分钟一、说明1.本题满分100分,完成时间15分钟。
2.要求独立完成,不得相互询问或讨论。
3.考核成绩为操作过程评分、操作结果评分和考核时间评分之和。
4.全部操作过程时间和操作后处理时间计入时间限额。
二、实验原理葡萄糖氧化酶(GOD )催化葡萄糖氧化成葡萄糖酸(D-葡萄糖酸δ内酯),并产生过氧化氢:葡萄糖+2H 2O+O 2葡萄糖酸+2H 2O 2在色原性氧受体(如联大茴香胺、4-氨基安替比林偶氮酚)的存在下,过氧化物酶催化过氧化氢,氧化色原物质,生成有色化合物。
其颜色深浅与血清葡萄糖含量成正比。
三、实验器材721分光光度计、小号试管≤7支、试管架1个、手动可调式移液器5~50μl 、1ml 和10ml 刻度吸管各两支、记号笔1支、水浴箱、擦镜纸,废物容器等。
四、实验试剂葡萄糖标准液、生理盐水、0.1mol/L 磷酸盐缓冲液、酶试剂、酚试剂等组成的试剂盒。
五、实验标本 新鲜血清。
六、操作步骤取3支小试管,按表进行操作。
先配制10ml 酶酚混合试剂(试剂比例R1:R2=9:1)。
加入物(ml ) 测定管 标准管 空白管 血清0.03 葡萄糖标准应用液0.03 生理盐水0.03 酶酚混合试剂3.03.03.0混匀,置37℃水浴中,保温10min ,分光光度计波长500nm ,比色杯光径1.0cm ,以空白管调零,分别读取标准管和测定管的吸光度。
七、数据记录与计算1.数据记录标准管吸光度测定管吸光度裁判签名确认2.计算血清葡萄糖浓度(写出计算公式及过程)裁判签名确认八、报告(如:血清葡萄糖:△.△△ mmol/L)。
裁判签名确认选手班级:选手姓名:选手学号:标本号:血糖测定(葡萄糖氧化酶法)评分标准序号项目考核内容分值扣分标准扣分得分1 准备工作报告参赛项目及序号5漏缺一项 1 着白大衣、帽、口罩、手套并按要求佩戴参赛签着装不整、漏缺某一项 1 仪表端庄、头发符合要求仪容、着装不整 1 小号试管、试管架、移液器及配套吸头、记号笔1支缺选及多选、台面凌乱 1检查仪器、试剂、标本状态并摆在操作台合理位置无检查 12 操作①试管编号②空白管加蒸馏水30µl③标准管加葡萄糖标准液30 µl④测定管加血清30µl⑤各管分别加酶工作液3ml⑥各管置37℃水浴保温10分钟取出比色12试管无编号 1空白管加蒸馏水 1标准管加葡萄糖标准 1测定管加血清30µl 1各管分别加酶工作液 1加液顺序错 1加错液 1加液量不准确 1无混匀 1混匀不规范 1无水浴 1水浴时间不准确 1反应液无倒回原试管 23 用722分光光度计比色①调波长(500nm)②清洗比色杯,将各管溶液倒入比色杯,置比色槽中③调“T 100%”和“T 0”,重复2次④进行比色测定,记录读数,反应液倒回原试管⑤回位至显示器出现“100.0”13无调波长 1调波长不准 1无调T100% 1无调T0 1分光光度计闭盖预热 1双手碰触变色杯光滑面 1比色后无洗干净比色杯 1未用擦镜纸擦拭比色杯 1比色杯未正确放比色槽 1显示器未稳定时读数 1⑥洗干净比色杯(三次)比色液体未达杯高一半 1反应液无倒回原试管 1读数不准确 14 数据与报告原始记录完整、规范5不完整、不规范 1 单位正确不正确 1 公式正确不正确 1 计算方法正确不正确 1 报告完整、正确不完整、不正确 15 计数结果相对误差=(︱X-T︱/T)(X为计数值,T为靶值)56(︱X-T︱/T)×56分、相对误差≥20%扣56分6 文明操作操作完成后台面的处理 1 摆放无序,用物随意放 1 将使用过的一次性物品弃入污物缸或指定位置。
English Answer:The standard operational procedure for blood glucose testing involves several key steps:1.Preparation:Gather the necessary equipment, including a lancet device, lancets, aglucose meter, test strips, and alcohol pads. Ensure that the glucose meter and test strips are compatible and are within their expiration dates.2.Cleaning:Use an alcohol pad to clean the finger or other test site to remove anydirt or debris.ncing:Adjust the lancet device to the desired depth setting and firmly press itagainst the prepared test site to obtain a blood sample.4.Applying Blood:Apply the blood sample to the test strip, following the instructionsprovided by the meter.5.Inserting the Test Strip:Insert the test strip into the glucose meter and wait for theresult. The meter will analyze the blood sample and display the glucose level.6.Recording the Result:Record the glucose level obtained, along with any relevantinformation such as the date, time, and any symptoms experienced.7.Disposal:Dispose of the used lancet, test strip, and alcohol pad properly to avoidcontamination.Remember to follow the manufacturer's instructions for the specific equipment being used and to sanitize hands and equipment before and after each use.中文回答:血糖检测的规范操作流程包括以下几个关键步骤:1.准备:准备必要的设备,包括采血笔、采血针、血糖仪、试纸和酒精棉片。
血糖常规检验标准操作规程1. 引言该操作规程旨在确保血糖常规检验的标准化操作流程,提高检验结果的准确性和一致性。
2. 术语定义- 血糖常规检验:通过测量血液中的血糖水平来评估机体糖代谢情况的检验方法。
- 操作规程:对某一具体操作的规范化描述和具体步骤的文档。
3. 检验设备和试剂准备1. 准备好血糖仪器和试剂。
2. 检查仪器和试剂的有效期和质量。
3. 校准血糖仪器。
4. 检验前准备1. 患者空腹8小时,禁止餐饮和饮水。
2. 每名患者应做好个人清洁,包括洗手和消毒。
5. 检验操作流程1. 工作人员应佩戴个人防护用品,包括手套、口罩和短袖。
2. 为患者消毒采血部位,并取血样。
3. 将采集的血样加入试剂管中。
4. 启动血糖仪器,按照仪器说明书操作。
5. 记录血糖检验结果。
6. 结果判读和报告1. 对于正常范围内的血糖水平,记为正常。
2. 对于异常范围内的血糖水平,参考相关标准进行评价和解释。
3. 将检验结果记录在病患的医疗档案中。
7. 质量控制1. 进行质量控制测试,确保仪器和试剂的准确性和可靠性。
2. 定期校准仪器,确保其精度。
8. 安全注意事项1. 操作过程中需遵守个人防护要求,如佩戴手套和口罩。
2. 处理废弃物时,采取正确的方法进行处理和处置。
9. 维护与故障排除1. 定期对仪器进行维护和清洁。
2. 出现故障时,及时联系维修技术人员进行排除。
10. 参考文献以上是血糖常规检验标准操作规程的简要说明,通过严格遵守操作流程和质量控制要求,可以提高检验的准确性和可靠性。
测血糖标准操作流程
1.清洁双手
用肥皂水洗净双手,并擦干。
2.安装试纸
从试纸瓶中取出试纸,避免接触试纸吸血口和电极区,试纸电极端平行插入血糖仪的试纸接口。
3.采血部位消毒
用酒精棉或75%酒精以采血点为中心,顺时针消毒,勿反复擦拭。
4.采血
待消毒部位完全干后进行采血,选择手指端两侧作为采血部位,尽量避免扎指端中间部位,因为中间神经丰富,痛感明显。
5.虹吸
第一滴血可能混有酒精会稀释血液,测量值不是很准确,用棉签剔除不要,出来的第二滴血,用已经安装好试纸条的血糖仪,在试纸条的采血端,通过“虹吸”效应,把这滴血吸入试纸条上面,避免血糖试纸紧贴皮肤,不利于“虹吸”。
把血糖仪平放,等待几秒钟后读数。
6.止血
用干净棉签止血,按压采血点1分钟左右,避免出现出血点(黑点)。
7.废物丢弃
将采血针、用过的试纸、酒精棉、棉签丢弃到垃圾桶。
建议用一个瓶子专门丢弃针头,避免垃圾分类时扎伤。
葡萄糖1.目的规范葡萄糖测定的操作程序,确保葡萄糖测定结果准确。
2.检验方法氧化酶法(两点终点法) 3.检验原理 葡萄糖+O 2+H 2O 葡萄糖酸+H 2O 2H 2O 2+4-AAP+苯酚红色醌亚胺+4H 2O在波长505nm 处比色,计算出GLU 浓度。
4.临床意义临床检测GLU 浓度主要用于糖尿病诊断、昏迷鉴别诊断及糖代谢研究。
血糖增高主要见于:①生理性高血糖:可见于饭后1-2小时或情绪紧张肾上腺分泌增加时;②病理性高血糖:可见于内分泌功能障碍引起的糖尿病、颅外伤、颅内出血、脑膜炎引起的颅内高压、脱水引起的高血糖、急慢性胰腺炎、肝功能障碍等。
血糖降低主要见于:①生理性低血糖:常见于饥饿和剧烈运动后。
②病理性低血糖:可见于高胰岛素血症、非胰性肿瘤、糖原累积症、自身免疫性疾病、半乳糖血症和肾脏疾病等。
5.标本采集及干扰因素 5.1标本要求空腹不抗凝静脉血2~3ml 。
5.2标本保存室温保存,及时送检。
血清ALT 在20-25℃可稳定24小时,在2-8℃储存可稳定7d ,-20℃冰冻保存可稳定30d 。
5.3注意事项推荐选用血清,避免溶血,红细胞中ALT 比血浆高约7倍,溶血时红细胞内ALT 可进入血浆,导致测定结果偏高。
标本应避免溶血、脂血、黄疸。
6.检测仪器日立7180全自动生化分析仪,使用具体要求和校准程序详见《仪器GOPOD操作规程》。
7.试剂 7.1试剂品牌美康生物 7.2包装规格200mL (R1:2×65mL+R2:1×70mL ) 7.3主要组成成分 试剂成分含量(最终反应体系) 试剂1 磷酸盐缓冲液 0.1mol/L 苯酚0.01×10-3mol/L 试剂2葡萄糖氧化酶 16000u/L 过氧化物酶 1000u/L 4-氨基安替比林0.02×10-3mol/L不同批次的试剂不推荐混合使用 7.4储存条件有效期试剂在2-8℃保存可稳定1年,试剂开瓶后于2-8℃可稳定1个月。
(完整版)血糖监测操作流程及评分标准血糖监测操作流程及评分标准评估准备核对医嘱,核对患者姓名、性别、年龄、检验申请单1.患者评估:患者病情、合作程度、意识状态、检验目的、血糖情况2.局部因素评估:选择合适采血部位自身评估:着装整洁,仪表端庄,符合操作要求;洗手,戴口罩用物评估:用物齐全,摆放有序,血糖仪性能良好,试纸在有效期内(血糖仪、血糖试纸、采血针、无菌棉签、75%酒精)携用物至床旁,核对患者信息,告知相关事项,取得合作,环境符合操作要求,保护患者隐私,选择适合采血的部位,核对血糖仪上的调码和试纸瓶上的号码是否一致,协助患者处舒适体位,再次检查局部皮肤情况,用2根酒精棉签消毒皮肤,待干,开机,当屏幕上闪现插入试纸提示时,可轻轻插入试纸实施再次核对患者姓名,将采血针固定在手指欲采血部位,快速刺入皮下,轻轻挤压手指,把第一滴血滴入试纸测试孔,测试孔应全部被血充满,(注意:血样不能重复添加,也不能涂抹到加样区,足够量的血正确滴入后,不要涂抹、移动试纸,等待屏幕上显示血糖的测定值,读取数据,再次查对并做好记录(检验报告申请报告单记录血糖测试结果)用干棉签按压采血针眼处至血止,告知患者测出的血糖结果。
评价整理床单位,清理用物,洗手,用物及垃圾按医疗垃圾分类处理。
健康宣教操作流程熟练,动作流畅,遵守无菌操作原则,操作方法正确,未引起操作相关并发症,患者感觉舒适,体现人性化关怀与患者进行有效交流、沟通,时间2min,超时终止操作。
常见血糖仪的操作规程和注意事项一、操作规程1、确认血糖仪和试纸条是否同一产家,代码是否一致;2、熟读血糖仪操作说明和注意事项;3、常用采血部位为中指或无名指的指腹4、消毒时用酒精,待酒精完全干燥后,方可采血,不要用碘酒或者碘伏消毒;5、注意观察不同患者的血液流动性,冬季可活活手指,确保采血部位有足够的吸血量;6、采血中不要过度挤压创口,以免组织液渗出,影响血糖结果。
二、注意事项:1、血糖仪试纸条是否过期;2、血糖仪是否存在环境污染;3、试纸条保存是否妥当,有些误差是由试纸条的变质引起的,试纸条用后应将试纸条储存在原装盒内密闭保存,避免其受到测试环境的温度、湿度、化学物质等的影响。
血糖仪标准操作规程(SOP)一、标本采集规程测试前的准备:·检查试纸和质控液储存是否恰当:试纸和质控液贮存温度2-30℃。
·检查试纸条的有效期及条码是否符合:包装完好的血糖试纸有效期为24个月,开瓶有效期为6个月。
无需条码。
·清洁血糖仪:确保血糖仪处于关闭状态,请使用CloroxHealthcare•Bleach•Germicidal•Wipes(漂白杀菌卫生湿巾)清洁血糖仪表面。
在擦拭血糖仪,请确保无液体流入血糖仪内部。
·检查质控品有效期:开瓶后6个月丢弃。
·请确保充足的采血量,否则会造成测试结果偏低。
采血前手臂下垂10秒,轻轻按摩手指,或在温水中浸泡2分钟,促进血液循环。
根据角质层厚度选择合适的采血针。
·不要用力挤压采血部位,以免造成组织液溢出,稀释血样,造成测试结果偏低。
·为了保证血样的纯正,请丢弃第一滴血,使用第二滴血监测1、用75%乙醇擦拭采血部位,待干后进行皮肤穿刺。
备注:采血部位通常采用指尖、足跟两侧等末梢毛细血管全血,水肿或感染的部位不宜采血;用酒精消毒,或用温水和肥皂水把手指洗干净,待手指完全干燥后开始采血,否则残留的水或酒精会稀释采得的血样,造成测试结果偏低;不可用碘酒或碘伏消毒,碘参与氧化还原反应,造成测试结果偏高。
2、将一次性采血装置(安全针)拧开,靠在手指一侧,用力使一次性采血装置(安全针)压向手指,针打出后血液自然流出。
备注:轻轻从手指根部向前按摩,促使血液流出,有助于采集血滴。
指尖采血建议擦掉第一滴血,使用第二滴血进行测试。
每次测试时选择不同的扎针点,以减少采血疼痛,以获得至少0.5微升的圆形血滴。
备注:根据《关于规范医疗机构临床使用便携式血糖检测仪采血笔的通知》(卫医发〔2008〕54号)的要求,对不同患者进行监测血糖采血操作时,必须使用一次性采血装置;使用后的一次性采血装置不得重复使用。
二、血糖检测规程1、准备好血糖仪,试纸,一次性采血装置(安全针),75%乙醇,棉签等必需品。