Engine exhaust from this product contains chemicals known to the State of California to cause cancer, birth defects, or other reproductive harm.Thank youThank you for selecting our products. Our personnel have proudly made every effort to ensure that your new pressure washer is of the quality you expect. But things do occasionally go wrong. This is why every pressure washer is covered by a limited warranty. Among other things, this warranty provides for the replacement of parts found to be defective during the operation of your new pressure washer. Please note that the owner/operator has certain obligations under the terms of the warranty. Be sure to read this manual for directions on proper installation, start-up, use, and stor-age of your pressure washer.Your new pressure washer was tested after production for proper pressure and flow.Please note that this process will sometimes leave a water residue in the pump. The dealer you have purchased your new machine from should review with you the prop-er installation, start-up, use, and storage. Most ‘big’problems occur when shortcuts are taken in one of these processes. If a problem occurs that you need some assis-tance with, please feel free to contact us at the listing below:Warranty Service Center520 Brooks RoadIowa Falls, IA 501261.800.648.6007Be familiar with the model plate located on your machine.Have the model and code number with you when you call for service. (Located on motor).Please make note of Model Identification Model #Code #Serial #Always have this information when callingWarranty Service Center.Statement of WarrantyThe manufacturer of this product agrees to repair or replace designated parts that prove defective within the warranty peri-od listed in the chart below. Specific limitations and exclusions apply. This warranty covers defects in material and workmanship and not failure due to normal wear, depreciation, abuse, accidental damage, negligence, improper use, maintenance or storage. To make claim under the terms of the warranty, all parts said to be defective must be returned to a designated Warranty Service Center for warranty inspection. The judgments and decisions of the factory-authorized personnel concerning the validity of war-ranty claims are final.Many components are covered by warranties given by their respective manufacturers. These warranties pass through to the end user. As a factory authorized and trained warranty service center the factory will honor the terms of all component warranties and satisfy claims of the appropriate warranty provisions.Normal wear items include but are not limited to: hoses, nozzles, filter, valves, seals and are not covered by this warranty.This warranty is in lieu of all other warranties, express or implied, including without limitation any warranties of mer-chantability or fitness for a particular purpose and all such warranties are hereby disclaimed and excluded by the Manufacturer. The Manufacturer's warranty obligation is limited to repair and replacement of defective products as provided herein and the Manufacturer shall not be liable for any further loss, damages or expenses, including damages from shipping, accident, abuse, acts of God, misuse or neglect. Neither is damage from repairs using parts not purchased from the Manufacturer or alterations performed by non-factory authorized personnel. Failure to install and operate equipment according to the guidelines put forth in the instruction manual shall void warranty.Accessories:Includes tips, guns, wands, hoses, injectors, unloaders, sandblasters, flat surface cleaners, hose reels, turbo nozzles, drain nozzles, brushes, foamers, GFCI units, thermal relief, filters, tanks, etc.90 daysWARNINGThe following warnings must be followed, Failure to follow these warnings could result in serious personal injury or death!Electrical equipment can cause shock and sparks.Do not bypass or remove the grounding prong in any electrical plug.Keep electrical plugs, connections and cords out of water and moisture.Refer to instructions prior to equipment operation.Disconnect from power source before servicing.Inspect and repair damaged or exposed electrical components prior to use.Never splice electrical cords on pressure washers.Never allow children or untrained personnel to operate machinery.High Temperature Water.Wear protective clothing and face shield.Do not direct water stream toward self or others.Do not spray electrical apparatus.High pressure water can cause death or serious injury.High pressure fluid can create a high pressure stream or ruptured vessel.Wear safety face shield.Relieve pressure before servicing.Do not modify/repair/rework vessel or change safety relief or pressure setting.Do not direct stream toward self or others.Pressurized fluid streams and rupturedpressure vessels can cause death or serious injury.Important Safety InstructionsWARNING - when using this product basic precautions should always be followed, including the following:1.Read all the instructions before using the product.2. To reduce the risk of injury, close supervision is necessary when a product is used near children.3.Know how to stop the product and bleed pressure quickly. Be thoroughly familiar with the controls.4.Stay alert - watch what you are doing.5.Do not operate the product when fatigued or under the influence of alcohol or drugs.6.Keep operating area clear of all persons.7.Do not overreach or stand on unstable support. Keep good footing and balance at all times.8.Follow the maintenance instructions specified in the manual.9.“WARNING - Risk of Injection or Injury - Do Not Direct Discharge Stream At Persons.”SAVE THESE INSTRUCTIONSGROUNDING INSTRUCTIONSThis product must be grounded. If it should malfunction or breakdown, grounding provides a path of least resis-tance for electric current to reduce the risk of electric shock. This product is equipped with a cord having an equipment-grounding conductor and a grounding plug. The plug must be plugged into appropriate outlet that is properly installed and grounded in accordance with all local codes and ordinances.DANGER - Improper connection of the equipment-grounding conductor can result in a risk of electrocution. Check with a qualified electrician or service personnel if you are in doubts to whether the outlet is properly grounded. Do not modify the plug provided with the product - if it will not fit the outlet, have a proper outlet installed by a qualified electrician. Do not use any type of adapter with this product.The pressure was set at the factory during the testing procedure, no adjustments to the machine should be required for operation. During operation do not leave the machine running for more than 2 minutes without the trigger gun being pulled. Although your machine has a by-pass valve on it and may have a thermal relief system, this can cause extensive pump damage. If machine will not be discharging water for more than 2 minutes, shut the machine off.Your machine is supplied with interchangeable spray tips. The black tip lowers your pressure and draws chemical. The other tips are for high pressure rinse at different spray angles. Red is 0 degree,yellow is 15 degree, green is 25 degree and white is 40 degree. The yellow tip is used for most stan-dard applications. Be sure the quick coupler is fully engaged before pulling the trigger gun. Failure to do may result in the tip becoming a projectile and may be lost or damage to property or persons may occur.Your pressure washer is supplied with a downstream chemical injector with dual chemical selector.The 1/4” clear vinyl tube is to be inserted into the desired chemical to apply. Be sure to use the black,low pressure nozzle to inject chemical. The chemical injector will only open up and allow chemical into the line when this tip is used. This tip enables the pressure to drop to approximately 250 PSI to draw chemical. The injector can be turned on and off or the rate of injection can also be set by turn-ing the knob that the clear vinyl tube attaches to. See calibration below.Be sure to flush injection system with clear water after use.The chart below is an average ratio observed using water as the test fluid to draw through the injector and serves only as a guide, your actual ratio will vary. Setting on valve Injection Ratio 96:188:1710:1613:1520:1428:1335:1250:1For better accuracy, calibrate using the following formula:(GPM x 128) ¸ (ounces drawn in 1 minute) = x:1.IE: If a 2.0 GPM machine draws 8 ounces of chemical in 1 minute: (2 x 128) 8 = 32:1Factory supplied hoses are sized in length and diameter for best operational performance and sized within the pressure capabilities. Additional hose added to the machine may change the performanceof the machine. Consult factory if you have any questions. When replacing or disconnecting thequick couplers be sure the machine is shut off and relieve the pressure from all hoses.DuringOperationNozzlesChemicalinjector useHoses andcouplers CalibrationLow NozzlePressure SoapInjectorNotWorkingProperly Surging Operation NoticeLow nozzle pressure is a common complaint. In a majority of instances, low nozzle pressure is gener-ally caused by one of the following:1. Plugged nozzle tip.2. Inlet screen plugged.3. Insufficient flow in gallons per minute (not pressure) to the pump.4. Unloader valve stuck open due to debris lodged under the check valve ball.5. Customer use of shut-off-type quick connectors.6. Plugged hose.7. Incorrect nozzle or soap tip installed.Another complaint is that pressure surges.That is, when the trigger is pulled, pressure is satisfactory for a moment then falls off. When the trigger is released, pressure builds up to normal levels. This isgenerally a sign that the water supply cannot provide the flow rate (gallons per minute) required by the pump. Following are some possible solutions:1. Make sure the supply is not restricted; that there are no under-sized fittings and the inlet screen is unobstructed.2. Make sure the flow rate of the water supply is sufficient for the pump. First, find the capacity of your pump in gallons per minute (gpm) as shown in the Water Supply section on Page 5. Then determine the flow rate of your supply by measuring the gallons that can be delivered in one minute. If your supply does not deliver the gpm your pump requires, do not use the pump. It will suck air, causing cavitation that can quickly damage pump components.3. Check for leaks in the supply fittings. Any leak will cause the pump to draw air and perform poorly. When a soap injector is not working properly, the problem is generally fairly easy to isolate. Check the following:1. Make sure the Black, soap tip is installed. Soap injectors will not work when high pressure nozzles are installed.2. Be sure that the soap injector valve is turned on, and turn selector valve to desired setting.3. A piece of debris may be caught in the injector valve, injector ball valve, or orifice. Disassemble and clean the injector.User maintenance procedures include replacing valves and seal. Unloader components are not user serviceable. Repairs involving unloaders and crankcase components should be referred to the factory or a factory authorized repair center.Mounting plateReplacement Quick Tips For All UnitsForward grip/soap control valve.Activates chemical injector by lowering pressure and allowing water to gothrough soap tip. (Turn handle counterclockwise to activate injector)Twist-fast socket(527800)Dual wand section(542002)Soap tip(4030 1/4meg)。
XXXXXXXXXX仪器设备标准操作规程1 目的:HH-S6电热恒温水浴锅标准操作规程。
2 范围:HH-S6电热恒温水浴锅。
3 责任者:检验操作人员。
4 规程内容:4.1 型号与操作条件:4.1.1 HH-S6型4.1.2 操作条件4.1.2.1 电源要求:220V50Hz。
4.1.2.2 控温范围:5O C~IOO O C4.2 操作说明:4.2.1 加水往水浴锅内加入一定量的蒸馏水,锅内水量不可低于二分之一,不得高于三分之二;4.2.2 通电插上电源,打开仪表盘电源开关,仪表通电后,上排显示InP,下排显示分度号,表示输入类型。
经过2秒钟后,上排显示量程上限,表示测量范围。
在经过2秒钟后,上排显示测量值,下排显示设定值,进入正常工作状态。
4.2.3 温度的设定:按功能键,上排显示sp。
按▲或▼键,使下排显示为所需要的设定温度值。
再按功能键,上排显示ST,按▼或▼键,使下排显示为所需要的定时时间值,再按设定键回到标准模式。
4.2.4 控制参数的设定:按功能键4秒钟以上,上显示控制参数的提示符(详见控制参数一览表),按▲或▼键,使下排显示为所需要的值。
继续按功能键,上排依次显示各参数的提示符,按▲或▼使各参数为所需要的值。
在按功能键4秒钟以上,回到标准模式。
(无按键下1分钟后自动返回标准模式)。
注:若上排数码管显示的下边出现000则说明热电偶接反,上边出现000 则说明热电偶∕热电阻>开路或温度超过测量范围。
4.2.5 定时功能:当ST=0时取消定时功能;当ST不等于零时,有定时功能,计时采用顺计时,当定时时间,下排数码管显示END,按增加键5秒以上可重新启按制器。
4.2.6 沸点功能:当选择水浴锅时,设定温度为>=99度时,加热全输出是水沸腾。
4.2.7 自整定功能:按▼键5 秒后运行指示灯闪烁,控制器开始自整定,自整定结束后运行指示灯长亮;得出一组最佳的PID参数,控制器按新PID参数进行控制。
HH-1S型数显恒温油浴锅
使用时必须先加油于锅内,接通电源,在温控仪上直接用拨码开关选择所需温度即可,温控仪会自动恒温。
绿色指示灯亮,表示加热器工作;红灯亮则处于恒温状态。
工作完毕,切断电源。
1.设定温度:按一下SET键可设定温度。
按SET键下排数码管数据闪动(上排正常测温),此时仪表进入温度设定状态,按△键设定值增加,按▽键设定值减小,再按一下SET键仪表回到正常工作状态温度设定完毕。
2.传感器误差的修正:按SET键3秒进入仪表内层菜单,选定参数SC后配合△或▽可修误差。
显示分辨率为0.1度传感器误差的修正的范围为±19.9,显示分辨率为1度的修正值为±50度。
完成后再按SET键3秒退出。
仪器出厂时修正值为0。
3.启动自整定:本产品能适应绝大多数控制系统,只有控制效果特别不理想时方可启动自整定。
方法一:仪表上电后按移位键3秒启动自整定。
方法二:按SET键3秒进入仪表内层菜单,选定参数At后按△键将下排调成1,再按SET键3秒可开启自整定。
自整定进行时Atu指示灯闪烁,其过程是仪表控制加热系统在设定点附近波动二次,当Atu指示灯灭时自整定结束,新的PID参数被锁存在芯片中。
超级恒温油浴602A超级恒温油浴是根据原来602的改进型,内胆采用优质不锈钢,一次性拉伸成型。
具有箱内液体通过外部循环再回流功能和不同性质液体通过循环加热器加热功能。
广泛用于干燥,浓缩,蒸馏,浸渍化学试剂,浸渍药品和生物药品,也可用于水浴恒温加热和其他温度实验,是生物,遗传,病毒,生产,环保,医药,卫生,化工实验室,分析室,教育科研的必备工具。
其主要热点:1、工作室油槽选材不锈钢有优越的抗腐蚀性能。
2、控温精确、数字显示、自动控温。
3、操作简单,使用安全。
一、产品技术参数1.1温度范围:RT+5~199℃1.2波动度:±1℃1.3显示精度:1℃1.4工作室尺寸:500*300*150MM(深*宽*高)1.5外形近似尺寸:750*400*450MM1.6升温速度:≤8℃/MiN1.7电源:220V 50HZ1.8加热功率:1500W1.9搁板架配置:1块1.10净重约:30KG1.11带木箱包装毛重约:40KG1.12包装尺寸约:800*500*550MM二、温度控制系统及特点2.1 温度测量:PT100高精度传感器2.2 控制系统:数显仪表2.3 循环系统:循环泵2.4 成套电器:数显智能仪表、循环泵、可控硅、熔断丝保险器三、箱体结构及用料3.1 油浴箱由冷板喷塑外壳,保温层,内胆组成,底部不锈钢加热器加热,保护搁板位于加热器上面,用于搁置试样物品,也可保护发热器。
箱体内循环泵利用箱内温度更均匀。
3.2 外箱体材料:外箱体采用优质冷板经磷酸皮膜盐处理后高温喷塑,上开门结构,颜色为电脑灰3.3 工作室材料:采用不锈钢内胆,不锈钢保护搁架及搁板1个3.4 测温口:PT100传感器从箱内保护式搁架下方进入3.5 电源线:电源线位于箱侧控制器位置3.6 控制器:位于箱体右侧面,内置控制系统,补偿导线连接PT100传感器3.7 加热元件:不锈钢电加热器3.8 整机重量:约25KG3.9 标准包装:木箱四、加热系统4.1 不锈钢加热器,位于箱内底部,由PT100传感器响应控制系统4.2 不锈钢加热器加热输出由数显智能仪表控制五、安全保护系统5.1 熔断式保护器5.2 工作室超温保护器,上限超温报警5.3 漏电保护,触电保护,确保使用安全性六、配备的技术资料及附件6.1 操作说明书6.2 产品保修卡6.3 合格证七、售后服务7.1 负责对用户进行远程技术指导7.2 及时提供设备的备件、配件7.3 提供设备使用过程中的技术咨询和支持7.4 接到客户故障通知8个工作时内立即响应八、主要元器件数显智能控制仪表PT100传感器不锈钢加热器。
HH系列电热恒温水浴锅作业指导书
1.目的
规范电热恒温水浴锅的操作方法。
2.适用范围
适用于HH系列电热恒温水浴锅的操作。
3.工作原理
通过智能数字温度控制电热管对水浴锅内进行加热。
4.操作步骤
4.1将水槽内加水高于隔板50mm
4.2接通电源,打开电源开关。
设定温度按SET键,然后按▲或▼键
调节所需温度,再按SET键确定温度。
4.3使用完毕,切断电源,将水全部放出,清洗干净,填写使用记录。
4.4异常处理:1)显示正常但不加热:由于设定值低于实际测量温度,查看设定值解决。
2)加热灯亮但不升温:由于加热管断线或输出继
电器损坏,检查更换。
3)显示数据暗淡且抖动:工作电压偏低。
5注意事项
5.1该产品使用220V交流电,电源采用三角安全插头,其中一只最
长的脚为安全接地脚,为此对应使用的三眼插座应妥善接地。
5.2水箱在加水时,水量不能少于最低水位(即不能使电热管露出水面),以免烧坏电热管。
造成漏电、漏水。
但也不要加的过多,以免沸腾时水量溢出表面。
5.3由于水箱内加温时水温上下温差较大,所以当需温度计测量时一定要待水温稳定状态
时测量。
5.4加热管之间有一根不锈钢金属棒做温度传感器,切勿碰撞,以免失灵。
5.5长期不用应将水槽内水放掉擦净存放在干燥处。
5.6仪器工作环境:温度-5℃—50℃,相对湿度35%—85%。
5.7注水时绝对不能将水流入控制箱内。
HHS系列电子恒温水浴锅使用说明
操作方法:
1 本产品使用电源必须符合交流220V50HZ的规定要求并应接装安全接地线。
2 先将清水注入浴锅内,至规定的刻度线,然后接通电源,打开电源开关,指示灯亮说明电路已正常接通。
3 旋转调温旋钮,就能调整所需浴温要求的规定温度,顺时针方向旋转为增加水浴温度,反正则为降低水浴温度,以表上读数为准。
4 在调整温度旋钮时,红绿指示灯会交替亮灭,当温度达到规定要求的温度时,红指示灯亮,表示加热电源已经切断,水浴停止加热,当浴温低于规定的要求温度时,绿指示灯亮加热电源自动接通继续进行水浴加热。
恒温油槽厂油浴锅安全操作及保养规程一、前言恒温油槽厂油浴锅是一种常见的实验仪器,常用于科研实验室及工业生产中。
正确的操作及保养可以提高设备的使用寿命,规避操作过程中的安全隐患。
本文旨在介绍恒温油槽厂油浴锅的安全操作及保养规程,帮助用户正确使用和保养设备。
二、安全操作规程2.1 设备安装恒温油槽厂油浴锅应安装在平稳的地面上,并确保地面的承重能力可以满足设备的重量。
2.2 电源连接恒温油槽厂油浴锅的电源线要接地,不能漏电。
接口要干燥,避免接触潮湿的环境。
2.3 操作方法1.在使用前,应验证设备温度显示是否准确。
可以使用温度计检测油浴锅中的油温,确保两者温度相同。
2.打开设备电源开关,并根据需要设置所需温度。
3.将待测样品放入油槽中,并根据需要调整样品的浸没深度。
4.正确安装温度传感器,确保温度传感器浸入样品中,并取代油槽温度。
5.注意不要使样品与油浴锅的加热管直接接触。
2.4 紧急处理恒温油槽厂油浴锅在操作过程中可能出现一些问题,例如超温报警、电源故障等。
当出现紧急情况时,应立即断电,并通知专业技术人员进行修理。
2.5 注意事项1.操作人员应戴上适当的防护手套和护目镜,以避免操作过程中的烫伤及眼睛受损。
2.在操作过程中,应避免油浴锅中的油溅到身上,以避免烫伤。
三、保养规程3.1 日常清洁1.切勿使用酸碱性溶液或锐器清洁油浴锅的表面。
2.每日下班之前应清洗油浴锅的内外表面,并将油槽中残留物清除干净。
3.温度传感器每天也应进行清洗。
3.2 维护保养1.定期检查油浴锅温度传感器和温度显示器的准确度,发现问题及时调整。
2.定期检查加热装置的工作状态,尤其是控制装置的故障。
3.3 长期闲置长期不使用油浴锅时,应将设备进行清洁和保养,并储存于干燥、阴凉的空间,避免掉漆、生锈等问题。
四、结尾恒温油槽厂油浴锅的安全操作及保养规程是使用和保养该设备的重要指南。
正确的操作和保养能够提高设备使用寿命,同时保证实验准确性和人员的安全性。
HYGY油介损测试仪使用说明书武汉市华英电力科技有限公司地址:武汉市东湖高新区光谷大道62号光谷总部国际7栋4层电话:400-027-0098*************传真:************E-mail:**************网址:目录一、概述 (3)二、控制面板 (4)1.主面板功能区 (4)三、油杯简介 (5)1.油杯结构 (5)2.油杯技术标准 (5)3.油杯使用 (5)4.拆装油杯电极 (6)5.装入油样 (6)6.油杯清洗 (6)7.油杯主要技术参数 (7)四、工作原理 (7)1.仪器内部功能构造框图 (7)2.介损测量原理图 (7)3.工作原理 (8)4.名词解释 (8)五、主要技术指标 (9)六、主机操作 (10)1.主菜单界面说明 (10)2.仪器操作说明及注意事项 (11)3.菜单操作详细说明 (12)七、仪器成套性 (16)八、运输、贮存 (16)九、开箱及检查 (17)十、其它 (17)一、概述HYGY-油介损测试仪是用于绝缘油等液体绝缘介质的介质损耗角及体积电阻率测试的一体化结构的高精密仪器。
内部集成了介损油杯、温控仪、温度传感器、介损测试电桥、交流试验电源、标准电容器、高阻计、直流高压源等主要部件。
其中加热部分采用了当前最为先进的高频感应加热方式,该加热方式具备油杯与加热体非接触、加热均匀、速度快、控制方便等优点。
交流试验电源采用AC-DC-AC转换方式,有效避免市电电压及频率波动对介损测试准确性影响,即便是发电机发电,该仪器也能正确运行。
内部标准电容器为SF6充气三极式电容,该电容的介损及电容量不受环境温度、湿度等影响,保证仪器长时间使用后仍然精度一致。
仪器内部采用全数字技术,全部智能自动化测量,多种模式测式,配备了大屏幕(320×240)触控式显示器,全中文菜单,每一步骤都有中文提示,测试结果可以打印输出,操作人员不需专业培训就能熟练使用。
郑重提醒:本公司提醒用户,该设备有高压输出,如果使用不当可能危及人身安全!操作员在未仔细阅读使用说明书之前,严禁使用本仪器!二、控制面板图一控制面板图1.主面板功能区①微型打印机:完成输出数据打印;②总电源开关:设备输入电源控制;③触控式液晶显示屏:主操作区,设备各项目的设定以及测试操作(具体操作见“操作”);④状态指示区;a)高压灯(红色):如果灯亮,表示油杯上已经带高压电;b)加热灯(绿色):如果灯亮,表示加热炉正在加热;如果灯闪烁,表示已接近设定温度,正在恒温;灯灭时同样要注意油杯上的高温;⑤复位:初始化整机的全部控制;三、油杯简介1.油杯结构图二油杯结构图①油杯杯体,测量加压极②油隙③油杯内电极,测量测试极④内电极固定钮⑤油杯内电极,测量屏蔽极⑥测试端⑦温度接口2.油杯技术标准油杯采用三极式结构,完全符合GB/T 5654-2007标准,极间间距2mm,可消除杂散电容及泻漏对介损测试结果的影响。
数显恒温油浴锅操作规程
一、操作方法:
1、温度设定:按SET键可设定温度,按SET键下排数码管数据闪动(上排正常测温),表示仪表进入温度设定状态,按△键设定值增加,按▽键设定值减小,持续按增加键或减少键设定值会快速变化,再按一下SET键仪表回到正常工作状态温度
2
3
1
2
3
4
5
6、 t 比例周期调节输出动作周期1-200秒
7、 Sc 测量值偏置传感器误差修正 -50℃-50℃
8、 Lck 参数锁0不锁定;1锁定除设定值外的参数;2锁定所有参数0-2
三、参数意义:
1、AL:超温报警参数,当温度超过AL值时ALM指示灯亮。
2、HY:主输出回差,防止继电器频繁动作。
3、P:为比例带,即比例控制值,仪表以设定值为中心P值分布在设定点两边,在P 值范围内仪表以PID方式工作。
4I:用于消除静差,太小会引起温度过冲。
一般10-40秒。
5、T:为加热输出循环周期,即继电器工作循环周期,T值小控制效果好,但T 值太小继电器会因频繁工作而减少寿命。
一般T值取20-60秒。
6
7
1
2
3
4
5。
单孔磁力搅拌油浴锅安全操作及保养规程单孔磁力搅拌油浴锅是一种常用于实验室的设备,它可以用来加热和搅拌油浴以进行化学反应和合成实验。
在使用单孔磁力搅拌油浴锅时,需要注意正确的操作和保养,以确保安全和使用寿命。
本文将介绍单孔磁力搅拌油浴锅的基本原理、安全操作规程、保养注意事项等方面的知识。
基本原理单孔磁力搅拌油浴锅的基本原理是利用电磁铁在水平方向上施加磁场,使得加热板上的磁力子能够旋转,从而带动磁力棒在容器内快速搅拌。
同时,通过控制加热板的温度,可以将油浴加热到所需的温度,从而实现化学反应和合成实验。
安全操作规程使用单孔磁力搅拌油浴锅需要遵循以下安全操作规程:1. 确保设备处于良好状态在使用单孔磁力搅拌油浴锅之前,需要检查设备是否处于良好状态。
首先,检查加热板和杯套是否平整,加热棒是否正常,磁力棒是否完好。
其次,检查电源线和插头是否完好,插头是否插紧。
2. 使用适当的容器在单孔磁力搅拌油浴锅中进行实验时,需要使用适合的容器。
一般来说,小型实验可以使用试管、烧杯等,在使用大容量容器时,需要选择大口径的容器以适应磁力棒的搅拌。
另外,需要注意容器底部是否平整,是否有裂纹或者缺口,以免引起事故。
3. 控制加热温度在使用单孔磁力搅拌油浴锅进行化学反应和合成实验时,需要控制加热温度。
一般来说,实验方案会给出所需的温度范围和控制方法。
需要根据实验情况选择合适的温控方式,保证加热温度的准确性和稳定性。
4. 避免过载操作单孔磁力搅拌油浴锅具有一定的负载能力,但是仍然需要避免过载操作。
在实验过程中需要根据实验方案和设备性能选择适当的容器和试剂量,避免过度负载导致设备失效或者事故发生。
5. 安全关闭设备在使用单孔磁力搅拌油浴锅结束后,需要采取安全关闭设备的措施。
首先,需要关闭加热器并切断电源,再将磁力棒和容器从加热板上取下,等待设备降温,才能进行拆卸、清理和保养。
保养注意事项除了正确的操作,单孔磁力搅拌油浴锅还需要进行保养,以延长设备的使用寿命和保证操作的安全性。
电热恒温油浴锅原理安全操作及保养规程电热恒温油浴锅是一种常用于化学实验中的设备,它采用电加热的方式控制油的温度,可以恒温加热样品,广泛应用于化学、医学等领域。
本文将介绍电热恒温油浴锅的工作原理,以及注意事项和保养规程。
原理电热恒温油浴锅以油为介质,通过加热杯底和外壳来加热油,油的温度通过恒温控制器来进行调节,以达到恒定温度的目的。
恒温控制器一般由温度传感器和恒温电路组成。
温度传感器负责检测油的温度,并将温度信号传递给恒温电路,恒温电路根据设置的温度和当前温度进行比较,并通过加热或冷却来维持恒定温度。
安全操作在使用电热恒温油浴锅时,需要注意以下安全事项:1.选择合适的介质油。
不同油的闪点和燃点不同,需要根据实验需求选择合适的介质油,并根据实验室的规定进行储存和处理。
2.不要加入过多的样品。
油浴锅需要预留一定空间来扩散热量,过多的样品会影响加热效果,并有可能溅出油,造成安全事故。
3.注意加热过程中的温度变化。
加热过程中,油的温度会逐渐升高,应在油温接近目标温度时逐渐降低加热功率,以免温度过高影响实验结果。
4.禁止使用损坏的设备。
如发现设备有损坏或异常,应立即停止使用,并进行维修或更换。
保养规程定期对电热恒温油浴锅进行保养,可以延长设备寿命,提高工作效率。
以下是常规维护和保养注意事项:1.清洁设备。
使用过后应及时清洗设备,去除残留物和污垢,并用干净的布擦干设备,防止生锈和损坏。
2.检查电源线和接线端子。
定期检查电源线和接线端子是否松动或有破损,避免电线断路,短路等电气故障。
3.检查恒温控制器。
定期检查恒温控制器的配件和连接线是否正常,如有破损或松动应及时更换或修理。
4.更换介质油。
不同样品的要求不同,使用前需要了解样品的所需的介质油,定期更换过期的油,防止污染和泄露。
总结电热恒温油浴锅是一种常用于化学实验的设备,掌握其工作原理以及注意事项和保养规程对于改善实验效果,保障实验安全具有至关重要的作用。
使用时需谨慎,特别注意安全事项,并定期进行保养和维护,以达到更好的效果。
HHS系列电子恒温不锈钢水浴锅说明书
一、概述
1、本产品设计结构合理,用料精良,温度恒定好,控温精度高,加热时间短,热能损
耗小,性能可靠,操作方便,外型美观,经久耐用,是医疗卫生单位,工矿企业科研系列大专院校化学实验室蒸馏、干燥、浓缩合成以及恒温加热化学药品,生物制品等理想的化验设备。
2、本产品内胆、顶盖等部件全采用不锈钢材质制造,最大浴孔直径120mm,每孔配有
四环一盖,采用管状电热元件直接加热无级调温,可在室温~100℃范围内任意选择某点恒温度作使用温度。
二、主要技术参数与规格
1、电源:200V 50Hz
2、控温范围:室温~100℃
三、使用方法及注意事项
1、水箱应放在固定的平台上,电源电压必须与本产品要求的电压相符,电源插座应采
用三孔安全插座,必须安装电线。
2、使用前应先将水加入箱内,水位必须高于隔板切勿将水或水位低于隔板加热,以防
损坏加热管。
3、插上电源,将控温设定旋纽调至所需要的温度刻度,绿灯亮表示加热升温,红灯亮
定温、保温。
盐浴炉、油浴炉安全操作规程(最新版)编制人:__________________审核人:__________________审批人:__________________编制单位:__________________编制时间:____年____月____日序言下载提示:该文档是本店铺精心编制而成的,希望大家下载后,能够帮助大家解决实际问题。
文档下载后可定制修改,请根据实际需要进行调整和使用,谢谢!并且,本店铺为大家提供各种类型的安全管理制度,如通用安全、交通运输、矿山安全、石油化工、建筑安全、机械安全、电力安全、其他安全等等制度,想了解不同制度格式和写法,敬请关注!Download tips: This document is carefully compiled by this editor. I hope that after you download it, it can help you solve practical problems. The document can be customized and modified after downloading, please adjust and use it according to actual needs, thank you!In addition, this shop provides you with various types of safety management systems, such as general safety, transportation, mine safety, petrochemical, construction safety, machinery safety, electrical safety, other safety, etc. systems, I want to know the format and writing of different systems ,stay tuned!盐浴炉、油浴炉安全操作规程1、开炉前必须先开通风机,凡有冷却设备的应先将冷却设备开好。
HH-W602高精度循环油浴锅说明书一、简介:
HH-W602高精度循环浴锅是我厂在恒温油浴锅基础上增加了循环功能,极大改善了以往油浴锅箱内各点油温不均匀的现象,同时增加了外循环接口,可以与仪器以外的实验设备相连接,如反应釜等。
HH-W602超级恒温油浴锅广泛应用于蒸馏、干燥、浓缩以及浸渍化学药品或生物制品。
油浴锅是各大中专院校、环保、科研、卫生、防疫、石油、冶金、化工、医疗等单位实验室化验人员必备的理想工具。
二、特点:
1、箱体内保温材料采用优质玻璃纤维作为隔热材料,确保仪器在高温使用时箱体外无明显温升。
2、加热器、油箱、循环泵全部采用优质不锈钢制造。
3、采用微电脑液晶温控仪,各种工作状态具有中文显示功能,控温精确,操作简单。
4、温控仪具有温度显示及恒温定时递减显示功能。
5、传感器故障代码提示功能。
6、显示温度与实际温度误差校正功能。
7、超温保护,超温蜂鸣器提示功能。
8、具有内、外循环双重功能。
9、循环泵定时功能。
三、技术参数:
1、控温范围:室温-300.0℃
2、温度波动:±0.1℃
3、温度均匀性:0.1℃
4、显示分辨率:0.1℃
5、显示方式:LCC液晶屏显示
6、控温方式:微电脑PID控制,PT100传感器
7、恒温定时:1分钟-99小时59分钟或常开
8、循环定时:1-120分钟或常开
9、循环流量:6L/min,功率40W
10、加热功率:2000W
11、内胆尺寸:400×300×180mm
12、开口尺寸:300×265mm 槽深180mm
13、电源:220v 50Hz
四、使用方法:
1、设备安装前应将油浴锅放置在平整的工作台上,先进行外观的检查:外
观应无破损,仪表外观完好,电源线无损伤,插头完好,电源开关灵活。
每台设备的电源线都接有单相三极插头,其中插头最上方的电极为接地线,用户使用时必须使用带有安全接地的电源插座。
2、通电前先向油箱内加入适量的油,以离上盖5-8cm为宜。
油的选择根据
不同使用温度而定:室温-150℃:甲基硅油; 100-250℃:苯基硅油;
200-300℃:350#高温导热油。
3、数字式温度控制仪表设定与使用:
(1)打开电源开关,电源指示灯亮,打开加热开关,温控仪亮,等待约20秒后,上排显示箱内实际温度,下排显示设定温度。
点击功能键,上排显示sp,下排显示设定温度,点击△或▽选择工作温度,再次点击功能键,上排显示st,下排显示需要加热的时间,点击△或▽选择恒温定时时间,如无需定时将时间设定为00.00即可设定完毕,再次点击功能键退出,此时将按设定参数运行。
(2)在箱体内放入需要加热的试剂瓶(如果使用外循环,请在加热前连接好循环管),打开循环开关,此时油在箱内循环流动,确保箱内各点油温的均匀性。
(3)实验完毕,先关闭循环开关,然后将温度设定为0,关闭加热开关,最后关闭电源开关,并切断电源。
五、注意事项:
1、本系列产品的各项技术指标出厂前均经过实际运行检验,如果用户因工艺要求需采用自备的温度计(具有一定准确度)进行测定时,以用户的温度计为准,如果温控仪表显示的数值与温度计有差异,用户通过温控仪内部此时将显示温度与实际温度调整一致。
2、超级恒温油浴锅使用时,必须先加油后通电,严禁干烧。
3、油浴锅使用时,必须有可靠的接地以确保使用安全。
4、油位低于电热管,不要通电使用,以免电热管爆裂损坏。
5、油位也不可过高,以免溢入电器箱损坏元件。
6、定期检查各接点螺丝是否松动,如有松动应加紧固,保持各电气接点接
触良好。
7、在使用时请戴好安全防护手套,以免烫伤。