Optical Detection of Two Intermediate Mass Binary Pulsar Companions
- 格式:pdf
- 大小:502.07 KB
- 文档页数:9
Customized. All distances. Immediate.* I n r e f e r e n c e t o c l i n i c a l o u t c o m e s a s c o m p a r e d t o m o n o v i s i o n . D a t a o n f i l e .Blending vision into better outcomes *.ZEISS PRESBYOND Laser Blended VisionZEISS PRESBYOND Laser Blended Vision A clear choice for patients with presbyopiaPRESBYOND ® Laser Blended Vision from ZEISS is an advanced method for treatingpatients with age-related loss of accommodation, also known as presbyopia. It offers the opportunity to achieve freedom from glasses by combining the simplicity and accuracy of corneal refractive surgery with the benefits of increased depth of field in retaining visual quality. As a surgical solution based on the naturally occurring spherical aberrations of the eye, this ZEISS software extends the scope of customized ablation beyond the limits of conventional monovision laser methods in several ways.Whether for its customized treatment profiles, its visual acuity at all distances, its indications range or its immediate impact, ZEISS PRESBYOND Laser Blended Vision is a clear treatment choice for the fast growing demographic of patients with presbyopia.ZEISS PRESBYOND Laser Blended Vision Customized. All distances. Immediate.ZEISS PRESBYOND Laser Blended Vision lets you greatly expand your ZEISS MEL ® 80 or MEL 90 excimer laser treatmentrepertoire and your patient base354Conventional monovisionWith conventional monovision treatment methods, the dominant eye is corrected for distance vision to almost plano while the non-dominant eye is corrected for near vision, usually up to -3.0 D. Optimal vision is achieved at distance and near range, requiring the brain to contend with two separate images at different levels of correction which not all patients cantolerate.Patients that do tolerate the method are left with an uncorrected gap in the intermediate range, the so-called “Blur Zone.” In addition to the fuzzy image, it can also cause other side effects such as reduced contrast sensitivity and stereoacuity.ZEISS PRESBYOND Laser Blended Vision As a physiologically optimized laser treatment method for patients with presbyopia, ZEISS PRESBYOND Laser Blended Vision represents the next stage in eye care excellence. Similar to conventional monovision, the dominant eye is corrected for distance vision to almost plano, whereas the non-dominant eye is corrected to be slightly myopic for near vision to -1.5 D. This micro-monovision strategy is further enhanced by a decisive difference: an increase in the depth of field of each eye using a wavefront-optimized ablation profile to create a continuous refractive power gradient for the whole optical zone of the cornea. This ZEISS software is an absolutely individualized treatment plan based on the preoperative spherical aberrations and the functional age of the eye. As a result, a customized fusion of the two images for near and distance vision is created for each patient – the so-called “Blend Zone.”Although similar to conventional monovision laser methods in terms of the workflow, ZEISS PRESBYOND ® Laser Blended Vision takes customized vision correction a step beyond, particularly with respect to the outcomes.Next-level vision correction beyond conventional monovisionThe unique Blend ZoneEssentially, the Blend Zone makes it easier for the brain to merge the images of both eyes, thereby achieving true binocular vision. In addition to excellent near and far vision, ZEISS PRESBYOND Laser Blended Vision patients also experience very good visual acuity and contrast sensitivity in the intermediate range.No increase in depth of fieldDominant eye±0.0 DNon-dominant eyeup to -3.0 DDISTANCEIncrease in depth of fieldIncrease in depth of fieldDominant eye±0.0 DNon-dominant eye-1.5 DNEARDISTANCENEAR67All distancesCustomizedImmediateOptimizing outcomes for patients with presbyopiaZEISS PRESBYOND Laser Blended VisionIndividualized ablationsZEISS PRESBYOND ® Laser Blended Vision is a truly customized solution for treating presbyopic patients. It incorporates preoperative wavefront data to fine-tune the depth of field for each eye individually. The functional age of the eye is also factored in. As a result, a personalized ablation profile is created per eye for optimized target refraction. The monovision component can be pre-adjusted for the patient’s tolerance level. Also, different optical zone sizes can be selected to account for the patient’s pupil size.Ideal for a growing demographic As an optimized laser method for age-related accommodation loss, ZEISS PRESBYOND Laser Blended Vision is ideally suited for serving the needs of patients 40-60 years of age – a fast-growing demographic group interested in sophisticated options. It is also one of the least invasive methods for addressing this target group.Familiar procedureFollowing the same workflow as conventional LASIK procedures, ZEISS PRESBYOND Laser Blended Vision combines the convenient binocular treatment planning of the CRS-Master ® with the proven comfort and workflow of the MEL ® 80 or MEL 90 excimer laser, all from ZEISS.Outstanding visual acuityBy customizing each eye individually,ZEISS PRESBYOND Laser Blended Vision provides excellent visual acuity for near and distance vision. Unlike traditional monovision methods, PRESBYOND Laser Blended Vision also offers good intermediate vision in the Blend Zone. According to clinical studies, there is virtually no loss of contrast sensitivity while stereoacuity is maintained. Also, side effects such as multiple images in one eye are almost eliminated.An all-natural approachZEISS PRESBYOND Laser Blended Vision is a physiologically optimized solution and a true binocular method for treating patients with presbyopia.Wide indication rangeZEISS PRESBYOND Laser Blended Vision is a proven and effective method for treating indications ranging from -8.0 D to +2.0 D, includingemmetropic and astigmatic patients (up to +2.0 cyl).Appropriate for most patientsA key advantage of ZEISS PRESBYOND Laser Blended Vision is that it is proven to be tolerated by more patients than conventional monovision. It is effective for treating up to 97% of all presbyopia-related forms of impairment as compared to only 59–67% for conventional monovision. Even patients with presbyopia also affected by emmetropia and astigmatism can be treated. In fact, it has the potential to achieve a far greater success rate than any comparable treatment along with giving patients the wow effect of being able to read without glasses the very same day. Thus, it positively impacts patients and refractive surgeons alike – visually for the former, economically for the latter.1-5 A competitive edgeZEISS PRESBYOND Laser Blended Vision allows practices already using a MEL 80 or MEL 90 excimer laser and CRS-Master from ZEISS to significantly expand their LASIK repertoire and increase the patient base. As such, this ZEISS software offers a decisive competitive advantage over other LASIK practices only specializing in monovision treatment methods.MEL 80 and MEL 90 from ZEISSType ArF excimer laser Wavelength 193 nmFrequency MEL 80: 250 HzMEL 90: FLEXIQUENCE ® 250 Hz / 500 Hz Dimensions (W x D x H)MEL 80:1550 mm x 800 mm x 1490 mmMEL 80 with patient supporting system: 3140 mm x 1800 mm x 1490 mm MEL 90:1360 mm x 730 mm x 1480 - 1700 mm MEL 90 with patient supporting system: 3230 mm x 2380 mm x 1700 mmSurgical microscope OPMI ® pico from ZEISS with integratedHD video camera Active eye trackerInfrared, pupil and limbus tracking, 1050 frames per second (fps), manual ablation center selection, automatic Pupil Center Shift Correction Beam dimensions0.7 mm FWHM (full width at half maximum), Gaussian beam profileVisuMax from ZEISSSystem components Patient supporting system, including platformIntegrated uninterruptible power supply (UPS)Surgical microscope with additional slit illumination Video camera with integrated recording Femtosecond laser parametersWavelength 1043 nm Laser pulse rate 500 kHzRecommended space requirements 180° setup with MEL 80 / MEL 90:4500 x 3800 mm90° setup with MEL 80 / MEL 90:4000 x 4000 mm89A perfect combination:The refractive system landscape of ZEISSTechnical dataDimensions (W x D x H)Max. 1060 x 420 x 1510 mmData transfer USB flash memory drive (USB memory stick)Data printoutVia network connection with Ethernet cable and optional network isolatorReferencesClaims made in this document are supported by information provided in the following publications:1. Reinstein DZ, Couch DG, Archer TJ. LASIK for Hyperopic Astigmatism and Presbyopia Using Micro-monovision With the Carl Zeiss Meditec MEL 80. J Refract Surg. 2009;25(1):37-58.2. Reinstein DZ, Archer TJ, Gobbe M. LASIK for Myopic Astigmatism and Presbyopia Using Non-Linear Aspheric Micro-Monovision with the Carl Zeiss Meditec MEL 80 Platform. J Refract Surg. 2011;27(1):23-37.3. Reinstein DZ, Carp GI, Archer TJ, Gobbe M. LASIK for the correction of presbyopia in emmetropic patients using aspheric ablation profiles and a micro-monovision protocol with the Carl Zeiss Meditec MEL 80 and VisuMax. J Refract Surg. 2012 [In Press].4. Reinstein DZ, Archer TJ, Gobbe M. Stereoacuity after Corneal Presbyopic LASIK in Myopic, Hyperopic and Emmetropic Patients. ESCRS Annual Meeting, Vienna, September 2011.5. Evans BJ. Monovision: a review. Ophthalmic Physiol Opt. 2007;27(5):417-439.CRS-Master from ZEISSSimple upgradePRESBYOND ® Laser Blended Vision is an optional software upgrade for the CRS-Master ® from ZEISS. It forms a perfect fit with the ZEISS MEL ® 80 or MEL 90 excimer laser, expanding the repertoire of customized refractive laser corrections far beyond the limits of conventional monovision methods.PRESBYOND Laser Blended Vision and CRS-Master from ZEISS are not intended for sale in the United States.Laser warning sign MEL 80/90Laser warning sign VisuMaxPrecise flapsThe ZEISS VisuMax ® creates flaps of a highly predictable thickness and of adjustable geometries for Femto-LASIK and ZEISS PRESBYOND Laser Blended Vision – the recommended treatment option for patients with presbyopia.Carl Zeiss Meditec AG Goeschwitzer Strasse 51–52 07745 JenaGermany/contacts /presbyond EN_34_1_3IIIPrintedinGermanyCZ-IX/217Thecontentsofthebrochuremaydifferfromthecurrentstatusofapprovaloftheproductorserviceofferinginyourcountry.Pleasecontactourregionalrepresentativesformoreinformation.Subjecttochangesindesignandscopeofdeliveryandduetoongoingtechnicaldevelopment.PRESBYOND,CRS-Master,MELandVisuMaxareeithertrademarksorregisteredtrademarksofCarlZeissMeditecAGorothercompaniesoftheZEISSGroup.©CarlZeissMeditecAG,217.Allrightsreserved.0297。
外文文献原稿和译文原稿Multiple single-chip microcomputer approach to fire detection and monitoring systemA.J. AI-Khalili, MSc, PhDD. AI-Khalili, MSc, PhDM.S. Khassem, MScIndexing term : Hazards, Design, Plant condition monitoringAbstract: A complete system for fire detection and alarm monitoring has been proposed for complex plants. The system uses multiple single chip architecture attached to a party line. The control algorithm is based on a two-level hierarchy of decision making, thus the complexity is distributed. A complete circuit diagram is given for the local and the central station with requirements for the software structure. The design is kept in general form such that it can be adapted to a multitude of plant configurations. It is particularly shown how new developments in technology, especially CMOS single chip devices, are incorporated in the system design to reduce the complexity of the overall hardware, e.g. by decomposing the system such that lower levels of hierarchy are able to have some autonomy in decision making, and thus a more complex decision is solved in a simple distributed method.1 Detection and alarm devicesA basic fire detection system consists of two parts, detection and annunciation. An automatic detection device, such as a heat, smoke or flame detector, ultraviolet orinfrared detectors or flame flicker, is based on detectingthe byproduct of a combustion. Smoke detectors, of both ionization and optical types, are the most commonly useddetector devices. When a typical detector of this type enters the alarm state its current consumption increasesfrom the pA to the mA range (say, from a mere 15pA in the dormant mode to 60 mA) in the active mode. Inmany detectors the detector output voltage is well defined under various operating conditions, such as thosegiven in Table 1. Themore sensitive the detector, themore susceptible it is to falsealarms. In order to control the detector precisely, either of the following methods is used: a coincidence technique can be built into the detector, or a filtering technique such that a logic circuit becomes active only if x alarms are detected within a time period T. The detection technique depends greatly on the location and plant being protected; smoke detectors are used for sleeping areas, infrared or ultraviolet radiation are used when flammable liquids are being handled, heat detectors are used for fire suppression or extinguishing systems. In general, life and property protection have different approaches.Alarm devices, apart from the usual audible or visible alarms, may incorporate solid state sound reproduction and emergency voice communication or printers that record time, date, location and other information required by the standard code of practice for fire protection for complex plants. Heaviside [4] has an excellent review of all types of detectors and extinguisher systems.1.1 Control philosophy and division of labourOur control philosophy is implemented hierarchically. Three levels of system hierarchy are implemented, with two levels of decision making. There is no communication between equipment on the same level. Interaction between levels occurs by upwards transfer of information regarding the status of the subsystems and downwards transfer of commands. This is shown in Fig. 1 where at level 1 is thecentral station microcomputer and is the ultimate decision maker (when not in manual mode). At level 2 are the local controllers, which reside in the local stations. At level 3 are the actual detectors and actuators. A manual mode of operation is provided at all levels.Information regarding the status of all detectors is transmitted on a per area basis to the local controllers. Their information is condensed and transmitted upward to the central microcomputer. Transfer of status is always unidirectional and upwards. Transfer of commands is always unidirectional and downwards, with expansion at the local control level. This approach preserves the strict rules of the hierarchy for exact monitoring detection and alarm systems associated with high risk plants.The classification of the two layers of controls is based upon layers of decision making, with respect to the facts that(a) When the decision time comes, the making and implementation of a decision cannot be postponed(b) The decisions have uncertainty(c) It will isolate local decisions (e.g. locally we might have an alarm although there may be a fault with the system)2 General hardwareI :Fig. 2 depicts our design in the simplest of forms. The system uses an open party line approach with four conductor cables going in a loop shared by all the remote devices and the control panel. This approach is simple in concept and is economically feasible. However, one major disadvantage is the dependency on a single cable for power and signaling. In cases where reliability is of extreme importance, two or even three cables taking differentroutes throughout the system may be connected in parallel. Fig. 3 gives the driver circuitry required to derive an expandable bus. This design takes advantage of recent advances in the single chip microcomputer technology to reduce the interface betweenthe central station and the local stations.2. 1 Central control taskA central unit provides a centralized point to monitor and control the systemactivities. In the system to be described the central control unit serves a fivefold purpose.(i) It receives information from the local stations and operates the alarms and other output devices.(ii) It notifies the operator in case of system malfunction.(iii) It provides an overall system control manual and automatic.(iu) It provides a system test point of local stations and itself.(u) It provides a central point for observation, learning and adaptation.2.2 Local stationsThe local stations can take local decisions regarding recognition of a risk situation, and act independently on local affairs. In this technique we depend on ‘load-type coordination’, e.g. the lower level units recognize the existence of other decision units on the same level; the central or the top level provides the lower units with a model of the relationship between its action and the response of the system.It is evident that a powerful machine is required at this stage so that all the required functions can be implemented. The availability of the new generation of microchips makes this architecture a feasible solution.A single chip microcomputer was chosen over discrete digital and analogue devices to interface to the field devices and to the central microcomputer. This is the main reason that previously this approach was not feasible.In selecting the microcomputer for the local stations, the criterion was the requirement for a chip which contains the most integration of the analogue and digital ports required for the interface and the utilization of CMOS technology owing to remoteness of the local stations. The choice was the Motorola 68HC11A4, for the following reasons:(a) It is CMOS technology; this reduces power consumption.(b) It has a UART on board; this facilitates serial communication.(e) It has an a/d converter on board; this eliminates an external A/D.(d) It has 4K of ROM, 256 bytes of RAM, 512 bytes of EERROM with 40 1/0 lines and a 16 bit timer; this satisfied all our memory and 1/0 requirements at the local station side.3 System implementationThe local station: Fig. 3 is the block diagram of the circuit used to utilize the MC68HCllA4 as a remote fire detecting circuit while Fig. 4 illustrates the samecircuit in an expanded form. It can be seen that the single microcontroller can be used to monitor more than one detector, thus reducing system cost.The loop power supply, which is usually between 28 and 26 V, is further regulated by a 5 V 100 mA monolithic low power voltage regulator to supply power to the microcontroller. The onboard oscillator, coupled with an external crystal of 2.4576 MHz, supplies the microcontroller with its timing signal which is divided internally by four to yield a processor frequency of 614.4 kHz, which is an even multiple of the RS 232 [7] baud rate generator. In this Section the term ‘supervised input or output’ will be used to mean that the function in question is monitored for open- and short-circuit conditions in addition to its other normal functions. More information can be found in Reference 9.4 Main loop5 ConclusionThis paper describes the development of a large scale fire detection and alarm system using multi-single chip microcomputers. The architecture used is a two-level hierarchy of decision making. This architecture is made possible by the new CMOS microcontrollers which represent a high packing density at a low power consumption yet are powerful in data processing and thus in decision making. Each local station could make an autonomous decision if the higher level of hierarchy allows it to do so. It has been tried to keep the system design in general format so it can be adapted to varying situations. A prototype of the described system has been built and tested [10]. The control part of the central station is implemented with a development card based on MC 68000 microprocessor (MEX 68KECB, by Motorola), which has a built-inmonitor called Tutor. The application programs were developed using the features provided by this monitor. The local stations’ controll ers were designed using the MC 68705R3, single-chip microcontroller.7 References1 ‘Fire protection guidelines for nuclear power plants’, US NRC Regulatory Guide 1.1202 BAGCHI, C.N.: ‘A multi-level distributed microprocessor system for a nuclear power pl ant fire protection system controls, monitoring, and communication’, IEEE Trans., 19823 PUCILL, P.M.: ‘Fire hazard protection, detection and monitoring systems’, Sea. Con, 2, Proceedings of Symposium on ADV in offshore and terminal measurement and control systems, Brighton, England, March 1979, pp. 353-3634 HEAVISID, L.: ‘Offshore fire and explosion detection and fixed fire’. Offshore Technological Conference, 12th Annual Proceedings, Houston, Texas, May 1980, pp. 509-5225 CELLENTANI, E.N., and HUMPHREY, W.Y.: ‘Coordinated detection/communication approach to fire protection’, Specify: Eng.,6 ‘Motorola Microprocessors Data Manual’ (Motorola Semiconductor Products, Austin, Texas, USA)7 Electronic Industries Association : ‘Interface between data terminal e quipment and data communication equipment employing serial binary data interchange’ (EIA Standard RS-232, Washington, DC, 1969)8 MESAROVIC, M.D., MACKO, D., TAKAHARA, Y.: ‘Theory of hierarchical multilevel systems’ (Academic Press, 1970)9 KASSEM, M.: ‘Fire alarm systems’, MSc. thesis, Dept. of Elec. & Comp. Eng., Concordia University, Montreal, Canada, 198510 LIE, P., and KOTAMARTI, U.: ‘The design of a fire alarm system using microprocessors’, C481 Project, Dept. of Elec. and Comp. Eng., Concordia Unive rsity, Montreal, Canada, 1986译文基于单片机的火灾探测和监控系统A.J. AI-Khalili, MSc, PhDD. AI-Khalili, MSc, PhDM.S. Khassem, MSc关键词:危险,设计,设备状态监测摘要:火灾探测及报警监控已成为一个复杂而完整的体系。
专利名称:SCANNING DEVICE发明人:DIETMAR UHDE,GERHARD WEISSMANN 申请号:AU5939090申请日:19900629公开号:AU5939090A公开日:19910222专利内容由知识产权出版社提供摘要:An optical scanning device with two holding devices for a compact disk player is fastened by means of one holding device in such a manner that the scanning device or its objective lens can move parallel to the optical axis and parallel to the surface of the disk. To ensure parallel guiding of maximum precision, the holding device is composed of two end parts (E), two intermediate parts (Z) and at least one central part (M). The axes of rotation of the joints (K1), each of which connects an end part (E) to an intermediate part (Z), are perpendicular to the axis of rotation of the joint (K2) which connects the central part (M) on both sides to an intermediate part (Z). The holding device can be made from a piece of plastic with notches at the joint points which form bending points that function as joints. Application to mechanical and optical scanning devices, for example, for compact disk players.申请人:DEUTSCHE THOMSON-BRANDT GMBH更多信息请下载全文后查看。
·140·2型糖尿病视网膜病变临床前期的微血管与神经组织异常表现王燕华 陈子林 广东医科大学 广东湛江 524000摘 要:糖尿病视网膜病变(diabetic…retinopathy,…DR)是糖尿病最常见及严重的并发症之一,是导致劳动力人群视力损害的首位原因。
大量研究表明,在发现临床上可见的糖尿病性视网膜病变之前,可能已经出现视网膜微血管损伤及视网膜糖尿病神经组织病变。
发现早期的隐匿病变,及早采取防治措施,可以延缓视功能损害,有助于改善糖尿病患者的生活质量,本文就2型糖尿病患者视网膜病变早期的微血管病变及神经组织病变的关系与异常表现的特点作综述分析。
关键词:2型糖尿病 临床前期糖尿病视网膜病变 视网膜微血管病变 视网膜神经组织病变DR是糖尿病最常见的并发症之一,是造成全球劳动人群视力不同程度损害的主要原因,也是我国防盲治盲的重点之一。
按2002年版DR的国际临床分级将散瞳眼底检查无异常定为无明显DR(NO-DR,NDR)。
研究表明,视网膜微血管损伤及视网膜糖尿病神经组织病变可能在临床上可见的糖尿病性视网膜病变之前已经发生并且二者关系紧密[1][2]。
目前临床上对DR的干预主要在出现视力损害的中后期阶段,当糖尿病患者出现临床上可见的视网膜病变后,视网膜结构和视力已经受损并难以逆转,且促使DR进展的风险也随之增加。
因此,更深入地了解2型糖尿病患者视网膜病变临床前期的微血管病变及神经组织病变可能会为DR提供更早和更有效的预防策略。
1糖尿病视网膜微血管病变与糖尿病视网膜神经变性的关系DR的发病机制目前尚不十分明确,大量研究表明,视网膜微血管损伤可能在临床上可见的糖尿病性视网膜病变之前已经发生[2]。
而在上世纪90年就有学者提出视网膜神经细胞凋亡发生在糖尿病的早期。
现在趋向认为DR是神经血管性疾病而非仅是微血管疾病,有学者提出了视网膜微血管元件这一概念,内皮细胞、周细胞与神经元及胶质细胞组成视网膜神经血管单位[3]。
M i c r o s c o p y f r o m C a r l Z e i s sSteREO Discovery.V8A New View of ThingsBrilliant Entry into the Class ofSophisticated StereomicroscopesSteREO Discovery.V8: Enhanced ViewingNew high-performance optics – this is the outstanding performance feature that Carl Zeiss has focusedon with the development of its latest entry levelmodel in the sophisticated stereomicroscope class.The SteR EO Discovery.V8 impresses with enhanced resolution, increased contrast and, most notably, a perceptibly improved stereoscopic impression. As aresult, it offers a image brilliance that is without equalin this class of instrument. For a visible increase in information in all biomedical and industrial applica-tions, the time has come for a new view of things.SteREO Discovery.V8at a glance:• Supreme ease of operation and ergonomicviewing posture• Zoom range of 8 : 1• Optimized optics design for a visible increasein image information• Manual focusing drive with adjustable clickstops• Manual and motorized stands with high stability• Generous specimen space with high workingdistance• Illumination and contrast methods based oncold light and LEDPCBLaterally grazing reflected light Objective: Plan S 1.0x Magnification: 16x*Mouth parts of the common houseflyOblique illumination in brightfieldtransmitted lightObjective: PlanApo S 1.0xMagnification: 80x*Wafer structureDarkfield reflected lightObjective: PlanApo S 1.5xMagnification: 120x*SteREO Discovery.V8stereoscopic impression of the microscopic image.Even on our SteREO Discovery.V8 entry level model.Another area we focused on during practical realiza-tion was the systematic minimization of stray light for the entire optical system. For exceptionally brilliant contrast and a new image quality with greater infor-mation content.The innovative simultaneous design process during optical modeling has resulted in a standardized opti-cal concept for all SteREO microscopes. For signifi-cantly improved resolution and a perceptibly betterThe Optical System: More Thanthe Sum of its Individual ComponentsToday, anyone developing a stereomicroscope that sets new standards with its optical system has to work constructively at the very limits of physical feasibility,taking full advantage of every new possibility offered by state-of-the-art optical design. With experience and innovativeness, you don’t have to look any further than the optical systems from Carl Zeiss.3 ranges of high-quality objectivesAchromat S: high-contrast images with a pronounced stereoscopic impression Plan S: flat,distortion-free object fieldsPlanApo S: precisely detailed resolution with no color fringes Parfocally harmonized for needle-sharp images over the entire magnification range from 1x to 8x: the new zoom body of SteREO Discovery.V8a.b.c.The Illumination: Show Your Specimen in a New Cold LightDesigned for slimline, space-saving light guides, opti-mized for flicker-free live images on the monitor,providing constant light output even if the line volt-age fluctuates, and with ventilation as quiet as a whisper – the high-intensity CL 1500 ECO cold lightThe quality of the illumination – this is all-important for contrasting in stereomicroscopy. The new fiber-optic CL 1500 ECO cold light source with its wide range of light guides and accessories offers you a variety of opportunities for highlighting your struc-tures perfectly.The fiber-optic CL 1500 ECO cold light source illuminates the specimen precisely with intensive infrared-free light.Here with a twin-arm goose neck for oblique reflected light with a targeted shadow effect.SMD-Board with white solder resist Reflected light with different light guides a.Linear slit light for lateral grazing light b.Fiber-optic annular slit illuminator for shadow-free 360°reflected lightc.Annular slit illuminator with polarization filter device to minimize reflective glare Objective: Plan S 1.0x Magnification: 15xContrasts with variable optimization in brightfield,darkfield and oblique light: the transmitted light equipment Ssource outperforms conventional fiber-optic systems thanks to several practical advantages, and offers excellent performance at a superb price!Incidentally, lamps and filters can be changed quick-ly and conveniently. Even when stacked.a.b.c.Or in a Completely Different Light: With White LEDsEach of the annular VisiLED illuminators is made up of eight LED segments that can be switched variably.A further advantage of noise-free light sources: upProviding the ability to change rapidly from shadow-free annular illumination to lateral oblique light, con-trast optimization through turning of oblique light around the specimen, continuous rotation of the illumi-nation for a stereoscopic impression of the object in the live image – and all this at the push of a button! The list of new contrasting possibilities offered by the VisiLED illumination system with its white LEDs is endless!The VisiLED HCT contrast stage offers a wide range of contrasts.It contains separate LED illuminators for brightfield and darkfield,and sliders for finely adjusta-ble oblique light.The LEDs are controlled using the MC1500 multicontroller.Mouth parts of the common houseflyTransmitted light with VisiLED HCT contrast stage a.Brightfieldteral darkfieldc.Oblique brightfield illumination Objective: PlanApo S 1.0x Magnification: 80xQuiet,durable and offering the best in daylight quality: VisiLED LED illuminationto 4 illumination settings can be stored and repro-duced again at any time.The MC1500 multicontroller of the VisiLED system allows control of reflected, transmitted and blended light.1.2.3.4.1. Interface with digitalimage worlds: documentationSteREO Discovery.V8 creates a connection for a variety of digital photo and video cameras,via various phototubes,with interface 60N.For the simple documentation of stereomicro-scopic images,consumer cameras,with their good price/per-formance ratio,are often recommended.Anyone wishing to satisfy higher demands should use the high-resolution AxioCam microscope cameras and the AxioVision imaging software from Carl Zeiss.2. Brilliant fluorescence: PentaFluar S PentaFluar S is the name of the retrofittable fluorescence equip-ment for stereomicroscopes belonging to the SteREO Discovery family.With up to five different filter blocks in the magazine and special high-performance light sources,this is an outstand-ing addition for contemporary fluorescence applications in stereomicroscopy.3. Better in position: the binocular ergo-phototube S 5-45°Ergonomics is also about choosing a relaxed sitting position when operating a microscope.The viewing angle and height have to coincide.The ergotube allows a free choice of viewing angle between 5 and 45 degrees.Intermediate tubes and two working positions for the eyepiece clamps vary the viewing height.4. Vertical 2D impression:the objective slideA must for documentation with subsequent image analysis,e.g.digital object measurements: the S/doc objective slide for the SteREO Discovery family of microscopes.Positioned directly beneath the zoom body,it enables the objective to be shifted precisely under one of the stereoscopic beam path for a vertical view of your specimen.The Expandable Platform:Flexible For a Variety of ApplicationsA typical feature of stereomicroscopes is their mod-ular system design. Equipped with intelligent inter-faces and fully integrated into the Carl Zeiss systems,SteREO Discovery.V8, with its comprehensive range of accessories, offers you a great deal of freedom in terms of organizing your workplace to suit your own practical needs.System OverviewSteREO Discovery.V8:The Technical DataObjectivesEyepiecesCarl Zeiss Micro scopy GmbH 07745 Jena, Germany ********************www.zeiss.de/stereo-discoverye c t t o c h a n g e .e d o n e n v i r o n m e n t a l l y -f r i e n d l y r ,b l e a c h e d w i t h o u t t h e u s e o f r i n e .。
MAIA 01MA N UA LE D ’U S O I NSTRU CTION MAN UALMO D E D ’EM P LO I GEB RAU C HSAN LEITU N G MAN UA L DE U SO Dispositivo Medico conforme alla direttiva 93/42/CE FARO SPA Ornago (Italy)L A M PA DA D E NTA L E A LE D DENTAL LED LIGHTMAIA01MA N UA LE D ’U S O I NSTRU CTION MAN UAL MO D E D ’EM P LO I GEB RAU C HSAN LEITU N G MAN UA L DE U SO Dispositivo Medico conforme L A M PA DA D E NTA L E A LE D DENTAL LED LIGHT 01MAN UA L DE U SO Dispositivo Medico conformealla direttiva 93/42/CE FARO SPA Ornago (Italy)LED口腔无影灯使用说明书医疗器械-符合93/42/CE标准 FARO SPA Ornago (Italy)始于1948年 经验与创新strumentidentali.itL a m p a d a d e n t a l e M A I Acontain instructions that must beAIAMelatnedadapmaLthe head allow the light beam to bedevices and precautions. Faro(for children) protective Faro eyeware are individual protection devices for protecting the eyes of the patientMAIA, DO NOT uSE detergents-SODIuM HYDrOXIDEL a m p a d a d e n t a l e M A I A- The doctor shall use disposable protection devices on the handles of the lamp or guarantee its sterilisation for versions that envisage this option.- Divide materials by type (e.g. ferrous, rubber, plastic, etc.). For scrapping and disposal of materials, comply with local regulations in force, even resorting to specialised firms that are recognised and authorised.- The packaging of the lamp is suitable to adequately protect it from penetration of external agents.- The lamp in its original packaging can be transported or kept in storage for a period of 15 weeks, if compliance with environmental conditions specified below is assured:- room temperature between -20° and 70°C - relative humidity between 10% and 90% - Atmospheric pressure between 500 a 1060 mBar- The lamp must be used in the following environmental conditions: - Temperature between 10 and 40°C - relative humidity between 30 and 75%-Atmospheric pressure between 700 and 1060 mBarstrumentidentali.itL a m p a d a d e n t a l e M A I AL a m p a d a d e n t a l e M A I AELEcTrOMAGNETIc IMMUNITYManufacturer’s guidelines and statement – Electromagnetic immunityThe lamp MAIA is designed to function in the electromagnetic environment specified below. The client or user must ensure its usein the said environment.Immunity test ComplianceElectromagnetic environment - Guidelines Electrostatic discharge (ESD)IEC/EN61000-4-2± 6kV contact± 8kV airThe floor must be in wood, concrete or ceramic. If the floor is covered with synthetic material, relative humidity should be at least 30%.The quality of supply network voltage should be typical of com -mercial or hospital environments.The quality of supply network voltage should be typical of com -mercial or hospital environments.The quality of supply network voltage should be typical of com -mercial or hospital environments. If the user of the lamp MAIA requires continuous use even without a supply network, use an uninterruptible power supply.Level of magnetic field at the network frequency typical of com -mercial or hospital environments.Electrical fast transient/burst IEC/EN61000-4-4SurgeIEC/EN61000-4-5V oltage dips, short interruption and voltage variation IEC/EN61000-4-11Power frequency magnetic field IEC/EN61000-4-8± 2kV power supply± 1kV for input/output lines ± 1kV differential mode ± 2kV common mode < 5% Ut for 0,5 cycle 40% Ut for 05 cycle 70% Ut for 25 cycle <5% Ut for 5 sec.3A/mConducted immunity IEC/EN61000-4-6Conducted immunity IEC/EN61000-4-63Vrms 150kHz to 80MHz (for non life-supportingequipment)3Vrms 80MHz to 2.5GHz (for non life-supportingequipment)Portable and mobile RF communication devices should not be used near any part of the dental unit, including cables, unless they comply with recommended distances calculated with the applicable equation for transmittor frequency.Recommended distances:d = 1.2√Pd = 1.2√P from 80 Mhz to 800 MHz d = 2.3√P from 800 Mhz to 2.5 GHzP is the maximum nominal power issued by the transmittor in Watts (W) depending on the manufacturer of the transmittor, and d is the recommended distance in metres (m).The intensity of the fixed RF transmittor field, as established in an electromagnetic investigation of site a, could be less than the compliance level of each frequency interval.There can be interference near devices marked with the following symbol:Note: Ut is the power-line voltageNote 1: The highest frequency interval is applied at 80 MHz and 800Mhz.Note 2: These guidelines might not apply to all situations. Electromagnetic propagation is influenced by absorption and reflection of structures, objects and persons.a) ISN bands (industrial, scientific and medical) between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz and 40.66 MHz to 40.70 MHz.b) Compliance levels in ISN bands between 150 kHz and 80 MHz and 80 MHz to 2.5 GHz present a decreasing probability of portable transmission devices causing interference if inadvertantly taken to the patient area.Therefore, an additional 10/3 factor has been incorporated into the formula used to calculate the distance between transmittors.c) Field intensities for fixed transmittors such as base stations for radiotelephones (mobiles and cordless) and cellular mobile radios on land, CB user equipment, AM and FM transmittors and TV transmittors cannot be theoretically estimated with precision. To establish an electromagnetic environment caused by fixed RF transmittors, an electromagnetic investigation of the site should be considered. If field intensity measured at the site of use of the dental unit exceeds the aforementioned applicable compliance level, normal function of the lamp should be monitored. If any abnormal performance is noticed, additional provisions such as a different orientation or position of the lamp might be necessary.d) The field intensity in an interval of frequencies from 150 kHz to 80 MHz should be less than 3 V/m.strumentidentali.itL a m p a d a d e n t a l e M A I Afield without 1fig. 1fig. 2fig. 45fig. 3L a m p a d a d e n t a l e M A I Alamp assembly, dental chair version- Install the lamp by inserting the terminal lamp pin into the specific hole on the dental chair.L a m p a d a d e n t a l e M A I A“MAIA” LAMP WITH SWITCH On/Off/Adjustment- To switch on or off, press and release the command lever to the left or right.- Adjustment:“MAIA” LAMP WITH PROXIMITY SWITCH On/Off- To turn the lamp on and off, place your hand close to the sensor, within a maximum distance of 3 cm. When the command is given, an acoustic signal will be heard (1 beep).- For reached, On/Off/Adjustment- To turn the lamp on and off, press and release button “A”. - Adjustment:a) When the minimum light intensity is obtained, you will hear an acoustic signal (1 beep).L a m p a d a d e n t a l e M A I AvIDEO-DIAGNOSTIC fuNCTIONThe Maia lamp has a function that enables it to be used when filming with a television camera and/or using diagnostic instruments (Diagnodent and laser, for example) without causing interference that could alter the diagnostic result.This function is only in manual switch equipped versions.Activation of the Video-Diagnostic function:1. Switch on the Maia dental lamp (a beep will be heard when the control is used).2. Release the control.3. Use the control again to reach the minimum light intensity (a beep will be heard when minimum intensity is reached) then without releasing the control keep it active for at least 4 seconds.4. A beep is emitted as confirmation, the light intensity rises to the maximum level and the Video-Diagnostic function is ACTIVE.If the lamp does not react as described in point 4 above, repeat the whole procedure from point 1.Deactivation of the Video-Diagnostic function:1. Switch on the Maia dental lamp (a beep will be heard when the control is used).2. Release the control.3. Use the control again to reach the minimum light intensity (a beep will be heard when minimum intensity is reached) then without releasing the control keep it active for at least 4 seconds.4. A beep is emitted as confirmation, the light intensity rises to the maximum level and the Video-Diagnostic function is DEACTIV ATED.If the lamp does not react as described in point 4 above, repeat the whole procedure from point 1.Dimming of the light intensity with the Video-Diagnostic function ACTIV ATED:With the Video-Diagnostic function activated, the regulation of the light intensity is modified from a continuous variation to a stepwise variation.Two intermediate levels of light intensity can be chosen between the maximum and minimum.Procedure:1. Switch on the Maia dental lamp (a beep will be heard when the control is used)2. Release the control.3. Use the control again to reduce the light intensity and release the control at the intensity desired.Note:• On reaching minimum intensity, a beep will be heard.• When the dental lamp is switched on again it will return to the maximum light intensity (a beep will be heard when the control is used).strumentidentali.itL a m p a d a d e n t a l e M A I Ado not use detergents-disinfectants SODIuM HYDrOXIDE cause: risk of breakage of plastic ClEANING THE DISHES “3”STErIlIZING THE HANDlESTo remove the handle, unscrew button at 121°/134° C for a total of 200 sterilisation cycles.OTHEr PArTS Of THE lIGHTL a m p a d a d e n t a l e M A I Asurfaces with the specific product p roduct f or c leaning o r d isinfection (yearly)2: <100AcOUSTIcL a m p a d a d e n t a l e M A I A: 17÷24v ac ± 10% - 50/60 Hz 22÷35v dc: 230 v 50/60 Hz : 9v A: 2 x T250mA1 250v : Class II deviceOPTICAl fEATurES Size of light spot : 170 mm x 85mmLux: 3000*-35000* lux @700mm Colour temperature: 5000 K。
眼用镜片论文:渐进多焦点眼镜片的优化设计【中文摘要】渐进多焦点眼用镜片(简称渐变镜片)与各类矫正近视或老视镜片相比具有许多优点。
渐进多焦点眼用镜片能同时满足视远与视近的需求,又避免了双光镜的视远与视近转换时视觉断裂等缺陷,目前渐进多焦点眼用镜片的应用日渐广泛。
在设计过程中,初始设计的渐变镜片往往在某些方面存在不足,如视远区不够开阔,通道过窄,像散过大等,本文提出两种对初步设计的渐变镜片的面形数据优化设计方法。
(一)解偏微分方程反演优化渐进多焦点眼用镜片的优化设计。
根据初始设计的渐变镜片面形数据反演得到镜片的光焦度分布,通过优化因子法和光焦度平移法得到优化后的光焦度分布,然后解偏微分方程得到优化后的镜片面形。
针对两种光焦度优化方法给出两个优化步骤和优化实例,结果表明:优化因子法主要能有效增加渐变通道的宽度。
光焦度平移优化法能明显扩展视远区的视觉清晰范围。
(二)平均曲率流优化渐进多焦点眼用镜片的优化设计。
提出了一种减小渐进多焦点眼用镜片像散的优化设计思想,即渐进镜片的渐变表面越接近球面,则渐进表面的像散越小。
阐述了平均曲率流原理并用平均曲率流使渐变表面的优化区域更加趋近球面。
给出了具体的优化算法、优化步骤和一个优化实例,结果表明:平均曲率流优化方法能够有效减少渐变表面的最大像散。
提出的优化设计方法既保留初始镜片中重要区域的光学性质不变又能减小镜片像散。
这些优化设计方法适于用任何设计方法设计的渐进多焦点眼镜片面形数据的优化。
【英文摘要】Since the progressive addition lens (PAL) has some advantages over bifocal and trifocal lens in correcting myopia and presbyopia in particular, it is more accepted by consumers. The PAL provides a continuous range of accommodative powers, and avoids having a visible dividing line or image jump like bifocal lens. The PAL has become a popular approach to the management of myopia and presbyopia in resent years. In the design process, the initial designed PAL often has some shortcomings, such as the clear region on distance-vision is not broader, the intermediate zone is narrower and the astigmatism is larger. Two optimizing design methods are proposed to optimize the preliminary designed surface shape data of the PAL.1. Optimizing the PAL by solving the partial differential equationThe power plot is obtained base on the inversion of the preliminary designed surface shape data. The power plot is changed by adjusting the optimizing factor of an optimization function. Otherwise, the optimized plot of the mean power can also be derived by shifting the mean power of one selected region to another and then by interpolating and smoothing. A partial differential equation of the elliptic type is founded based on the changed mean power. The optimized vector heights of the surface are solved out. Compared with theoriginal lens, the clear region on the intermediate zone become wider by the optimization factor method and the clear region on distance-vision is broaden by the shifting power method.2. Optimizing the PAL by mean curvature flowAn optimizing design concept is presented that the more the surface of the progressive lens becomes closer to spherical by the process of mean curvature flow, the more the astigmatism is smaller. The principle of mean curvature flow operator and the optimizing algorithm are illustrated. An initial designed progressive lens and an optimized lens by this method are manufactured and tested. Compared with the original lens, the optimized lens has smaller maximum astigmatism and larger clear region on the distance-vision zone.All the optimizing methods presented in this paper can reduce the undesirable astigmatism in some regions of the progressive lens surface while retaining desirable optical features of the progressive lens. And the optimizing approaches can optimize the progressive lens given by any design method.【关键词】眼用镜片渐进多焦点优化像散【备注】索购全文在线加好友:1.3.9.9.3.8848同时提供论文写作一对一指导和论文发表委托服务【英文关键词】lenses progressive additionoptimization astigmatism【目录】渐进多焦点眼镜片的优化设计中文提要4-5Abstract5-6第一章引言9-201.1 人眼的结构和功能9-111.1.1 人眼的结构9-101.1.2 人眼的功能10-111.2 人眼屈光不正的类型及矫正原理11-171.2.1 远视及其矫正原理12-131.2.2 近视及其矫正原理13-141.2.3 散光及其矫正原理14-151.2.4 老视及其矫正原理15-171.3 视力矫正的一般方法17-181.4 课题研究的意义18-20第二章渐进多焦点眼镜片的基本理论20-262.1 渐进多焦点眼镜片的基本结构和功能20-222.2 渐进多焦点眼镜片的性能评价方法22-232.3 渐进多焦点眼用镜片设计要求和设计原则232.4 渐进多焦点眼镜片的主要设计方法概述23-252.4.1 渐进多焦点眼镜片的设计分类23-242.4.2 当前渐进多焦点眼镜片的一些主要设计方法24-252.5 渐进多焦点眼镜片设计存在的问题25-26第三章解偏微分方程反演优化渐进多焦点眼镜片26-403.1 反演优化法的基本原理26-273.2 优化因子法27-343.2.1 优化因子函数27-283.2.2 优化因子法优化实例及结果分析28-343.3 光焦度平移法34-393.3.1 光焦度平移法的理论基础34-363.3.2 光焦度平移法优化实例及结果分析36-393.5 本章小结39-40第四章平均曲率流优化渐进多焦点眼镜片40-524.1 平均曲率流原理40-434.2 平均曲率流优化渐进多焦点眼镜片的方法设计43-464.3 优化实例及结果分析46-504.3.1 优化前后镜片面形数据光焦度和像散分析46-494.3.2 优化前和优化后镜片加工样品光学参数测量结果分析49-504.4 本章小结50-52第五章总结与展望52-545.1 总结525.2 创新点52-535.3 展望53-54参考文献54-56攻读学位期间公开发表的论文及科研成果56-57致谢57-58。
COLLECTING AND HANDLING OF NATURAL GAS SAMPLES FOR CUSTODY TRANSFERChristopher L. Grant, Dr. Darin L. George, and Jacob L. ThorsonSouthwest Research InstituteINTRODUCTIONThe American Petroleum Institute (API) Manual of Petroleum Measurement Standards (MPMS) Chapter 14.1, Collecting and Handling of Natural Gas Samples for Custody Transfer, provides practical guidance for gas sampling in custody transfer applications. Though gas sampling should not be performed without fully reading the standard, this paper is designed to provide supplemental information, illustrative examples, and guidelines for how best to use API 14.1. Specific sections of the standard are highlighted and broadened with additional detail. Special emphasis is put on the accurate sampling of an unknown gas stream.THE IMPORTANCE OF THE HYDROCARBON DEW POINTIn natural gas sampling applications, it is important to be familiar with the hydrocarbon dew point (HDP) and to be aware of how it can affect your sample’s accuracy. This is different from the water dew point, which is another topic of concern, but will not be discussed in this paper. An example image of the HDP as viewed in a chilled mirror device is shown in Figure 1. The HDP is defined as the temperature for a given pressure at which hydrocarbon condensation begins (1). The HDP is often plotted on a temperature versus pressure chart as shown with the dashed blue line in Figure 2. To the right of the HDP curve and above the critical temperature, no liquids will be present. As the pressure-temperature state moves to the left of the HDP curve, liquids will condense, and a natural gas sample will contain gas and liquid phases simultaneously.Figure 1. Gas at the Hydrocarbon Dew Point in a Chilled Mirror DeviceIn the image above, faint droplets are visible on the mirror with an iridescent ring around the perimeter of the mirror. This indicates thatthe HDP has been reached (1).Note that the curve passes through or near common pipeline operating temperatures and pressures in a variety of locations. A common process that causes a gas to condense is known as the Joule-Thomson (J-T) effect and is caused by a gas cooling as its pressure drops. This process can be encountered in sampling systems if a gas sample flows through a restriction such as a partially open needle valve. If this is the case, the gas and equipment must be warmed enough to counteract the J-T cooling effect. Many of the guidelines outlined in API Chapter 14.1 are aimed at avoiding this transition during sampling.The HDP curve shown in Figure 2 below is an approximation, and the true HDP curve can be difficult to predict accurately for some pipeline gases. Because different components condense at different temperatures and rates, crossing the HDP curve will change the density, heating value, and many other properties of the remaining gas. Generally, heavy hydrocarbons condense before lighter components causing a drop in the measured heating value and density – two key measurements in custody transfer applications.Figure 2. Pressure vs Temperature Plot of a Hydrocarbon Dew Point Curve An example plot of a phase boundary curve for a typical natural gas mixture is shown. The blue line represents the HDP curve (2). Care must be taken in the handling of a sample after it is collected to avoid condensation and distortions in the sample properties. If a gas sample has changed phase within the sample cylinder, and the cylinder has not been opened, the condensation process may be reversed. This is accomplished by heating the sample cylinder above the predicted HDP for enough time to revaporize all of the condensation. The standard requires that the sample cylinder be held at 30°F above the HDP for at least two hours. This revaporization must be conducted before any liquid or gas has been removed from the sample cylinder, or the gas sampled by the GC and that remaining in the cylinder will both be distorted.As an example of the effects of condensation, consider a mix of 1,500 Btu/scf gas with the components shown in Table 1. This gas is rich but is well within the range of natural gases found upstream of processing stations. If this gas were at 75 psia, the HDP would be roughly 91°F. A drop of 50°F below the hydrocarbon dew point would cause condensation and would cause the remaining gas to have a heating value 70 Btu/scf lower than the sampled gas stream. This would coincide with only a 3% drop in vapor fraction. If this occurred in a 300 cc cylinder at 75 psia, the condensed liquid would be less than 1/1,000 of a pound, a small enough amount to easily avoid detection (2). If this condensation were to occur in a crevice or other difficult to clean area, it could contaminate a later sample and increase its measured heating value (3).Table 1. Composition of an Example Gas Mixture with a Heating Value of 1,500 Btu/scfPercentComponent MoleMethane 64.107Ethane 10.330Propane 7.128Iso-butane 2.174Normal butane 6.386Iso-pentane 1.874Normal pentane 2.307Normal hexane 0.538Normal heptane 0.187Normal octane 0.086Normal nonane 0.023Normal decane 0.016Nitrogen 3.939Carbon dioxide 0.906Total 100.001GUIDELINES FOR INITIAL SAMPLING OF A GAS STREAM OF UNKNOWN HYDROCARBON DEW POINT AND COMPOSITIONSampling an unknown gas stream for custody transfer poses unique risks. Without knowledge of the gas stream composition, samples may be collected improperly, and J-T cooling or exposure to ambient temperatures can drop the sample temperature below its HDP. A gas stream below its HDP can lose heavy hydrocarbons through condensation, and the liquids can be trapped in the sample connections or in the cylinder itself. This can lead to several issues:∙ A sample analysis that indicates a lower (or higher) energy content than the actual gas stream, leading to lower (or higher) custody transfer revenue.∙Pipeline natural gas that violates custody transfer tariffs, but is not recognized as being in violation.∙Equipment problems such as damaged turbines, flooding of burners, and poor combustion.In short, sampling an unknown gas stream requires care and attention to avoid potentially significant costs to the operator and/or the end user of the natural gas.Measuring HDPs before SamplingWhen faced with sampling an unknown gas stream, the preferred technique is to measure the HDP and keep the temperature of the sample (and the sampling equipment) above the HDP. HDP measurement can be done manually using chilled mirror devices or automatically with a variety of automated analyzers.Manual HDP measurement has been standardized in Annex G of the seventh edition of API Chapter 14.1 (1). The annex references the ASTM D1142 standard for measuring water dew points using a chilled mirror device (8). The annex expands on this standard by providing guidelines for measuring the HDP of natural gas mixtures using the same device. The procedures were developed using a combination of practical industry knowledge and applied research. Notably, the annex provides uncertainty values for HDP measurements made with chilled mirrors (5).Annex G also illustrates different types and amounts of condensation that an operator could see within a chilled mirror device during normal use and while diagnosing problems. A few of these illustrations are shown in Figure 3. It should be noted that manual measurement, though standardized, is still subjective, since different operators can obtain different results on the same gas stream.Figure 3. Example Images of Condensation Observed on a Chilled Mirror Device Clockwise from the top left, these images show hydrocarbon condensation, water condensation, glycol contamination, and alcoholcondensation (5).Automated HDP measurement devices can vary greatly in operating methods. Examples include optical detection of condensate on a chilled mirror, spectroscopic analysis of a gas stream and correlation to HDP, and gas chromatographic analysis and equation of state HDP calculation. Automated HDP measurement devices are objective and are often more repeatable than manual measurements. However, they can suffer from measurement interference and other sources of bias.Sampling Without Direct HDP MeasurementsWhen the HDP is impractical to measure, gas samples can be taken, but the samples should be checked to confirm that the gas in the pipeline is above its HDP. API Chapter 14.1 describes various approaches to collecting samples when the HDP is unknown. Most of these approaches require the use of an equation of state (EOS) to predict the HDP from an analysis of the sample. From most to least preferred, the recommended sampling methods are listed below.1.Take a constant pressure spot sample at or above the flowing gas temperature, perform an extended analysis, andcalculate the HDP temperature using the analysis and an EOS.e a pressure-reducing sampling method, perform an extended analysis, and calculate the HDP temperature using anEOS.e historical information, including past analyses and dew point measurements from a similar source.4.Take a spot sample at line pressure, heat the sample gas to at least 30ºF above the flowing temperature at the time ofthe sample, perform an extended analysis, and calculate the HDP temperature using an EOS.If the HDP temperature calculated from one of these samples is above the flowing temperature, the sample is suspect and cannot be considered representative of the gas stream. Direct measurement of the HDP is recommended before any further gas samples are taken.USE OF EQUATIONS OF STATE (EOS)As noted above, EOSs may be used with sample analyses to predict the HDP of a gas stream. Several EOSs are used by the natural gas industry to predict the properties of natural gases with a given composition at a known pressure. The Peng-Robinson (P-R) EOS and Soave-Redlich-Kwong (SRK) EOS are two EOSs commonly used by the natural gas industry to predict the HDP of natural gases. These have been found to predict the HDP well for leaner gases, but often under-predict the HDP for richer gases, particularly at higher pressures. This is illustrated in Figure 4, where both EOSs are compared to experimental data for a representative production gas with a heating value of 1,325 Btu/scf.A more recent EOS, developed at Ruhr-Universität Bochum (the University of Bochum) in Germany, is known as GERG-2008(10). This equation uses a thermodynamic property known as the Helmholtz free energy to predict various properties of gas mixtures, including HDPs and phase boundaries. Limited tests of this equation have shown better agreement with existing HDP data than the P-R and SRK EOSs at higher pressures and for richer gases, as shown in Figure 4.Figure 4. Predicted versus Experimental Hydrocarbon Dew PointThe above plot compares experimentally measured hydrocarbon dew points to the P-R EOS (black line), the SRK EOS (pink line), and the GERG-2008 EOS (green line). This demonstrates the tendency of the P-R and SRK EOSs to under-predict the HDP of rich gases, such asproduction gases, at higher pressures.Heavy Hydrocarbon “Lumping” ModelsOne persistent challenge in the use of EOSs to predict HDPs (and other natural gas properties) is the need for data on hydrocarbons heavier than hexane. Often, these components are not measured individually by a field GC, but are reported as a combined total, or a “C6+ fraction.” These hydrocarbons comprise a small portion of common natural gas mixtures but have a significant impact on the mixture’s density, heating value, and HDP. Therefore, various models have been developed for characterizing these components.Research has shown that currently there is no single characterization method that works best for predicting the HDP of all gas streams (5). The research did show that treating the C6+ fraction as normal hexane did not accurately predict the HDP and could cause the HDP to be under-predicted by as much as 70°F. Many field GCs use the GPA 60/30/10 method of characterizing the C6+ fraction (60% n-hexane, 30% n-heptane, and 10% n-octane). This approach, and similar standard fraction distributions, can accurately characterize the density, heating value, and many other properties of many natural gas streams. However, these same distributions generally predict the HDP of typical gas streams only to within ±25°F (5). There are other characterization models available, but questions remain about how well they work when used with different EOS to predict HDPs (9). Where possible, a periodic extended analysis of a gas stream to at least C9 is recommended for accurate HDP predictions.Differences between Calculated and Measured HDPsUsers should be aware of the differences between HDPs predicted by an EOS and measured HDPs. An EOS predicts the temperature at which the first few heavy hydrocarbon molecules in the gas stream theoretically condense out to form liquid. A device used to measure the HDP will register the temperature at which the smallest amount of condensate can be detected. In the case of a Bureau of Mines chilled mirror device, the HDP will depend on the skill and eyesight of the operator.Both HDPs predicted by EOS and HDP measurements are subject to biases. For HDPs calculated from a sample analysis, these can include biases in the sampling procedure, biases in the sample analysis, and biases in the EOS and its parameters. HDP measurement biases are commonly related to the detection method, but may also be related to the method of transporting the sample to the instrument.Measured HDP temperatures tend to be a few tenths to a few degrees Fahrenheit less than a calculated HDP, since an instrument requires more than a few molecules to detect condensation. In general, as the gas composition becomes richer, the differences between measured and calculated HDPs tend to decrease, because more heavy hydrocarbons will condense out from a richer gas as the temperature falls below the HDP.EQUIPMENT HEATING REQUIREMENTSTo avoid sample condensation and the errors discussed above, API Chapter 14.1 requires that sampling equipment be kept at least 30°F above the predicted HDP. It does allow operators to use a lower margin if the difference between experimental and predicted hydrocarbon dew points has been shown to be less than 30°F for the gas of interest. This requirement is separate from the heating requirements imposed by Joule-Thomson cooling. The requirement covers all equipment that comes in contact with the gas sample, and is intended as a safety margin to ensure that the gas stays above its HDP as it travels to the sample cylinder. API Chapter 14.1 also gives guidance on how to properly heat or insulate sampling equipment to consistently keep it at the required temperature.SAMPLE PROBE LENGTH AND LOCATIONBesides precautions to avoid sample condensation, the length and location of the sample probe should also be examined. As natural gas production has increased, flow rates through existing pipelines have correspondingly increased. As a consequence, industry has witnessed large diameter pipelines flowing gas at higher pressures and velocities than ever before. This combination of longer probes required by larger diameter pipelines and higher velocities has increased the fatigue loading on probes as they begin to resonate. If not accounted for by the probe designer, fatigue loading can cause probes to fail catastrophically and to be swept downstream into other equipment.API Chapter 14.1 gives equations and other guidance for selecting sample probes appropriately to avoid these failures. For example, in Table 2, the maximum length for probes is recommended based on common probe diameters (see Figure 5). API Chapter 14.1 gives several other guidelines for probe installation.∙Probes should be mounted vertically at the top of a straight run of pipe.∙If the gas is not near its HDP, the probe may be placed at any axial location in a meter run that doesn’t interfere with the performance of the primary metering element (1).∙If the gas is at or near its HDP, the probe should be at least five nominal diameters downstream from any major disturbances. This is designed to avoid ingesting liquid droplets that could be condensed out of the gas or swept into the gas in the wake of the disturbance. Some of the major disturbances listed are orifice plates, elbows, tees, and flow conditioners (1).ProbeODProbe LengthFigure 5. Two Example Sample ProbesSample probes are shown with beveled (left) and straight-cut (right) ends. Straight-cut probes are preferred over beveled probes (1).Recommended values for the labeled dimensions are shown in Table 2.Table 2. Maximum Probe Length Recommendations for Common Probe DiametersProbe Outer Diameter (in) Recommended Max ProbeLength (in).250 2.00.375 3.25.500 4.25.750 6.50SAMPLING METHODSAPI 14.1 references GPA Standard 2166 (6) regarding spot sampling methods, and the reader is referred to that standard for details of the various methods. One change of note to the GPA standard that may not yet be commonly used is related to the fill-and-empty method of sampling. In this method, a length of tube must be installed downstream of the sample cylinder with a flow restriction at the end of the tube. This flow restriction forces the pressure drop to occur at the orifice instead of inside of the sample cylinder and reduces the risk of condensation inside the cylinder itself.The previous edition of GPA 2166 required that this flow restriction be a drilled plug. The standard now allows for other flow restrictions and specifically discusses the use of devices with a flow coefficient (Cv) between 0.09 and 0.53. One device specifically mentioned that can meet this specification is a partially open needle valve, although any device is allowed as long as it meets the requirements for thermal isolation and throttling.GUIDELINES FOR LABORATORY ANALYSISTo help ensure that natural gas samples are both properly collected and analyzed, API Chapter 14.1 gives guidance on laboratory sample analysis, with specific guidance to the preparation of gas chromatograph (GC) calibration standards. For example, it requires that laboratories meet the GPA operational requirements laid out in GPA Standard 2198 (7). API Chapter 14.1 also requires that GC calibration standards be prepared according to GPA 2198 and calls out specific requirements from that standard. It requires that gases must be gravimetrically prepared; that is, each component must be weighed as it is added to the mixture. It requires that those measurements be traceable back to NIST or an equivalent standards body.API Chapter 14.1 also requires that each component of a given GC calibration mixture be screened for impurities and that any impurities be accounted for in the final composition. It also calls out the required accuracy of the composition as shown in Table 3. This guideline requires that the accuracy of each component’s concentration fall within the specified ranges based on the nominal concentration.Table 3. Required Blending AccuracyThe table below illustrates the required accuracy for GC calibration standards. These guide the required accuracy for each componentbased on what percent of the total composition it comprises (7).Percent AccuracyPercent Concentration(mole %)0 to 0.099% 5%0.10 to 9.999% 2%10.0% to 100% 1%AUTO-IGNITIONAPI Chapter 14.1 Section 16 reflects industry concerns regarding auto-ignition of natural gas in sample containers. There is a theoretical possibility of auto-ignition if a sample cylinder is not properly purged and filled. Specifically, if a sample cylinder is stored at a low pressure and then is rapidly pressurized with gas, a shockwave could occur within the cylinder. This shockwave would compress the gas at its forefront and correspondingly heat it. If this heating brought the gas above its auto-ignition temperature and the cylinder had also contained oxygen before the rapid filling process, a combustion process could occur. This process would require an inlet valve with a large flow area that was opened very quickly, as with a large quarter turn valve. API Chapter 14.1 notes that API is not aware of this actually occurring in the field, but the process is possible in theory (1).CHECKLIST FOR INSPECTING FIELD SAMPLING LOCATIONS AND PROCEDURESThe seventh edition of API Chapter 14.1 includes a checklist (Annex H) for inspecting field sites where natural gas is sampled, the sampling methods used, and the procedures used to analyze samples in the lab. The checklist is recommended for use by field personnel, company auditors, and training instructors to ensure that API Chapter 14.1 guidelines are followed and that natural gas samples are collected according to API Chapter 14.1 requirements. In 2017, the Bureau of Land Management (BLM) referenced API Chapter 14.1, Annex H in its regulations for the measurement of natural gas produced on federal lands, and this checklist will be used by BLM inspectors as well.CONCLUSIONAPI 14.1 serves the natural gas industry as the standard for best practices in the collection and handling of gas samples in custody transfer. In support, this paper has expanded on several important sections of the standard, providing background information, examples, and guidelines.REFERENCES1.API Manual of Petroleum Measurement Standards, Chapter 14 – Natural Gas Fluids Measurement, Section 1 – Collectingand Handling of Natural Gas Samples for Custody Transfer, Seventh Edition, American Petroleum Institute, WashingtonD.C., May 2016.2.George, D. L. and Kelner E., Lessons Learned from the API MPMS, Chapter 14.1 Gas Sampling Research Project,Proceedings of the 2014 American School of Gas Measurement Technology, Houston, TX,.3.Metering Research Facility Program: Natural Gas Sample Collection and Handling-Phase I, Behring, K.A. III and Kelner,E., GRI Topical Report No. GRI-99/0194, Gas Technology Institute, Des Plaines, Illinois, August 1999.4.Metering Research Facility Program: Natural Gas Sample Collection and Handling-Phase V, George, D. L., Burkey, R.C., and Morrow, T. B., GRI Topical Report No. GRI-05/0134, Gas Technology Institute, Des Plaines, Illinois, March 2005.5.Metering Research Facility Program: Natural Gas Sample Collection and Handling-Phase IV, George, D. L., Barajas, A.M., Kelner, E., and Nored, M., GRI Topical Report No. GRI-03/0049, Gas Technology Institute, Des Plaines, Illinois, January 2005.6.GPA Standard 2166, Obtaining Natural Gas Samples for Analysis by Gas Chromatography, Gas Processors Association,Tulsa, Oklahoma, 2005 (reaffirmed 2017).7.GPA Standard 2198, Selection, Preparation, Validation, Care and Storage of Natural Gas and Natural Gas LiquidsReference Standard Blends, Gas Processors Association, Tulsa, Oklahoma, 2016.8.ASTM Standard D1142, Standard Test Method for Water Vapor Content of Gaseous Fuels by Measurement of Dew PointTemperature, ASTM International, West Conshohocken, PA, 1995 (reaffirmed 2012).ughton, A., Use of the GERG-2008 Equation of State for Hydrocarbon Dew Point Calculation, Proceedings of the 2015American Gas Association Operations Conference, Grapevine, TX.10.Kunz, O., Klimeck, R., Wagner, W., and Jaeschke, M., The GERG-2004 Wide-Range Equation of State for Natural Gasesand Other Mixtures, GERG TM- 15, 2007.11.Metering Research Facility Program: Natural Gas Sample Collection and Handling-Phase II, Kelner, E., Britton, C. L.,Behring, K.A. III and Sparks, C. R., GRI Topical Report No. GRI-01/0069, Gas Technology Institute, Des Plaines, Illinois, January 2003.12.Metering Research Facility Program: Natural Gas Sample Collection and Handling-Phase III, Kelner, E., Sparks, C. R.,and Behring, K.A. III, GRI Topical Report No. GRI-01/0070, Gas Technology Institute, Des Plaines, Illinois, August 2002.13.GPA Standard 2172, Calculation of Gross Heating Value, Relative Density, Compressibility and Theoretical HydrocarbonLiquid Content for Natural Gas Mixtures for Custody Transfer, Third Edition, Gas Processors Association, Tulsa, Oklahoma, 2014.14.Title 43, Code of Federal Regulations, Part 3170, Onshore Oil and Gas Production, Subpart 3175, Measurement of Gas,November 17, 2016.。
小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小木虫论坛,打造学术最前沿 电镜常用英文词汇500个(美语版)- A - Abbe, Ernst (1840-1905) German mathematician and physicist, professor at Jean, and inventor of much opticalapparatus at the Zeiss works. His inventions include the apochromatic objective, the compensating ocular, the Abbe condenser, a well corrected oil-immersion achromatic condenser, the immersion objective, Abbe apertometer, Abbe refractometer, and the drawing camera; he evolved the Abbe theory of resolution and microscope imagery, the numerical aperture formula, and other optical theories. Abbe apertometer In microscopy, a device for measuring simultaneously, the numerical and angularapertures of an objective or condenser. The back focal plane of the objective is viewed within auxiliary lens, and the device is set to show the position of an indicator just cutting into two opposite edges of the conoscopic field of view. Abbe's law of limiting resolution For a periodic structure of units separated by distance d and obliquely illuminated by the unrefracted ray and one of the two diffracted rays (extremely oblique illumination), Abby applied the law of diffraction: d = 0.5 lambda /NA, where: lambda = wavelength of themonochromic light or shortest of mixed wavelengths NA = the limiting numerical aperture (NA) (q.v.) of objective or condenser.Abbe theory of image formation see Abbe theory of image formation Abbe limit Ernst Abbe's specification for the limit of resolution of a diffraction-limited micro-scope. According to Abbe, a detail with a particular spacing in the specimen is resolved when the NA of the objective lens is large enough to capture the first-order diffraction patternproduced by the detail at the wavelength employed. See also Rayleigh criterion, Sparrow limit.Abbe substage apparatusIncludes a rack and pinion for horizontal displacement of an iris diaphram to obtain oblique lighting. Abbe test plateA long, wedge-shaped coverslip about 0.20 mm thick at one end and 0.10 to 0.12 mm at the other end coated chemically with a silver film on which are ruled horizontal lines. At each variation in thickness of 0.01 mm there are vertical lines. By means of oblique illumination and by focusing on different portions of the plate, it is possible to determine the optimum coverslip thickness for any objective and also, for microscopes with drawtubes, the tube length for best objective performance. The approximate freedom from小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小木虫论坛,打造学术最前沿 spherical and chromatic aberrations can also be estimated. Small isolated bits of silver near the edges of the lines form good objects for the star testAbbe theory of image formation Abbe's theory is based on the fact that a non-self-luminous particle, which is illuminatedby an extraneous source, gives rise to diffracted light rays, in addition to the dioptric pencil. He stated that to form a good microscopical image as many of the diffracted rays as possible should be intercepted by the objective. With closely ruled lines, his theory is easily demonstrated by observing the back lens of the objective, for here the diffracted rays can be observed directly if the aperture diaphram is closed. It can be shown that, when the illumination is arranged to exclude the diffracted images, resolution is lost. aberration any error that results in image degradation. Such errors may be chromatic, spherical, astigmatic chromatic, distortion, or curvature of field: and can result from design or execution, or both. aberrationFailure of an optical or electron-optical lens to produce exact geometrical (and chromatic) correspondence between an object and its image. In a video camera tube or cathode-ray tube, aberrations anse when the (electrostatic or electromagnetic) lens does not bring the electron beam to sharply focused points uniformly on the target or screen, or to correct geometrical positions, as the beam is deflected. aberration In an optical system, any defect that degrades its performance from that which could beachieved with a perfect lens, e.g., failure of an object point to be imaged as a point. aberration, chromaticA defect in a lens or optical system due to the greater refraction of shorter wavelengths over that of loner ones at a lens surface. Hence the focal length of a simple lens is shorter for blue than for red rays. This dispersion of the wave-lengths will cause color fringes in the image field of a lens with such an aberration. aberration, sphericalA lens defect whereby image forming rays of one color, passing through the outer zones of a lens come to focus at a different distance from the lens than do those of more central rays. With a simple spherical (or plano-spherical) lens the outer rays always meet the axis closer to the lens than do more central rays and the lens is uncorrected or "undercorrected". When the reverse is true the lens has been "overcorrected". abrasive mediaExtremely hard materials (Diamond, SiC, Al2O3.etc.), usually in a very fine particulate form (<15 micrometers), used in the initial stages of specimen preparation to grind and polish samples to the desired thickness or finish.accommodation (of the eye)The act of adjusting the eye, to bring objects that are closer to the eye in focus.小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小木虫论坛,打造学术最前沿 Contraction of the ciliary muscles relaxes the tension on the crystalline lens, which rounds up by virtue of its elasticity while also moving forward slightly. The net effect isto rescue the overall focal length of the eye.achromatic literally, color-free. A lens or prism is said to be achromatic when corrected for two colors. The remaining color seen in an image formed by such a lens is said to be secondary chromatic aberration. achromaticLiterally, color free. A truly achromatic lens would transmit light without decomposing it spectrally and there would be no color fringing in the image. A doublet, composed of a positive and negative lens element, can be made achromatic for two colors which much improves the performance for most of the others. achromatic aplanatic condenser A well-corrected microscope condenser lens; corrected for chromatic and spherical aberrations and satisfying the sine condition.achromatic lensalso achromatic A lens cluster whose foci and power are made the same for two wavelengths (commonly for the red hydrogen C line, A = 6563 A, and blue F line, A =4861 A). The simplest achromatic is a doublet that combines two single lenses with different dispersions and curvatures to achromatize the combination. Even the simplest "achromatic 'microscope objectives contain two such doublets. achromatic objectivean objective that is corrected chromatic for two colors, and spherically for one, usually in the yellow-green part of the spectrum.air-lock An intermediate, enclosed chamber of a vacuum or pressure system through which an object may be passed without effectively changing the vacuum or pressure of the main body of the system. Reference here is to the vacuum of all electron microscopes. Airy diskthe image of a bright point object as focused by a lens system. With monochromatic light, it consists of a central point of maximum intensity surrounded by alternate circles of light and darkness caused by the reinforcement and interference of diffracted rays. The light areas are called maxima and the dark areas minima. The distribution of light from the center to the outer areas of the figure was investigated mathematically by Sir George Airy. The diffraction disk forms a basis for determining the resolving power of an ideal lens system. The diameter of the disk depends largely on the aperture of the lens. The diffraction of light causing the Airy disk is a factor limiting the resolution of a well corrected optical system. airy diskThe bright disk of light (surrounded by alternating dark and bright diffraction rings)that is小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小木虫论坛,打造学术最前沿 formed by a perfect diffraction-limited lens, focusing an image of an infinitely small source of light. For a minute absorbing spot, the diffraction pattern is a dark Airy disksurrounded by brighter and darker diffraction rings. Since the Airy disk is the smallest unit that makes up the image of a luminous or absorbing object (formed by a properly corrected microscope lens in focus), the radius of the disk determines the limit of resolution of the microscope. aliasing A pattern of image sampling error in digital systems. Aliasing forces spatial frequencycomponents higher than a critical value (the Nyquist frequency) to be displayed at progressively lower frequencies. Aliasing introduces an undesirable moiré pattern when the spatial frequency of the signal exceeds the sampling rate in a digitizer. amplifier (optical) When the projected image of the microscope is projected less than 250 mm the objective may be moved enough from its usual position (for visual observation) to degrade the image. It can be restored by adding a lens above the ocular to bring the image into focuson the desired plane but with the objective still in its position of best-corrected image. A similar system requires an "ocular" in which the same corrections are made. Such lens systems are termed "amplifiers" (e.g., Homalsof Zeiss, Amphiplans of B & L, etc.). analyzer an optical device, capable of producing plane polarized light, used for detecting the state of polarization. analyzer A second polarizing element inserted beyond a preparation. When its vibration directionis at right angles to the vibration direction of the polarize, the field becomes black if no anisotropic specimen is on the stage or when viewing an anisotropic substance in an extinction position or directly down an optic axis of an anisotropic crystal. (See polarizer) Å: Angstrom A metric unit of length measurement= 1x10 -10 meters or 0.0000000001 meters 4x10 -9 inches or 0.000000004 inchesAtomic diameters range between 1 and 2 Angstroms ångström unita unit of linear measure named after A. J. Ångström. It is 1 x 10-10 metres: 1 um = 10.000 Å. It is generally abbreviated as A. in the United States: elsewhere, it is variously abbreviated A, A., A.U., Å., or ÅU.angular aperture see aperture, angular anisotropicQuality of a transparent material having different refractive indices depending on the vibration direction of the transmitted light, hence any material that affects polarized light differently according to its direction through the material.anisotropy小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小木虫论坛,打造学术最前沿 The difference in physical properties such as velocity of wave propagation, elasticity, refractive index, absorbance, etc., for the same material measured along differentdirectionsannular illumination The result of placing a stop in the first focal plane of the condenser to produce an illuminating cone of light with all the light flux near the surface of the cone. The central part of the cone will be dark. This arrangement is often used for a darkfield effect in low-power work by transmitted light. The condenser must be correctly focused andcentered. By reflected light, darkfield illumination is attained with an annular con-denser fitted around the objective for opaque objects. Annular illumination implies that the object is lighted from all sides. annular stop The opaque ring-shaped stop with a small central opening usually placed in the objective back focal plane to provide dispersion staining. anodeThe electrode to which a major flow of electrons takes place internally (as in a cathode-ray tube) or to which an external positive voltage supply is connected.aperture A thin (<100 micrometer thick) disk or strip of metal (usually Pt) with a small (2-100 micrometers) circular through-hole. Used to restrict electron beams and filter out unwanted scattered electrons before image formation. apertureIn optics an aperture is an opening that restricts the size of the light beam that enters or leaves a lens or lens system and is often controlled by an iris aperture diaphram. Its correct placement can be a very critical matter. The aperture can, of course, be filled by a lens; it can be at a focal plane of the lens as is that of the substage condenser of a microscope. Opening this aperture wider increases image brightness and definition(resolving power) but decreases depth of field and contrast. Aperture size is measured variously according to the field of use of the lens, often by its simple diameter as is that of the telescope. (But see: angular aperture , relative aperture, numerical aperture ). The free aperture of a lens is that which is unrestricted except by its own diameter. aperture In a video camera tube or monitor, the aperture refers to the size and shape of the electron beam that lands on the target or phosphor. In an optical instrument, the opening of a lensor aperture stop.aperture for electron microscopyanode aperture: The opening in the accelerating voltage anode shield of the electron gun through which the electrons must pass to irradiate the specimen. condenser aperture: An opening in the condenser lens controlling the number of electrons entering the lens and the angular aperture of the electron beam. The angular aperture can also be controlled by小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小木虫论坛,打造学术最前沿 the condenser lens current. physical objective aperture: A metallic diaphram, with a small central hole, used to limit the cone of electrons accepted by the objective lens. Thisimproves image-contrast since highly scattered electrons are prevented from arriving at the Gaussian image plane and therefore cannot contribute to background fog. aplanatic. Free from spherical aberration and coma. aperture correction An electronic process used in some high-resolution video cameras and monitors that compensates for the loss in sharpness of detail due to the finite dimensions (aperture) ofthe scanning beam.aperture functionIn a diffraction-limited optical system, the function that determines the relationship between the image and each point in the object. Modifying the aperture function changes the image according to the modified Fourier-filtering (or optical filtration) property of the aperture. aperture planeIn a microscope adjusted for Koehler illumination, the conjugate planes that include the light source, the condenser iris diaphram, the objective lens back aperture, and the eye point. Spaces in the aperture planes are the reciprocal of those in the field planes. aperture, angular the angle between the most divergent rays that can pass through a lens to form the image of an object. aperture, angular (AA)The angle subtended between the axis of a lens and the largest accepted angle of the image-forming rays. With microscope objectives the trigonometric sine of this angle is used to define numerical aperture (NA)(q.v.) but as measured from the axial object point. (Note: Some commercial advertising literature has erroneously used the full angle in defining NA.)- B - backscattered electronProduced by an incident electron colliding with the nucleus of an atom in the specimen. The incident electron is then scattered "backward" about 180 degrees with no appreciable loss of energy, an elastic collision. backscattered electron imagingThe production of backscattered electrons from a sample varies directly with the specimen's average atomic number; higher atomic number elements produce more backscattered electrons than lower atomic number ones. Detection of Backscattered Electrons is achieved by using a "donut" shaped solid state semiconductor device mounted on the bottom of the objective lens. When Backscattered Electrons strike the小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小木虫论坛,打造学术最前沿 detector electron-hole pairs are created which are then counted. This quantity is translated into a pixel intensity and displayed on the CRT, forming the image. By splitting thedetector into halves (or quadrants) differences in the signal level on the individual detector segments provide surface topography information. back aperture The exit pupil of a microscope objective lens. The objective lens back aperture, which can be examined with a phase telescope or by inserting a Bertrand lens, displays the conoscopic interference figure and diffraction patterns.back focal lengthAs measured on the principal axis, from the second lens vertex to the back focal point of the lens. It is not the equivalent of the focal length. back focal plane The plane, normal to the lens axis, situated at the back focus of a lens. back lens In any compound lens (a lens system composed of more than one lens element), the lastlens through which the light passes is called the back lens. It may be a single simple lens, a doublet, or triplet. See front lensbalsam, Canada a resin from the balsam fir Abies balsamea . Dissolved in xylene, toluene, or benzene it is used as a mountant for permanent microscopical preparations. Its refractive index may vary from 1.530 to l.545 and its softening point from room temperature to 100deg.C, these properties varying with age and solvent content. If impure it discolors with age.Bertrand lens see lens, Bertrand.barrel distortionA distortion, or aberration, base microscopical The microscope's supporting structure such as the horseshoe type. Heavier bases, some containing the entire illuminating system, began to appear after World War II. Becke lineWhen the liquid phase of a microscopical mount has arefractive index different from that of the solid phase, a line or narrow band of light can be observed around or just within the outlines of the specimen as the microscope tube is raised or lowered from its position of best focus. The presence of the line indicates the difference in index referred to, and its absence, there-fore, indicates similarity of index between the specimen and its mounting fluid. The Becke line is useful in determining the refractive index of transparent, microscopic particles. See refractive index (n) by Becke line. bellows lengthThe distance from the eye point to the image plane in a photomicrographic apparatus. Bertrand lens小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小木虫论坛,打造学术最前沿 A small, low-power lens, usually on a slide for insertion into the drawtube between analyzer and ocular. It is used to observe the back focal plane of the objective so as toexamine interference figures or as an aid in achieving interference figures. It is apt to be strongly astigmatic. It is used to image the lamp filament in setting up Kohler illumination ( q.v.) as well as for centering dispersion staining stops to the substage aperture diaphram. biaxial crystals Anisotropic crystals in the orthorhombic, monoclinic and triclinic systems. They havethree principal refractive indices alpha, beta, and gamma, and two isotropic directions, i.e., optic axes. bifilar eyepiece. An ocular with two crossed hairs, wires, filaments or threads each of which has perpendicular motion. binocular microscope A microscope fitted with double eyepieces for vision with both eyes. The purpose in dividing the same image from a single objective of the usual compound micro-scope is to reduce eyestrain and muscular fatigue which may result from monocular, high-powermicroscopy. The purpose in obtaining a different image for each of two oculars is to provide stereoscopy by means of two different angles of view. There are two kinds of stereoscopic microscopes: binobjective (Greenough) older type and monobjective (common main objective, CMO) newer type. (See stereo microscope, Greenough microscope, etc.) birefringence The numerical difference in refractive indices for a substance. In a given crystal view, the interference color (retardation) between crossed polars depends on the birefringence andthickness: Retardation (nm) = 1000 x thickness (m) x birefringencebisectrix, acutein biaxial crystals, that principal axis of the ellipsoid of indexes which bisects the smaller angle between the optic axes. bisectrix, obtuse in biaxial crystals, that principal axis of the ellipsoid of indexes which bisects the larger angle between the optic axes.bright field imagingAn imaging mode in a TEM that uses only unscattered Electrons to form the image. Contrast in such an image is due entirely to mass-thickness variations in amorphous samples, and may include diffraction contrast in crystalline samples. bright field illuminationThe method of lighting the specimen with a solid cone of rays. Transmitted bright field illumination is performed by a substage condenser. Reflected bright field illumination is performed by a vertical illuminator. Compare dark field illuminationbrightnessThe brightness of an extended luminous source is termed the intensity per unit area小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小木虫论坛,打造学术最前沿 measured in candles per cm 2. Also, the intensity of reflection; it may be measured in lamberts; one lambert equals 1 lumen/cm 2.- C -calcite a doubly refracting mineral used in the manufacture of polarizing prisms. It is uniaxialnegative and in the trigonal diversion of the hexagonal system of crystals. Its indexes are epsilon =1.486, omega =1.658; its hardness is 3 on the Mohr scale and specific gravity2.711. Canada balsam see balsam, Canada. cardioid dark field condenser A condenser designed with two reflecting surfaces; the first, a spherical surface which reflects the rays to a second, cardioid (heart-shaped) surface. The virtue in such an arrangement is that, if the cardioid surface is of true figure, the lens is both achromaticand aplanatic. It has a limiting numerical aperture of about 1.0. Thus objectives of a greater numerical aperture cannot be used successfully with it. A "true" cardioid figure is the trace of a point on the circumference of a circle rolling around an equal, fixed circle. Cassegrainian darkfield condenser Named after Cassegrain, astronomer of the 17th century. A high-power, dark field con-denser to be used with objective apertures as high as 1.3. It is sometimes called the luminous spot ring condenser (Zeiss). catadioptricAn optical system in which both reflecting and refracting curved surfaces are used to form an image. Some ''reflecting" objective lenses, as well as video projection systems, are catadioptric; the latter uses a Schmidt plate to correct the spherical aberration introduced by the spherical reflecting mirror. central stop (for condenser) An opaque disk placed in the ring carrier or diaphragm carrier of the substage apparatus. It excludes the central rays and is used for dark field work at low magnification. Variablestops can be contracted or expanded as desired. For colored stops see optical staining characteristic X-rayAn X-ray having a unique energy that is emitted by an atom in the sample during its de-energization after ionization of one of its electrons by an incident electron. chromatic aberrationa defect in a lens or lens system as a result of which the lens possesses different focal lengths for radiation of different wavelengths.collimating lensA lens used to produce a collimated beam from a light source. With a true point source on the axis the beam would be an axial tube of parallel rays With practical sources having小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小木虫论坛,打造学术最前沿 real area the beam will be a diverging cone of parallel rays from each light point; this beam still has special properties and can be collected again by a good lens. collimationthe operation of controlling a beam of radiation so that if the light source were a point, the light rays would become parallel. The total bundle of rays diverge as the source size increases. colloid A suspension of sub-light-microscopic particles. This definition arbitrarily limits the sizeof the particles to 0.1 - 0.005mm. Such particles may be studied by dark field illumination, particularly with the light ultramicroscope or by means of an electron microscope. coma a lens aberration occurring in that part of the image field that is some distance from the principal axis of the system. It results from different magnification in the various lens zones. Extra-axial object points appear as short comet-like images with the brighter small head toward the center of the field (positive coma) or away from the center (negativecoma).comaA lens aberration in which the off-axis beams do not form single focused spots but rather comet-shaped patterns . coma A lens aberration which occurs in part of the image field which is slightly away from the principal axis of the system. It results from different magnifications in the various lenszones. It causes extra-axial object points to appear as short comet-like images, with the tail either toward the center of the field (positive coma) or away from the center (negative coma). Coma is fundamentally due to faulty position of the principal points of the lens. compensating eyepieces those designed for use with objectives such as apochromats in order to correct chromatic aberration . condenser or condenser lensa term applied to lenses or mirrors designed to collect, control, and concentrate radiation in an illumination system.condenser, Abbe originally a two-lens substage condenser combination designed by Ernst Abbe. It lacks chromatic correction though designed for a minimum of spherical aberration and has onlya very low-angle aplanatic cone. It may be rated with a numerical aperture as high as 1.3. condenser circleThe image of the aperture iris diaphragm of the substage condenser (q.v.) as seen in the back focal plane of the objective.condenser (substage)小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小 木 虫 论 坛 精 彩 奉 献 打造学术最前沿 小木虫论坛,打造学术最前沿 In microscopy, the lens mounted before the microscope stage, which transmits light to the object. There are two main categories of condensers: (1) bright field and (2) dark field.Bright field condensers are of four distinct types: (a) Abbe condenser ,(q.v.), an uncorrected condenser composed of two separable lenses; (b) aplanatic condenser; (c) achromatic condenser which has full corrections for color and spherical aberrations; (d) aplanatic achromatic condenser. the dark field condenser for low powers may be nothing more than a low-power bright field condenser with a central stop . Medium- or high-powered dark field condensers are usually of the cardioid (q.v.) or paraboloid (q.v.) type. The lamp lens is loosely called a condenser lens, but light-collecting lens is a more definite term. All microscope condensers must be carefully focused and aligned for best results. condenser, dark field a condenser forming a hollow cone of light with its apex (or focal point) in the plane of the specimen. When used with an objective having a numerical aperture lower than the minimum numerical aperture of the hollow cone, only light deviated by the specimen enters the objective. Objects are seen as bright images against a dark background.condenser, dark fieldThe ordinary bright field condenser of low power, used with a central stop, makes a good dark field condenser. See special dark field condensers: paraboloid , cardioid and Cassegrainian . They all form a dark field while illuminating the specimen with a hollow cone of light. The lower limiting aperture of the condenser must be greater than the NA of the objective with which it is to be used. Thus, no direct light enters the objective; the specimen is seen by reflected or scattered light on a dark background. See condensers condenser, darkfield, bispherica dark field condenser consisting of a convex spherical reflector mounted concentric with a larger concave reflector. The rays are formed into a diverging cone by the convex reflector. The annular concave reflector then forms a hollow converging cone which is focused on the subject. See condenser, darkfield, paraboloid a dark field condenser consisting of a reflecting surface in the form of a segment of aparaboloid of revolution. Parallel rays entering the condenser around the periphery of the central stop are reflected from the curved surfaces and converge at the focus of the paraboloid. Seecondenser, variable-focus essentially an Abbe condenser in which the upper lens element is fixed and the lower movable. The lower lens may be used to focus the illumination between the elements sothat it emerges from the stationary lens as a large diameter parallel bundle. The field of low-power objectives may thus be filled without removing the top element. At the opposite extreme it can be adjusted to have a numerical aperture as high as 1.3. critical illumination see illumination, critical .confocal optics。
Design and Characterization of Single Photon APD Detector forQKD ApplicationAbstractModeling and design of a single photon detector and its various characteristics are presented. It is a type of avalanche photo diode (APD) designed to suit the requirements of a Quantum Key Distribution (QKD) detection system. The device is modeled to operate in a gated mode at liquid nitrogen temperature for minimum noise and maximum gain. Different types of APDs are compared for best performance. The APD is part of an optical communication link, which is a private channel to transmit the key signal. The encrypted message is sent via a public channel. The optical link operates at a wavelength of 1.55μm. The design is based on InGaAs with a quantum efficiency of more than 75% and a multiplication factor of 1000. The calculated dark current is below 10-12A with an overall signal to noise ratio better than 18dB. The device sensitivity is better than -40dBm, which is more than an order of magnitude higher than the dark current, corresponding to a detection sensitivity of two photons in pico-second pulses.I. INTRODUCTIONPhoton detectors sensitive to extremely low light levels are needed in a variety of applications. It was not possible to introduce these devices commercially several years ago because of the stringent requirements of QKD. Research efforts however resulted in photon detectors with reasonably good performance characteristics. The objective here is to model a single photon detector of high sensitivity, suitable for a QKD system. The detector is basically an APD, which needs cooling to very low temperature (77K) for the dark current to be low. The wavelength of interest is 1.55μm. Different applications may impose different requirements, and hence the dependence of the various parameters on wavelength, temperature, responsivity, dark current, noise etc, are modeled. Comparison of the results from calculations based on a suitable model provides amenable grounds to determine the suitability of each type of APD for a specific application.Attacks on communication systems in recent years have become a main concern accompanying the technological advances. The measures and counter measures against attacks have driven research effort towards security techniques that aim at absolute infallibility. Quantum Mechanics is considered one of the answers, due to inherent physical phenomena. QKD systems which depend on entangled pairs orpolarization states will inevitably require the usage of APDs in photon detection systems. How suitable these detectors may be, depends on their ability to detect low light level signals, in other words “photon counting”. It is therefore anticipated that the application of high security systems will be in high demand in a variety of fields such as banking sector, military, medical care, e-commerce, e-government etc.Ⅱ. AV ALANCHE PHOTO DIODEA. Structure of the APDFig. 1 shows a schematic diagram of the structure of an APD. The APD is a photodiode with a built-in amplification mechanism. The applied reverse potential difference causes accelerates photo-generated carriers to very high speeds so that a transfer of momentum occurs upon collisions, which liberates other electrons. Secondary electrons are accelerated in turn and the result is an avalanche process. The photo generated carriers traverse the high electric field region causing further ionization by releasing bound electrons in the valence band upon collision. This carrier generation mechanism is known as impact ionization. When carriers collide with the crystal lattice, they lose some energy to the crystal. If the kinetic energy of a carrier is greater than the band-gap, the collision will free a bound electron. The free electrons and holes so created also acquire enough energy to cause further impact ionization. The result is an avalanche, where the number of free carriers grows exponentially as the process continues.The number of ionization collisions per unit length for holes and electrons is designated ionization coefficients αn and αp, respectively. The type of materials and their band structures are responsible for the variation in αn and αp. Ionization coefficients also depend on the applied electric field according tothe following relationship:,exp[]n p b a Eαα=- (1) For αn = αp = α, the multiplication factor, M takes the form11aW M -= (2)W is the width of the depletion region. It can be observed that M tends to ∞ when αW →1, whichsignifies the condition for junction breakdown. Therefore, the high values of M can be obtained whenthe APD is biased close to the breakdown region.The thickness of the multiplication region for M = 1000, has been calculated and compared withthose found by other workers and the results are shown in Table 1. The layer thickness for undoped InPis 10μm, for a substrate thickness of 100μm .The photon-generated electron-hole pairs in the absorption layer are accelerated under theinfluence of an electric field of 3.105V/cm. The acceleration process is constantly interrupted by randomcollisions with the lattice. The two competing processes will continue until eventually an averagesaturation velocity is reached. Secondary electron-hole pairs are generated at any time during theprocess, when they acquire energy larger than the band gap Eg. The electrons are then accelerated andmay cause further impact ionization.Impact ionization of holes due to bound electrons is not as effective as that due to free electrons.Hence, most of the ionization is achieved by free electrons. The avalanche process then proceedsprincipally from the p to the n side of the device. It terminates after a certain time, when the electronsarrive at the n side of the depletion layer. Holes moving to the left create electrons that move to the right,which in turn generate further holes moving to the left in a possibly unending circulation. Although this feedback process increases the gain of the device, it is nevertheless undesirable for several reasons. First, it is time consuming and reduces the device bandwidth. Second, it is a random process and therefore increases the noise in the device. Third, it is unstable, which may cause avalanche breakdown.It may be desirable to fabricate APDs from materials that permit impact ionization by only one type of carriers either electrons or holes. Photo detector materials generally exhibit different ionization rates for electrons and holes. The ratio ofthe two ionization rates k = βi/αi is a measure of the photodiode performance. If for example, electrons have higher ionization coefficient, optimal behavior is achieved by injecting electrons of photo-carrier pairs at the p-type edge of the depletion layer and by using a material with k value as small as possible. If holes are injected, they should be injected at the n-type edge of the depletion layer and k should be as large as possible. Single-carrier multiplication is achieved ideally, when k = 0 with electrons or with k = ∞for holes.B.Geiger ModeGeiger mode (GM) operation means that the diode is operated slightly above the breakdown threshold voltage, where a single electron–hole pair can trigger a strong avalanche. In the case of such an event, the electronics reduce the diode voltage to below the threshold value for a short time called “dead time”, during which the avalanche is stopped and the detector is made ready to detect the next batch of photons. GM operation is one of the basic of Quantum Counting techniques when utilizing an avalanche process (APD) that increases the detector efficiency significantly.There are a number of parameters related to Geiger mode. The general idea however is to temporarily disturb the equilibrium inside the APD.The Geiger mode is placing the APD in a gated regime and the bias is raised above the breakdownvoltage for a short period of time. Fig. 2 shows the parameters characterizing the Geiger operation. The rise and fall times of the edges are neglected because they are made fast. Detection of single photons occurs during the gate window.作者:Khalid A. S. Al-Khateeb, Nazmus Shaker Nafi, Khalid Hasan国籍:美国出处:Computer and Communication Engineering (ICCCE), 2010 International Conference on 11-12 May 2010用于量子密钥的单光子APD探测器设计摘要本文提出的是单光子探测器及其各种特性的建模与设计。
差动位移传感器原理及公式Differential displacement sensors measure the position of an object by detecting the difference in movement between two separate points. 差动位移传感器通过检测两个不同位置之间的运动差异来测量物体的位置。
This type of sensor is commonly used in industrial automation, machine tool positioning, and other applications where precise measurement of displacement is critical. 这种传感器通常被用于工业自动化、机床定位和其他需要精确位移测量的应用中。
The basic principle of operation involves comparing the movement of two mechanical elements that are connected to the object being measured. 工作原理的基本原理是比较两个机械元件的运动,这两个机械元件与被测量的物体相连。
The difference in displacement between the two points is then converted into an electrical signal, which can be used to determine the position of the object. 两点之间的位移差异随后被转换成一个电信号,这个信号可以用来确认物体的位置。
The formula for calculating the displacement measured by a differential displacement sensor is Δx = x2 - x1. 差动位移传感器测量的位移计算公式是Δx = x2 - x1。
a rXiv:as tr o-ph/6251v123Fe b26Optical Detection of Two Intermediate Mass Binary Pulsar Companions B.A.Jacoby 1,2,D.Chakrabarty 3,M.H.van Kerkwijk 4,S.R.Kulkarni 1,and D.L.Kaplan 1,5ABSTRACT We report the detection of probable optical counterparts for two Intermediate Mass Binary Pulsar (IMBP)systems,PSR J1528−3146and PSR J1757−5322.Recent radio pulsar surveys have uncovered a handful of these systems with putative massive white dwarf companions,thought to have an evolutionary his-tory different from that of the more numerous class of Low Mass Binary Pulsars (LMBPs)with He white dwarf companions.The study of IMBP companions via optical observations offers us several new diagnostics:the evolution of main sequence stars near the white-dwarf-neutron star boundary,the physics of white dwarfs close to the Chandrasekhar limit,and insights into the recycling pro-cess by which old pulsars are spun up to high rotation frequencies.We were unsuccessful in our attempt to detect optical counterparts of PSR J1141−6545,PSR J1157−5112,PSR J1435−6100,and PSR J1454−5846.Subject headings:binaries:close —pulsars:general —stars:neutron —white dwarfs1.IntroductionThe majority of recycled pulsars are in low mass binary pulsar(LMBP)systems,con-sisting of a neutron star and a low-mass white dwarf.The LMBPs are widely considered to be descendants of the Low-Mass X-ray Binaries(LMXBs).The progenitors are thus a massive star primary(which gives rise to the neutron star)and a low mass( 1M⊙)sec-ondary.In contrast,double neutron star binaries,exemplified by PSR B1913+16,descend from binaries in which both the primary and secondary are massive stars,each forming a neutron star.Over the past few years,astronomers have come to appreciate the existence of another class of binary pulsars,the so-called intermediate mass binary pulsars(IMBPs)with massive C-O or O-Ne-Mg white dwarf companions.As suggested by their name,IMPBs are thought to descend from binary star systems with a massive primary and a secondary which is inter-mediate in mass.First,the primary becomes a neutron star through a supernova explosion. Later,the secondary evolves into a massive white dwarf,transferring matter to and recycling the pulsar in the process(van den Heuvel1994;Tauris et al.2000;Taam et al.2000).As in LMBP systems,tidal damping circularizes the orbit because the supernova occurs before the companion becomes a compact object.16candidate IMBP systems are currently known.Not all pulsars with massive white dwarf companions share this evolutionary path. PSR B2303+46(Stokes et al.1985)and PSR J1141−6545(Kaspi et al.2000)have compan-ions with masses similar to the IMBP systems;however,these slowly-rotating pulsars appear to be unrecycled and their orbits are eccentric.In systems such as these it is thought that neither the primary nor the secondary was initially massive enough to form a neutron star. As the primary evolved into a massive white dwarf it transferred matter to the secondary, thereby making the secondary massive enough to eventually become a neutron star.Here again thefinal outcome is a massive white dwarf and a neutron star,but because the su-pernova occurs after the primary has become a white dwarf the orbit remains eccentric and the neutron star is not recycled(Tauris&Sennels2000).Though only two such systems are known,they may exist in numbers greater than neutron star binary systems(Portegies Zwart &Yungelson1999).The detection of the white dwarf companion via optical observations can help clarify this interesting evolutionary path(van Kerkwijk&Kulkarni1999).Apart from these tests of binary evolution,these systems may offer us new insights into the physics of how neutron stars are spun up by accretion.It is clear that the mass transfer of the recycling process results in a decreased magneticfield,as well as an increased rotation rate for the neutron star.The spin period at the end of the spin-up phase,P0,is a critical input to pulsar recycling models.A comparison of the white dwarf age from cooling models with the pulsar spin-down age(which assumes that P0is much smaller than the current spinperiod)can,in principle,allow the determination of P0(Camilo et al.1994).2.ObservationsWe have obtained optical observations offields containing six IMBP systems discoveredin recent radio pulsar surveys with the Parkes radio telescope(Fig.1;Tab.1;Camiloet al.,2001;Edwards&Bailes,2001A;Kaspi et al.,2000).We observed PSR J1141−6545, PSR J1157−5112,PSR J1435−6100,PSR J1528−3146,PSR J1454−5846,and PSR J1757−5322 in R band on the nights of6–8August2002with the the Magellan Instant Camera(MagIC) on the6.5m Baade telescope at Magellan Observatory.Seeing was generally good,but some targets were observed at high airmass,giving a broader point spread function.Conditions were photometric on6and8August,but there were clouds present on7August.Each of our six targets was observed for two10-minute exposures on one of the photometric nights except for PSR J1528−3146.These data were reduced following standard practices(bias subtraction,flatfielding with domeflats),photometrically calibrated with observations of the Stetson standard star L112-805,and astrometrically calibrated using the USNO B-1.0 catalog.The astrometric uncertainty in all observations presented here is dominated by thetie between the USNO-B1.0system and the International Celestial Reference Frame(∼0.′′2in each coordinate).On the night of4June2003,we observed PSR J1528−3146once again with MagIC. Conditions were not photometric,but better than on our previous attempt.We obtained2 exposures of5minutes each in R and2exposures of10minutes each in B,reduced in the standard manner as before.A rough photometric calibration was obtained using stars from the USNO B-1.0catalog,which also provided the astrometric calibration.Table2gives the relevant parameters of the best imaging observations in each band for each target.For each image,a model point spread function was constructed based on several stars in thefield using the daophot package in iraf.Limiting magnitudes were determinedby placing a number of artificial stars of a given magnitude in thefield and measuring their magnitudes with aperture photometry.This process was repeated tofind the input artificial star magnitude that resulted in a standard deviation of∼0.3in the measured magnitude, corresponding to a3σdetection.The second attempt at imaging the PSR J1528−3146field revealed a faint object in theR band image atαJ2000=15h28m34.s955,δJ2000=−31◦46′06.′′73,and in the B band imageatαJ2000=15h28m34.s945,δJ2000=−31◦46′06.′′71,consistent with the pulsar timing position. This potential counterpart is faint;we estimate R∼24.2and B∼24.5,but this photometryis somewhat uncertain due to calibration with the USNO B-1.0photographic magnitudes. This object is blue relative to most other stars in thefield.Our observation of PSR J1757−5322showed a possible object at the radio pulsar’s timing position,but it was difficult to see in the glare of a brighter star.Subtraction of the brighter star from the image using the daophot substar task reveals a faint object with R∼24.6atαJ2000=17h57m15.s174,δJ2000=−53◦22′26.′′17,consistent with the pulsar timing position.We subsequently obtained a near-IR image of thefield with PANIC on the 6.5m Clay telescope at Magellan Observatory on18April2003,observing for a total of72 minutes in K s band.We subtracted dark frames,then produced a sky frame for subtraction by taking a sliding box-car window of4exposures on either side of a reference exposure.We then added the exposures together,identified all the stars,and produced masks for the stars that were used to improve the sky frames in a second round of sky subtraction.Astrometry was again provided by the USNO B-1.0catalog,and photometric calibration by comparison with several2MASS stars in thefield.There is no object present at the pulsar’s position to the detection limit of the image,K s=20.8.The implied limit on the color corresponds to a main sequence spectral type of∼M4or earlier,and is thus consistent with a white dwarf.Several of thesefields are rather crowded;this was especially problematic in the case of PSR J1435−6100,whose position overlaps with three blended objects in our image.On the night of6June2003,we obtained a spectrum of the bright object near the pulsar position with the LDSS2on the Clay telescope,and determined that it is a reddened F-type main sequence star and thus not associated with the pulsar.We used the daophot allstar task to subtract stars near the positions of PSR J1157−5112and PSR J1435−6100,eliminating the possibility of fainter counterparts hidden by the nearby brighter objects in these cases.3.Discussion and ConclusionsWe detected optical counterparts for two out of the six IMBP systems we studied, PSR J1528−3146and PSR J1757−5322.From Table1,one sees that these are the two nearest targets.Thus,it is quite possible that deeper observations would reveal the coun-terparts in the remaining binaries as well.In Figure2,we show cooling curves for hydrogen atmosphere white dwarfs with masses from0.5M⊙to1.2M⊙,along with the observationally-inferred absolute R magnitudes of massive white dwarf pulsar companions versus the spin-down ages of their pulsars.The absolute magnitudes have large uncertainties which are difficult to quantify because the only constraint on the pulsar distances is based on dispersion measure and a model of the galacticelectron distribution(Cordes&Lazio2002);however,this exercise is still instructive.We note that in all cases where optical observations failed to detect an IMBP counterpart,the predicted magnitude is fainter than the observation’s detection threshold.As previously mentioned,it is thought that the companion stars in the PSR J1141−6545 and PSR B2303+46systems must have been fully evolved by the time the pulsars formed. Therefore,in these systems,the pulsar age does not constrain the white dwarf age and the failure to detect the PSR J1141−6545companion is not troubling.The detected optical counterpart of PSR B2303+46(van Kerkwijk&Kulkarni1999)is significantly fainter than predicted by the cooling model based on the pulsar’s spin-down age.In addition to the expectation that the white dwarf is older than the pulsar,this object has the largest z-distance from the galactic plane in this sample;it is above much of the ionized gas in the galactic disk,so the dispersion measure-based distance estimate could be significantly smaller than the true distance.In all of the other systems,the neutron star formedfirst and the pulsar’s spin-down age should,in principle,correspond to the time since the end of the companion’s evolution.The otherfive detected objects are all brighter than predicted by the cooling curves if they are as old as their pulsars’characteristic ages.Although there is a large uncertainty associated with the absolute magnitude of each object,as a group,they suggest that the standard spin-down model for pulsars may in fact significantly overestimate the pulsar age in these cases,possibly because P0was not much smaller than the current spin period.BAJ and SRK thank NSF and NASA for supporting their research.MHvK acknowl-edges support by the National Sciences and Engineering Research Council of Canada.DLK thanks the Fannie&John Hertz Foundation for its support.BAJ holds a National Research Council Research Associateship Award at the Naval Research Laboratory.Basic research in astronomy at NRL is supported by the Office of Naval Research.REFERENCESBailes,M.,Ord,S.M.,Knight,H.S.,&Hotan,A.W.2003,ApJ,595,L49 Benvenuto,O.G.&Althaus,L.G.1999,MNRAS,303,30Bergeron,P.,Wesemael,F.,&Beauchamp,A.1995,PASP,107,1047Camilo,F.,Lyne,A.G.,Manchester,R.N.,Bell,J.F.,Stairs,I.H.,D’Amico,N.,Kaspi, V.M.,Possenti,I.,Crawford,F.,&McKay,N.P.F.2001,ApJ,548,L187Camilo,F.,Thorsett,S.E.,&Kulkarni,S.R.1994,ApJ,421,L15Cordes,J.M.&Lazio,T.J.W.2002,astro-ph/0207156Edwards,R.T.&Bailes,M.2001a,ApJ,547,L37—.2001b,ApJ,553,801Kaspi,V.M.,Lyne,A.G.,Manchester,R.N.,Crawford,F.,Camilo,F.,Bell,J.F.,D’Amico, N.,Stairs,I.H.,McKay,N.P.F.,Morris,D.J.,&Possenti,A.2000,ApJ,543,321 Kulkarni,S.R.1986,ApJ,306,L85Lundgren,S.C.,Foster,R.S.,&Camilo,F.1996,in ASP Conf.Ser.105:IAU Colloq.160: Pulsars:Problems and Progress,ed.S.Johnston,M.A.Walker,&M.Bailes(San Francisco:ASP)497Neckel,T.&Klare,G.1980,A&AS,42,251Portegies Zwart,S.F.&Yungelson,L.R.1999,MNRAS,309,26Schlegel,D.J.,Finkbeiner,D.P.,&Davis,M.1998,ApJ,500,525Stokes,G.H.,Taylor,J.,&Dewey,R.J.1985,ApJ,294,L21Taam,R.E.,King,A.R.,&Ritter,H.2000,ApJ,541,329Tauris,T.M.&Sennels,T.2000,A&A,355,236Tauris,T.M.,van den Heuvel,E.P.J.,&Savonije,G.J.2000,ApJ,530,L93van den Heuvel,E.P.J.1994,A&A,291,L39van Kerkwijk,M.&Kulkarni,S.R.1999,ApJ,516,L25van Kerkwijk,M.H.,Bassa,C.G.,Jacoby,B.A.,&Jonker,P.G.2005,in ASP Conf.Ser.328,Radio Pulsars,ed.F.A.Rasio&I.H.Stairs(San Francisco:ASP)357Table1.Parameters of six target massive white dwarf binary systems Pulsar P Bτc P b e m c min d a Reference (ms)(109G)(Gyr)(d)(M⊙)(kpc)a Distance estimated from dispersion measure using model of Cordes&Lazio(2002) References.—(1)Kaspi et al.,2000;(2)Edwards&Bailes,2001b;(3)Camilo et al.,2001;(4)Jacoby et al.,in prep.Table2.Observations of massive white dwarf binary systems Pulsar Filter seeing Detection Limit Potential Counterpart a(arcsec)(magnitudes)(magnitudes)a Figures in parenthesis are uncertainties in the last digit quoted.Fig. 1.—Images of fields of PSR J1141−6545,PSR J1157−5112,PSR J1435−6100,PSR J1528−3146,PSR J1454−5846,and PSR J1757−5322.Circles indicate the 3σun-certainty in the pulsar position;tick marks show the pulsar position where a plausible coun-terpart was rge images are in R band.For PSR J1528−3146,the inset shows the B band image.For PSR J1757−5322,the inset at upper left shows the R band image after the subtraction of bright star near pulsar position,and the inset at lower left shows the K s image.For PSR J1141−6545,the timing position from Bailes et al.(2003)was used;in all other cases positions were taken from the references in Table 1.Fig.2.—White dwarf cooling curves and observations of massive white dwarf pulsar com-panions.The curves show the absolute R magnitude versus age for massive white dwarfs with hydrogen atmospheres.Points show the observationally-derived M R versus pulsar char-acteristic age for massive white dwarf pulsar companions with arrows indicating upper limits from non-detections.The diameter of each point is proportional to the most likely mass of the white dwarf,assuming a1.35M⊙pulsar and60◦orbital inclination with these exceptions: the companion of PSR J1141−6545has a mass of(0.99±0.02)M⊙(Bailes et al.2003);the most likely masses of the companions of PSR B2303+46and PSR J1157−5112are greater than the Chandrasekhar mass,so we have assigned them diameters corresponding to1.4M⊙. The curves are based on the luminosity–age relation for0.5,0.7,and1.0M⊙white dwarfs with hydrogen fractions of10−4,and a1.2M⊙white dwarf with a hydrogen fraction of10−6, all with zero metallicity for the envelope(Benvenuto&Althaus1999).To these cooling relations,we applied bolometric corrections and colors as a function of temperature for a log(g)=8white dwarf with a hydrogen atmosphere(Bergeron et al.1995).Apparent R magnitudes were converted to absolute R magnitudes using the dispersion measure-distance model of Cordes&Lazio(2002),with an extinction correction from Neckel&Klare(1980) for objects within5◦of the galactic plane,and from Schlegel et al.(1998)for higher latitude pulsars.In the cases of PSR J2145−0750and PSR B0655+64,R was calculated based on the measured V and inferred temperature of Lundgren et al.(1996),using colors from Bergeron et al.(1995).Photometry for PSR J0621+1002is from Kulkarni(1986);PSR B2303+46from van Kerkwijk&Kulkarni(1999);and PSR J0621+1002from van Kerkwijk et al.(2004).。