Single point load cells

Special features

? Accuracy class C3MR and

C6MR with OIML-R60 test report ? Off center load compensated (OIML R 76)

? Degree of protection IP67 (according to EN 60 529) ? Shielded connection cable ? Suitable for multi range applications

Optional:

? Explosion proof versions ? Different cable lenghts

? Aligned output, suitable for connection in parallel

Dimensions (in mm; 1 mm= 0.03937 inches)

Load direction

Cover with ATEX, Zone 1 and 21; Zone 2 and 22 (31 mm x 25 mm)

x4

SP4C3-MR 20kg

Connection cable

(see page 3)

v

year week

12345678

Plug

(see page 3)

M6x0.5, 16 deep

34

M6x1-6H 35

6-M6 1-6H, 16 deep

Round cable 3000"50 mm

Mounting:

19.1

99,6. 150

19.1 6.1

Max. capacitiesx 36 kg:

Cylindrical head screw M6-8.8 Tightening torque: 6 N ⊕m * SP 4/1 kg: 18 mm

Max. capacitiesy 50 kg:

Cylindrical head screw M6-10.9 Tightening torque: 14 N ⊕m

B2610-1.0 en

Specifications

Type

SP4M... Accuracy class 1)

C3MR Max. number of load cell intervals (n LC Maximum capacity (E max )2)

) 3000

kg g

1 3 5 7 10 1

15 20 2

30 2

50 5

75 5

100 10

150 10

200 20

Minimum LC verification interval (v min ) 0.1

0.2

0.5

0.5

1

Temperature effect on zero balance (TK 0)

% v. C n

/10 K

Maximum platform size Sensitivity (C n )

mm 300 300 450 450 600 600

Maximum capacity 1 kg

1.8 +15 %; ?10 % (Option 6: A1 = 1.8 mV/V "0.1 %)

2.0"10 % (Option 6: A = 2mV/V"0.1 %)

0"0.1

Maximum capacities 3 kg ...200 kg

mV/V

Zero balance

Temperature effect on sensitivity (TK C )3) Temperature, range +20...+40?C [+68...+104?F]

?10...+20?C [+14...+68?F]

% of C n /10 K

"0.0175 "0.0117 Non-linearity (d lin )3) "0.0166

"0.0166 "0.0166 "0.0233

300 (500)

Hysteresis error (d hy )3)

% of C n

Minimum dead load output return (MDLOR) Off center load error 4) Input resistance (R LC ) W

Output resistance (R 0)

300...500 (Option 6: A = 410W"0.2W)

Reference excitation voltage (U ref ) 5

Nom. range of excitation voltage (B U ) Max. excitation voltage

V 1 ... 12 15

Isolation resistance (R is ) at 100 V DC

Nom. (rated) range of ambient temperature (B T ) Operating temperature range (B tu ) Storage temperature range (B tl ) Limit load (E L )

GW >2

?10 ... +40 [+14 (104)

?C [?F] ?10 ... +50 [+14...+122] ?25 ... +70 *?13 (158)

150

300 300

% of E max Lateral load limit (E lq ), static Breaking load (E d )

Nominal (rated) displacement at E max (s nom approx.

), < 0.5

< 0.3

< 0.25

mm kg

Weight (G), approx. 0.45 IP67 Degree of protection acc. to EN 60 529 (IEC 529)

Material: Measuring body

Application protection Cable sheath

Aluminum Silicone rubber

PVC

1) According to OIMLR60 with P LC = 0.7. 2) Max. eccentric loading according to OIML R76.

3) The values for linearity deviation (d ), relative reversibility error (d ) and temperature effect on sensitivity (TK C ) are recommended values. lin hy

The sum of these values remain within the cumulated error limit acc. to OIML R60.

4) According to OIML R76.

HBM

2

B2610-1.0 en

Specifications (continuation)

Type

SP4MC6MR Accuracy class 1)

C6MR 6000 15 Maximum number of load cell intervals (n LC Maximum capacity (E max ) 2) Minimum LC verification interval (v min ) kg g 7 10 18 1 36 2

) 0.5 0.5

1

Temperature effect on zero balance (TK 0) % of C n

/ 10 K Max. platform size Sensitivity (C n ) Zero signal

mm 300 x 300

2.0"10 %

450 x 450 1.8"10 %

2.4"10 %

mV/V

0"0.10

Temperat. effect on sensitivity (TK C ) in the temperature range 3)

+20...+40?C [+68...+104?F] ?10...+20?C [+14 ... +68?F]

% of C n / 10 K

"0.0087 "0.0058 "0.0083 "0.0166

Relative reversibility error (d hy ) 3) Non-linearity (d lin ) 3)

Minimum dead load output return (MDLOR)

% of C n

"0.0083 "0.0116

Off-center load error 4)

1) According to OIMLR60 with P LC = 0.7

2) Max. eccentric loading according to OIML R76.

3) The values for linearity deviation (d ), relative reversibility error (d ) and temperature effect on sensitivity (TK C ) are recommended values. lin hy

The sum of these values remain within the cumulated error limit acc. to OIML R60. 4) According to OIML R76.

For more specifications, see table SP4M..., accuracy class C3MR (page 2)

Wiring code

Connection with 6 wire cable (24 AWG, cable length, selectable: 1.5 m; 3 m; 6 m; 12 m) Detailed description of the Pancon plug (CE100F26-6), 6-pole

Shield (yellow) = Shield connected to load cell body Contact 1 (white) = Signal (+) Contact 2 (red) = Signal (?) Contact 3 (black) = Excitation (?) Contact 4 (blue) = Excitation (+) Contact 5 (green) = Sense (+) Contact 6 (gray) = Sense (?) blue marking

B26091.0 de