当前位置:文档之家› Comparison of Droplet Digital PCR to Real-Time PCR for Quantitative Detection of Cytomegalovirus

Comparison of Droplet Digital PCR to Real-Time PCR for Quantitative Detection of Cytomegalovirus

Comparison of Droplet Digital PCR to Real-Time PCR for Quantitative Detection of Cytomegalovirus
Comparison of Droplet Digital PCR to Real-Time PCR for Quantitative Detection of Cytomegalovirus

1

Comparison of Droplet Digital PCR to Real-Time PCR for Quantitative Detection of Cytomegalovirus 1

Hayden RT 1, Gu Z 1, Ingersoll J 3, Abdul-Ali D 3, Shi L 2, Pounds S 2, Caliendo AM 3 2

3

1Departments of Pathology and 2Biostatistics, St. Jude Children’s Research Hospital, Memphis, TN; 4

3Department of Pathology and Laboratory Medicine, Emory University School of Medicine, Atlanta, GA 5

and Emory Center for AIDS Research, Emory University, Atlanta, GA 6

7

8

*Corresponding author: 9

Department of Pathology 10

St. Jude Children's Research Hospital 11

262 Danny Thomas Place 12

Rm C5025, Mailstop #250 13

Memphis, TN 38105-2794 14 Randall.Hayden@https://www.doczj.com/doc/823237107.html, 15

16

Running title: Comparison of Digital to Real-Time PCR 17

Key words: Real-time PCR, Digital PCR, Viral load, Cytomegalovirus

18

Copyright ? 2012, American Society for Microbiology. All Rights Reserved.J. Clin. Microbiol. doi:10.1128/JCM.02620-12 JCM Accepts, published online ahead of print on 5 December 2012

2

Abstract 19

Quantitative real-time PCR has been widely implemented for clinical viral load testing, but a lack of 20

standardization and relatively poor precision has hindered its usefulness. Digital PCR offers highly 21

precise, direct quantification without requiring a calibration curve. Performance characteristics of real-22

time PCR were compared to those of droplet digital PCR (ddPCR) for cytomegalovirus (CMV) viral load 23

testing. Ten-fold serial dilutions of the World Health Organization (WHO) and the National Institute of 24

Standards and Technology (NIST) CMV quantitative standards were tested, together with the 25

AcroMetrix? CMV tc Panel (Life Technologies, Carlsbad, CA) and 50 human plasma specimens. Each 26

method was evaluated using all three standards for quantitative linearity, lower limit of detection (LOD), 27

and accuracy. Quantitative correlation, mean viral load, and variability were compared. Real-time PCR 28

showed somewhat higher sensitivity than ddPCR (LOD of 3 log 10versus 4 log 10copies and IU/mL for NIST 29

and WHO standards). Both methods showed a high degree of linearity and quantitative correlation, for 30

standards (R 2≥ 0.98 in each of 6 regression models) and clinical samples (R 2=0.93) across their detectable 31

ranges. For higher concentrations, ddPCR showed less variability than RT-PCR for the WHO standards 32

and Acrometrix standards (p< 0.05). RT-PCR showed less variability and greater sensitivity than did 33

ddPCR in clinical samples. Both digital and real-time PCR provide accurate CMV viral load data over a 34

wide linear dynamic range. Digital PCR may provide an opportunity to reduce quantitative variability 35

currently seen using real-time PCR, but methods need to be further optimized to match the sensitivity of 36

real-time PCR. 37

38

39

40

3

Introduction 41

Over the past several years, viral load testing has evolved from a highly complex and labor-intensive 42

procedure to a routine part of patient care. Such methods are now integral to a diverse range of clinical 43

practice settings and diagnostic and treatment guidelines. These include their implementation in 44

patients with HIV or hepatitis, and post-transplant monitoring for cytomegalovirus, Epstein-Barr virus, 45

adenovirus, and BK virus, among others. Quantitative values have been used to follow efficacy of anti-46

viral therapy and to help determine changes in that therapy. Rising viral burden has been used as a 47

trigger for pre-emptive treatment, to prevent symptomatic infection. CMV testing is the archetype for 48

the latter application; viral load testing by real-time PCR in particular has changed the epidemiology of 49

CMV disease in transplant patients, and is now a standard part of post-transplant care (1, 4, 15, 16, 18, 50

30). 51

Despite these advances, challenges remain in viral load testing which relate primarily to intrinsic 52

limitations of current methodology and center on the issues of accuracy, standardization, and precision 53

(9, 19, 22, 29). Quantitative determinations by real-time PCR are indirect, depending on relationship of 54

cycle threshold of a test sample to a calibration curve. The latter, in turn, is typically generated by 55

testing a series of known standards, across the linear range of the assay. However, marked variation in 56

assay performance characteristics and in materials used as calibration standards may prevent 57

agreement between different laboratories, even when testing identical material. The use of 58

international standards, such as has been made available by the WHO (7, 11, 17), has helped mitigate 59

this issue, but has not resolved the problem. Standards are available only for a few of the most common 60

target analytes. Even when available, they may not be fully commutable, and thus may behave 61

differently, depending on the assay system utilized (5, 25). 62

63

4

Perhaps even more problematic is the poor reproducibility often seen among quantitative molecular 64

tests. Within and between lab precision, even when using identical methodology, can be surprisingly 65

poor (9, 13, 19). Variations in any aspect of these complex methods are magnified by the multistep 66

nature of the process and by the reliance on calibration curves, which themselves may vary over time. 67

The very nature of relying on the measure of a dynamic process, such as rate of target amplification, 68

carries with it intrinsic fluctuations that one could not expect to fully eliminate. This suggests that a 69

more direct method of quantification may be of value. 70

Droplet digital PCR is such a direct method (3, 12, 14, 20, 21, 26). It relies on limiting partition of PCR 71

reaction volume, such that a positive result in any of a large number of micro-reactions (in this case, 72

20,000) indicates the presences of a single target molecule in a given reaction. The number of positive 73

reactions, together with Poisson’s distribution, can then be used to produce a direct, high confidence 74

measurement of original target concentration. This methodology removes both the reliance on rate-75

based measurements (Ct values) and the need for the use of calibration curves. Studies targeting low-76

copy number genes, typically in the field of molecular oncology, have demonstrated a high degree of 77

sensitivity and precision of digital PCR compared to real-time PCR (6, 10, 23, 27). To-date, limited data 78

exists regarding the application of this new technology to viral load testing, but it promises to markedly 79

improve our ability to reproducibly quantify viruses and to obviate the need for developing a costly 80

series of international standards. 81

Herein, we compared ddPCR to real-time PCR for quantitative detection of CMV in both artificially 82

seeded samples and clinical plasma samples, the former including recently developed international 83

standard materials. 84

85

5

Materials and Methods 86

87

Reference materials 88

World Health Organization International (WHO) Standard for human cytomegalovirus (hCMV) 89

WHO-hCMV was purchased from the National Institute for Biological Standards and Control (NIBSC, 90

Potters Bar Hertfordshire, EN6 3QG United Kingdom) and was prepared from whole virus of the HCMV 91

Merlin strain. It was supplied as in a lyophilized format containing a total of 6.7 log 10 International Units 92

(IU) (5x106IU). One vial of the WHO standard was reconstituted with 1.0 mL of nuclease free water to a 93

concentration of 6.7 log 10 IU/mL (5x106 IU/mL). Dilutions were made in CMV negative human plasma to 94

obtain a 10-fold concentration gradient from 1-6 log 10IU/mL. For each concentration, 6 aliquots of 200 95

μL were extracted on the Qiagen EZ1 XL extractor using the Qiagen EZ1 Virus Mini Kit v2.0 (Qiagen, Inc., 96

Valencia, CA) following the manufacturer’s recommended protocol which includes the addition of an 97

internal control (IC) (7.2 μL). The specimens were eluted in 60 μL, pooled, aliquoted, and stored at -80o C. 98

99

National Institute of Standard & Technology (NIST) Standard Reference Material 100

NIST-HCMV was purchased from NIST (Gaithersburg, MD 20899) and prepared from a bacterial artificial 101

chromosome of CMV Towne ?147 BAC containing the genome of the Towne strain of CMV. The standard 102

was supplied in a unit of three component DNA materials, each component containing 150 μL of purified 103

DNA solution at a different concentration of CMV genome/μL. NIST Component C (127 μL, with a given 104

concentration of 19,641 copies/μL (4.29 log 10 copies /μL) was diluted with 373uL of TE for a 105

concentration of 6.7 log 10 copies/mL (5x106 copies/mL). Dilutions were made in TE to obtain a 10-fold

106

6

concentration gradient from 1-6 log 10 copies/mL and stored at -80o C. As the NIST was received as 107

purified nucleic acid; it was not extracted prior to amplification. 108

109

AcroMetrix? CMV tc Panel 110

The AcroMetrix? CMV tc Panel was purchased from AcroMetrix (Life Technologies, Carlsbad, CA), 111

containing intact, encapsulated viral particles of hCMV strain AD169. The panel consisted of a normal 112

human plasma (tested non-reactive for CMV DNA) and four members over a 10-fold concentration 113

gradient from 2.5 to 5.5 log 10 IU/mL (2.9x102 to 2.9x105 (IU/mL) in human plasma. For each 114

concentration, 200 μL was extracted and eluted in 60 μl on the Qiagen EZ1 XL extractor using the Qiagen 115

EZ1 Virus Mini Kit v2.0 (Qiagen, Inc., Valencia, CA) following the manufacturer’s recommended protocol. 116

Internal control (IC) (7.2 μL) was added pre-extraction. Six aliquots of 200 μL from each concentration 117

were processed, extracts were pooled, aliquoted, and stored at -80?C until usage. 118

119

Patient specimens 120

50 de-identified human plasma specimens were used in the study. Samples were collected from 121

residual material, originally collected for clinical testing at Emory University Hospital from June through 122

August of 2011. Samples were chosen to represent a wide spectrum of CMV concentrations, based on 123

clinical test results. For each clinical specimen, three aliquots of 200 μL were extracted on the Qiagen 124

EZ1 XL extractor using the Qiagen EZ1 Virus Mini Kit v2.0 (Qiagen, Inc., Valencia, CA) following the 125

manufacturer’s recommended protocol which includes the addition of an internal control (IC) (7.2 μL). 126

The specimens were eluted in 60 μL, pooled, aliquoted and stored at -80o C. This protocol was approved 127

by the Emory University Institutional Review Board (IRB).

128

7

129

Droplet Digital PCR (ddPCR) 130

A laboratory developed test (LDT) for CMV detection was used, together with the QX100? Droplet 131

Digital? PCR System (Bio-Rad, Pleasanton, CA). The ddPCR reaction mixture consisted of 10 μL of a 2x 132

ddPCR Mastermix (Bio-Rad), 2 μL of CMV primer/probe mix (artus CMV PCR analyte specific reagent, 133

Qiagen, Valencia, CA), and 5 μL of sample nucleic acid solution in a final volume of 20 μL. The entire 134

reaction mixture was loaded into a disposable plastic cartridge (Bio-Rad) together with 70 μL of droplet 135

generation oil (Bio-Rad) and placed in the droplet generator (Bio-Rad). After processing, the droplets 136

generated from each sample were transferred to a 96-well PCR plate (Eppendorf, Germany). PCR 137

amplification was carried out on a T100 Thermal Cycler (Bio-Rad) using thermal profile of beginning at 138

95?C for 10 minutes, followed by 40 cycles of 94?C for 30 seconds and 60?C for 60 seconds and 1 cycle of 139

98?C for 10 minutes, and ending at 12?C. After amplification, the plate was loaded on the Droplet Reader 140

(Bio-Rad) and the droplets from each well of the plate were read automatically at a rate of 32 wells per 141

hour. ddPCR data was analyzed with QuantaSoft analysis software (Bio-Rad) and the quantification of 142

target molecule was presented as copies per μL of PCR. 143

144

Quantitative real-time PCR 145

Quantitative CMV PCR testing was performed using an LDT incorporating artus CMV PCR analyte specific 146

reagents (Qiagen, Inc., Valencia, CA) on the Qiagen Rotor Gene instrument. For each amplification 147

reaction, 20 μL of purified nucleic acid was added to 30 μL of a reaction mixture of master mix and 148

magnesium. The thermo-cycler parameters were as follows: hold 10 minutes at 95?C followed by 15 149

seconds at 95?C, 30 seconds at 65?C and 20 seconds at 72?C for 45 cycles. A four point standard curve as

150

8

well as a positive and a negative control were included on all runs. The laboratory-determined limit of 151

detection (LOD) in this assay using plasma was 2.3 log 10 IU/mL. 152

Statistical analysis 153

For each standard’s dilution series, the concentration was log-transformed as log 10(concentration+1) for 154

purposes of subsequent statistical analyses. Digital PCR measurements were transformed as 155

log 10(copies/ml+1) and real-time PCR measurements were transformed as log 10(IU/1.38 +1) for purposes 156

of subsequent statistical analyses. The factor of 1.38 was used in the transformation of real-time PCR 157

measurements to convert IU/ml to the copies/ml units reported by digital PCR. One was added prior to 158

log-transformation to avoid obtaining undefined values by taking the log of zero for negative samples. 159

For each technology and standard, the limit of detection (LOD) was defined as the smallest nonzero 160

concentration at which all replicates gave a positive qualitative result. For each standard and 161

technology, the log-transformed measurements at or above the LOD were regressed against the log-162

transformed concentrations by simple least squares. Bartlett’s test(2) was used to compare standard 163

deviations of log-transformed results across technologies for each concentration and standard. 164

For clinical samples, the mean log-transformed measurement was computed for each technology and 165

each clinical sample. Simple least squares regression was used to evaluate the quantitative agreement 166

of the mean values of the two technologies. For each sample, the number of repeats giving a positive 167

result was determined for each technology. The number of samples with a difference in the number of 168

positive results was determined. A binomial test was used to determine the significance of the number 169

of samples with fewer positive results by digital PCR than by real-time PCR. Similarly, the standard 170

deviation of results was computed for each sample and technology. A binomial test was used to 171

determine he statistical significance of the number of samples with a smaller standard deviation by real-172

time PCR than by digital PCR.

173

9

Results 174

Evaluation using quantitative standards 175

The three quantitative standards (WHO, NIST, Acrometrix) were tested as unknowns by both real-time 176

PCR and ddPCR to generate analytical operating characteristics for each method. Each concentration of 177

each standard was run in triplicate on three separate runs (total of nine results per sample). Analysis 178

was performed to determine limit of detection, quantitative linearity, quantitative agreement compared 179

both to nominal concentration and to results from real-time PCR, and reproducibility of results using 180

each methodology. 181

182

As shown in Table 1, each method had the same LOD when using Acrometrix standards (2.48 log IU/mL). 183

However, real-time PCR demonstrated a lower LOD by one log unit dilution for both the WHO and NIST 184

standards (10-fold greater sensitivity than ddPCR). Digital PCR (dPCR) detected 100% of samples at log 3 185

IU/mL using WHO material and at log 4 copies/mL using NIST material. 186

187

Both methods showed excellent linearity above the LOD. For each technology and standard, the 188

estimated slope coefficients were close to 1 (0.93-1.04) and R-squared values were very close to 1 189

(≥0.98), indicating both linearity and correlation between nominal and measured values (Figure 1). 190

Quantitative correlation was also shown between methods (Figure 2). Correlation was reduced, 191

particularly for ddPCR, at the low end of each dilution series, corresponding to analyte concentrations 192

with less than 100% detection. Regression model equations for Figures 1 and 2 are shown in 193

Supplemental Tables 1 and 2, respectively.

194

10

The lower sensitivity of ddPCR may be largely attributed to use of a 4-fold smaller input volume. When 195

we plotted the probability of detection against the nominal total number of input viral copies 196

(copies/mL × input volume in mL), the two technologies showed very similar performance 197

(Supplemental Figure S1). This indicated that the sensitivity difference was largely attributable to the 198

input volume differences. 199

Table 1 shows the standard deviation and mean log 10 transformed copies for each technology, standard 200

and concentration. Only matched concentrations of each standard run with both methods are reported. 201

Overall, the methods produced similar results. The maximum differences between mean measurement 202

values of the two technologies for all concentrations tested were 0.12, 0.37, and 1.27 log 10 copies/mL 203

for Acrometrix, NIST, and WHO standards, respectively. Among concentrations above the LOD for both 204

methods, the maximum differences were 0.12, 0.37, and 0.45. Digital PCR showed significantly less 205

variability than real-time PCR for Acrometrix log 10 concentrations ≥ 4.48 and for WHO log 10 206

concentrations ≥ 4. Real-time PCR showed significantly less variability than digital PCR at NIST log 10 207

concentration = 3 and WHO log 10 concentration = 2. The two technologies did not show significantly 208

different variability for the other concentrations of the standards. 209

210

Evaluation using clinical samples 211

Clinical samples were each tested in duplicate on three separate PCR runs (total of six results per 212

sample, Table S3). Mean viral load and result variability was compared for each method. Figure 3 shows 213

linear regression and Bland-Altman plots comparing the mean quantitative values for ddPCR with those 214

of real-time PCR, demonstrating close agreement between the two systems (mean difference=-0.247 215

log 10 copies/ml; SD=0.336 log 10 copies/ml).

216

11

Digital PCR had a significantly lower positivity rate than real-time PCR. Digital PCR obtained fewer 217

positive results among its six repeated replicates than did real-time PCR for 14 samples but the converse 218

was not true for any samples (p = 0.0001). Additionally, the variability of real-time results was less than 219 that of digital PCR results for 34 of 50 (68%) samples (95% CI = 53%- 81%; p = 0.015). 220

221

12

Discussion 222

This study offers one of the earliest published assessments of digital PCR as a viable means for 223

quantitative detection of DNA viremia in clinical samples. Previous work with this relatively new 224

technology has focused on its application in molecular oncology (6, 21). However, the utility of a direct 225

(rather than relative) measure of quantification for the field of clinical molecular virology cannot be 226

overstated, particularly if demonstrated to be accurate, sensitive, and precise. Findings here show 227

ddPCR to have such potential for the measurement of CMV. While the current assay showed somewhat 228

lower sensitivity than real-time methodology, other operating characteristics were comparable. 229

Quantitative accuracy was high and generated viral loads that matched well with nominal values of 230

international and commercial quantitative standards. Linearity and quantitative correlation with real-231

time PCR results were high; and precision matched or exceeded that of real-time PCR, when assessed 232

within the analytical measurement range of the test. Furthermore, our data suggest that the lower 233

sensitivity and greater variability of this ddPCR may be attributed to a lower input volume; however, our 234

experiment was not designed to definitively determine the impact of input volume on sensitivity. 235

Another experiment designed explicitly for that purpose may be necessary to confirm our 236

interpretation. Clinical samples showed greater variability with ddPCR than with real-time PCR, while 237

the converse was true for reference materials. Samples detected near the LOD of either method showed 238

greater variability. Because the LOD of ddPCR was somewhat higher than for real-time PCR, the results 239

of clinical samples disproportionately fell close to the LOD of ddPCR, potentially accounting for the 240

increased variability (since clinical samples tended to have a lower viral load overall, while reference 241

materials had their concentrations evenly distributed along the linear range of the assays). 242

243

13

The disadvantages of real-time PCR for such measurements have been well-documented. Several 244

authors have now shown a high degree of quantitative variability among such assays, not only for those 245

targeting CMV, but also for other blood-borne viruses (8, 9, 19, 22). Laboratory developed assays are 246

particularly prone to such variability, and a wide range of factors have been shown to affect both 247

accuracy and precision of these tests (9). The development of international standards for CMV and other 248

such viruses promises to mitigate such problems (particularly that of accuracy), but numerous issues 249

remain. These include the difficulty of generating widely commutable standards and other aspects of 250

assay design that impinge on amplification efficiency, all integrally related to the generation of 251

reproducible results when measurement depends on relative rates of amplification, as with real-time 252

methods. In addition, gains seen in the development of international standards and automated, FDA 253

cleared assays are unlikely to be duplicated for the wide range of viruses currently of clinical interest 254

and may be limited only to those with a strong commercial market. 255

The clinical import of improved assay characteristics is multifold. Present methods for measuring CMV 256

load may show quantitative variability among laboratories exceeding four orders of magnitude (9, 19). 257

Quantitative accuracy and precision are crucial to assay interpretation and to the ability to set 258

meaningful, universal treatment thresholds, both for clinical disease attribution and for pre-emptive 259

therapeutic strategies. In addition, the portability of results can have direct implications for patient care. 260

Currently, ongoing treatment of patients who move from one institution to another may be challenging 261

and require new baseline viral loads. Comparability of clinical study results that rely on such 262

measurements is a final issue of importance, fundamental to advancing patient care. While the recent 263

availability of an FDA-cleared, commercial assay may improve the situation for CMV testing, its impact 264

remains uncertain. Digital PCR may address all of these issues, utilizing end-point PCR for direct 265

measurement of viral loads, potentially improving accuracy and precision over current, real-time 266

methods.

267

14

Since it was initially described in the 1990’s, other authors have shown digital PCR to have advantages in 268

several early manuscripts, focused primarily on research applications in molecular genetics and 269

oncology. Specifically, it has been used for mutational analysis, assessment of allelic imbalance, cancer 270

detection, and allelic expression analysis (3, 6, 21). Its ability to detect and accurately quantify minority 271

mutations in a predominating background of normal sequence has proven advantageous over other PCR 272

methodologies. This capability has also been exploited in the area of prenatal diagnostics, where its use 273

for the detection of fetal aneuploidies and single gene Mendelian disorders using maternal plasma 274

samples has also been described (31). Few authors, however, have yet investigated the application of 275

this technology to infectious disease diagnostics. Experimental systems have demonstrated proof-of-276

principle for quantification of adenoviral genome copies, GB virus in transfected cell culture lines, serial 277

dilutions of hepatitis C virus and human immunodeficiency virus (HIV) RNA, and purified, serially diluted 278

HIV RNA from two clinical samples (14, 24, 28). The latter studies demonstrated a 4 log 10 linear range 279

and variable agreement with real-time PCR methods. None of these studies examined a commercially 280

available digital PCR system, amenable to routine clinical use; and none utilized large numbers of clinical 281

samples in a clinical laboratory setting. This report supports the clinical applicability of ddPCR for 282

routine use in clinical diagnostic molecular virology. The system used here has a relatively small 283

footprint (requiring less than two linear feet of bench space for both the droplet generator and reader, 284

together). It can be used at roughly the same cost as real-time PCR (as most reagents used here were 285

identical in the two systems) and has scalable, rapid throughput (6.5 hours for 96 reactions and 2.2 286

hours for 8 reactions, compared to 4 hours for 33 real-time PCR reactions). Hands on time for ddPCR 287

was 25 minutes for 8 and 120 minutes for 96 reactions, while for real-time PCR it was 85 minutes for 33 288

reactions. This now makes digital PCR practical for routine use in clinical laboratories, adding relevance 289

to the results shown above.

290

15

The conclusions here are of course limited based on the number and genetic heterogeneity of clinical 291

samples available. In addition, the conclusions here are drawn based on comparison using only one 292

assay design, targeting only one of a number of viruses for which such a method may have clinical utility. 293

Indeed, work ongoing in our laboratory seeks to extend the present study by looking at other 294

comparator assays and other viruses, particularly those with RNA genomes. In addition, the 295

development and use of other digital platforms will be crucial to building upon this work. The 296

somewhat higher LOD, compared to real-time PCR seen can be accounted for entirely by the small total 297

sample volume, compared to the real-time system used (5μL for ddPCR, compared to 20μL for real-time 298

PCR, statistical analysis not shown). Increased reaction volume for ddPCR, currently limited by system 299

design, might also lead to improved quantitative correlation between methods at low concentrations of 300

analyte and might extend the precision advantage of ddPCR, already seen higher concentrations. Finally, 301

future studies comparing ddPCR to the recently released FDA-approved test for CMV viral load testing 302

(and others, as they become available) may prove valuable in assessing the future roles for each 303

technology. 304

305

The data shown here confirm earlier work in experimental oncology and show proof of principle for the 306

application of ddPCR for routine use in clinical, molecular virology. The advantages of direct 307

quantification by end-point, limiting partition PCR could obviate the need for development and purchase 308

of costly quantitative calibrators. This paradigm shift could reduce testing costs, improve accuracy and 309

precision of quantitative viral load assays. Such improvements will have direct clinical utility, particularly 310

in diagnostic microbiology and virology laboratories serving immunocompromised patients, where the 311

number of viruses of importance continues to outstrip our ability to develop methods for quantitative 312

detection, and where our present diagnostic tools remain severely limited in utility due to suboptimal 313

operating characteristics. These findings support both the continued development of this technology

314

16

and further work to explore its value for routine use in clinical diagnostic testing for infectious 315

pathogens. 316

317

Acknowledgements: 318

Artus CMV RG PCR reagents were graciously provided by Qiagen, Inc. (Valencia, CA). This work was 319

supported in part by the Anderson Charitable Foundation, the American Lebanese Syrian Associated 320

Charities (ALSAC), and the Emory Center for AIDS Research (P30 AI050409). Dr. Caliendo is a co-321

investigator on a CMV viral load clinical trial that is funded by Qiagen. 322

323

324

325

326

327

References 328

1. Barker, J. N., R. E. Hough, J. A. van Burik, T. E. DeFor, M. L. MacMillan, M. R. O'Brien, and J. E. 329

Wagner. 2005. Serious infections after unrelated donor transplantation in 136 children: impact 330

of stem cell source. Biol Blood Marrow Transplant 11:362-370. 331

2. Bartlett, M. S. 1937. Properties of sufficiency and statistical tests. Proceedings of the Royal 332

Society of London Series A 160:268–282.

333

17

3. Bhat, S., J. Herrmann, P. Armishaw, P. Corbisier, and K. R. Emslie. 2009. Single molecule 334

detection in nanofluidic digital array enables accurate measurement of DNA copy number. Anal 335

Bioanal Chem 394:457-467. 336

4. Boeckh, M., and P. Ljungman. 2009. How we treat cytomegalovirus in hematopoietic cell 337

transplant recipients. Blood 113:5711-5719. 338

5. Caliendo, A. M., M. D. Shahbazian, C. Schaper, J. Ingersoll, D. Abdul-Ali, J. Boonyaratanakornkit, 339

X. L. Pang, J. Fox, J. Preiksaitis, and E. R. Schonbrunner. 2009. A commutable cytomegalovirus 340

calibrator is required to improve the agreement of viral load values between laboratories. Clin 341

Chem 55:1701-1710. 342

6. Diehl, F., and L. A. Diaz, Jr. 200

7. Digital quantification of mutant DNA in cancer patients. Curr 343

Opin Oncol 19:36-42. 344

7. Fryer, J. F. H., A. B.; Anderson, R.; Minor, P. D.;the Collaborative Study Group. 2010. 345

Collaborative Study to Evaluate the PRoposed 1st WHO International Standard for Human 346

Cytomegalovirus (HCMV) for Nucleic Acid Amplification (NAT)-Based Assays. National Institute 347

for Biological Standards and Control. 348

8. Hayden, R. T., K. M. Hokanson, S. B. Pounds, M. J. Bankowski, S. W. Belzer, J. Carr, D. Diorio, M. 349

S. Forman, Y. Joshi, D. Hillyard, R. L. Hodinka, M. N. Nikiforova, C. A. Romain, J. Stevenson, A. 350

Valsamakis, H. H. Balfour, Jr., and U. S. E. W. Group. 2008. Multicenter comparison of different 351

real-time PCR assays for quantitative detection of Epstein-Barr virus. J Clin Microbiol 46:157-352

163. 353

9. Hayden, R. T., X. Yan, M. T. Wick, A. B. Rodriguez, X. Xiong, C. C. Ginocchio, M. J. Mitchell, and A. 354

M. Caliendo. 2012. Factors contributing to variability of quantitative viral PCR results in 355

proficiency testing samples: a multivariate analysis. J Clin Microbiol 50:337-345.

356

18

10. Hindson, B. J., K. D. Ness, D. A. Masquelier, P. Belgrader, N. J. Heredia, A. J. Makarewicz, I. J. 357

Bright, M. Y. Lucero, A. L. Hiddessen, T. C. Legler, T. K. Kitano, M. R. Hodel, J. F. Petersen, P. W. 358

Wyatt, E. R. Steenblock, P. H. Shah, L. J. Bousse, C. B. Troup, J. C. Mellen, D. K. Wittmann, N. G. 359

Erndt, T. H. Cauley, R. T. Koehler, A. P. So, S. Dube, K. A. Rose, L. Montesclaros, S. Wang, D. P. 360

Stumbo, S. P. Hodges, S. Romine, F. P. Milanovich, H. E. White, J. F. Regan, G. A. Karlin-361

Neumann, C. M. Hindson, S. Saxonov, and B. W. Colston. 2011. High-throughput droplet digital 362

PCR system for absolute quantitation of DNA copy number. Anal Chem 83:8604-8610. 363

11. Holden, M. J., R. M. Madej, P. Minor, and L. V. Kalman. 2011. Molecular diagnostics: 364

harmonization through reference materials, documentary standards and proficiency testing. 365

Expert Rev Mol Diagn 11:741-755. 366

12. Kiss, M. M., L. Ortoleva-Donnelly, N. R. Beer, J. Warner, C. G. Bailey, B. W. Colston, J. M. 367

Rothberg, D. R. Link, and J. H. Leamon. 2008. High-throughput quantitative polymerase chain 368

reaction in picoliter droplets. Anal Chem 80:8975-8981. 369

13. Kraft, C. S., W. S. Armstrong, and A. M. Caliendo. 2012. Interpreting quantitative 370

cytomegalovirus DNA testing: understanding the laboratory perspective. Clin Infect Dis 54:1793-371

1797. 372

14. Kreutz, J. E., T. Munson, T. Huynh, F. Shen, W. Du, and R. F. Ismagilov. 2011. Theoretical design 373

and analysis of multivolume digital assays with wide dynamic range validated experimentally 374

with microfluidic digital PCR. Anal Chem 83:8158-8168. 375

15. Ljungman, P. 2010. Molecular monitoring of viral infections after hematopoietic stem cell 376

transplantation. Int J Hematol 91:596-601. 377

16. Ljungman, P., M. Hakki, and M. Boeckh. 2011. Cytomegalovirus in hematopoietic stem cell 378

transplant recipients. Hematol Oncol Clin North Am 25:151-169.

379

19

17. Madej, R. M., J. Davis, M. J. Holden, S. Kwang, E. Labourier, and G. J. Schneider. 2010. 380

International standards and reference materials for quantitative molecular infectious disease 381

testing. J Mol Diagn 12:133-143. 382

18. Mori, T., and J. Kato. 2010. Cytomegalovirus infection/disease after hematopoietic stem cell 383

transplantation. Int J Hematol 91:588-595. 384

19. Pang, X. L., J. D. Fox, J. M. Fenton, G. G. Miller, A. M. Caliendo, and J. K. Preiksaitis. 2009. 385

Interlaboratory comparison of cytomegalovirus viral load assays. Am J Transplant 9:258-268. 386

20. Pinheiro, L. B., V. A. Coleman, C. M. Hindson, J. Herrmann, B. J. Hindson, S. Bhat, and K. R. 387

Emslie. 2012. Evaluation of a droplet digital polymerase chain reaction format for DNA copy 388

number quantification. Anal Chem 84:1003-1011. 389

21. Pohl, G., and M. Shih Ie. 2004. Principle and applications of digital PCR. Expert Rev Mol Diagn 390

4:41-47. 391

22. Preiksaitis, J. K., X. L. Pang, J. D. Fox, J. M. Fenton, A. M. Caliendo, and G. G. Miller. 2009. 392

Interlaboratory comparison of epstein-barr virus viral load assays. Am J Transplant 9:269-279. 393

23. Sanders, R., J. F. Huggett, C. A. Bushell, S. Cowen, D. J. Scott, and C. A. Foy. 2011. Evaluation of 394

digital PCR for absolute DNA quantification. Anal Chem 83:6474-6484. 395

24. Shen, F., B. Sun, J. E. Kreutz, E. K. Davydova, W. Du, P. L. Reddy, L. J. Joseph, and R. F. Ismagilov. 396

2011. Multiplexed quantification of nucleic acids with large dynamic range using multivolume 397

digital RT-PCR on a rotational SlipChip tested with HIV and hepatitis C viral load. J Am Chem Soc 398

133:17705-17712. 399

25. Vesper, H. W., W. G. Miller, and G. L. Myers. 2007. Reference materials and commutability. The 400

Clinical biochemist Reviews / Australian Association of Clinical Biochemists 28:139-147. 401

26. Vogelstein, B., and K. W. Kinzler. 1999. Digital PCR. Proc Natl Acad Sci U S A 96:9236-9241.

402

20

27. Whale, A. S., J. F. Huggett, S. Cowen, V. Speirs, J. Shaw, S. Ellison, C. A. Foy, and D. J. Scott. 2012. 403

Comparison of microfluidic digital PCR and conventional quantitative PCR for measuring copy 404

number variation. Nucleic acids research 40:e82. 405

28. White, R. A., 3rd, S. R. Quake, and K. Curr. 2012. Digital PCR provides absolute quantitation of 406

viral load for an occult RNA virus. J Virol Methods 179:45-50. 407

29. Wolff, D. J., D. L. Heaney, P. D. Neuwald, K. A. Stellrecht, and R. D. Press. 2009. Multi-Site PCR-408

based CMV viral load assessment-assays demonstrate linearity and precision, but lack numeric 409

standardization: a report of the association for molecular pathology. J Mol Diagn 11:87-92. 410

30. Zhang, L. F., Y. T. Wang, J. H. Tian, K. H. Yang, and J. Q. Wang. 2011. Preemptive versus 411

prophylactic protocol to prevent cytomegalovirus infection after renal transplantation: a meta-412

analysis and systematic review of randomized controlled trials. Transpl Infect Dis 13:622-632. 413

31. Zimmermann, B. G., S. Grill, W. Holzgreve, X. Y. Zhong, L. G. Jackson, and S. Hahn. 2008. Digital 414

PCR: a powerful new tool for noninvasive prenatal diagnosis? Prenat Diagn 28:1087-1093. 415

416

417

相关主题
文本预览
相关文档 最新文档